The C-Boot™ is a prescriptive device that induces controlled compression of the calf, the foot or a combined compression of both. The C-Boot™ is intended to assist patients, suffering from lymphatic or venous disorders, by treating many conditions including: - Prevention of deep vein thrombosis (DVT) - Enhancement of blood circulation - Reduction of post-operative pain and swelling - Reduction of wound-healing time - Stasis dermatitis - Treatment and assist healing of cutaneous ulceration - Venous stasis ulcers - Leg ulcers - Chronic venous insufficiencies - Reduction of edema - Prevent pooling of fluids in limbs - Lymphedema
Device Story
Wearable pneumatic compression boot for peripheral vascular and lymph disorders; utilizes mechanical energy generated by patient movement and weight transfer during walking to inflate sleeves; provides sequential, cyclical compression to calf/foot; replaces electrical pumps found in traditional systems with self-powered mechanical actuation; operates via patient mobility; enhances lymphatic/venous return; reduces edema and promotes healing; used by patients to manage chronic conditions; clinical benefit derived from harnessing gravitational force and body weight to drive therapeutic massaging effect on vessel walls.
Clinical Evidence
Bench testing only. Functional, performance, and safety testing conducted; comparative analysis performed against predicate devices (WizAir and Lympha Press) to demonstrate substantial equivalence.
Technological Characteristics
Wearable pneumatic compression boot; mechanical energy source (patient movement/weight transfer) vs electrical; sequential, intermittent compression; 1-2 cycles per minute; fabric construction; non-networked; non-electronic.
Indications for Use
Indicated for patients with lymphatic or venous disorders requiring compression therapy for DVT prevention, circulation enhancement, edema reduction, wound healing, and management of venous/leg ulcers or chronic venous insufficiency. Contraindicated in patients with advanced peripheral arterial occlusive disease, decompensated heart disease, specific phlebitis, phlegmasia coerulea dolens, sensory loss/neuropathy in lower limbs, supportive dermatosis, or primary chronic arthritis.
Regulatory Classification
Identification
A compressible limb sleeve is a device that is used to prevent pooling of blood in a limb by inflating periodically a sleeve around the limb.
Predicate Devices
WizAir™ Ambulant Pneumatic Compression System (K002287)
{0}------------------------------------------------
K041659
Image /page/0/Picture/2 description: The image shows a logo with a stylized design. On the left side, there is a shape resembling a quarter circle with a square inside. To the right of this shape, the letters 'DOT' are written in a simple, sans-serif font. The logo has a clean and modern appearance.
## 510(k) Summary C-Boot™ - Lymph and Venous-Insufficiency System
#### 1. Applicant
C-Boot Ltd. Name: Address: 10 Mahanain st., Haifa 34481, Israel +972-54-5369919 Phone: Fax: +972-4-9849078
#### 2. Contact Person
| Name: | Avi Abraham |
|----------|------------------------------------|
| Company: | Avnet Engineering |
| Address: | 8 Marva st., Carmiel 21691, Israel |
| Phone: | +972-54-4949699 |
| Fax: | +972-4-9589111 |
| Email: | avi_abraham@yahoo.com |
#### Date Prepared 3.
Date: June 08,2004
#### Device Name ধ
Common Name:Lymph and Venous-Insufficiency System C-Boot™ Trade Name:
#### Classification 5.
Classification Name: Compressible Limb Sleeve Device Class: Class II Panel: Cardiovascular JOW Product Code: 870.5800 Requlation Code:
#### 6. U.S. Agent
Steve Doppelt C-Boot® Marketing Consultant 5 Wyndstone Way Framingham, MA 01701 508 820 0707 Phone: 508 820 9971 Fax: doppmed@aol.com E-Mail:
C-Boot 510(k)
June 2004
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------Page 40 of 51
Lymph and Venous Insufficiency System
{1}------------------------------------------------
## 7. Predicate Devices
The "C-Boot® Lymph and Venous-Insufficiency System" is substantially equivalent to the commercially available "WizAir™ Ambulant Pneumatic Compression System" (of MCS), 510(k) number K002287 cleared on 25-Aug-2000:
It is also substantially equivalent to the commercially available Lympha Press® (of Mego-Afek), 510(k) number K810338 cleared on 20-Oct-1981.
## 8. Device Description
C-Boot Ltd. introduces a wearable pneumatic compression technology embodied in a therapeutic boot that has the potential to increase treatment efficacy and improve the quality of life for peripheral vascular and lymph distorders patients.
C-Boot's solution integrates the technological advantages of dynamic compression pumps, that enable preset sequential pressure control to individualize treatment regimes, with the mobility benefit of static compression stockings, that activate the calf muscle pump and strengthened the body's natural tendency to enhance lymphatic and venous return.
The device relates to a self-powered compression devices and methods for promoting circulation and therapeutic compression: It is comprising a pump, a plurality of sleeves or limb massager, which utilizes sequential, cyclical pressure, energy or vector forces, to aid circulation in a body part such as the limbs. The inflatable sleeves are filled with air by the pump and transmit massaging movement to the vessel walls whenever the patient wearing the device - walks or transmits body's weight from one leg to another. The pressure generated is a function of the user's body weight and gravitational force. The device provides methods to harness energy generated during movement and weight transfer and uses the energy for compressing to create a massaging effect on the circulation and the muscular-skeletal system.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows a logo with the letters "DDT" to the right of a circular shape. The circular shape is bisected vertically, with the left half being a square and the right half being an arrow pointing to the right. The logo is black and white.
## 9. Indications for Use
The C-Boot™ is a prescriptive device that induces controlled compression of the calf, the foot or a combined compression of both.
The C-Boot™ is intended to assist patients, suffering from lymphatic or venous disorders, by treating many conditions including:
- ਕ. Prevention of deep vein thrombosis (DVT)
- Enhancement of blood circulation b.
- Reduction of post-operative pain and swelling C.
- d. Reduction of wound-healing time
- e. Stasis dermatitis
- f. Treatment and assist healing of cutaneous ulceration
- g . Venous stasis ulcers
- Leg ulcers h.
- Chronic venous insufficiencies i.
- i. Reduction of edema
- Prevent pooling of fluids in limbs k.
- -Lymphedema
## 10. Contraindications
The C-Boot™ should not be used in the following cases: Advanced peripheral arterial occlusive diseases (calves pain while walking), Decompensated heart disease, Specific phlebitis, Phlegmasia coerulea dolens (swollen blue limb), Diabetic patients with lack of reduced sensation in lower limb or foot, Supportive dermatosis, Intolerance of C-Boot™ fabric, sensory disturbances of the limb, Advance peripheral neuropathy, Primary chronic arthritis.
{3}------------------------------------------------
## 11. Technological Characteristics and Substantial Equivalence
The C-Boot™ has combined indications of its two (2) predicate devices - WizAir™ and Lympha Press®- into one product. All three (3) devices have same shape of surrounding sleeves, inflation mode (sequential), mode of Compression (intermittent), number of Cycles (1-2 per minute) and are made all of basically same materials.
The main differences are the dynamic compression pumps, the overlapping-massaging-movements and the mechanical power source of C-Boot™ (compared to an electrical source of the other predicate devices), which raise neither safety nor effectiveness questions, as C-Boot™ functions within same boundaries of performances.
## 12. Performance Data
A series of functional, performance and safety testing, including comparative analysis to Mego-Afek® and WizAir™ - demonstrates that C-Boot™ is substantially equivalent to Mego-Afek® and WizAir™, without raising new safety and/or effectiveness issues.
{4}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Image /page/4/Picture/2 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular with text around the perimeter. In the center of the seal is a stylized image of an eagle.
OCT 2 8 2004
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
C-Boot Ltd. c/o Mr. Avi Abraham Avnet Engineering 8 Marva Street Carmiel 21691, Israel
Re: K041659
C-Boot™ Regulation Number: 21 CFR 870.5800 Regulation Name: Compressible Limb Sleeve Regulatory Class: Class II (two) Product Code: JOW Dated: August 12, 2004 Received: August 16, 2004
Dear Mr. Abraham:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for are waver to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Mr. Avi Abraham
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
ourna R. la chrien
( Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows a logo with the letters "BDDT" in a stylized font. The letters are arranged horizontally, with the first letter "B" being the largest and partially obscured by a semi-circular shape on the left. The semi-circular shape is filled with a checkered pattern. The letters "DDT" are smaller and appear to be in a bold, sans-serif font.
# 10. Indications for Use Statement
### 510(k) Number (if known): K041659
#### Device Name: C-Boot™
#### Indications for Use:
The C-Boot™ is a prescriptive device that induces controlled compression of the calf, the foot or a combined compression of both.
The C-Boot™ is intended to assist patients, suffering from lymphatic or venous disorders, by treating many conditions including:
- Prevention of deep vein thrombosis (DVT) 1.
- Enhancement of blood circulation 2.
- Reduction of post-operative pain and swelling 3.
- Reduction of wound-healing time 4.
- క్. Stasis dermatitis
- Treatment and assist healing of cutaneous ulceration 6.
- Venous stasis uicers 7.
- Leg ulcers 8.
- Chronic venous insufficiencies ு.
- 10. Reduction of edema
- 11. Prevent pooling of fluids in limbs
- 12. Lymphedema
_____ Prescription Use - لا__ AND/OR (Per 21 CFR 801 Subpart D)
Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE) Mina L. Value (Division Sign-Off) Division of Cardiovascular Devices 510(k) Number_ko41659 .............................................................................................................................................................................. Page 1 of 1 Lymph and Venous Insufficiency System
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.