N APOLIPOPROTEIN STANDARD SERUM

K041870 · Dade Behring, Inc. · JIX · Jul 27, 2004 · Clinical Chemistry

Device Facts

Record IDK041870
Device NameN APOLIPOPROTEIN STANDARD SERUM
ApplicantDade Behring, Inc.
Product CodeJIX · Clinical Chemistry
Decision DateJul 27, 2004
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1150
Device ClassClass 2

Intended Use

For calibration of the Apolipoprotein A-I and Apolipoprotein B assays on BN™ Systems.

Device Story

N Apolipoprotein Standard Serum is a lyophilized human serum-based calibrator containing stabilizers and preservatives. It is used in clinical laboratory settings to establish reference curves for the quantitative determination of Apolipoprotein A-I and Apolipoprotein B assays on BN™ Systems. The calibrator is reconstituted by laboratory personnel and used to calibrate the analytical system, ensuring accurate measurement of patient samples. This process supports clinical decision-making regarding lipid metabolism and cardiovascular risk assessment.

Clinical Evidence

No clinical data. Performance was established via bench testing, specifically stability studies. Stability was evaluated over three lots at 2° to 8° C, demonstrating no significant change in recovery for at least 36 months (unopened) and 15 days once reconstituted.

Technological Characteristics

Lyophilized reagent; human serum matrix; immunonephelometric assay principle; calibrated against WHO reference preparations SP1-01 and SP3-07; compatible with Dade Behring BN Systems.

Indications for Use

Indicated for the calibration of Apolipoprotein A-I and Apolipoprotein B assays on Dade Behring BN Systems.

Regulatory Classification

Identification

A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)

Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # 510(k) Summary N Apolipoprotein Standard Serum ## Manufacturer's Name, Address, Telephone, and Contact Person, Date of 1. Preparation: Manufacturer: Dade Behring Marburg GmbH Emil-von-Behring Str. 76 Marburg/Germany Contact Information: Dade Behring Inc. Glasgow Site P.O. Box 6101 Newark, Delaware 19714 Attn: Donna Wolf Tel: 302-631-0384 July 8, 2004 KO41870 Preparation date: 2. # Device Name/ Classification: N Apolipoprotein Standard Serum / Calibrator, Class II (862.1150) ### 3. Identification of the Legally Marketed Device: Randox Apolipoprotein Calibrator (K023158) ### 4. Device Description: N Apolipoprotein Standard Serum is a lyophilized calibrator prepared from human serum with stabilizers and preservative. It is intended to establish reference curves for the quantitative determination of Apolipoprotein A-I and Apolipoprotein B assays on BNTM Systems. #### 5. Device Intended Use: For calibration of the Apolipoprotein A-I and Apolipoprotein B assays on BN™ Systems. ### 6. Medical device to which equivalence is claimed and comparison information: There are a number of in vitro diagnostic products that are used for the establishment of reference curves. One such product is the Randox Laboratories Apolipoprotein Calibrator (K023158). Both products are lyophilized humanbased calibrators intended for the calibration of Apolipoprotein A-1 and B. ### 7. Device Performance Characteristics: ## Stability: Stability was evaluated by testing N Apolipoprotein Standard Serum in duplicate at each time point for a total of three lots. The standard was evaluated at the recommended storage temperature of 2° to 8° C. Stability testing supports no significant change in recovery for at least 36 months, and for 15 days, once reconstituted. The N Apolipoprotein Standard Serum is substantially equivalent to other comparable calibrator products in commercial distribution. {1}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Image /page/1/Picture/2 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, with three intertwined snakes around a staff. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the caduceus. # JUL 27 2004 Food and Drug Administration 2098 Gaither Road Rockville MD 20850 Ms. Donna Wolf Regulatory Affairs and Compliance Manger Dade Behring Inc. P.O. Box 6101 Newark, DE 19714 k041870 Re: Trade/Device Name: N Apolipoprotein Standard Serum Regulation Number: 21 CFR 862.1150 Regulation Name: Calibrator Regulatory Class: Class II Product Code: JIX Dated: July 8, 2004 Received: July 13, 2004 Dear Ms. Wolf: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). {2}------------------------------------------------ ## Page 2 This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours, Jean M. Cooper, MS, DVM Jean M. Cooper, MS, D.V.M. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Indications for Use 510(k) Number (if known):_____________________________________________________________________________________________________________________________________________________ Device Name:___ N Apolipoprotein Standard Serum Indications For Use: For calibration of the Apolipoprotein A-I and Apolipoprotein B assays on BN™ Systems. Prescription Use AND/OR Over-The-Counter Use (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) Alberto Sating Division Sign-Off Dev 510(k) k041870 Page 1 of
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...