C-BOOT

K041659 · C-Boot , Ltd. · JOW · Oct 28, 2004 · Cardiovascular

Device Facts

Record IDK041659
Device NameC-BOOT
ApplicantC-Boot , Ltd.
Product CodeJOW · Cardiovascular
Decision DateOct 28, 2004
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.5800
Device ClassClass 2
AttributesTherapeutic

Intended Use

The C-Boot™ is a prescriptive device that induces controlled compression of the calf, the foot or a combined compression of both. The C-Boot™ is intended to assist patients, suffering from lymphatic or venous disorders, by treating many conditions including: - Prevention of deep vein thrombosis (DVT) - Enhancement of blood circulation - Reduction of post-operative pain and swelling - Reduction of wound-healing time - Stasis dermatitis - Treatment and assist healing of cutaneous ulceration - Venous stasis ulcers - Leg ulcers - Chronic venous insufficiencies - Reduction of edema - Prevent pooling of fluids in limbs - Lymphedema

Device Story

Wearable pneumatic compression boot for peripheral vascular and lymph disorders; utilizes mechanical energy generated by patient movement and weight transfer during walking to inflate sleeves; provides sequential, cyclical compression to calf/foot; replaces electrical pumps found in traditional systems with self-powered mechanical actuation; operates via patient mobility; enhances lymphatic/venous return; reduces edema and promotes healing; used by patients to manage chronic conditions; clinical benefit derived from harnessing gravitational force and body weight to drive therapeutic massaging effect on vessel walls.

Clinical Evidence

Bench testing only. Functional, performance, and safety testing conducted; comparative analysis performed against predicate devices (WizAir and Lympha Press) to demonstrate substantial equivalence.

Technological Characteristics

Wearable pneumatic compression boot; mechanical energy source (patient movement/weight transfer) vs electrical; sequential, intermittent compression; 1-2 cycles per minute; fabric construction; non-networked; non-electronic.

Indications for Use

Indicated for patients with lymphatic or venous disorders requiring compression therapy for DVT prevention, circulation enhancement, edema reduction, wound healing, and management of venous/leg ulcers or chronic venous insufficiency. Contraindicated in patients with advanced peripheral arterial occlusive disease, decompensated heart disease, specific phlebitis, phlegmasia coerulea dolens, sensory loss/neuropathy in lower limbs, supportive dermatosis, or primary chronic arthritis.

Regulatory Classification

Identification

A compressible limb sleeve is a device that is used to prevent pooling of blood in a limb by inflating periodically a sleeve around the limb.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K041659 Image /page/0/Picture/2 description: The image shows a logo with a stylized design. On the left side, there is a shape resembling a quarter circle with a square inside. To the right of this shape, the letters 'DOT' are written in a simple, sans-serif font. The logo has a clean and modern appearance. ## 510(k) Summary C-Boot™ - Lymph and Venous-Insufficiency System #### 1. Applicant C-Boot Ltd. Name: Address: 10 Mahanain st., Haifa 34481, Israel +972-54-5369919 Phone: Fax: +972-4-9849078 #### 2. Contact Person | Name: | Avi Abraham | |----------|------------------------------------| | Company: | Avnet Engineering | | Address: | 8 Marva st., Carmiel 21691, Israel | | Phone: | +972-54-4949699 | | Fax: | +972-4-9589111 | | Email: | avi_abraham@yahoo.com | #### Date Prepared 3. Date: June 08,2004 #### Device Name ধ Common Name:Lymph and Venous-Insufficiency System C-Boot™ Trade Name: #### Classification 5. Classification Name: Compressible Limb Sleeve Device Class: Class II Panel: Cardiovascular JOW Product Code: 870.5800 Requlation Code: #### 6. U.S. Agent Steve Doppelt C-Boot® Marketing Consultant 5 Wyndstone Way Framingham, MA 01701 508 820 0707 Phone: 508 820 9971 Fax: doppmed@aol.com E-Mail: C-Boot 510(k) June 2004 ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------Page 40 of 51 Lymph and Venous Insufficiency System {1}------------------------------------------------ ## 7. Predicate Devices The "C-Boot® Lymph and Venous-Insufficiency System" is substantially equivalent to the commercially available "WizAir™ Ambulant Pneumatic Compression System" (of MCS), 510(k) number K002287 cleared on 25-Aug-2000: It is also substantially equivalent to the commercially available Lympha Press® (of Mego-Afek), 510(k) number K810338 cleared on 20-Oct-1981. ## 8. Device Description C-Boot Ltd. introduces a wearable pneumatic compression technology embodied in a therapeutic boot that has the potential to increase treatment efficacy and improve the quality of life for peripheral vascular and lymph distorders patients. C-Boot's solution integrates the technological advantages of dynamic compression pumps, that enable preset sequential pressure control to individualize treatment regimes, with the mobility benefit of static compression stockings, that activate the calf muscle pump and strengthened the body's natural tendency to enhance lymphatic and venous return. The device relates to a self-powered compression devices and methods for promoting circulation and therapeutic compression: It is comprising a pump, a plurality of sleeves or limb massager, which utilizes sequential, cyclical pressure, energy or vector forces, to aid circulation in a body part such as the limbs. The inflatable sleeves are filled with air by the pump and transmit massaging movement to the vessel walls whenever the patient wearing the device - walks or transmits body's weight from one leg to another. The pressure generated is a function of the user's body weight and gravitational force. The device provides methods to harness energy generated during movement and weight transfer and uses the energy for compressing to create a massaging effect on the circulation and the muscular-skeletal system. {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows a logo with the letters "DDT" to the right of a circular shape. The circular shape is bisected vertically, with the left half being a square and the right half being an arrow pointing to the right. The logo is black and white. ## 9. Indications for Use The C-Boot™ is a prescriptive device that induces controlled compression of the calf, the foot or a combined compression of both. The C-Boot™ is intended to assist patients, suffering from lymphatic or venous disorders, by treating many conditions including: - ਕ. Prevention of deep vein thrombosis (DVT) - Enhancement of blood circulation b. - Reduction of post-operative pain and swelling C. - d. Reduction of wound-healing time - e. Stasis dermatitis - f. Treatment and assist healing of cutaneous ulceration - g . Venous stasis ulcers - Leg ulcers h. - Chronic venous insufficiencies i. - i. Reduction of edema - Prevent pooling of fluids in limbs k. - -Lymphedema ## 10. Contraindications The C-Boot™ should not be used in the following cases: Advanced peripheral arterial occlusive diseases (calves pain while walking), Decompensated heart disease, Specific phlebitis, Phlegmasia coerulea dolens (swollen blue limb), Diabetic patients with lack of reduced sensation in lower limb or foot, Supportive dermatosis, Intolerance of C-Boot™ fabric, sensory disturbances of the limb, Advance peripheral neuropathy, Primary chronic arthritis. {3}------------------------------------------------ ## 11. Technological Characteristics and Substantial Equivalence The C-Boot™ has combined indications of its two (2) predicate devices - WizAir™ and Lympha Press®- into one product. All three (3) devices have same shape of surrounding sleeves, inflation mode (sequential), mode of Compression (intermittent), number of Cycles (1-2 per minute) and are made all of basically same materials. The main differences are the dynamic compression pumps, the overlapping-massaging-movements and the mechanical power source of C-Boot™ (compared to an electrical source of the other predicate devices), which raise neither safety nor effectiveness questions, as C-Boot™ functions within same boundaries of performances. ## 12. Performance Data A series of functional, performance and safety testing, including comparative analysis to Mego-Afek® and WizAir™ - demonstrates that C-Boot™ is substantially equivalent to Mego-Afek® and WizAir™, without raising new safety and/or effectiveness issues. {4}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Image /page/4/Picture/2 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular with text around the perimeter. In the center of the seal is a stylized image of an eagle. OCT 2 8 2004 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 C-Boot Ltd. c/o Mr. Avi Abraham Avnet Engineering 8 Marva Street Carmiel 21691, Israel Re: K041659 C-Boot™ Regulation Number: 21 CFR 870.5800 Regulation Name: Compressible Limb Sleeve Regulatory Class: Class II (two) Product Code: JOW Dated: August 12, 2004 Received: August 16, 2004 Dear Mr. Abraham: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for are waver to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {5}------------------------------------------------ Page 2 - Mr. Avi Abraham Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, ourna R. la chrien ( Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ Image /page/6/Picture/0 description: The image shows a logo with the letters "BDDT" in a stylized font. The letters are arranged horizontally, with the first letter "B" being the largest and partially obscured by a semi-circular shape on the left. The semi-circular shape is filled with a checkered pattern. The letters "DDT" are smaller and appear to be in a bold, sans-serif font. # 10. Indications for Use Statement ### 510(k) Number (if known): K041659 #### Device Name: C-Boot™ #### Indications for Use: The C-Boot™ is a prescriptive device that induces controlled compression of the calf, the foot or a combined compression of both. The C-Boot™ is intended to assist patients, suffering from lymphatic or venous disorders, by treating many conditions including: - Prevention of deep vein thrombosis (DVT) 1. - Enhancement of blood circulation 2. - Reduction of post-operative pain and swelling 3. - Reduction of wound-healing time 4. - క్. Stasis dermatitis - Treatment and assist healing of cutaneous ulceration 6. - Venous stasis uicers 7. - Leg ulcers 8. - Chronic venous insufficiencies ு. - 10. Reduction of edema - 11. Prevent pooling of fluids in limbs - 12. Lymphedema _____ Prescription Use - لا__ AND/OR (Per 21 CFR 801 Subpart D) Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Mina L. Value (Division Sign-Off) Division of Cardiovascular Devices 510(k) Number_ko41659 .............................................................................................................................................................................. Page 1 of 1 Lymph and Venous Insufficiency System
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...