K042009 · Usa Instruments, Inc. · MOS · Aug 11, 2004 · Radiology
Device Facts
Record ID
K042009
Device Name
MARK III PHASED ARRAY SHOULDER COIL
Applicant
Usa Instruments, Inc.
Product Code
MOS · Radiology
Decision Date
Aug 11, 2004
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1000
Device Class
Class 2
Indications for Use
The Mark III Phased Array Shoulder Coil is a receive-only phased array RF coil, used for obtaining diagnostic images of the shoulder and adjacent regions in Magnetic Resonance Imaging systems. The indications for use are the same as for standard MR Imaging. The Mark III Phased Array Coil is designed for use with the Signa EXCITE 3T MRI scanner manufactured by GE Medical Systems, Inc.
Device Story
Mark III Phased Array Shoulder Coil is a receive-only RF coil for MRI systems; designed for use with GE Signa EXCITE 3T scanner. Device consists of three volume RF coil elements in phased array configuration; enclosed in rigid housing and vinyl-coated PVC foam. Positioned on patient shoulder during imaging; captures RF signals for diagnostic image reconstruction. Does not transmit RF power; decoupling via switching diodes isolates elements during RF transmission to prevent RF burns. Used in clinical MRI settings by trained radiology personnel. Benefits include high-quality diagnostic imaging of shoulder anatomy.
Clinical Evidence
Bench testing only. No clinical data provided. Safety and effectiveness supported by comparison of design, materials, and operational characteristics to legally marketed predicate devices.
Indicated for use in standard MR imaging of the shoulder and adjacent regions for diagnostic purposes.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
K042009
Special 510(k) Premarket Notification Special 510(K) Premarket 110 Shoulder Coil July 23, 2004
Image /page/0/Picture/2 description: The image shows the logo for USA Instruments, Inc. The logo consists of the company name in a simple, sans-serif font, with the letters "USA" slightly larger than "Instruments, Inc." To the left of the company name is a circular symbol with the letters "US" inside. The symbol appears to be a certification mark.
### AUG 1 1 2004 SUMMARY OF SAFETY AND EFFECTIVENESS
| 1. Device Name : | Magnetic Resonance Imaging Accessory |
|-----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 2. Proprietary Name : | Mark III Phased Array Shoulder Coil |
| 3. Classification : | Class II |
| 4. Establishment Registration #: | 1529041 |
| 5. Manufacture Facility Location: | USA Instruments, Inc., 1515 Danner Drive,<br>Aurora, Ohio 44202, USA<br>Telephone: 330-995-8500; Fax: 330-562-1422. |
| 6. Performance Standard: | No applicable performance standards have been issued<br>under Section 514 of the Food, Drug and Cosmetic Act. |
| 7. Intended Use: | The Mark III Phased Array Shoulder Coil is a receive-only phased<br>array RF coil, used for obtaining diagnostic images of the shoulder<br>and adjacent regions in Magnetic Resonance Imaging systems. The<br>indications for use are the same as for standard MR Imaging. The<br>Mark III Phased Array Coil is designed for use with the Signa<br>EXCITE 3T MRI scanner manufactured by GE Medical Systems,<br>Inc. |
| 8. Device Description : | The Mark III Phased Array Coil consists of three volume RF coil<br>elements in a phased array design. The coil elements and<br>associated circuitry are enclosed to prevent any exposure to the<br>patient or environment. The coil electronics are enclosed in both<br>the rigid housing and the vinyl coated PVC foam. The coil is<br>positioned on the patient's shoulder during imaging. |
{1}------------------------------------------------
Special 310(k) Premarket Notification Special - STU(K) Prematics - Mark III Shoulder Coil
USA Instruments - Mark III Shoulder Coil
July 23, 2004
Image /page/1/Picture/1 description: The image shows the logo for USA Instruments, Inc. The logo consists of a circular symbol with the letters "US" inside, followed by the text "USA Instruments, Inc." The text is in a simple, sans-serif font and is aligned to the right of the circular symbol. The logo is black and white.
# 9. Safety and Effectiveness:
| Mark III Phased Array Shoulder Coil<br>Product Features | Comparison to Predicate or other 510(k) cleared<br>products |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use<br>Shoulder imaging applications | -Similar to the Mark 9000 Shoulder Coil manufactured by<br>USA Instruments Inc. (K010946)<br>-Similar to the Phased Array Shoulder Coil manufactured<br>by Medical Advances Inc. (K945778) |
| Indications for Use<br>Identical to routine MRI imaging | -Similar to the Mark 9000 Shoulder Coil manufactured by<br>USA Instruments Inc. (K010946)<br>-Similar to the Phased Array Shoulder Coil manufactured<br>by Medical Advances Inc. (K945778) |
| Coil Enclosure Material<br>Vinyl coated PVC foam<br>Flame retardent Polyurethane Plastic<br>PVC Plastic | -Similar to the Mark 9000 Shoulder Coil manufactured by<br>USA Instruments Inc. (K010946)<br>-Similar to the Magna 5000 Phased Array CTL Spine Coil<br>manufactured by USA Instruments, Inc. (K994345) |
| Coil Design<br>Receive-only phased array design | -Similar to the Mark 9000 Shoulder Coil manufactured by<br>USA Instruments Inc. (K010946) |
| Decoupling<br>Switching diode decoupling | -Similar to the Mark 9000 Shoulder Coil manufactured by<br>USA Instruments Inc. (K010946) |
| Prevention of RF Burns<br>Does not transmit RF power; decoupling<br>isolates the coil elements from RF fields<br>during RF transmission; coil elements and<br>circuitry are enclosed in a non-conductive<br>housing. | -Similar to the Mark 9000 Shoulder Coil manufactured by<br>USA Instruments Inc. (K010946)<br>-Similar to the Excalibur 9000 TotalSENSE™ Quad Array-<br>Head Coil manufactured by USA Instruments, Inc.<br>(K022582) |
| Radio Frequency Absorption<br>Coil is a receive only coil and does not<br>transmit RF power; power deposition during<br>imaging is limited by SAR algorithm | -Similar to the Mark 9000 Shoulder Coil manufactured by<br>USA Instruments Inc. (K010946) |
| Formation of Resonant Loop<br>Decoupling isolates the coil elements from RF<br>fields during RF transmission; length of cable<br>and stiffness does not permit looping | -Similar to the Mark 9000 Shoulder Coil manufactured by<br>USA Instruments Inc. (K010946) |
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services (HHS). The logo features a stylized depiction of an eagle with three stripes forming its wing and tail feathers. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the eagle symbol.
od are: Dring Administration 9200 Corporate Boulevard Rockville MD 20850
# AUG 1 1 2004
Mr. James Wrenn QA and Regulatory Manager USA Instruments, Inc. 1515 Danner Drive AURORA OH 44202
Re: K042009 Trade/Device Name: Mark III Phased Array Shoulder Coil Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: 90 MOS Dated: July 23, 2004
Received: July 26, 2004
Dear Mr. Wrenn:
We have reviewed your Section 510(k) premarket notification of intent to market the device WC nave reviewed your beceived the device is substantially equivalent (for the indications ferenced adove and nave betermile gally marketed predicate devices marketed in interstate for use stated in the enorosars) to togers the Medical Device Amendments, or to continer of pror to may 20, 2011 accordance with the provisions of the Federal Food, Drug, devices that have been require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The r ou may, therefore, marks of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean I hat FDA has made a determination that your device complies with other requirements of the Act that I DA has made a ceveres and regulations administered by other Federal agencies. You must comply or any I each states and intents, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product quality by sections (Sections 531-542 of the Act), 21 CFR 1000-1050.
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) I his letter will anow you to begin harmening of substantial equivalence of your device to a legally premarket nothcation. The PDA miding or baseman for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific advice to: your down general with the following numbers, based on the regulation number at the top of the letter:
| 8xx.1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Additionally, for questions on the prossesson Also, please note the regulation entitled, "Misbranding Other of Complianoo ut (201) 974) 974 Part 807.97) you may obtain. Other general by relected to premarket nonmanton - (r the Act may be obtained from the Division of Small information on your responsional and Consumer Assistance at its toll-free number (800) 638-2041 or Manufacturers, International and Selless http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Brigdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Special 510(k) Premarket Nouncation USA Instruments - Mark III Shoulder Coil July 23, 2004
Image /page/4/Picture/1 description: The image shows the logo for USA Instruments, Inc. The logo consists of a circular symbol with the letters "us" inside, along with the company name written in a simple, sans-serif font. The logo is clean and professional, suggesting a company that specializes in precision instruments. The text is black on a white background.
#### STATEMENT OF INTENDED USE
510(k) Number (if known): __ / < 0 Y 2 ou 9
Device Name: Mark III Phased Array Shoulder Coil
Indications for Use:
The Mark III Phased Array Shoulder Coil is a receive-only PF coil, used for obtaining
The Mark III Phased Array Massactic Researce Imaging systems Th The Mark III Phased Array Shoulder Coll is a receive Resonance Imagner Imaging systems. The diagnostic images of the shoulder and adjacent regions in Nil Phased Array Coll is designed for
indications for use are the same as for standard MR Inaging. The Mark II Phase indications for use are the same as for standard First hanges
use with the Signa EXCITE 3.0T MRI scanner manufactured by GE Medical Systems, Inc
# (PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
# Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801-109)
OR
Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________
Nancy Brogdon
Division of Reproduct and Radiological Devi 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.