ACU-STRAP MOTION SICKNESS BAND

K041877 · Health Ent., Inc. · MVV · Oct 8, 2004 · NE

Device Facts

Record IDK041877
Device NameACU-STRAP MOTION SICKNESS BAND
ApplicantHealth Ent., Inc.
Product CodeMVV · NE
Decision DateOct 8, 2004
DecisionSESE
Submission TypeTraditional
Device ClassClass U
AttributesTherapeutic

Indications for Use

The Acu-Strap™ Travel and Motion Sickness Band is intended for the relief of nausca. Nausea may be experienced due to Travel (Motion Sickness), Pregnancy (Morning Sickness), Anesthesia or Chemotherapy.

Device Story

Acu-Strap™ is a pair of stretchable fabric wristbands featuring an integrated hard plastic button. Device operates via mechanical acupressure; user places band on wrist so button applies constant pressure to P6 (Nei-Kuan) acupressure point. Intended for patient self-use to alleviate nausea associated with travel, pregnancy, anesthesia, or chemotherapy. No external power, software, or complex processing required.

Clinical Evidence

Bench testing only. Testing performed to verify dimensions, elasticity, and relative pressure characteristics.

Technological Characteristics

Stretchable fabric wristband with integrated hard plastic button. Mechanical acupressure sensing/actuation principle. No energy source. Dimensions and materials are similar to predicate. Non-sterile.

Indications for Use

Indicated for relief of nausea in patients experiencing motion sickness, morning sickness, or nausea secondary to anesthesia or chemotherapy. Available for over-the-counter use.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ### OCT 8 - 2004 # 510(k) Summary for Acu-Strap™ Travel and Motion Sickness Band K041877 #### 1. SPONSOR Health Enterprises, Incorporated 90 George Leven Drive North Attleboro, MA 02760 (508) 695-0727 Contact Person: Glenn Leman, Chief Executive Officer #### 2. DEVICE NAME Proprietary Name: Acu-Strap Travel and Motion Sickness Band Common/Usual Name: Motion sickness band Classification Information: Motion sickness bands have not yet been classified by FDA, but have been given the following classification name and product code: | Name | Product Code | 21 CFR Ref. | Panel | |--------------------|--------------|------------------|-----------| | Acupressure device | MVV | To Be Determined | Neurology | #### 3. PREDICATE DEVICES Acu-Strap™ Travel and Motion Sickness Band is substantially equivalent to the Sea-Band, 510(k) No. K033268. #### 4. DEVICE DESCRIPTION The Acu-Strap™ Travel and Motion Sickness Band product comprises a pair of wristbands, each with a round button added to the interior face of the band. Because the Acu-Strap™ Travel and Motion Sickness Band is stretchable, it will be sold in only one size. The Acu-Strap™ Travel and Motion Sickness Band operates by placing a constant pressure on an acupressure point (the P6 or "Nei-Kuan" point) near the wrist. {1}------------------------------------------------ #### 5. INTENDED USE The Acu-Strap™ Travel and Motion Sickness Band is intended for the relief of nausca. Nausea may be experienced due to Travel (Motion Sickness), Pregnancy (Morning Sickness), Anesthesia or Chemotherapy. #### 6. TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE Acu-Strap™ Travel and Motion Sickness Band is made of a stretchable fabric material with a hard plastic button that is snapped in place during manufacturing process. It is similar to the predicate device in terms of dimensions, materials, and mechanical characteristics. #### 7 PERFORMANCE TESTING Acu-Strap™ Travel and Motion Sickness Band has been subjected to testing for dimensions, elasticity, and relative pressure. {2}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features a stylized eagle with three stripes representing the three levels of government: federal, state, and local. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 OCT 8 - 2004 Health Enterprises, Inc. C/o Mr. Daniel J. Dillon Medical Device Consultants, Inc. 49 Plain Street North Attleboro, Massachusetts 02760 Re: K041877 Trade/Device Name: Acu-Strap™ Travel and Motion Sickness Band Regulatory Name: Acupressure Device Regulatory Class: Unclassified Product Code: MVV Dated: September 20, 2004 Received: September 22, 2004 Dear Mr. Dillon: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ Page 2 - Mr. Daniel J. Dillon This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours. Mark A. Millmann Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Indications for Use 510(k) Number (if known): K041877 Acu-Strap™ Travel and Motion Sickness Band Device Name: Indications for Use: The Acu-Strap™ Travel and Motion Sickness Band is intended for the relief of nausea. Nausea may be experienced due to Travel (Motion Sickness), Pregnancy (Morning Sickness), Anesthesia or Chemotherapy. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-the-Counter Use X (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) for Mark N Milkers Sion-Division of General, Restorative, and Neurological Devices 510(k) Number K041877
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...