MODIFICATION TO DIMENSION NT-PROBNP FLEX REAGENT CARTRIDGE METHOD

K042347 · Dade Behring, Inc. · NBC · Nov 19, 2004 · Clinical Chemistry

Device Facts

Record IDK042347
Device NameMODIFICATION TO DIMENSION NT-PROBNP FLEX REAGENT CARTRIDGE METHOD
ApplicantDade Behring, Inc.
Product CodeNBC · Clinical Chemistry
Decision DateNov 19, 2004
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1117
Device ClassClass 2

Indications for Use

The PBNP assay used on the Dimension® clinical chemistry system with the heterogeneous immunoassay module is an in vitro diagnostic assay for the quantitative determination of N-terminal pro-brain natriuretic peptide (NT-proBNP) in human plasma. In individuals suspected of having congestive heart failure (CHF), measurements of NT-proBNP are used as an aid in the diagnosis and assessment of severity. The test is further indicated for the risk stratification of patients with acute coronary syndrome and heart failure.

Device Story

In vitro diagnostic assay; quantitative determination of NT-proBNP in human plasma; used on Dimension clinical chemistry systems. Principle: one-step sandwich enzyme immunoassay; sample incubated with chromium dioxide particles coated with polyclonal antibodies; conjugate reagent (alkaline phosphatase-labeled polyclonal antibody) added; magnetic separation/washing removes unbound conjugate; sandwich-bound ALP triggers amplification cascade; produces colored product absorbing at 510 nm; color intensity proportional to NT-proBNP concentration. Used in clinical laboratories; operated by trained personnel. Output: numerical concentration value (pg/mL). Clinical utility: aids diagnosis of CHF; assesses disease severity (NYHA classes); risk stratifies acute coronary syndrome and heart failure patients. Benefits: provides objective biomarker data to support clinical decision-making regarding heart failure management.

Clinical Evidence

Clinical study evaluated 308 individuals without CHF and 227 patients with CHF. Primary endpoints: clinical sensitivity and specificity using age-stratified cut-offs (125 pg/mL for <75 yrs; 450 pg/mL for ≥75 yrs). Results: AUC 0.921 (95% CI 0.898-0.943). Sensitivity ranged 77-92%; specificity 77-97% across age/gender subgroups. Data demonstrated relationship between NT-proBNP levels and NYHA functional classes (I-IV), supporting use for severity assessment.

Technological Characteristics

One-step sandwich enzyme immunoassay; chromium dioxide particles; alkaline phosphatase (ALP) conjugate; synthetic flavin adenine dinucleotide phosphate (FADP) amplification; 510 nm absorbance measurement. Compatible with Dade Behring Dimension RxL Max, RxL, and Xpand systems. Sample types: lithium heparin, sodium heparin, EDTA plasma. Standards: NCCLS EP 5-A.

Indications for Use

Indicated for quantitative determination of NT-proBNP in human plasma to aid in diagnosis of suspected congestive heart failure and for risk stratification and severity assessment in patients with acute coronary syndrome and heart failure.

Regulatory Classification

Identification

The B-type natriuretic peptide (BNP) test system is an in vitro diagnostic device intended to measure BNP in whole blood and plasma. Measurements of BNP are used as an aid in the diagnosis of patients with congestive heart failure.

Special Controls

*Classification.* Class II (special controls). The special control is “Class II Special Control Guidance Document for B-Type Natriuretic Peptide Premarket Notifications; Final Guidance for Industry and FDA Reviewers.”

Predicate Devices

Submission Summary (Full Text)

{0} 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY DEVICE ONLY TEMPLATE A. 510(k) Number: k042347 B. Purpose for Submission: Modification to previously cleared device (k041417). Indications for Use modified to add that the test is for “risk stratification and severity assessment of patients with acute coronary syndrome and heart failure.” C. Analyte: B-type natriuretic peptide test system (BNP) D. Type of Test: Quantitative E. Applicant Dade Behring, Inc. F. Proprietary and Established Names: Dimension® NT-proBNP (PBNP) Flex® reagent cartridge method G. Regulatory Information: 1. Regulation section: 21 CFR 862.1117, B-type natriuretic peptide test system 2. Classification: Class II 3. Product Code: NBC 4. Panel: 75 H. Intended Use: 1. Intended use(s): The PBNP assay used on the Dimension® clinical chemistry system with the heterogeneous immunoassay module is an in vitro diagnostic assay for the quantitative determination of N-terminal pro-brain natriuretic peptide (NT-proBNP) in human plasma. In individuals suspected of having congestive heart failure (CHF), measurements of NT-proBNP are used as an aid in the diagnosis and assessment of severity. The test is further indicated for the risk stratification of patients with acute coronary syndrome and heart failure. 2. Indication(s) for use: The Dimension® PBNP Flex® method is an in vitro diagnostic assay for the quantitative determination of N-terminal pro-brain natriuretic peptide (NT-proBNP) in human plasma. Measurements of NT-proBNP are used as an aid in the diagnosis of individuals suspected of having congestive heart failure and for risk stratification and severity assessment of patients with acute coronary syndrome and heart failure. 3. Special condition for use statement(s): Prescription use {1} Page 2 of 10 4. Special Instrument Requirements: Dade Behring Dimension RxL Max™, RxL, and Xpand® I. Device Description: The Dade Behring Dimension® PBNP Flex® reagent cartridge method is an in vitro diagnostic test that consists of prepackaged reagents in a flexible plastic cartridge for use only on the Dimension® clinical chemistry system. J. Substantial Equivalence Information: 1. Predicate device name(s): Roche Diagnostics Elecsys® proBNP Immunoassay 2. Predicate K number(s): k022516 3. Comparison with predicate: | Similarities | | | | --- | --- | --- | | Item | Dimension NT-proBNP | Roche NT-proBNP | | Assay type | Immunoassay | Immunoassay | | Antibody | Polyclonal sheep antibody | Polyclonal sheep antibody | | Cut-off | 125 pg/mL for patients <75 years | 125 pg/mL for patients <75 years | | | 450 pg/mL for patients ≥ 75 years | 450 pg/mL for patients ≥ 75 years | | Reference | Roche purified synthetic NT-proBNP | Roche purified synthetic NT-proBNP | | Differences | | | | Item | Dimension NT-proBNP | Roche NT-proBNP | | Indications for Use | For the in vitro quantitative determination of N-terminal pro-brain natriuretic peptide in human plasma as an aid in the diagnosis and assessment of severity of individuals suspected of having congestive heart failure. The test is further indicated for the risk stratification of patients with acute coronary syndrome and heart failure. | in vitro quantitative determination of NT-proBNP in human serum and plasma as an aid in the diagnosis of individuals suspected of having congestive heart failure. The test is further indicated for the risk stratification of patients with acute coronary syndrome and congestive heart failure. | | Reportable range | 10-30,000 pg/mL | 5-35,000 pg/mL | | Analytical sensitivity | ≤ 10 pg/mL | 5 pg/mL | | Functional sensitivity | ≤ 30 pg/mL | < 50 pg/mL | | Sample volume | 50 μL | 20 μL | {2} Page 3 of 10 K. Standard/Guidance Document Referenced (if applicable): NCCLS EP 5-A, and Class II Special Controls Guidance Document for B-Type Natriuretic Peptide Premarket Notifications: Final Guidance for Industry and FDA Reviewers (11/30/2000). L. Test Principle: The PBNP method is a one-step enzyme immunoassay based on the “sandwich” principle. Sample is incubated with chromium dioxide particles coated with polyclonal antibodies which recognize an epitope located in the N-terminal part of proBNP, and a conjugate reagent [alkaline phosphatase (ALP)] labeled polyclonal antibody specific for a second independent epitope on NT-proBNP, to form a particle/NT-proBNP/ conjugate sandwich. Unbound conjugate is removed by magnetic separation and washing. After separation and washing, the conjugate sandwich is transferred to the cuvette where the sandwich-bound ALP triggers an amplification cascade. ALP dephosphorylates synthetic flavin adenine dinucleotide phosphate (FADP) to produce FAD. FAD binds to apo D-amino acid oxidase and converts it to active holo D-amino acid oxidase. Each molecule of holo D-amino acid oxidase produces multiple molecules of hydrogen peroxide (H₂O₂). H₂O₂ in the presence of horseradish peroxidase (HRP), converts 3,5-dichloro-2-hydroxybenzenesulfonic acid (DCHBS) and 4-aminoantipyrine(4-AAP) to a colored product that absorbs at 510 nm. The color change measured is directly proportional to the concentration of NT-proBNP present in the patient sample. M. Performance Characteristics (if/when applicable): 1. Analytical performance: a. Precision/Reproducibility: | Sample | Mean (pg/mL) | Within-Run Precision | | Total Precision | | | --- | --- | --- | --- | --- | --- | | | | SD (pg/mL) | % CV | SD (pg/mL) | % CV | | Human plasma pool 1 | 159 | 3.4 | 2.2 | 9.1 | 5.7 | | Internal QC Pool 1 | 449.5 | 8.1 | 1.8 | 16.6 | 3.7 | | Internal QC Pool 2 | 956.7 | 15.3 | 1.6 | 34.9 | 3.6 | | Control Level 1 | 175.5 | 2.0 | 1.1 | 6.8 | 3.8 | | Control Level 2 | 3733.8 | 71.8 | 1.9 | 115.9 | 3.1 | Precision testing was done in accordance with NCCLS EP5-A. Specimens at each level were analyzed in duplicate once per day for 20 days. b. Linearity/assay reportable range: The reportable range of the assay is from 10-30,000 pg/mL. A high PBNP plasma pool (PBNP = 35979.5 pg/mL) was diluted with a low PBNP pool (PBNP = 6.8 pg/mL) to produce 6 levels of PBNP. High range linearity was evaluated by comparing observed vs. expected values obtained with the PBNP method. A linear regression analysis was then performed on the data to yield the following: slope = 0.954, r=0.999, intercept = 512 pg/mL. Lower range linearity was evaluated by diluting two patient samples (Sample 1 PBNP = 4616 pg/mL, Sample 2 PBNP = 1037 pg/mL) to produce 6 levels of PBNP {3} Page 4 of 10 for each sample. Linear regression analyses were performed on the data to yield the following: Sample 1: slope = 0.990, r = 1.000, intercept = -8.50 Sample 2: slope = 1.030, r = 0.999, intercept = -1.63 Hook effect was evaluated using samples containing NT-proBNP concentrations ranging from 0 to 300,000 pg/mL. Data indicated no hook effect up to 300,000 pg/mL. c. Traceability (controls, calibrators, or method): The assay is referenced to Roche purified synthetic NT-proBNP. The assigned values for the Dimension® PBNP Calibrator are referenced to a master pool containing synthetic human N-terminal pro-brain natriuretic peptide. d. Detection limit: The analytical sensitivity for the PBNP assay is ≤ 10 pg/mL. This is defined as the concentration at two standard deviations (n = 20) of a sample devoid of NT-proBNP. Functional sensitivity was determined by performing a 20 day ANOVA experiment using samples prepared with Roche synthetic antigen and the Dimension® PBNP calibrator base (zero concentration level) targeted at low NT-proBNP concentrations. Total % CV was plotted versus the concentration for each sample. The functional sensitivity was determined to be ≤ 30 pg/mL. e. Analytical specificity: No significant interference was found for bilirubin (conjugated) up to 60 mg/dL, bilirubin (unconjugated) up to 20 mg/dL, hemoglobin up to 1000 mg/dL, triglycerides up to 3000 mg/dL, and rheumatoid factor up to 500 IU/mL. The pharmaceutical Natrecor® shows no significant cross reactivity at 0 and 125 pg/mL NT-proBNP. An extensive list of other compounds was evaluated for interference and was found to have no significant interference or cross reactivity. A list of these compounds is present in the PBNP labeling. f. Assay cut-off: See k041417 - The recommended medical decision thresholds by age group are: Patients &lt; 75 years 125 pg/mL Patients ≥ 75 years 450 pg/mL 2. Comparison studies: a. Method comparison with predicate device: Comparison using split patient heparinized plasma samples between the Dade Behring Dimension® PBNP Flex method and the predicate Roche Elecsys® proBNP method demonstrated the following method comparison using samples with values ranging from 16.1 – 29,893.1 pg/mL: | Comparative Method | Slope | Intercept (pg/mL) | Correlation Coefficient | n | | --- | --- | --- | --- | --- | | Roche Elecsys® proBNP | 0.90 | -15.4 | 0.985 | 352 | {4} Page 5 of 10 b. Matrix comparison: Plasma specimens (lithium heparin, sodium heparin, and EDTA) may be used for the PBNP assay. Serum samples should not be used with the PBNP assay. Lithium heparin samples (n = 55) ranging from 13 to 29, 221 pg/mL when compared to sodium heparin and EDTA samples gave slopes of 0.95 and 0.96, correlation coefficients of 0.998 and 0.998, and intercepts of 0.9 and 10.9 respectively using Passing-Bablok regression statistics. 3. Clinical studies: a. Clinical sensitivity: Clinical Studies: For the Reference Study Group, NT-proBNP concentrations were determined in 308 individuals without congestive heart failure (163 women and 145 men). This population included apparently healthy individuals and individuals with diabetes, hypertension, and pulmonary disease. For the Disease Study Group, NT-proBNP concentrations were determined in 227 patients diagnosed with congestive heart failure (CHF). This population included 69 women and 158 men. The tables below show the clinical sensitivity and specificity of the Dimension® PBNP assay using a cutoff of 125 pg/mL for patients younger than 75 years and 450 pg/mL for patients 75 years or older. | Males | | | | --- | --- | --- | | | <75 yrs | ≥75 yrs | | % Sensitivity | 84 % | 92% | | 95% Confidence Interval | 77 – 91 | 84 – 99 | | % Specificity | 95% | 77% | | 95% Confidence Interval | 90 - 99 | 67 – 88 | | Females | | | | --- | --- | --- | | | <75 yrs | ≥75 yrs | | % Sensitivity | 77% | 91% | | 95% Confidence Interval | 64 - 89 | 79 – 100 | | % Specificity | 97% | 88% | | 95% Confidence Interval | 89 - 98 | 80 - 96 | b. Clinical specificity: See Clinical Sensitivity above c. Other clinical supportive data (when a and b are not applicable): 4. Clinical cut-off: The Receiver Operator Curve (ROC) presents the clinical sensitivity and specificity at various cutoffs for the 227 patients diagnosed with CHF and 308 patients without CHF. The ROC curve for the Dimension® PBNP assay is {5} shown below. The area under ROC curve (AUC) for the Dimension® PBNP assay is 0.921 with a 95% confidence interval of 0.898 to 0.943. An age-matched ROC analysis of the clinical data was performed via the weighted method described in Kondratovich, M (2002), Matched Receiver Operating Characteristic (ROC) analysis and propensity scores, Proceedings of the 2002 Joint Statistical Meeting, Biopharmaceutical Section, New York, NY. The resulting AUC is 0.925 with a 95% confidence interval of 0.903 to 0.946. ![img-0.jpeg](img-0.jpeg) A box and whiskers plot of the clinical study population is presented below. Recommended clinical thresholds are 125 pg/mL for patients younger than 75 years and 450 pg/mL for patients 75 years and older. Three disease group samples with values above the assay range are not displayed in the plot. {6} Page 7 of 10 ![img-1.jpeg](img-1.jpeg) 5. Expected values/Reference range: NT-proBNP concentrations in the Reference Group are shown in the following tables. The recommended medical decision thresholds, by age group, are: Patients &lt; 75 years: 125 pg/mL [14.8 pmol/L] Patients ≥75 years: 450 pg/mL [53.2 pmol/L] Reference Study Group NT-proBNP concentrations were determined in 308 individuals without congestive heart failure (163 women and 145 men). This population included apparently health individuals and individuals with diabetes, hypertension, and pulmonary disease. The statistics for NT-proBNP concentrations in the reference study group are shown in the following table. Reference Study Group All | | <55 yrs | 55 - 64 yrs | 65 - 74 yrs | ≥ 75 yrs | | --- | --- | --- | --- | --- | | Mean | 32.1 | 39.1 | 83.1 | 353.6 | | SD | 38.2 | 46.2 | 54.6 | 775.8 | | Median | 17.9 | 18.8 | 59.5 | 125.6 | | 95th Percentile | 81.9 | 163.1 | 202.6 | 1372.8 | | % < 125 pg/mL | 97% | 93% | 88% | - | | % < 450 pg/mL | - | - | - | 83% | | N | 163 | 15 | 8 | 122 | {7} Page 8 of 10 ## Males | | <55 yrs | 55 - 64 yrs | 65 - 74 yrs | ≥ 75 yrs | | --- | --- | --- | --- | --- | | Mean | 29.7 | 37.9 | 76.0 | 414.3 | | SD | 42.5 | 41.6 | - | 889.0 | | Median | 13.4 | 13.4 | 76.0 | 111.8 | | 95^{th} Percentile | 134.4 | 114.3 | 76.0 | 1475.7 | | % < 125 pg/mL | 95% | 100% | 100% | - | | % < 450 pg/mL | - | - | - | 77% | | N | 76 | 6 | 1 | 62 | ## Females | | <55 yrs | 55 - 64 yrs | 65 - 74 yrs | ≥ 75 yrs | | --- | --- | --- | --- | --- | | Mean | 34.2 | 39.9 | 84.1 | 291.0 | | SD | 34.0 | 51.6 | 58.9 | 639.7 | | Median | 23.7 | 18.8 | 59.5 | 131.3 | | 95^{th} Percentile | 75.8 | 163.1 | 202.6 | 1080.7 | | % < 125 pg/mL | 99% | 89% | 86% | - | | % < 450 pg/mL | - | - | - | 88% | | N | 87 | 9 | 7 | 60 | ## Disease Study Group Blood samples were obtained from 227 patients diagnosed with congestive heart failure (CHF). The population included 69 women and 158 men. The descriptive statistics and New York Heart Association (NYHA) functional classes are provided below. ### CHF Population – All | | <55 yrs | 55 - 64 yrs | 65 - 74 yrs | ≥75 yrs | | --- | --- | --- | --- | --- | | Mean | 2400.1 | 3120.7 | 4167.4 | 4579.3 | | SD | 3855.9 | 8809.1 | 8550.7 | 8721.2 | | Median | 660.5 | 533.3 | 1506.7 | 2513.4 | | 95^{th} Percentile | 10961.2 | 16817.2 | 10889.3 | 11398.6 | | % > 125 pg/mL | 79% | 73% | 90% | - | | % > 450 pg/mL | - | - | - | 91% | | N | 48 | 49 | 61 | 69 | ### CHF Population – Males | | <55 yrs | 55 - 64 yrs | 65 - 74 yrs | ≥75 yrs | | --- | --- | --- | --- | --- | | Mean | 2748.2 | 3475.1 | 4427.8 | 5773.0 | | SD | 4179.3 | 10332.0 | 9669.3 | 10238.5 | | Median | 1148.2 | 821.7 | 1273.4 | 3086.6 | | 95^{th} Percentile | 12127.1 | 16817.2 | 10889.3 | 14118.8 | | % > 125 pg/mL | 77% | 84% | 89% | - | | % > 450 pg/mL | - | - | - | 92% | | N | 34 | 32 | 45 | 47 | {8} Page 9 of 10 ## CHF Population – Females | | <55 yrs | 55 – 64 yrs | 65 – 74 yrs | ≥75 yrs | | --- | --- | --- | --- | --- | | Mean | 1554.7 | 2453.7 | 3435.2 | 2028.9 | | SD | 2886.2 | 5024.5 | 4175.0 | 2581.3 | | Median | 257.8 | 492.9 | 1828.4 | 1077.1 | | 95^{th} Percentile | 10961.2 | 20855.8 | 13287.1 | 8891.2 | | % > 125 pg/mL | 86% | 53% | 94% | - | | % > 450 pg/mL | - | - | - | 91% | | N | 14 | 17 | 16 | 22 | ## CHF Population – All | NYHA Functional Class | | | | | | | --- | --- | --- | --- | --- | --- | | | All CHF | NYHA I | NYHA II | NYHA III | NYHA IV | | Mean | 3693.0 | 1492.9 | 2124.5 | 5423.2 | 7134.5 | | SD | 7911.8 | 2260.2 | 2946.1 | 10692.5 | 11900.1 | | Median | 1422.0 | 659.2 | 1077.1 | 2636.5 | 2754.5 | | 5^{th} Percentile | 59.3 | 47.9 | 29.4 | 117.4 | 76.4 | | 95^{th} Percentile | 11398.6 | 5748.8 | 8344.4 | 13287.1 | 20855.8 | | % > Cutoff | 85% | 76% | 83% | 90% | 90% | | Minimum | 17.6 | 21.7 | 17.6 | 60.3 | 50.6 | | Maximum | 70025.3 | 10613.6 | 16817.2 | 70025.3 | 63515.3 | | N | 227 | 53 | 72 | 71 | 31 | ## CHF Population – Males | NYHA Functional Class | | | | | | | --- | --- | --- | --- | --- | --- | | | All CHF | NYHA I | NYHA II | NYHA III | NYHA IV | | Mean | 4273.6 | 1706.9 | 2582.3 | 6273.3 | 9300.3 | | SD | 9109.3 | 2242.9 | 3238.0 | 12099.8 | 15619.8 | | Median | 1710.7 | 938.6 | 1273.4 | 3269.9 | 2717.1 | | 5^{th} Percentile | 60.3 | 57.8 | 33.1 | 122.1 | 76.4 | | 95^{th} Percentile | 13600.1 | 5748.8 | 8344.4 | 15686.3 | 63515.3 | | % > Cutoff | 86% | 81% | 83% | 93% | 88% | | Minimum | 21.7 | 21.7 | 27.3 | 60.3 | 76.4 | | Maximum | 70025.3 | 10613.6 | 16817.2 | 70025.3 | 63515.3 | | N | 158 | 41 | 48 | 53 | 16 | ## CHF Population – Females | NYHA Functional Class | | | | | | | --- | --- | --- | --- | --- | --- | | | All CHF | NYHA I | NYHA II | NYHA III | NYHA IV | | Mean | 2363.5 | 761.6 | 1208.8 | 2920.1 | 4824.3 | | SD | 3725.4 | 1301.7 | 2013.8 | 3749.6 | 5565.2 | {9} | Median | 933.5 | 232.6 | 550.2 | 1933.5 | 2754.5 | | --- | --- | --- | --- | --- | --- | | 5thPercentile | 50.6 | 47.9 | 24.4 | 103.2 | 50.6 | | 95thPercentile | 10375.0 | 4723.5 | 2524.6 | 13287.1 | 20855.8 | | % > Cutoff | 81% | 58% | 83% | 83% | 93% | | Minimum | 17.6 | 47.9 | 17.6 | 103.2 | 50.6 | | Maximum | 20855.8 | 4723.5 | 10029.4 | 13287.1 | 20855.8 | | N | 69 | 12 | 24 | 18 | 15 | These results show that there is a relationship between the severity of the clinical signs and symptoms of CHF and the median NT-proBNP concentration, demonstrating that the Dimension® PBNP Method can be used as an aid in the diagnosis of all degrees of CHF severity including asymptomatic patients. # N. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...