K042369 · Sonoscape Company Limited · IYN · Oct 20, 2004 · Radiology
Device Facts
Record ID
K042369
Device Name
SONOSCAPE ULTRASOUND SYSTEM SSI-1000/SSI-5000
Applicant
Sonoscape Company Limited
Product Code
IYN · Radiology
Decision Date
Oct 20, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1550
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The device is a general-purpose ultrasonic imaging instrument intended for use by a qualified physician for evaluation of Abdomen; Pediatric; Small Organ (breast, tests, thyroid); heart soft tissue; Peripheral Vascular, Musculo-skeletal (conventional) and Urology.
Device Story
SonoScape SSI-1000/SSI-5000 is a diagnostic ultrasound system; available as portable (SSI-1000) or roll-around (SSI-5000) units. System uses phased, linear, and curved linear array transducers to acquire ultrasound data via piezoelectric effect. Operates in B-Mode, M-Mode, Color-Flow, Pulsed, Power, and CW Doppler modes. Features 64-channel digital beam former, progressive dynamic receive focusing, and native frequency digital scan converter. Operated by physicians in clinical settings. Output displayed on LCD (SSI-1000) or CRT (SSI-5000) monitors. Provides real-time imaging and measurements (distance, area, velocity, etc.) to assist in clinical diagnosis of abdominal, cardiac, vascular, and musculoskeletal conditions. Includes LAN port for image management and USB 2.0 for data transport.
Clinical Evidence
Bench testing only. No clinical data provided. System performance validated through acoustic output measurements, safety compliance testing (IEC601-1, IEC60601-2), and comparison of imaging capabilities to predicate devices.
Technological Characteristics
All-digital architecture; 64-channel beam former; progressive dynamic receive focusing. Supports phased, linear, and curved linear array probes. Modes: B, M, Color/Power/PW/CW Doppler, Tissue Harmonic Imaging. Connectivity: LAN, USB 2.0. Operating system: Linux. Safety: IEC601-1, IEC60601-2. Acoustic output: Track 3, MI max 1.9, Ispta max 720mW/cm2.
Indications for Use
Indicated for diagnostic ultrasound imaging or fluid flow analysis in patients for applications including Abdominal, Ob/GYN, Small Organ (breast, thyroid, testes), Cardiac, Peripheral Vascular, and Musculo-skeletal (conventional).
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{1}------------------------------------------------
### OCT 2 0 2004 PREMARKET NOTIFICATION [510(k)] Summary
| Trade Name: | SonoScape Ultrasound System, SSI-1000/SSI-5000TM<br>with PA2.5 MHz Phased Array, LA7.5 MHz Linear Array, and<br>CLA3.5 MHz Curved Linear Array. |
|----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Diagnostic Ultrasound System |
| Classification Name: | Ultrasonic Pulsed Echo Imaging System, 90 IYO |
| Manufacturer's Name: | SonoScape Company Limited<br>4/F., Yizhe Building, Yuquan Road,<br>Nanshan, 518051, Shenzhen, China |
| Contact: | Mr. Jinzhong Yao, President<br>Telephone: (86) 755-26722890<br>Fax: (86) 755-26722850 |
| U.S. Agent: | Bob Leiker<br>Quality & Regulatory Services, Inc.<br>Dublin, CA 94568 |
GE Logiq 500 - K970901, K991611, and K010329 Predicate Device:
The SonoScape SSI-1000/SSI-5000 ultrasound system is an Device Description: integrated preprogrammed color ultrasound imaging system, capable of producing high detail resolution intended for clinical diagnostic imaging applications.
The all digital architecture with progressive dynamic receive focusing allows the system to maximize the utility of all imaging transducers to enhance the diagnostic utility and confidence provided by the system. The exam dependent default setting allows the user to have minimum adjustment for imaging the patient, while the in depth soft-menu control allows the advanced user to set the system
Image /page/0/Picture/7 description: The image shows a portable computer with a screen and keyboard. The computer is labeled "SSI-1000" at the bottom left of the image. The computer has a handle on top, suggesting it is designed to be carried around. The screen is dark, and the keyboard is visible below it.
for different situations. The architecture allows cost-effective system integration to a variety of upgrade-able options and features.
{2}------------------------------------------------
The SonoScape System can be configured either as a portable (SSI-1000) model, or as a roll-around model on wheels (SSI-5000). These systems are designed with
the latest technology, using the same quality procedures as ultrasound systems which have been available in the market for vears.
This SonoScape system is a general purpose, software controlled, diagnostic ultrasound system. Its basic function is to acquire ultrasound data and display the image in B-Mode (including Tissue Harmonic Image), M-Mode, Color-Flow Doppler, Pulsed Doppler and Power Doppler, or a combination of these modes.
The major features of the SonoScape SSI-1000/SSI-5000:
- 64 Channel all digital beam former .
- Progressive dynamic receive focusing .
- Wide band all digital demodulation .
- Native frequency digital scan converter .
- SSI-1000 can be hand carried for portable use .
- Remote access image management through LAN port .
- USB2.0 flash drive for image transport and software upgrade .
- Support for Phase array, Linear array, and Curve Linear array probes .
- Based on Linux operating system .
- Supports 2D B-mode, M-mode, Harmonic Image, Color, Power Doppler, Pulse . wave Doppler, and CW.
The SSI-1000 utilizes an LCD viewing monitor, and the SSI-5000 uses a CRT monitor. The following drawings are provided for illustration;
Image /page/1/Figure/17 description: The image shows three different models of medical equipment, labeled as SSI-1000 and SSI-5000. The SSI-1000 model is a boxy unit with horizontal ventilation slats. The SSI-5000 model has a monitor on top and an articulated arm extending from the side, and it is mounted on a wheeled base. The middle model has a monitor on top and is mounted on a wheeled base.
Premarket Notification 510(k) Summary
Image /page/1/Picture/21 description: The image shows a black and white picture of an SSI-5000 ultrasound machine. The machine has a monitor on top, a keyboard, and a control panel. The machine is on wheels, and there are cables hanging from the side. The image is a close-up of the machine, and the background is blurred.
{3}------------------------------------------------
Intended Use: The device is a general-purpose ultrasonic imaging instrument intended for use by a qualified physician for evaluation of Abdomen; Pediatric; Small Organ (breast, tests, thyroid); heart soft tissue; Peripheral Vascular, Musculo-skeletal (conventional) and Urology.
## Technological Characteristics
| Display Modes | Single and dual 2-D; Display of Duplex 2-D/M-mode; 2-D/Pulsed Doppler and<br>Triplex 2-D/CD/Pulsed Doppler image formats; Dual B and Color in real time |
|---------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Measurements | Distance; area; circumference; calipers; velocity, PI, RI. Cardiac. OB and Vascular<br>package |
| Principle of<br>Operation | Applying high voltage burst to the Piezoelectric material in the transducer and<br>detect the reflected echo to construct the 2-D B-mode, Doppler color, and Doppler<br>spectrum image for diagnostic purpose. |
| Operating<br>Controls | TGC 8 slider, +/- 24dB Depth Range: 3 to 24 cm Image sector size: 32 lines to full B (256 lines) Image Sector position: Steering within full maximum B orientation flip: L/R key with marking on the screen B Dynamic range control: preset 5 curves over 50-90 dB Gray Scale Control: 8 Settings Focal Number: 16 focal zone setting B persistence: 30-90% recursive Image Processing: Smoothing, edge enhancement PW sweeping speed 2,4,8 sec over display. PW Wall filter setting: 16 settings, 0.25 to 20% of PRF PW sample volume: 0.5 to 10mm with 0.5mm step size. PW/B update: with UPDATE key PW cursor steering: Steer soft key PW angle correction: 0 to 70 degree user control PW trace: Peak, Mean PW spectrum dynamic range: 5 preset curve over 15-48 dB Spectrum baseline shift and invert Color ROI setting: trackball and set key to control size and position Color steering on flat probe: +, 0, - Color Wall Filter: Color wall filter with 16 selection, 0.25-20% of PRF Color & B priority: C-B priority soft menu Color Packet size: preset per Exam range from 8 to 12 Color spatial filter: preset per Exam, horizontal, vertical, off Zoom factor: 1 to 10 continuously Freeze control: Toggling freeze key Cine control: step, play backward, play continuously |
| Acoustic<br>Output | Track 3; MI, TIS, TIC, TIB<br>Derated Ispta: 720mW/cm2 maximum, TIS/TIB/TIC:0.1-4.0 Range, Mechanical<br>Index: 1.9 Maximum, or Derated Isppa: 190 W/cm² max |
| Safety<br>Compliance | IEC601-1 International Electrotechnical Comission; Medical Electrical Equipment<br>IEC60601-2 International Electrotechnical Comission; Electromagnetic<br>Compatibility |
{4}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three swooping lines representing its wings. The eagle is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA", which is arranged in a circular fashion around the left side of the eagle.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OCT 2 0 2004
SonoScape Company, Ltd. % Mr. Bob Leiker Quality & Regulatory Services 7263 Cronin Circle DUBLIN CA 94568-2330
Re: K042369
י
;
Trade Name: SonoScape SSI-1000/SSI-5000 Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic pulsed echo imaging system Regulation Number: 21 CFR 892.1570 Regulation Name: Diagnostic ultrasonic transducer Regulatory Class: II Product Code: 90 IYN, IYO, and ITX Dated: October 11, 2004 Received: October 12, 2004
Dear Mr. Leiker:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use with the SonoScape SSI-1000/SSI-5000 Diagnostic Ultrasound System, as described in your premarket notification:
Transducer Model Number
2P1, PA2.5 MHz Phased Array L741, LA7.5 MHz Linear Array C344, CLA3.5 MHz Curved Linear Array
{5}------------------------------------------------
Page 2 - Mr. Leiker - - - -
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), W it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act, or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This determination of substantial equivalence is granted on the condition that prior to shipping the first device, you submit a postclearance special report. This report should contain complete information, including acoustic output measurements based on production line devices, requested in Appendix G, (enclosed) of the Center's September 30, 1997 "Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers." If the special report is incomplete or contains unacceptable values (e.g., acoustic output greater than approved levels), then the 510(k) clearance may not apply to the production units which as a result may be considered adulterated or misbranded.
The special report should reference the manufacturer's 510(k) number. It should be clearly and . prominently marked "ADD-TO-FILE" and should be submitted in duplicate to:
> Food and Drug Administration Center for Devices and Radiological Health Document Mail Center (HFZ-401) 9200 Corporate Boulevard Rockville, Maryland 20850
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to · proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801, please contact the Office of Compliance at (301) 594-4591. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597 or at its Internet address {"http://www.fda.gov/cdrh/dsmamain.html".
{6}------------------------------------------------
Page 3 - Mr. Leiker
Ìf you have any questions regarding the content of this letter, please contact Rodrigo C. Perez at (301) 594-1212. .
:
· Sincerely yours,
Nancy C Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices : Office of Device Evaluation Center for Devices and Radiological Health .
.
.
.
・
・
-
:
.
.
.
.
.
.
:
:
:
,
.
. . . . .
: .
,
·
:
:
.
: : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : :
:
: :
.
.
Enclosure(s)
:
:
1
:
.
. . . .
・・・・・・・・・・
:
1. . . . . . . . . . .
:
and the production of the comments of
. . . . . . . . .
. . . . . . . . . . . . .
:
.
,一个时间:
.
.
、一个
.
: .
:
.
.
,
{7}------------------------------------------------
510(k) Number:________________________________________________________________________________________________________________________________________________________________
Device Name: SonoScape SSI-1000/SSI-5000
Diagnostic Ultrasound Pulsed Echo System Diagnostic Ultrasound Pulsed Doppler Imaging System
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: Mode of Oneration
| Mode of Operation | | | | | | | | | | |
|------------------------------------------|---|---|---|-----|-----|---------------|---------------------------|------------------------|----------|-------------------------|
| Clinical Application | A | B | M | PWD | CWD | Color Doppler | Power (Amplitude) Doppler | Color Velocity Imaging | Combined | Tissue Harmonic Imaging |
| Opthalmic | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | N | N | N | | N | N | | Note 1 | N |
| Ob/GYN | | N | N | N | | N | N | | Note 1 | N |
| IntraOperative<br>Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ (breast,<br>thyroid, testes) | | N | N | N | | N | N | | Note 1 | N |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | N | N | N | N | N | N | | Note 1 | N |
| Transesophageal | | | | | | | | | | |
| Trans-Rectal | | | | | | | | | | |
| Trans-Vaginal | | | | | | | | | | |
| Trans-Urethral | | | | | | | | | | |
| Intra-Vascular | | | | | | | | | | |
| Peripheral Vascular | | N | N | N | | N | N | | Note 1 | N |
| Laparascopic | | | | | | | | | | |
| Muscular-Skeletal<br>Conventional | | N | N | N | | N | N | | Note 1 | N |
| Muscular-Skeletal<br>Superficial | | | | | | | | | | |
| Others (Specify) | | | | | | | | | | |
P = previously cleared by FDA E = added under Appendix E N = new indication
Note 1: Combined includes: B/M; B/PWD; B/Color Doppler; B/Power Doppler; B/Color Doppler/PWD and B/Power Doppler/PWD
Additional Comments:
Concurrence of CDRH, Office of Device
Nancy C. Brogdon
(Division Sign-Off)
Division of Reproductive. Abdominal,
and Radiological Devices
Page 2 of 5 Uno
Prescription Use (Per 21 CFR 801.109)
Indications For Use 510(k) Number ____________________________________________________________________________________________________________________________________________
{8}------------------------------------------------
510(k) Number:________________________________________________________________________________________________________________________________________________________________
Device Name: 2P1, PA2.5 MHz Phased Array Diagnostic Ultrasound Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Mode of Operation | | | | | | | | | | |
|------------------------------------------|---|---|---|-----|-----|------------------|---------------------------------|------------------------------|----------|-------------------------------|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Color<br>Velocity<br>Imaging | Combined | Tissue<br>Harmonic<br>Imaging |
| Opthalmic | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Ob/GYN | | | | | | | | | | |
| IntraOperative | | | | | | | | | | |
| Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ (breast,<br>thyroid, testes) | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | N | N | N | N | N | N | | Note 1 | N |
| Transesophageal | | | | | | | | | | |
| Trans-Rectal | | | | | | | | | | |
| Trans-Vaginal | | | | | | | | | | |
| Trans-Urethral | | | | | | | | | | |
| Intra-Vascular | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | |
| Laparascopic | | | | | | | | | | |
| Muscular-Skeletal<br>Conventional | | | | | | | | | | |
| Muscular-Skeletal<br>Superficial | | | | | | | | | | |
| Others (Specify) | | | | | | | | | | |
P = previously cleared by FDA E = added under Appendix E N = new indication
Note 1: Combined includes: B/M; B/PWD; B/Color Doppler; B/Power Doppler; B/Color Doppler/PWD and B/Power Doppler/PWD
Additional Comments:
Concurrence of CDRH, Office of D
Nancy C. Bergdon
(Division S Division of Reproductive and Radiological Device 510(k) Number _
Prescription Use (Per 21 CFR 801.109)
Indications For Use
{9}------------------------------------------------
.
510(k) Number:________________________________________________________________________________________________________________________________________________________________
Device Name: L741, LA7.5 MHz Linear Array Diagnostic Ultrasound Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: Mode of Oneration
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Color<br>Velocity<br>Imaging | Combined | Tissue<br>Harmonic<br>Imaging |
|------------------------------------------|---|---|---|-----|-----|------------------|---------------------------------|------------------------------|----------|-------------------------------|
| Opthalmic | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Ob/GYN | | | | | | | | | | |
| IntraOperative<br>Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ (breast,<br>thyroid, testes) | N | N | N | | N | N | N | | Note 1 | N |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Trans-Rectal | | | | | | | | | | |
| Trans-Vaginal | | | | | | | | | | |
| Trans-Urethral | | | | | | | | | | |
| Intra-Vascular | | | | | | | | | | |
| Peripheral Vascular | N | N | N | | N | N | N | | Note 1 | N |
| Laparascopic | | | | | | | | | | |
| Muscular-Skeletal<br>Conventional | N | N | N | | N | N | N | | Note 1 | N |
| Muscular-Skeletal<br>Superficial | | | | | | | | | | |
| Others (Specify) | | | | | | | | | | |
P = previously cleared by FDA E = added under Appendix E N = new indication
Note 1: Combined includes: B/M; B/PWD; B/Color Doppler; B/Power Doppler; B/Color Doppler/PWD and B/Power Doppler/PWD
Additional Comments:
Concurrence of CDRH, Office of Device Evaluation (ODE
Nancy C. hogdon
(Division Sign-Off)
Prescription Use (Per 21 CFR 801.109)
Division of Reproductive, Abdominal, and Radiological Dovices
Indications For Use I - Indications For Use
{10}------------------------------------------------
510(k) Number:________________________________________________________________________________________________________________________________________________________________
Device Name: C344, CLA3.5 MHz Curved Linear Array Diagnostic Ultrasound Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Mode of Operation | | | | | | | | | | |
|------------------------------------------|---|---|---|-----|-----|------------------|---------------------------------|------------------------------|----------|-------------------------------|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Color<br>Velocity<br>Imaging | Combined | Tissue<br>Harmonic<br>Imaging |
| Opthalmic | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | N | N | N | | N | N | | Note 1 | N |
| Ob/GYN | | N | N | N | | N | N | | Note 1 | N |
| IntraOperative | | | | | | | | | | |
| Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ (breast,<br>thyroid, testes) | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Trans-Rectal | | | | | | | | | | |
| Trans-Vaginal | | | | | | | | | | |
| Trans-Urethral | | | | | | | | | | |
| Intra-Vascular | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | |
| Laparascopic | | | | | | | | | | |
| Muscular-Skeletal<br>Conventional | | | | | | | | | | |
| Muscular-Skeletal<br>Superficial | | | | | | | | | | |
| Others (Specify) | | | | | | | | | | |
P = previously cleared by FDA E = added under Appendix E N = new indication
Note 1: Combined includes: B/M; B/PWD; B/Color Doppler; B/Power Doppler; B/Color Doppler/PWD and B/Power Doppler/PWD
Additional Comments:
Concurrence of CDRH, Office of Device Evaluation (OD
Nancy C brondon
(Division Sign-Off Division of Reproductive, Abdo and Radiological Devices 510/k) Number _________
Prescription Use (Per 21 CFR 801.109)
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.