← Product Code [EZL](/productcode/EZL) · K041983

# OPTION-VM URINARY CATHETER/ADPATER, MODELS MV39016, MVA39016 (K041983)

_Opticon Medical · EZL · Nov 17, 2004 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K041983

## Device Facts

- **Applicant:** Opticon Medical
- **Product Code:** [EZL](/productcode/EZL.md)
- **Decision Date:** Nov 17, 2004
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.5130
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The OPTION-vm is intended to provide drainage of the urinary bladder. The OPTION-vm is indicated for use only for urinary bladder drainage in male patients: 1) who have acute conditions that require short-term (14 days or less) urinary management; 2) who are capable of operating the device in accordance with its instructions for use; and 3) for whom normal bladder cycling is not contraindicated.

## Device Story

Indwelling urinary catheter for bladder drainage; sterile, single-use, disposable. Composed of biocompatible silicone elastomers; features flexible shaft with two proximal eyelets, dual lumens (drainage and balloon inflation), retention balloon, self-sealing microvalve port, and urine discharge bulb with integral valve. Used with optional Continuous Drainage Adaptor for passive drainage into standard collection bags. Prescribed by physician; inserted/removed by healthcare professional; operated by patient for bladder cycling. Provides controlled drainage; benefits patients requiring short-term urinary management.

## Clinical Evidence

Bench testing only.

## Technological Characteristics

Biocompatible silicone elastomers; dual-lumen shaft; retention balloon; self-sealing microvalve; integral valve assembly for drainage control. Sterile, single-use, disposable.

## Regulatory Identification

A urological catheter and accessories is a flexible tubular device that is inserted through the urethra and used to pass fluids to or from the urinary tract. This generic type of device includes radiopaque urological catheters, ureteral catheters, urethral catheters, coudé catheters, balloon retention type catheters, straight catheters, upper urinary tract catheters, double lumen female urethrographic catheters, disposable ureteral catheters, male urethrographic catheters, and urological catheter accessories including ureteral catheter stylets, ureteral catheter adapters, ureteral catheter holders, ureteral catheter stylets, ureteral catheterization trays, and the gastro-urological irrigation tray (for urological use).

## Predicate Devices

- OPTION-vf™ Urinary Catheter ([K023090](/device/K023090.md))
- OPTION-vf™ Urinary Catheter with Adaptor ([K033830](/device/K033830.md))
- Bardex® Silicone Foley Catheter ([K760093](/device/K760093.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

NOV 1 7 2004

K04/1983

page 1 of 2

#### 510(k) SUMMARY

#### OPTICON MEDICAL

## OPTION-vm™ Urinary Catheter

#### Submitter Information

Opticon Medical 7001 Post Road, Suite 100 Dublin, OH 43016

Primary Contact: Glenn D. Brunner, President Phone Number: (614) 366-2000 Fax Number: (614) 336-2059

Date Prepared: July 21, 2004

## Device Name

Trade / Proprietary Name: OPTION-vm™ Urinary Catheter
OPTION-vm™ Urinary Catheter with Adaptor

Common / Usual Name: urinary catheter or Foley catheter

Classification Name: Catheter, Retention type, Balloon (product code EZL; 21 CFR 876.5130)

#### Predicate Device

- K023090, OPTION-vf" Urinary Catheter .
- K033830, OPTION-vf" Urinary Catheter with Adaptor .
- K760093, Bardex® Silicone Foley Catheter .

#### Intended Use

The OPTION-vm is intended to provide drainage of the urinary bladder.

The OPTION-vm is indicated for use only for urinary bladder drainage in male patients: 1) who have acute conditions that require short-term (14 days or less) urinary management; 2) who are capable of operating the device in accordance with its instructions for use; and 3) for whom normal bladder cycling is not contraindicated.

{1}------------------------------------------------

K041985

Page 2 of 2

# Device Description

The OPTION-vm is an indwelling catheter that provides drainage of the urinary bladder. It is a sterile, single-use, disposable device that is to be prescribed by a physician and inserted and removed by an appropriate health care professional. The device is composed of biocompatible silicone elastomers, and consists of: a flexible shaft with two opposing eyelets in the proximal tip for urine entry, one internal lumen for urine drainage and a second lumen for balloon inflation; a retention balloon; a self-sealing balloon inflation microvalve port; and a urine discharge bulb with integral valve. The catheter may be used with the Continuous Drainage Adaptor accessory to provide continuous urinary drainage.

## Comparison to Predicate Devices

The OPTION-vm and the predicate devices are all indwelling urinary catheters, which are used in the same anatomic location for the same purpose. The OPTION-vm is similar in design, materials of construction, physical dimensions, methods of insertion, maintenance and removal, etc. to the predicate catheters. It also follows equivalent principles of operation as the OPTION-vf; it uses the exact same components to form the identical integral valve assembly that allows the user to control bladder drainage, thus it can be initially used with the same Continuous Drainage Adaptor to provide continuous passive drainage of the urinary bladder into a standard urine drainage bag.

## Supporting Information

Verification test results reported in this 510(k) application substantiate equivalence to the predicate devices. Thus, the OPTION-vm does not raise any new questions of safety or efficacy.

## Conclusion

The OPTION-vm Urinary Catheter is substantially equivalent to the predicate devices.

9-003

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird symbol, with three curved lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

NOV 1 7 2004

Mr. Glenn D. Brunner President Opticon Medical 7001 Post Road, Suite 100 DUBLIN OH 43016

Re: K041983

Trade/Device Name: OPTION-vm™ Urinary Catheter Regulation Number: 21 CFR §876.5130 Regulation Name: Urological catheter and accessories Regulatory Class: II Product Code: 78 EZL Dated: October 6, 2004 Received: October 7, 2004

Dear Mr. Brunner:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the do ice, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:

| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology)          | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology)                      | 240-276-0120 |
| Other           |                                  | 240-276-0100 |

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsmaldsmamain.html

Sincerely yours,

Nancy C. Brogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

510(k) Number (if known): K041983

Device Name: OPTION-vm™ Urinary Catheter

The OPTION-vm is indicated for use only for urinary bladder drainage in male Indications for Use: patients: 1) who have acute conditions that require short-term (14 days or less) urinary management; 2) who are capable of operating the device in accordance with its instructions for use; and 3) for whom normal bladder cycling is not contraindicated.

(PLEASE DONOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use
(Per 21 CFR 801.109)

OR

Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________

Daniel h. Syprom

---

(Division Sign Off)

(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number.

4-002

(Optional Format 1-2-96)

---

**Source:** [https://fda.innolitics.com/device/K041983](https://fda.innolitics.com/device/K041983)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
