← Product Code [KDJ](/productcode/KDJ) · K041792

# FRESENIUS STAYSAFE PATIENT CONNECTORS, MODEL 050-87211 (K041792)

_Fresenius Medical Care North America · KDJ · Sep 24, 2004 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K041792

## Device Facts

- **Applicant:** Fresenius Medical Care North America
- **Product Code:** [KDJ](/productcode/KDJ.md)
- **Decision Date:** Sep 24, 2004
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.5630
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The Fresenius StaySafe Patient Connectors are intended for use with a peritoneal cycler for drainage and infusion of PD solution during a peritoneal dialysis exchanges. The Fresenius StaySafe Patient Connectors are indicated for use in acute and chronic peritoneal dialysis.

## Device Story

Fresenius STAYSAFE Patient Connectors facilitate cycler-assisted peritoneal dialysis; device incorporates pin and push-button/plunger mechanism to maintain closed system upon patient disconnection; pin engages patient catheter during PAUSE or treatment end; includes accessories like adaptors, tubing sets, and extension sets; provided sterile for single use; used in clinical or home settings by patients or caregivers; ensures sterile connection/disconnection during PD exchanges; prevents contamination; supports fluid management during dialysis.

## Clinical Evidence

No clinical data; bench testing only.

## Technological Characteristics

Disposable peritoneal dialysis tubing set and connector; pin and push-button/plunger mechanism; sterile; single-use; mechanical operation.

## Regulatory Identification

(1) A peritoneal dialysis system and accessories is a device that is used as an artificial kidney system for the treatment of patients with renal failure or toxemic conditions, and that consists of a peritoneal access device, an administration set for peritoneal dialysis, a source of dialysate, and, in some cases, a water purification mechanism. After the dialysate is instilled into the patient's peritoneal cavity, it is allowed to dwell there so that undesirable substances from the patient's blood pass through the lining membrane of the peritoneal cavity into this dialysate. These substances are then removed when the dialysate is drained from the patient. The peritoneal dialysis system may regulate and monitor the dialysate temperature, volume, and delivery rate together with the time course of each cycle of filling, dwell time, and draining of the peritoneal cavity or manual controls may be used. This generic device includes the semiautomatic and the automatic peritoneal delivery system.(2) The peritoneal access device is a flexible tube that is implanted through the abdominal wall into the peritoneal cavity and that may have attached cuffs to provide anchoring and a skin seal. The device is either a single use peritioneal catheter, intended to remain in the peritoneal cavity for less than 30 days, or a long term peritoneal catheter. Accessories include stylets and trocars to aid in the insertion of the catheter and an obturator to maintain the patency of the surgical fistula in the abdominal wall between treatments. (3) The disposable administration set for peritoneal dialysis consists of tubing, an optional reservoir bag, and appropriate connectors. It may include a peritoneal dialysate filter to trap and remove contaminating particles. (4) The source of dialysate may be sterile prepackaged dialysate (for semiautomatic peritoneal dialysate delivery systems or “cycler systems”) or dialysate prepared from dialysate concentrate and sterile purified water (for automatic peritoneal dialysate delivery systems or “reverse osmosis” systems). Prepackaged dialysate intended for use with either of the peritoneal dialysate delivery systems is regulated by FDA as a drug.

## Special Controls

*Classification.* Class II (special controls). The following accessories are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9: A catheter finger grip that is non-patient contacting and intended for single use with a peritoneal catheter; a continuous ambulatory peritoneal dialysis (CAPD) belt; and a catheter stand that does not include weigh scales.

## Predicate Devices

- Fresenius AMP 80/2 Pediatric Cycler System ([K811986](/device/K811986.md))
- Fresenius CAPD StaySafe Disposable Administration Sets with StaySafe Connector ([K022412](/device/K022412.md))
- Fresenius CAPD Safelock Transfer Set ([K904806](/device/K904806.md))
- Closed Disconnect System ([K900106](/device/K900106.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# SEP 2 4 2004

K041792
Page 1 of 2

## Fresenius STAYSAFE Patient Connectors 510(k) Premarket Notification

## 510K Summary

In accordance with the Food and Drug Administration Rule to implement provisions of the Safe Medical Devices Act of 1990 and in conformance with 21 CFR 807, this information serves as a Summary of Safety and Effectiveness for the Fresenius STAYSAFE Patient Connectors.

| Company:                            | Fresenius Medical Care North America<br>95 Hayden Ave.<br>Lexington, MA 02420                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
|-------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date:                               | July 1, 2004                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Contact Person:                     | Arthur Eilinsfeld                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Trade Name:                         | Fresenius STAYSAFE Patient Connectors                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Common Name:                        | Disposable Peritoneal Dialysis Tubing Sets and<br>Accessories                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Classification Name and Reference:  | 21 CFR §876.5630 Set, Administration, for Peritoneal<br>Dialysis, Disposable - Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Device Product Code and Panel Code: | 78/KDJ/Gastroenterology-Urology                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Predicate Device(s):                | Fresenius AMP 80/2 Pediatric Cycler System (now<br>Marketed under ECO™ Peritoneal Dialysis 4 Lead<br>Prefilled Set with Snap Disconnect) (#K811986 - SE<br>08/31/81);<br>Fresenius CAPD StaySafe Disposable Administration<br>Sets with StaySafe Connector (#K022412 - SE<br>03/05/03);<br>Fresenius AMP 80/2 Pediatric Cycler System (now<br>Marketed under ECO™ Peritoneal Dialysis 4 Lead<br>Prefilled Set with Snap Disconnect) (#K811986 - SE<br>08/31/81);<br>Fresenius CAPD Safelock Transfer Set (#K904806 -<br>SE 01/11/91);<br>Closed Disconnect System (#K900106 SE - 07/16/90);<br>submitted under Delmed, now owned by Fresenius<br>Medical Care. |

# 000035

{1}------------------------------------------------

### Fresenius STAYSAFE Patient Connectors 510(k) Premarket Notification

## 510K Summary

#### DEVICE INFORMATION:

#### A. DESCRIPTION

The Fresenius StaySafe Patient Connectors are designed for use during cycler assisted peritoneal dialysis. The device incorporates a pin and push button/plunger mechanism that provides a closed system after patient disconnection. The pin engages the patient's catheter during a PAUSE or at then end of the treatment. The device also has several accessories associated with its use: StaySafe adaptors. Multiple tubing sets, and extension sets. The device is provided sterile and is for single use only.

#### B. INTENDED USE

The Fresenius StaySafe Patient Connectors are intended for use with a peritoneal cycler for drainage and infusion of PD solution during peritoneal dialysis exchanges. The Fresenius StaySafe Patient Connectors are indicated acute and chronic peritoneal dialysis.

#### C. SUBSTANCIAL EQUIVALENCE INFORMATION

The indications for use, technological characteristics, design features, and materials are identical to the predicate devices. The STAYSAFE Patient Connectors are substantially equivalent to the predicate devices supported by the information, materials data, device description, and performance testing submitted and/or described within this 510K.

000036

741792

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird with three stylized wing-like shapes above a wavy line, representing water or movement.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

SEP 2 4 2004

Ms. Nichole Riek Regulatory Affairs Supervisor Fresenius Medical Care - North America Dialysis Products Division 95 Hayden Avenue LEXINGTON MA 02420-9192

Re: K041792

Trade/Device Name: STAYSAFE Patient Connectors Regulation Number: 21 CFR §876.5630 Regulation Name: Peritoneal dialysis system and accessories Regulatory Class: II Product Code: 78 KDJ Dated: July 1, 2004 Received: July 2, 2004

Dear Ms. Riek:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) I his ieter will anow you to ocgin mating of substantial equivalence of your device to a legally premarket nothcation. The PDF missing cation for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific advice to your as the following numbers, based on the regulation number at the top of the letter:

| 8xx.1xxx                         | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4654 |
| Other                            | (301) 594-4692 |

Additionally, for questions on the promotion and advertising of your device, please contact the Additionally, 101 questions on the promote the regulation entitled, "Misbranding Other of Complance at (2017 97 - 105 7.97) you may obtain. Other general by relevence to premarked includion (2 obtained from the Division from the Division of Small information on your responsionalist and Consumer Assistance at its toll-free number (800) 638-2041 or Manufacturers, International and Sollass http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Nancy C. Brogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

l'age 2

{4}------------------------------------------------

K041792

Image /page/4/Picture/1 description: The image shows a black and white illustration of a funnel. The funnel is composed of several trapezoidal shapes stacked on top of each other, decreasing in size from top to bottom. The top of the funnel is the widest, and the bottom is the narrowest, forming a point.

# Fresenius Medical Care

# Indications for Use Statement

Device Name:

Fresenius STAYSAFE Patient Connectors

Indications for Use:

The Fresenius StaySafe Patient Connectors are intended for use with a peritoneal cycler for drainage and infusion of PD solution during a peritoneal dialysis exchanges. The Fresenius StaySafe Patient Connectors are indicated for use in acute and chronic peritoneal dialysis.

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED

| Concurrence of CDRH, Office of Device Evaluation (ODE)        |                                                     |
|---------------------------------------------------------------|-----------------------------------------------------|
| Prescription Use (Per 21 CFR 801.109)                         | OR Over-The-Counter Use                             |
| Fresenius Medical Care North America                          |                                                     |
| Corporate Headquarters:                                       | 95 Hayden Avenue Lexington, MA 02420 (781) 402-9000 |
| (Division Sign-Off)                                           |                                                     |
| Division of Reproductive, Abdominal, and Radiological Devices |                                                     |
| 510(k) Number                                                 | K041792                                             |

000037

---

**Source:** [https://fda.innolitics.com/device/K041792](https://fda.innolitics.com/device/K041792)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
