AMERICAN BANTEX, MODEL MS 3

K042104 · American Bantex Corp. · INI · Oct 21, 2004 · Physical Medicine

Device Facts

Record IDK042104
Device NameAMERICAN BANTEX, MODEL MS 3
ApplicantAmerican Bantex Corp.
Product CodeINI · Physical Medicine
Decision DateOct 21, 2004
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 890.3800
Device ClassClass 2

Indications for Use

The MS 3 Mini Scooter has been designed to help people who have a difficult time walking. The MS 3 Mini Scooter and upper-body mobility. People who have some mobility, but cannot walk for long distances or people that may need r copic who halk will generally purchase this scooter. The MS 3 Mini Scooter is designed for both indoor and limited outdoor use in clean and dry conditions.

Device Story

The American Bantex MS 3 Mini Scooter is a motorized 3-wheeled vehicle designed to assist individuals with limited walking ability. The device features a rear-wheel drive system with a single front wheel for steering. Users operate the scooter via a tiller-mounted variable speed dial and a VSI joystick controller. The body is constructed from rigid Acrylonitrile-Butadiene-Styrene (ABS) copolymer. The scooter is powered by a 24V sealed transaxle motor and includes a regenerative electric braking system supplemented by a secondary parking brake. The device is designed to break down into components for tool-less assembly and transport. It is intended for use in indoor and limited outdoor environments. By providing motorized mobility, the device enables users to navigate distances they would otherwise be unable to walk, thereby increasing independence and mobility.

Clinical Evidence

Bench testing only. Performance testing included speed tests (max speed), braking distance measurements, and acceleration/deceleration calculations. Testing confirmed the device met performance requirements with no degradation detected. All test instruments were calibrated and traceable to NIST.

Technological Characteristics

Materials: Acrylonitrile-Butadiene-Styrene (ABS) copolymer shell. Energy: 24V sealed transaxle motor; 110-120V AC integrated charger. Control: VSI joystick controller; tiller-mounted variable speed dial. Braking: Regenerative electric motor brake and secondary parking brake. Form factor: 3-wheeled, modular/disassemblable. Connectivity: None.

Indications for Use

Indicated for individuals with limited mobility or difficulty walking long distances who retain sufficient upper-body mobility to operate the device. Intended for indoor and limited outdoor use in clean, dry conditions.

Regulatory Classification

Identification

A motorized three-wheeled vehicle is a gasoline-fueled or battery-powered device intended for medical purposes that is used for outside transportation by disabled persons.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ K042104 # OCT 2 1 2004 # AMERICAN BANTEX CORPORATION 1815 Rollins Road Burlingame, California (650) 697 - 3545 (650) 697 - 3596 Tracy S. Best, Regulatory Affairs Consultant Preparation Date: July 30, 2004 ### Summary of Safety and Effectiveness for the: | Trade Name: | American Bantex MS 3, Mini Scooter | |-------------------------------|------------------------------------------------| | Common Name: | Scooter, 3-wheeled, Powered | | Classification Number & Name: | 89 INI (890.3800) Vehicle, Motorized 3-Wheeled | ## Legally Marketed Predicate Devices for Substantial Equivalence: *Mega Motion, Inc. - Mega 3 Scooter. K982145 Rationale for SE: The Mega Motion, Mega 3 is a Class II device that is intended to provide assistance and help people lead a more productive and mobile lifestyle. The 3-wheeled scooter provides a stable ride while the single forward wheel steers the scooter. Mega 3 is an outdoor and indoor product that is capable of traveling on bumpy or uneven terrain. The pneumatic tires and sealed transaxle contribute to the quiet operation of the scooter. The Mega 3 has a built-in charger that plugs into a standard 110-120V~ outlet. The dynamic regenerative brake (electric motor) and secondary parking brake are redundant to one another. ## Description of Submitted Device: The American Bantex MS 3 Mini Scooter is a Class II device that also is intended to provide assistance to individuals that are unable to walk long distances. The rear-wheel drive and single wheeled steering abilities allow the user to effortlessly control this device. The body (shell) is made of rigid Acrylonitrile-Butadiene-Styrene Copolymer. The shell is available to the consumer in two colors. The variable speed dial on the tiller column enables users single-handed operation while moving forward. The sealed (24 V) transaxle motor is quiet to operate. As with the predicate device, the MS 3 breaks-down into several components for easy, tool less assembly, disassembly and transportation. The MS 3 is substantially equivalent in safety, efficacy, technology, and intended use to the Mega 3, marketed by Mega Motions, Inc. American Bantex affirms that we contract with a fully operational quality manufacturing system, which conforms to the OSR requirements of 21 CFR Part 820, as well as the Quality clements of the Furopean Medical Device Directive, 93/42/FIEC for CE Marking. Design History Files are maintained for the development and distribution of products and they are tested using Independent Testing Services prior to marketing them. Summary of Safety and Effectiveness Page 1 of 1 00 {2}------------------------------------------------ ## ATTACHMENT - SUMMARY MATRIX | PROCEDURE: | For Running Speed Test: | |------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | The EUT was set up according to fully operation. The Speed Meter put on the<br>Equipment Under Test (EUT)'s wheel. Let EUT runs 50 meter to find a max<br>speed data. | | | For Break Test: | | | Let EUT running with fully speed for 10 meter and release the controller. After<br>the EUT stop, measure the distance between the stop point and release<br>controller point. | | | For Acceleration & Deceleration: | | | Use A=V/T equation to find the data. A=acceleration rate; V=meter; T=Second2 | | RESULTS: | The EUT meets the requirements, There was no performance degradation<br>detected during this test. | | CHANGES OR<br>MODIFICATIONS: | There were no modifications performed by test personnel. | | Test Equipment | Manufacturer | Model No. | Last Cal. | Cal. Due | | | | |--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|-------------|------------|------------|--|--|--| | Speed Meter | SMART | Q809-09(18) | 08/16/2003 | 08/16/2004 | | | | | Note: All testing were performed using internationally recognized standards. All test instruments were<br>calibrated and traosable to the National Institute of Standards and Technology (NIST). | | | | | | | | All parameters relating to wheelchair motion are controlled by the VSI (joystick) controller which was tested with the device as submitted as part of the 510(k). : {3}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is a stylized image of an eagle with three lines representing its wings. #### Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 # OCT 2 1 2004 American Bantex Corporation C/o Mr. Tracy S. Best Regulatory Affairs Consultant 994 North Main Street Bountiful, Utah 84010 Re: K042104 Trade/Device Name: MS 3 Mini Scooter Regulation Number: 21 CFR 890.3800 Regulation Name: Motorized three-wheeled vehicle Regulatory Class: II Product Code: INI Dated: October 4, 2004 Received: October 7, 2004 Dear Mr. Best: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The recaral controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can he found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {4}------------------------------------------------ This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Mark N Millman Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # Indications for Use 510(k) Number (if known): Device Name: MS 3 Mini Scooter ## Indications for Use: The MS 3 Mini Scooter has been designed to help people who have a difficult time walking. The THE MO 3 Mini Socctor and upper-body mobility. People who have some mobility, but cannot walk for long distances or people that may need r copic who halk will generally purchase this scooter. The MS 3 Mini Scooter is designed for both indoor and limited outdoor use in clean and dry conditions. Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use _ (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of D valuation (ODE) Mark N. Mcllhenser Restorative. and Neurological De **510(k) Number** K042104
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...