TONOPACH ULTRASONIC TONOMETER/PACHYMETER MODEL P-201 (THE TONOPACH)
K042099 · Retinapharma Technologies, Inc. · HKX · Dec 10, 2004 · Ophthalmic
Device Facts
Record ID
K042099
Device Name
TONOPACH ULTRASONIC TONOMETER/PACHYMETER MODEL P-201 (THE TONOPACH)
Applicant
Retinapharma Technologies, Inc.
Product Code
HKX · Ophthalmic
Decision Date
Dec 10, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.1930
Device Class
Class 2
Indications for Use
The TonoPach Ultrasonic Tonometer/Pachymeter Model P-201 is a diagnostic instrument that is intended for use in the measurement of both intraocular pressure and corneal thickness. It is intended for use during or following surgery and as a screening, monitoring or diagnostic aid in patients with normal, disease-induced or surgically altered anatomy, or in glaucoma suspects under current therapeutic guidelines.
Device Story
TonoPach P-201 is a portable, battery-operated, handheld ophthalmic instrument measuring intraocular pressure (IOP) and corneal thickness (CT) simultaneously at the same corneal locus. IOP measurement uses applanation pneumatometry (micro strain gauge); CT measurement uses reflection pulsed ultrasound. System includes electronics, transducer handpiece, protective membranes, and optional foot pedal. Used in office, bedside, or remote locations by clinicians. Output displayed on LCD; provides statistical storage. Enables simultaneous assessment of IOP and CT, aiding glaucoma management and surgical monitoring; benefits patients by providing integrated diagnostic data in a single procedure.
Clinical Evidence
Clinical trial evaluated 347 eyes of clinic volunteers. Tonometry performance compared against Goldmann tonometry. Analysis included scatter plots, Bland & Altman plots, mean, standard deviation, and variance. Results met or exceeded specifications. Additional bench testing performed on water-filled spherical membranes (tonometry) and polycarbonate wafers (pachymetry). Animal studies (rabbit and porcine eyes) verified accuracy and reproducibility.
Technological Characteristics
Handheld, battery-operated ophthalmic instrument. Sensing: micro strain gauge for applanation tonometry; reflection pulsed ultrasound for pachymetry. Measurement range: 5-80 mmHg (IOP), 125-1300 um (CT). Connectivity: LCD display, print output. System includes transducer handpiece and electronics. Sterilization: protective membranes used.
Indications for Use
Indicated for measurement of intraocular pressure and corneal thickness in patients with normal, disease-induced, or surgically altered ocular anatomy, or glaucoma suspects. Intended for use during or following surgery, screening, monitoring, or diagnostic aid.
Regulatory Classification
Identification
A tonometer and accessories is a manual device intended to measure intraocular pressure by applying a known force on the globe of the eye and measuring the amount of indentation produced (Schiotz type) or to measure intraocular tension by applanation (applying a small flat disk to the cornea). Accessories for the device may include a tonometer calibrator or a tonograph recording system. The device is intended for use in the diagnosis of glaucoma.
{0}------------------------------------------------
K042099
### 510(k) SUMMARY
| 1 | Submitter (Contact) Information | |
|---|---------------------------------|-----------------------------------------------------------|
| A | Company Name/Address: | RetinaPharma Technologies, Inc. |
| | | 944 Morgan Road |
| | | Jenkintown, PA 19046 |
| B | Company Phone: | (215) 885-4558 |
| C | Company Fax: | (215) 885-2788 |
| D | Contact Person: | Terry A. Fuller, Ph.D. |
| | | President |
| E | Manufacturing Location: | The Company has not established a manufacturing facility. |
| | | It will notify the Agency upon its establishment. |
- 2 Device Information
A Device Trade Name: TonoPach™ Ultrasonic Tonometer/Pachymeter Model P-201 (the "TonoPach™")
# B Device Classification:
The TonoPach is a combination of two devices, a pneumatometer and a pachymeter, which have already been determined to be substantially equivalent to legally marketed predicate devices with regard to safety, effectiveness and intended use. The device classifications are presented in the following table.
| | Pneumatometer | Pachymeter |
|------------------------------------------------|-------------------------------------------------|------------------------------------------------------------------|
| Common Name: | Tonometer for Measuring<br>Intraocular Pressure | Ultrasonic Pachymeter<br>for Measurement of<br>Corneal Thickness |
| Classification Name: | Tonometer, AC Powered | Ultrasonic Pulsed Echo<br>Imaging System |
| Product Code: | HKX | 90-IYO |
| Regulatory Class & Tier: | II | II |
| Regulatory Number: | 21 CFR 866.1930 | 21 CFR 892.1560 |
| Classification & Review<br>Advisory Committees | Ophthalmic | Radiology |
#### 3 Substantial Equivalence
The RetinaPharma TonoPach Ultrasonic Tonometer/Pachymeter is judged to be substantially equivalent in safety, effectiveness and intended use to the following legally marketed devices:
{1}------------------------------------------------
| Predicate Device | Manufacturer | 510(k) | Date Cleared |
|---------------------------------|-----------------------|----------|-----------------|
| TonoPen ophthalmic<br>tonometer | Oculab, Inc. | K 852774 | August 20, 1985 |
| Pocket Ultrasonic<br>Pachymeter | Quantel Medical, Inc. | K993674 | July 13, 2000 |
| Echoscan Model US-1800 | Nidek, Inc. | K020876 | May 3, 2002 |
## I ist of predicate devices
#### Device Description. ব
The TonoPach is a portable, battery operated, handheld ophthalmic instrument which measures both intraocular pressure (IOP) and corneal thickness (CT) simultaneously and at the same locus on the cornea. The device uses the principles of applanation pneumatometery to measure IOP and the principles of reflection pulsed ultrasound pachymetry to measure CT. The system is comprised of the TonoPach electronics, transducer handpiece, protective membranes and optionally, a foot pedal.
#### ട Intended Use
The TonoPach Ultrasonic Tonometer/Pachymeter Model P-201 is a diagnostic instrument that is intended for use in the measurement of both intraocular pressure and corneal thickness. It is intended for use during or following surgery and as a screening, monitoring or diagnostic aid in patients with normal, disease-induced or surgically altered anatomy, or in glaucoma suspects under current therapeutic guidelines.
#### 6 Technological Equivalence
Table 6.1 provides a list of predicate devices and the "k" number. In Tables 6.2, 6.3 and 6.4 the TonoPach Pachymeter portion and the Ultrasonic Tonometer are compared, respectively, to predicate devices to show the equivalence. Description statements were relied on to ascertain the intended use and technological features of legally marketed devices, and the substantial equivalence to the TonoPach to such legally marketed devices. The comparison of the intended use and technological features of this device to other legally marketed devices indicates that this device is substantially equivalent to legally marketed predicate devices with regard to safety, effectiveness and intended use. Note: areas of difference between the two devices are indicated in the tables. RetinaPharma believes that these differences are minor and should not raise any concerns regarding the overall safety and efficacy of the TonoPach.
| Table 6.1: List of predicate devices | | | |
|----------------------------------------------------------------------|--------------------------------------|--------------------|------------------------------|
| Predicate Device | Manufacturer | 510(k) | Date Cleared |
| Tonometer<br>TonoPen ophthalmic tonometer | Oculab, Inc. | K 852774 | August 20, 1985 |
| Pachymeter<br>Pocket Ultrasonic Pachymeter<br>Echoscan Model US-1800 | Quantel Medical, Inc.<br>Nidek, Inc. | K993674<br>K020876 | July 13, 2000<br>May 3, 2002 |
{2}------------------------------------------------
| Table 6.2: Comparison of Tonometer Characteristics | | |
|------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | TonoPen vs. TonoPach P 201 | |
| | Predicate Device<br>Mentor TonoPen 3 | RetinaPharma<br>TonoPach P 201 |
| Measurement | IOP | IOP |
| Pressure measurement device | Micro strain gauge | Micro strain gauge |
| Measurement technique | Applanation | Applanation |
| Pressure range | 5-80 mm Hg | 5-80 mm Hg |
| Pressure contact area* | 1.5 mm | 1.8 mm |
| Measurement frequency | 500 reading per second | 500 readings per second |
| Display/measure | LCD / 2 digit reading | LCD / 2 digit reading |
| Statistical storage | Yes | Yes |
| Pressure traceability | Manometer & Goldmann | Manometer & Goldmann |
| Usage | Handheld | Handheld or Slitlamp mounted |
| Power supply | Battery | Battery |
| System configuration* | Single unit; transducer and<br>system electronics combined | Tethered unit; transducer is<br>tethered from the system<br>electronics |
| Versatility | Can be used with the patient in<br>any position, making it suitable<br>for use in the office, bedside or in<br>remote locations. Suitable for use<br>in cases of their regular or Hg<br>corneal astigmatism. | Can be used with the patient in<br>any position, making it suitable<br>for use in the office, bedside or in<br>remote locations. Suitable for us<br>in cases of their regular or Hg<br>corneal astigmatism. |
| *Indicates differences between the predicate device and TonoPach | | |
| Table 6.3: Comparison of Pachymeter Characteristics<br>"Pocket" Ultrasonic Pachymeter and Echoscan US-1800 vs. TonoPach P 201 | | | |
|-------------------------------------------------------------------------------------------------------------------------------|--------------------------------|--------------------------------------|--------------------------------|
| | Predicate Device<br>"Pocket" | Predicate Device<br>Echoscan US-1800 | RetinaPharma<br>TonoPach P 201 |
| Measurement | Corneal thickness | Corneal thickness | Corneal thickness |
| Measurement system | Ultrasonic pulse<br>reflection | Ultrasonic pulse<br>reflection | Ultrasonic pulse<br>reflection |
| Probe type | Solid,<br>Angled | Solid,<br>Straight, Angled | Solid<br>Straight, Angled |
| Measurement range* | 100 to 1300 um | 200 to 1300 um | 125 to 1300 um |
| Clinical accuracy | +/- 5 um | +/- 5 um | +/- 5 um |
| Least significant display<br>digit | 1 um | 1 um | 1 um |
| Default velocity* | 1620 m/sec | 1640 m/sec,<br>changeable | 1640 m/sec |
| Exam Mode(s) | Single point, 1 map | Single point, 3 maps | Single point, 1 map |
| Data Output | LCD, Print | LCD, Print | LCD, Print |
| Power requirements | Battery | 120 VAC, 60 Hz | Battery |
| Weight | 0.46 kg | 6 kg | 0.75 kg |
| *Pachymeter mode. Indicates difference between the predicate device and the TonoPach | | | |
{3}------------------------------------------------
#### 7 TonoPach Testing
Substantial testing has been performed using the TonoPach. It includes inanimate, animal and human evaluation.
- A Inanimate: Models for both tonometry and pachymetry have been used.
- l ) The tonometry model consists of a water-filled spherical membrane attached to a manometer. When the TonoPach handpiece tip touches the sphere, it mimics intraocular pressure.
- 2) The pachymetry model consists of polycarbonate wafers of varying thicknesses. When the TonoPach handpiece tip touches a wet wafer, it is suitable to mimic the cornea for thickness measurement purposes.
- B Animal: Live rabbit eyes and freshly enucleated porcine eyes were used for several studies.
- l) Rabbit eyes were used to test the tonometer and pachymeter portions of the TonoPach. They were used to evaluate the TonoPach's performance on a live eye. Also, they were used to assist in evaluation of the System's numerical accuracy and reproducibility.
- 2) Pig eyes were used predominately to verify and validate tonometer accuracy and reproducibility. One study was used to demonstrate repeatability. Three porcine eyes were studied at each 10 mmHg pressure over a pressure range of 5 to 80 mmHg with 10 independent measurements per pressure. The data exceeds product specifications.
- C Clinical Crial: The TonoPach was evaluated on eye clinic volunteers who required tonometery. All patients who agreed to volunteer were admitted to the study.
- 1) Tonometry testing was the predominated the data collected. 347 eyes were tested for comparison to Goldmann tonometery. Scatter and Bland & Altman plots as well as mean, standard deviation and variance around the mean reveal that the TonoPach meets or exceeds specifications.
8 Statement of Confidentiality
Information contained in this letter and its attachments is considered confidential, proprietary information and a trade secret until RetinaPharma Technologies, Inc. distributes such information publicly. The release of information contained herein is to be governed by applicable provisions of the Food, Drug and Cosmetic Act, the Freedom of Information Act and pertinent FDA regulations in 21 CFR Part 20.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling a stylized caduceus or a bird in flight, composed of three curved lines.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC 1 0 2004
RetinaPharma Technologies, Inc. % Terry Fuller, Ph.D. 944 Morgan Road Jenkintown, PA 19046
Re: K042099
Trade/Device Name: TonoPach Ultrasonic Tonometer/Pachymeter Model P-201 Regulation Number: 21 CFR 886.1930 Regulation Name: Tonometer and accessories Regulatory Class: Class II Product Code: HKX Dated: July 30, 2004 Received: August 4, 2004
Dear Dr. Fuller:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
Page 2 - Terry Fuller
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4613. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
A halyl Rosenthal
A. Ralph Rosenthal, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# 3) Diagnostic Ultrasound Indications for Use Form
### Page 1 of 1
510(k) Number (if known):
Device name: RetinaPharma Model P 201, TonoPachTM System
Intended Use: The TonoPach Ultrasonic TonometerPachymeter Model P-201 is a diagnostic instrument that is intended for use in the measurement of both intraccular pressure and corneal thickness. It is intended for use during or following surgery and as a screening, monitoring or diagnostic aid in patients with normal, disease-induced or surgically altered anatomy, or in glaucoma suspects under current therapeutic guidelines.
| Clinical Application | Mode of Operation | | | | | | | | | |
|----------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) |
| Ophthalmic | N | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Intraoperative (specify) | | | | | | | | | | |
| Intraoperative<br>Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ (specify) | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | |
| Musculo-skeletal<br>Superficial | | | | | | | | | | |
| Other (specify)* | | | | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Additional Comments:
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use | <div style="display:inline-block; vertical-align:middle;"> <img alt="Checkmark" src="checkmark.png" style="width:15px; height:15px; vertical-align:middle;"/> </div> (Per 21 CFR 801.109) |
|------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Clay R. Butterman |
|--|-----------------------------|
| | (Division Sign-Off) |
| | Division of Ophthalmic Ear, |
| | Nose and Throat Devises |
510(k) TonoPach Indications for Use modified 041210.doc
From 510(k) 040729_Cover D.doc
| 510(k) Number | K042099 |
|---------------|---------|
|---------------|---------|
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.