POLARIS PRESSURE ADJUSTABLE VALVE SYSTEM, MODELS SPV-140, SPV-300, SPV-400

K042481 · Sophysa SA · JXG · Dec 14, 2004 · Neurology

Device Facts

Record IDK042481
Device NamePOLARIS PRESSURE ADJUSTABLE VALVE SYSTEM, MODELS SPV-140, SPV-300, SPV-400
ApplicantSophysa SA
Product CodeJXG · Neurology
Decision DateDec 14, 2004
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.5550
Device ClassClass 2
AttributesTherapeutic, Pediatric

Intended Use

To drain cerebrospinal fluid (CSF) for the management of hydrocephalus.

Device Story

Implantable pressure-adjustable valve system for hydrocephalus management; shunts CSF from cerebral ventricles to abdominal cavity or right atrium. Features ball-in-cone valve seat for constant intraventricular pressure maintenance. Pressure settings adjustable transcutaneously. Incorporates self-locking rotor-shuttle micro-magnet system; prevents unintended pressure changes from unidirectional magnetic fields or MRI exposure by ensuring rotor remains fixed. Used by neurosurgeons in clinical settings.

Clinical Evidence

No clinical data provided; substantial equivalence based on design, performance, and functional characteristics.

Technological Characteristics

Implantable pressure-adjustable valve; ball-in-cone valve seat design; self-locking rotor-shuttle micro-magnet system for MRI compatibility; pressure-adjustable transcutaneously; models SPV-140, SPV-300, SPV-400 with varying pressure ranges and adjustment increments.

Indications for Use

Indicated for adult and pediatric patients requiring cerebrospinal fluid (CSF) drainage for the management of hydrocephalus.

Regulatory Classification

Identification

A central nervous system fluid shunt is a device or combination of devices used to divert fluid from the brain or other part of the central nervous system to an internal delivery site or an external receptacle for the purpose of relieving elevated intracranial pressure or fluid volume (e.g., due to hydrocephalus). Components of a central nervous system shunt include catheters, valved catheters, valves, connectors, and other accessory components intended to facilitate use of the shunt or evaluation of a patient with a shunt.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ 042481 ## 10. SMDA Summary of Safety and Effectiveness SMDA Summary of Safety and Effectiveness ### 510(k) Summary Polaris® SPV-140, SPV-300, SPV-400 Pressure Adjustable Valve A. Submittor Information Sponsor: Manufacturer: SOPHYSA SA SOPHYSA SA C/o Interactive Consulting Inc. 22 rue Jean Rostand 70 Walnut Street Parc Club Orsay Université Wellesley, MA 02481 91893 ORSAY Cedex, France Tel: (781) 239-8108 Tel: 011-331-69 35 35 00 Fax: (781) 239-8005 Fax: 011 331 69 35 36 90 Website : www.sophysa.com | Contact Person: | Jean-Christophe Audras, Regulatory Affairs | |-----------------|--------------------------------------------| | Date Prepared: | September 10, 2004 | B. Device Identification Common/Usual Name: Proprietary Name: Regulatory Class: Hydrocephalus Shunt Polaris® SPV-140, SPV-300, SPV-400 Pressure Adjustable Valve System Class II by 21 CFR 882,5550 C. Identification of Predicate Device(s) The Polaris® Pressure Adjustable Valve models SPV-140, SPV-300, SPV-400 is substantially equivalent to the SOPHYSA Sophy® Polaris® SPV Valve (K031097) previously cleared and currently marketed. #### D. Device Description The Polaris® Pressure Adjustable Valve models SPV-140, SPV-300, SPV-400 is an implantable device designed for the treatment of hydrocephalus in adult and pediatric patients by shunting, thereby providing continuous, controlled intraventricular pressure and CSF drainage from the cerebral ventricles. Intraventricular pressure is maintained at a constant level by the device's ballin-cone valve seat design, and the value is pressure-adjustable transcutaneously. Drroinage is directed to the abdominal cavity or to the right atrium of the heart. The basic settings of the Polaris® SPV-140 Valve are 10, 40, 80, 110 and 140 mm H₂O; adjustments to intermediate pressures are made manually in 30 mm H20 increments (decrements). The basic settings of the Polaris® SPV-300 Valve are 50, 100, 150, 220 and 300 mm H2O; adjustments to intermediate pressures are made manually in 50-70 mm H20 increments (decrements). The basic settings of the Polaris® SPV-400 Valve are 80, 150, 230, 330 and 400 mm H2O; adjustments to intermediate pressures are made manually in 70-100 mm H20 increments (decrements). The specific feature of the self-locking rotor-shuttle micro-magnet system of the passive Polaris® SPV Valve is that the adjustment position of each pressure setting cannot be changed by a unidirectional magnetic field. A domestic magnetic field or an exposure to MRI attracts the Image /page/0/Picture/18 description: The image shows the number 20020. The numbers are written in a simple, sans-serif font. The numbers are slightly faded, but still legible. The image is a close-up of the numbers. {1}------------------------------------------------ shuttles in the same direction and thus cannot unlock them simultaneously, therefore the rotor cannot be mobilized and the pressure setting remains fixed and constant. ### E. Substantial Equivalence The Polaris® Pressure Adjustable Valve models SPV-140, SPV-300, SPV-400 is substantially equivalent to the Polaris® SPV Pressure Adjustable Valve System (K031097) in terms of intended use, materials, biocompatibility, design, performance, function, and operating characteristics. #### F. Indications for Use To drain cerebrospinal fluid (CSF) for the management of hydrocephalus. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized symbol of an eagle or bird with three wing-like shapes. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 DEC 1 4 2004 Sophysa SA c/o Ms. Jackie Masse Interactive Consulting, Inc. 70 Walnut Street Wellesley, Massachusetts 02481 Re: K042481 Trade/Device Name: Polaris® Pressure Adjustable Valve System Models Polaris® SPV-140, SPV-300, SPV-400 Regulation Number: 21 CFR 882.5550 Regulation Name: Central nervous system fluid shunt and components Regulatory Class: II Product Code: JXG Dated: November 8, 2004 Received: November 9, 2004 Dear Ms. Masse: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ Page 2 - Ms. Jackie Masse This letter will allow you to begin marketing your device as described in your Section 510(k) I mis lotter will and h yourse of substantial equivalence of your device to a legally premailed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Miriam C Provost Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Indications for Use K042481 510(k) Number (if known): Device Name: Polaris® Pressure Adjustable Valve System Models Polaris® SPV-140, SPV-300, SPV-400 Indications For Use: To drain cerebrospinal fluid (CSF) for the management of hydrocephalus Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use _ (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Miriam C. Provost (Division Sign-Off Division of General, Restorative, and Neurological Devices 510(k) Number_________________________________________________________________________________________________________________________________________________________________ Page 1 of 1
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...