K041710 · Hoya Conbio · GEX · Aug 3, 2004 · General, Plastic Surgery
Device Facts
Record ID
K041710
Device Name
VERSAWAVE DENTAL ER:YAG LASER SYSTEM
Applicant
Hoya Conbio
Product Code
GEX · General, Plastic Surgery
Decision Date
Aug 3, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The VersaWave Dental Er:YAG Laser System is intended for Hard Tissue General Indications: Class I, II, III, IV and V cavity preparation. Carries removal. Hard tissue surface roughening or etching. Enameloplasty, excavation of pits and fissures for placement of sealants. Root Canal Hard Tissue Indications: Tooth preparation to obtain access to root canal. Root canal preparation including enlargement. Root canal debridement and cleaning. Bone Surgical Indications: Cutting, shaving, contouring and resection of oral osseous tissues (bone). Osteotomy. Endodontic Surgery (Root Amputation) Indications: Flap preparation incision of soft tissue to prepare a flap and expose the bone. Cutting bone to prepare a window access to the apex (apices) of the root(s). Apicoectomy amputation of the root end. Root end preparation for retrofill amalgam or composite. Removal of pathological tissues (i.e. cysts, neoplasm or abscess) and hyperplastic tissues (i.e. granulation tissue) from around the apex. Laser Periodontal Procedures: Full thickness flap. Partial thickness flap. Split thickness flap. Laser soft tissue curettage. Laser removal of diseased, infected, inflamed, and necrosed soft tissue within the periodontal pocket. Removal of highly inflamed edematous tissue affected by bacteria penetration of the pocket lining and junctional epithelium. Removal of granulation tissue from bony defects. Sulcular debridement. Osteoplasty and osseous recontouring. Ostectomy. Osseous crown lengthening. Soft Tissue Indications Including Pulpal Tissues: Incision, excision, vaporization, ablation, and coagulation of oral soft tissues, including: Excisional and incisional biopsies. Exposure of unerupted teeth. Fibroma removal. Flap preparation-incision of soft tissue to prepare a flap and expose the bone. Flap preparation-incision of soft tissue to prepare a flap and expose the unerupted teeth (hard and soft tissue impactions). Frenectomy and frenotomy. Gingival troughing for crown impressions. Gingivectomy. Gingivoplasty. Gingival incision and excision. Hemostasis and coagulation. Implant recovery. Incision and drainage of abscesses. Incision and drainage of periapical abscesses. Laser soft tissue curettage of the post-extraction tooth sockets and the periapical area during apical surgery. Leukoplakia. Operculectomy. Oral papillectomies. Pulpotomy. Pulp extirpation. Pulpotomy as an adjunct to root canal therapy. Reduction of gingival hypertrophy. Removal of pathological tissues (i.e. cysts, neoplasm, abscess) and hyperplastic tissues (i.e. granulation tissue). Root canal debridement and cleaning. Soft tissue crown lengthening. Treatment of canker sores, herpetic and aphthous ulcers of the oral mucosa. Vestibuloplaty. *For use on adult and pediatric patients.
Device Story
VersaWave Dental Er:YAG Laser System is a medical laser for dental applications. Device uses Er:YAG laser energy to perform hard and soft tissue procedures, including cavity preparation, bone resection, and soft tissue ablation/coagulation. Operated by dental professionals in clinical settings. Laser energy interacts with target tissues to achieve cutting, vaporization, or hemostasis. Output is controlled by the clinician to facilitate surgical access, tissue removal, or therapeutic treatment. Benefits include precise tissue management and reduced need for traditional mechanical instruments. Device is used for both adult and pediatric patients.
Clinical Evidence
Bench testing only; no clinical data provided.
Technological Characteristics
Er:YAG laser surgical instrument. Operates as a powered laser system for hard and soft tissue dental applications. Class II device.
Indications for Use
Indicated for adult and pediatric patients requiring oral hard tissue procedures (cavity prep, caries removal, bone surgery, endodontic surgery), periodontal procedures (flap, curettage, debridement, osseous recontouring), and soft tissue procedures (incision, excision, ablation, coagulation, biopsy, frenectomy, pulpotomy).
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
DELight Dental Er: YAG Laser System
WaterLase, Millennium Dental Laser System
Premier Centauri Dental Laser System
Submission Summary (Full Text)
{0}------------------------------------------------
K041710 1/3
### 9. 510(k) Summary
## AUG - 3 2004
| Company: | HOYA ConBio (formerly Continuum Electro-Optics, Inc.)<br>47733 Fremont Blvd<br>Fremont, CA 94538<br>(800) 532-1064 phone<br>(510) 445-4550 fax |
|----------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Jim Green<br>Vice President of Engineering |
| Device Trade Name: | VersaWave Dental Er:YAG Laser System |
| Common Name: | Medical Laser System |
| Classification Name:<br>Classification Code: | Instrument, surgical, powered, laser<br>79-GEX |
| Equivalent Device(s): | DELight Dental Er: YAG Laser System<br>WaterLase, Millennium Dental Laser System<br>Premier Centauri Dental Laser System |
| Intended Use: | The VersaWave Dental Er:YAG Laser System is intended for |
### Hard Tissue General Indications*
- Class I, II, III, IV and V cavity preparation .
- Carries removal .
- Hard tissue surface roughening or etching .
- Enameloplasty, excavation of pits and fissures for placement of sealants .
*For use on adult and pediatric patients
## Root Canal Hard Tissue Indications
- Tooth preparation to obtain access to root canal .
- Root canal preparation including enlargement .
- Root canal debridement and cleaning .
## Root Canal Hard Tissue Indications
- Tooth preparation to obtain access to root canal ●
- Root canal preparation including enlargement .
- Root canal debridement and cleaning .
### Bone Surgical Indications
- Cutting, shaving, contouring and resection of oral osseous tissues (bone) .
- . Osteotomy
## Endodontic Surgery (Root Amputation) Indications
- Flap preparation incision of soft tissue to prepare a flap and expose the bone .
{1}------------------------------------------------
$$
\zeta \circ \mathcal{A} \ne \emptyset$$
- Cutting bone to prepare a window access to the apex (apices) of the root(s). .
- Apicoectomy amputation of the root end. .
- Root end preparation for retrofill amalgam or composite .
- Removal of pathological tissues (i.e. cysts, neoplasm or abscess) and hyperplastic . Removal of pathological nesues (rom around the apex. Note: Any tissue growth (i.e. tissues (1.6. grandiaren besions) must be submitted to a qualified laboratory for histopathological evaluation.
### Laser Periodontal Procedures
- Full thickness flap .
- Partial thickness flap .
- Split thickness flap .
- Laser soft tissue curettage .
- Laser son ubsed of our and of the . periodontal pocket
- perfouontal pother
Removal of highly inflamed edematous tissue affected by bacteria penetration of the . pocket lining and junctional epithelium
- Removal of granulation tissue from bony defects .
- Sulcular debridement (removal of diseased, inflamed or necrosed soft tissue . in the periodontal pocket to improve clinical indices including gingival index, gingival bleeding index, probe depth, attachment loss and tooth mobility)
- gingi rai otooming masti, master, many of bone to correct osseous defects and . create physiological osseous contours)
- Ostectomy (resection of bone to restore bony architecture, resection of bone for . grafting, etc.)
- Osseous crown lengthening .
## Soft Tissue Indications Including Pulpal Tissues*
Incision, excision, vaporization, ablation, and coagulation of oral soft tissues, including:
- Excisional and incisional biopsies .
- Exposure of unerupted teeth .
- Fibroma removal .
- Flap preparation-incision of soft tissue to prepare a flap and expose the bone. .
- Flap preparation-incision of soft tissue to prepare a flap and expose the unerupted . teeth (hard and soft tissue impactions)
- Frenectomy and frenotomy .
- Gingival troughing for crown impressions .
- Gingivectomy .
- . Gingivoplasty
- Gingival incision and excision .
- Hemostasis and coagulation .
- Implant recovery ●
- Incision and drainage of abscesses .
- Incision and drainage of periapical abscesses .
- Laser soft tissue curettage of the post-extraction tooth sockets and the periapical area . during apical surgery
{2}------------------------------------------------
$$<\mathcal{V} \not\rightarrow \mathcal{V} \not\rightarrow \mathcal{V} \mathcal{C} \equiv \mathcal{V} \mathcal{C}$$
- . Leukoplakia
- Operculectomy .
- Oral papillectomies .
- Pulpotomy .
- . Pulp extirpation
- Pulpotomy as an adjunct to root canal therapy .
- Reduction of gingival hypertrophy .
- Removal of pathological tissues (i.e. cysts, neoplasm, abscess) and hyperplastic . tissues (i.e. granulation tissue). Note: any tissue growth (i.e., cyst, neoplasm or other tissues (f.c. grandration aboratory for histopathological evaluation.
- Root canal debridement and cleaning .
- Soft tissue crown lengthening .
- Treatment of canker sores, herpetic and aphthous ulcers of the oral mucosa .
- Vestibuloplaty .
*For use on adult and pediatric patients
The VersaWave Dental Er: YAG Laser System, the DELight Er: Comparison: Y AG Laser System, the WaterLase, Millennium Dental Laser System, and the Premier Centauri Dental Laser System are equivalent in operating parameters, physical characteristics, and intended uses.
| Nonclinical Performance<br>Data: | None |
|----------------------------------|------|
| Clinical Performance Data: | None |
| Additional Information: | None |
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
AUG - 3 2004
Mr. Jim Green Vice President of Engineering Hoya ConBio 47733 Fremont Boulevard Fremont, California 94538
Re: K041710 K0+1710
Trade/Device Name: VersaWave Dental Er:YAG Laser System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: GEX Dated: June 18, 2004 Received: June 29, 2004
Dear Mr. Green:
We have reviewed your Section 510(k) premarket notification of intent to market the devices indication we nave reviewed your bectorit 310(t) presidentially equivalent (for the indications
referenced above and have determined the device is substantially equivalent felerenced above and nave determined with marketed predicate devices marketed in interstate for use stated in the encrosule) to regarly manner date of the Medical Device Amendments, or to commerce prior to May 20, 1776, the enaounce with the provisions of the Federal Food, Drug, devices that have been recalismed in accessfire approval of a premarket approval application (PMA).
and Cosmetic Act (Act) that do not require approval of a premarket approva and Cosmetic Act (71ct) that Go not request of the general controls provisions of the Act. The You may, therefore, market the act include requirements for annual registration, listing of general controls provisions of the free labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is classified (300 a0070) and existing major regulations affecting your device can
may be subject to such additional controls. Existing major may be subject to such additional controliations, Title 21, Parts 800 to 898. In addition, FDA may be found in the Outs ocements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not nean Ficase oe advised that I DTT 3188aa.co brour device complies with other requirements of the Act that IDA has made a decormination administered by other Federal agencies. You must of ally I cut all the Act's requirements, including, but not limited to: registration and listing (21 Compry with an the Ace 3 requirements, one . enomantfacturing practice requirements as set CTN in the quality systems (QS) regulation (21 CFR Part 820); and if applicable the electronic forth in the quality systems (Sections 531-542 of the Act): 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2 - Mr. Jim Green
This letter will allow you to begin marketing your device as described in your Section 510(k) This iction will and will and of substantial equivalence of your device to a legally premaired predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific at no 101 ) at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small other general international and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Miriam C. Provost
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# 6. Indications for Use Statement
K041710 510(k) Number:
Device Name:
VersaWave Dental Er:YAG Laser System
Indications for Use:
### Hard Tissue
#### General Indications*
- Class I, II, III, IV and V cavity preparation .
- Carries removal .
- Hard tissue surface roughening or etching .
- Flard doods carrer and fissures for placement of sealants .
*For use on adult and pediatric patients
## Root Canal Hard Tissue Indications
- Tooth preparation to obtain access to root canal .
- Root canal preparation including enlargement .
- Root canal debridement and cleaning .
## Root Canal Hard Tissue Indications
- Tooth preparation to obtain access to root canal .
- Root canal preparation including enlargement .
- Root canal debridement and cleaning .
### Bone Surgical Indications
- Cutting, shaving, contouring and resection of oral osseous tissues (bone) .
- Osteotomy .
## Endodontic Surgery (Root Amputation) Indications
- offer Burger J (1000) is incision of soft tissue to prepare a flap and expose the bone .
- Cutting bone to prepare a window access to the apex (apices) of the root(s). .
- Apicoectomy amputation of the root end. .
- Root end preparation for retrofill amalgam or composite .
- Removal of pathological tissues (i.e. cysts, neoplasm or abscess) and hyperplastic . tissues (i.e. granulation tissue) from around the apex. Note: Any tissue growth (i.e. trists, neoplasm, or other lesions) must be submitted to a qualified laboratory for histopathological evaluation.
### Laser Periodontal Procedures
- Full thickness flap .
- Partial thickness flap .
- Split thickness flap .
- Laser soft tissue curettage .
{6}------------------------------------------------
- Laser removal of diseased, infected, inflamed, and necrosed soft tissue within the . periodontal pocket
- periodomial pocket
Removal of highly inflamed edematous tissue affected by bacterna penetration of the . pocket lining and junctional epithelium
- Removal of granulation tissue from bony defects .
- Removal of granufation assub from Seny and features or necrosed soft tissue . Suicular debrident (removal of electional indices including gingival index.
in the periodontal pocket to improve clinical indices including gingival index. In the perfodomal pooker to maps.
- gifigival bleeding moce uspan, in
Osteoplasty and osseous recontouring (removal of bone to correct osseous defects and . create physiological osseous contours)
- Ostectorysionogical 665008 between bony architecture, resection of bone for . grafting, etc.)
- Osseous crown lengthening .
## Soft Tissue Indications Including Pulpal Tissues*
Incision, excision, vaporization, ablation, and coagulation of oral soft tissues, including:
- Excisional and incisional biopsies .
- Exposure of unerupted teeth .
- Fibroma removal .
- Flap preparation-incision of soft tissue to prepare a flap and expose the bone. .
- I Tap preparation incision of soft tissue to prepare a flap and expose the unerupted . teeth (hard and soft tissue impactions)
- Frenectomy and frenotomy .
- Gingival troughing for crown impressions .
- Gingivectomy .
- Gingivoplasty .
- Gingival incision and excision .
- Hemostasis and coagulation ●
- Implant recovery ●
- Incision and drainage of abscesses ●
- Incision and drainage of periapical abscesses .
- Incision and Gramage of the post-extraction tooth sockets and the periapical area . during apical surgery
- Leukoplakia .
- Operculectomy .
- Oral papillectomies .
- Pulpotomy .
- Pulp extirpation .
- Pulpotomy as an adjunct to root canal therapy .
- Reduction of gingival hypertrophy .
- Recuenon of gingival tissues (i.e. cysts, neoplasm, abscess) and hyperplastic . tissues (i.e. granulation tissue). Note: any tissue growth (i.e., cyst, neoplasm or other tissues (I.C. grandhation assac). Trevel in the many for histopathological evaluation.
I sions) must be submitted to a qualified laboratory for histopathological evaluation.
- Root canal debridement and cleaning .
- Soft tissue crown lengthening .
{7}------------------------------------------------
3/3
- Treatment of canker sores, herpetic and aphthous ulcers of the oral mucosa ●
- Vestibuloplaty .
*For use on adult and pediatric patients
Prescription Use _____________________________________________________________________________________________________________________________________________________________ X (21 CFR 801 Subpart D)
OR
Over-the-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Miriam C. Provost
(Division Sign-Off) Division of General, Restorative, and Neurological Devices
K041710 510(k) Number_
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.