K041732 · Blue Torch Medical Technologies · FGM · Sep 21, 2004 · Gastroenterology, Urology
Device Facts
Record ID
K041732
Device Name
CAVERMAP SURGICAL AID
Applicant
Blue Torch Medical Technologies
Product Code
FGM · Gastroenterology, Urology
Decision Date
Sep 21, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.4730
Device Class
Class 1
Indications for Use
The Blue Torch Medical Technologies CaverMap Surgical Aid is intended to provide stimulation to the body to locate and identify nerves and to test their excitability.
Device Story
The CaverMap Surgical Aid Perineal Tip is an accessory electrode for the CaverMap Surgical Aid System. It consists of a surgical stainless steel electrode encapsulated in dielectric insulating tubing, with an exposed tip for nerve stimulation. The device is used by surgeons during perineal radical prostatectomy, prostate brachytherapy, and prostate cryotherapy. The physician uses the probe to deliver electrical stimulation to tissues; the device aids in locating and identifying cavernosal and parasympathetic nerves by testing their excitability. The device is an adjunct to nerve-sparing surgical techniques; it does not replace the surgeon's clinical judgment or expertise in identifying neurovascular bundles. The system includes a reusable control handle for connection to the probe tip. The device provides the same charge density profile as previously marketed electrodes.
Clinical Evidence
Bench testing only. The device relies on design controls and performance testing to demonstrate substantial equivalence to the predicate device. No clinical data was provided.
Technological Characteristics
Materials: Surgical stainless steel electrode, dielectric insulating tubing. Energy: Electrical nerve stimulation. Form factor: Perineal probe tip with graduated length markings. Connectivity: Reusable control handle. Standards: 21 CFR 820, ISO 46001, EN 60601-1, ISO 10993.
Indications for Use
Indicated for stimulation of cavernosal and associated parasympathetic nerves in males during open or laparoscopic prostatectomy, prostate brachytherapy, prostate cryotherapy, and open colorectal procedures. Used as an adjunct to nerve-sparing techniques to aid the physician in nerve location. Not intended to replace surgeon expertise in mapping neurovascular bundles.
Regulatory Classification
Identification
A manual gastroenterology-urology surgical instrument and accessories is a device designed to be used for gastroenterological and urological surgical procedures. The device may be nonpowered, hand-held, or hand-manipulated. Manual gastroenterology-urology surgical instruments include the biopsy forceps cover, biopsy tray without biopsy instruments, line clamp, nonpowered rectal probe, nonelectrical clamp, colostomy spur-crushers, locking device for intestinal clamp, needle holder, gastro-urology hook, gastro-urology probe and director, nonself-retaining retractor, laparotomy rings, nonelectrical snare, rectal specula, bladder neck spreader, self-retaining retractor, and scoop. A manual surgical instrument that is intended specifically for use as an aid in the insertion, placement, fixation, or anchoring of surgical mesh during urogynecologic procedures are classified under § 884.4910 of this chapter.
Predicate Devices
Blue Torch Medical Technologies CaverMap Surgical Aid (K970971)
Submission Summary (Full Text)
{1}------------------------------------------------
K041732
page 1 of 2
# SEP 2 1 2004 510(k) Summary of Safety and Effectiveness Blue Torch Medical Technologies CaverMap Surgical Aid Additional Perineal Tip
## Company Name
Blue Torch Medical Technologies
# Official Contact
Frederick Tobia
## Device Name
| Proprietary Name: | Blue Torch Medical Technologies CaverMap® Surgical Aid Perineal<br>Tip |
|-------------------------|---------------------------------------------------------------------------------------------|
| Common Name: | Nerve Stimulator/Locator |
| Classification Name(s): | 21 CFR § 874.1820 Stimulator, Nerve<br>21 CFR § 876.4730 Probe And Director, Gastro-Urology |
## Predicate Devices used for Substantial Equivalence
Blue Torch Medical Technologies CaverMap Surgical Aid K970971
## Intended Use
The Blue Torch Medical Technologies CaverMap Surgical Aid is intended to provide stimulation to the body to locate and identify nerves and to test their excitability.
### Indications for Use
The Blue Torch Medical Technologies CaverMap® Surgical Aid Perineal Tip is an accessory stimulating tip to be used for a perineal approach to nerve stimulation. It is an addition to the Blue Torch Medical Technologies CaverMap® Surgical Aid System. The system is indicated for use in the stimulation of the cavemosal and associated parasympathetic nerves during open or laproscopic prostatectorny, prostate brachytherapy placement, prostate cryotherapy, and open colorectal (surgical) procedures in males. The device aids the physician in locating these nerves. The device is designed as an adjunct to the current open or laproscopic prostatectomy, prostate brachytherapy placement, prostate cryotherapy, and open colorectal procedures in which a nerve sparing technique is used. The Surgical Aid is not designed to replace the surgeon's expertise in mapping out the neurovascular bundles. Each physician's skill determines whether these nerves are spared regardless of any aid..
PAGE 30 OF 36 510(K) CAVERMAP SURGICAL AID PERINEAL TIP
{2}------------------------------------------------
K041732
Page 2 of 2
### Description
The CaverMap Surgical Aid Perineal Tip has been designed specifically for use during perineal radical prostatectomy, prostate brachytherapy, and prostate cryotherapy procedures, where a nerve sparing technique is employed.
The perineal probe tip is designed as a single electrode made from surgical stainless steel. The probe tip is encapsulated in dielectrical insulating tubing, with the exception of the tip which is exposed to deliver the stimulation to the nerves. The tip also includes graduated length markings below the insulating tubing. The surface area of the stimulating (exposed) trocar tip is the same as the current electrodes of the currently marketed probe tip, giving the electrode the same charge density profile as the currently marketed probe tips.
A new reusable control handle was also designed in order to allow for a proper connection to the new probe tip. The materials used to manufacture the new handle are exactly the same as the currently marketed product.
The perineal probe tip functions exactly the same as the existing probe tip. The new tip was designed for use during nerve sparing procedures where a perineal approach to stimulating the cavernosal nerves is indicated.
## Summary of Standards Achieved
FDA Quality Systems Regulation 21 CFR § 820 ISO 46001: Quality System EN 60601-1 ISO 10993
#### Summary
In summary, the Blue Torch Medical Technologies CaverMap Surgical Aid Perineal Tip is substantially equivalent to legally marketed devices. Quality System & Design Controls assure the device is substantially equivalent to the predicate devices with respect to its performance, safety, and effectiveness.
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird with three curved lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the bird symbol.
SEP 2 1 2004
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Frederick Tobia Regulatory Consultant Blue Torch Medical Technologies, Inc. 200 Homer Avenue Ashland Technical Center ASHLAND MA 01721
Re: K041732
Ro41732
Trade/Device Name: Blue Torch Medical Technologies CaverMap Surgical Aid Perineal Probe Tip Regulation Number: 21 CFR §874.1820 Regulation Name: Surgical nerve stimulator/locator Product Code: 77 ETN Regulation Number: 21 CFR §876.4730 Regulation Name: Manual gastroenterology-urology surgical instrument and accessories Product Code: 78 FGM Regulatory Class: II Dated: August 23, 2004 Received: August 25, 2004
Dear Mr. Tobia:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2
This letter will allow you to begin marketing your device as described in your Section 510(k) I his letter will anow you to obgin maining of substantial equivalence of your device to a legally prematication. The PDF mailing of and in the your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific advice for your de of the following numbers, based on the regulation number at the top of the letter:
| 8xx. 1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Additionally, 101 questions on the process. Also, please note the regulation entitled, "Misbranding Onlice of Complance at (2017 27) 21 11 (21 CFR Part 807.97) you may obtain. Other general by relevence to premarked nonmoution (er the Act may be obtained from the Division of Small Information on your responsionalist and Consumer Assistance at its toll-free number (800) 638-2041 or Manblacturers, International and Solass
(301) 443-6597 or at its Internet address http://www.fda.gov/cdch/dsmaldsmamain.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
### Page _
510(k) Number (if known):
K041932
Device Name:
Blue Torch Medical Technologies CaverMap® Surgical Aid Perineal Probe Tip
Indications for Use: The Blue Torch Medical Technologies CaverMap® Surgical Aid Perineal Tip is an accessory stimulating tip to be used for a perineal approach to nerve stimulation. It is an is all accessory of manang Medical Technologies CaverMap® Surgical Aid System. The system is indicated for use in the stimulation of the cavemosal and associated parasympathetic nerves during moreated for are in the outsecures, prostate brachytherapy placement, prostate cryotherapy, and open of laptoseopic producedem), provenles. The device aids the physician in locating these nerves open colored is designed as an adjunct to the current open or laproscopic prostatectomy, prostate I he uctier is deagered as arraste cryotherapy, and open colorectal procedures in which a nerve oracing technique is used. The Surgical Aid is not designed to replace the surgeon's expertise in sparing to the neurovascular bundles. Each physician's skill determines whether these nerves are spared regardless of any aid.
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
#### Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use: _ (Per 21 CFR 801.109)
PAGE 35 OF 36
OR
Over-The-Counter Use*_*_
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number: K041737
CAVERMAP SURGICAL AID PERINEAL TIP
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.