SD360 DIGITAL RECORDER/SD360 HOLTER DIGITAL RECORDER
K041901 · North East Monitoring, Inc. · MWJ · Aug 31, 2004 · Cardiovascular
Device Facts
Record ID
K041901
Device Name
SD360 DIGITAL RECORDER/SD360 HOLTER DIGITAL RECORDER
Applicant
North East Monitoring, Inc.
Product Code
MWJ · Cardiovascular
Decision Date
Aug 31, 2004
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 870.2800
Device Class
Class 2
Indications for Use
Detection of Arrhythmias: The NorthEast Monitoring, Inc. SD360 Digital Recorder is indicated for use in continuous monitoring of cardiac rhythm when intermittent arrhythmia are suspected due to patient symptoms such as palpitations, transient ischemic attacks (TIAs), syncope (fainting), or other such symptoms as determined by the physician. Efficacy of Treatment: The NorthEast Monitoring Inc. SD360 Digital Recorder is indicated for use to determine whether current pharmacological treatment(s) of known arrhythmia is effective by measuring the frequency and duration of the arrhythmia compared to the frequency and duration prior to treatment. Pacemaker Evaluation: The NorthEast Monitoring Inc. SD360 Digital Recorder is indicated for use to evaluate the function of implanted pacemakers to insure that the pacemaker is functioning within prescribed limits. The NorthEast Monitoring SD360 Digital Recorder is to be used by or on the order of a physician.
Device Story
Ambulatory Holter monitor records cardiac rhythm data for up to 72 hours. Physician/technician applies electrodes and wires to patient; patient maintains activity diary and uses event button to mark symptoms. Device records ECG signals onto removable SD card; no real-time analysis performed. Post-recording, data downloaded to compatible NorthEast Monitoring Holter analysis system for review. Device features small LCD displaying time, error messages, or lead quality. Powered by single AA battery. Used in clinical or home settings to assist physicians in diagnosing arrhythmias, evaluating treatment efficacy, or checking pacemaker function. Benefits include capturing intermittent cardiac events for clinical decision-making.
Clinical Evidence
Bench testing only. Device passed electrical safety and EMC testing requirements. No clinical data provided.
Technological Characteristics
Ambulatory ECG recorder; 3-channel recording; 0.05-70 Hz or 0.05-150 Hz bandwidth; 180/360/720 samples/sec sampling rates. Uses standard Ag/AgCl electrodes. Storage via removable SD card. Powered by single 1.5V AA alkaline, NiMH, or Lithium L91 battery. Dimensions: 8.7cm x 6.5cm x 2cm; Weight: 99.3g with battery. Connectivity: SD card data transfer; optional PC/PDA interface for patient data entry.
Indications for Use
Indicated for patients requiring continuous cardiac rhythm monitoring to detect intermittent arrhythmias (e.g., palpitations, TIAs, syncope), evaluate pharmacological treatment efficacy for known arrhythmias, or assess implanted pacemaker function. For use by or on the order of a physician.
Regulatory Classification
Identification
A medical magnetic tape recorder is a device used to record and play back signals from, for example, physiological amplifiers, signal conditioners, or computers.
Predicate Devices
NorthEast Monitoring DR180 II Holter Recorder (K001288)
Submission Summary (Full Text)
{0}------------------------------------------------
K041901
p. 1 of 4
# 510(k) Summary AUG 3 1 2004
Date Prepared [21 CFR 807.92(a)(1)]
July 9, 2004
#### Submitter's Information [21 CFR 807.92(a)(1)]
Joseph M. Azary C/o NorthEast Monitoring Inc. 543 Long Hill Avenue Shelton, CT. 06484
Azary Technologies has received authorization to submit this 510(k) on behalf of the sponsor NorthEast Monitoring Inc. NorthEast Monitoring Inc. located at Two Clock Tower Place, Suite 360, Maynard, MA 01754, is an FDA-registered medical device under establishment# 1224919.
#### Trade Name, Common Name, Classification [21 CFR 807.92(2)(2)]
Device trade names are: NorthEast Monitoring SD360 Digital Recorder, SD360 Digital Holter Recorder Common Name: Ambulatory ECG Recorder, Ambulatory Electrocardiograph (without analysis) Classification: Class II, 21 CFR 870.2800, MWJ
#### Predicate Device |21 CFR 807.92(a)(3)|
- NorthEast Monitoring DR180 II Holter Recorder K001288 .
The subject device has the same indications for use as the predicate. Both the subject device and predicate device can be used for 3 channel recording. The main differences between the subject device and predicate device are as follows:
- . The subject device is smaller and thinner.
- The subject device weighs less. .
- The subject device requires only one AA battery (as opposed to two AA batteries required by the . predicate).
- . The subject device does not have 12 lead capabilities (this was not a heavily used option in the predicate device).
- . The LCD screen in the subject device is smaller. The LCD screen of predicate device showed waveforms, whereas the LCD screen of the subject device shows lead quality as a numeric value.
- The subject device uses SD cards instead of flash cards for memory. The SD cards are smaller, but . have the same memory capabilities as the flash cards.
{1}------------------------------------------------
#### Description of the Device [21 CFR 807.92(a)(4)]
An ambulatory monitor, sometimes called a Holter, is a painless method to monitor the heart beat for a All anloulatory moments, osmethics or 72 hours). The Holter is a small recording device that records the heart beat while being worn by the patient.
The physician or technician places electrodes and wires on the patient. The wires are connected to the I he prysicial or connectar patient is asked to write down a diary of daily activity including the time and character of any symptoms.
The patient can push a patient event button to mark an event. The patient must write down details about the event in their diary so that physician or technician can relate the event to specific symptoms or activities !
The NorthEast Monitoring SD360 Digital Recorder is a holter monitor designed to facilitate the ambulatory cardiac monitoring, on the order of a physician, of those patients who may benefit from such monitoring including but not limited to those with complaints of palpitations, syncope, chest pains, shortness of breath, or those who need to be monitored to judge their current cardiac function, such as patients who have recently received pacemakers.
The SD360 Digital Recorder package includes:
- . SD360 Digital Recorder
- . Operation Manual
- SD Card .
- Patient Cable
- Pouch ●
The data obtained by monitoring is not analyzed at the time of recording. After the recording is complete, the data must later be downloaded to a compatible NorthEast Monitoring holter analysis system to be analyzed. The Holter Analysis Software was cleared by FDA under K930564.
The SD360 is not intended to replace real time telemetry monitoring for patients suspected of having life-threatening arrhythmias.
The SD360 digital recorder is powered by one 1.5 volt AA alkaline battery (MN1500 or the equivalent), one AA rechargeable NiMH (nickel metal hydride) battery, or one AA Eveready Lithium L91 battery. Batteries should not be re-used for a second patient. The batteries are not included; users are instructed to purchase 2 AA batteries.
{2}------------------------------------------------
The device is compatible with standard silver / silver chloride ECG electrodes. Electrodes are not provided with the subject device. The user is instructed to purchase standard silver / silver chloride ECG electrodes.
The SD360 digital recorder uses NorthEast Monitoring SD360 patient cables with either seven leads or five leads for a 3-channel holter recording. The patient cable connects to the recorder via a 9-pin female connector on the recorder. A patient cable is provided with the SD360 Digital Recorder.
The SD360 has a small LCD that is used to display either time of day (during the recording), error messages (during the hookup procedure or during recording), or lead quality (during the hookup procedure).
The data collected by the SD360 digital recorder is stored on a removable SD Card. To store 24 hours in normal mode, the minimum capacity of the SD Card should be 28 megabytes; 56 megabytes are required for 24 hours in high resolution mode. To store 24 hours in 360 samples/sec mode, 112 megabytes are required. To stored 24 hours in 720 samples/sec, 224 megabytes are required. Double all storage requirements for 48 hour recordings, and triple them for 72 hour recordings. The SD360 is provided with an SD memory card with at least 32 megabytes.
The SD360 is packaged in a plastic bag in a cardboard shipping carton. The shipping carton will also include a patient cable and a pouch. The pouch is used by the patient to hold the digital recorder while in use.
The physician or technician can optionally use a PC as an interface to key in patient information. If a PC were used, the patient information would be keyed into the SD card using the NorthEast Monitoring Holter Analysis software (cleared under a separate 510k).
Another option for entering patient information is through the use of a PDA (i.e. Palm Pilot). Patient information such as patient name, sex, date of birth, identification number, scan number, hookup tech name or initials, physician name, indications, and medications.
{3}------------------------------------------------
| Characteristic | Specification |
|--------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Dimensions | 8.7cm (length) x 6.5cm (width) x 2cm (depth) |
| Weight | 70.9 grams (2.5 oz) without battery |
| | 99.3 grams (3.5 oz) with battery |
| Recording Bandwidth | 0.05 to 70 hertz in 180 samples/sec mode; 0.05 to 150 hertz<br>in 360 or 720 samples/sec mode. |
| Prefilter Sampling Rate | 720 samples/sec |
| Data Stored | In 180 samples/sec mode, data stored at 180 samples/sec (4<br>sample average), in 360 samples/sec mode, data stored at 360<br>samples/sec (2 sample average), in 720 samples/sec mode.<br>data stored at 720 samples/sec. |
| Pacemaker Sensitivity | 2 millivolts |
| Pacemaker Pulse Duration | 100 to 2500 microseconds |
# Physical and Electrical Specifications:
## Intended Use [21 CFR 807.92(a)(5)]
- Detection of Arrhythmias: The NorthEast Monitoring, Inc. SD360 Digital Recorder is indicated for use న్న in continuous monitoring of cardiac rhythm when intermittent arrhythmia are suspected due to patient symptoms such as palpitations, transient ischemic attacks (TIAs), syncope (fainting), or other such symptoms as determined by the physician.
- Efficacy of Treatment: The NorthEast Monitoring Inc. SD360 Digital Recorder is indicated for use to 6. determine whether current pharmacological treatment(s) of known arrhythmia is effective by measuring the frequency and duration of the arrhythmia compared to the frequency and duration prior to treatment.
- Pacemaker Evaluation: The NorthEast Monitoring Inc. SD360 Digital Recorder is indicated for use to 7. evaluate the function of implanted pacemakers to insure that the pacemaker is functioning within prescribed limits.
- The NorthEast Monitoring SD360 Digital Recorder is to be used by or on the order of a physician. 8.
## Technological Characteristics [21 CFR 807.92(a)(6)]
NorthEast Monitoring, Inc. believes that the subject device is substantially equivalent to the predicate device. The subject device has the same indications for use as the predicate.
The main technological difference is the device is smaller, thinner, weighs less, uses only one battery, does not have 12 lead capabilities, uses SD cards instead of flash cards for memory, and can be used with a Palm Pilot to enter patient information.
## Performance Data [21 CFR 807.92(b)(1)|
The subject device has been subjected to and passed electrical safety and EMC testing requirements.
## Conclusion [21 CFR 807.92(b)(3)]
We believe the changes are minor and conclude that the subject is as safe and effective as the predicate devices.
{4}------------------------------------------------
Image /page/4/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an abstract symbol that resembles an eagle or other bird-like figure.
AUG 3 1 2004
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Azary Technologies, LLC c/o Mr. Joseph M. Azary President 543 Long Hill Avenue Shelton, CT 06484
Re: K041901
K041901
Trade Name: Northeast Monitoring SD360 Digital Recorder Regulation Number: 21 CFR 870.2800 Regulation Name: Medical Magnetic Tape Recorder Regulatory Class: II (two) Product Code: MWJ Dated: July 14, 2004 Received: July 14, 2004
Dear Mr. Azary:
We have reviewed your Section 510(k) premarket notification of intent to market the device We have reviewed your Security (10(x) premaince in substantially equivalent (for the indications
referenced above and have determined the device is substants and in interstat referenced above and nave decembied the arresed predicate devices marketed in interstate for use stated in the enclosure) to legally mancede produced Device Amendments, or to
commerce prior to May 28, 1976, the enactment of the Federal Food. Drug commerce prior to May 28, 1970, the chance with the provisions of the Federal Food. Drug, devices that have been reculted in accordance with as proval approval application (PMA).
and Cosmetic Act (Act) that do not require approval assessor oval application of the and Cosmetic Act (Act) that do not require approvate or are al controls provisions of the Act. The
You may, therefore, market the device, subject to the general posignisti You may, therefore, market the device, sayeer to the governments for annual registration, listing of
general controls provisions of the Act include requirements misbranding a general controls provisions of the riot net merator required in the branding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is classified (see above) mis existing major regulations affecting your device can
may be subject to such additional controls. Existing major regulation EDA may be subject to such additional controlist - Existing major of 998. In addition, FDA may
be found in the Code of Federal Regulations, Title 21, Parts 800 Register be found in the Code of Federal Regarmers
publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 – Mr. Joseph M. Azary
Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please be advised that FDA s issualles of a succession complies with other requirements of the Act
that FDA has made a determination that your device complies . You must that FDA has made a decernmation that your as res by other Federal agencies. You must or any Federal statutes and regulations administered of excistration and listing (21 list.).
comply with all the Act's requirements, including, but not limited to: registrame comply with all the Act s requirements, more and manufacturing practice requirements as setting CFK Part 807), labeling (21 CFR Part 800); good manat 820); and if applicable, the electronic forth in the quality systems (QD) regalication (Sections 531-542 of the Act); 21 CFR 100-1050. product radiation control provisions (Decions over device as described in your Section 5 10(k)
This letter will allow you to begin marketing your device to a logic I his letter will anow you to begin maneting your antial equivalence of your dever to a legally premarket nothcation. The PDA mixing of backandal square and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Patt 801), please If you desire specific advice for your active of case ante the regulation entitled, contact the Office of Companies at (301) sifection" (21CFR Part 807.97). You may obtain Milsoraliums by reference to premainters with the Act from the Division of Small other geleral Informational and Consumer Assistance at its toll-free number (800) 638-2041 or Manufacturers, International and Gozess http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Nul R.L. Ogden
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use
510(k) Number (if known):_____________________________________________________________________________________________________________________________________________________ K041901
Device Name: NorthEast Monitoring Inc. SD360 Digital Recorder
Indications For Use:
- Detection of Arrhythmias: The NorthEast Monitoring, Inc. SD360 Digital Recorder is indicated 1. Detection of Arrily annual Price Northern when intermittent arrhythmia are suspected for use in continuous montoring of sitations, transient ischemic attacks (TIAs), syncope (fainting), or other such symptoms as determined by the physician.
- (1amiling), or other bases of the NorthEast Monitoring Inc. SD360 Digital Recorder is indicated for 2. Ellicacy of Treatment. The Northback Homent(s) of known arrhythmia is effective use to determine when barrent pharmator together arrhythmia compared to the frequency and duration prior to treatment.
- Pacemaker Evaluation: The NorthEast Monitoring Inc. SD360 Digital Recorder is indicated for 3. r acemaker Lyanaton. The NorthLasted pacemakers to insure that the pacemaker is functioning within prescribed limits.
- The NorthEast Monitoring SD360 Digital Recorder is to be used by or on the order of a 4. physician.
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use _ (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Neil Rpogh
(Division Sign-Off)
Division of Cardiovascular Devices
fw 892
510(k) Number_
Page 1 of L
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.