← Product Code [JWH](/productcode/JWH) · K041591

# SCORPIO-FLEX POSTERIOR STABILIZED TIBIAL INSERT COMPONENTS (K041591)

_Howmedica Osteonics Corp. · JWH · Sep 23, 2004 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K041591

## Device Facts

- **Applicant:** Howmedica Osteonics Corp.
- **Product Code:** [JWH](/productcode/JWH.md)
- **Decision Date:** Sep 23, 2004
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3560
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Scorpio -Flex Posterior Stabilized Tibial Insert components are for use in total knee arthroplasty for painful, disabling joint disease of the knee resulting from degenerative arthritis, rheumatoid arthritis or post-traumatic arthritis; post-traumatic loss of knee joint configuration and function; moderate varus, valgus or flexion deformity in with the ligamentous structures can be returned to adequate function and stability; and/or revision of previous unsuccessful knee replacement or other procedure, ligamentous instability requiring implant bearing surface geometries with increased constraint; and absent or non-functioning posterior cruciate ligament.

## Device Story

Scorpio-Flex Posterior Stabilized Tibial Insert is a component of the Scorpio Total Knee System; redesigned to provide improved hyperextension. Device is a semi-constrained, cemented prosthesis used in total knee arthroplasty. Operates as a mechanical bearing surface to replace damaged knee joint components. Used in clinical/surgical settings by orthopedic surgeons. Output is a physical implant providing joint stability and range of motion. Benefits include restoration of knee function and pain relief for patients with degenerative or post-traumatic joint disease.

## Clinical Evidence

Bench testing only. No clinical data provided. Equivalence established through mechanical testing including range of constraint, post stress, and contact stress/area analysis.

## Technological Characteristics

Semi-constrained, cemented knee prosthesis. Materials: polymer/metal/polymer. Design modification focused on tibial insert geometry to improve hyperextension. Class II device per 21 CFR 888.3560.

## Regulatory Identification

A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).

## Predicate Devices

- Scorpio Total Knee System

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K 041/591
Special 510(k) Premarket Notification

# Special 510(k) Summary of Safety and Effectiveness: Design Modification to the Scorpio -Flex Posterior Stabilized Tibial Insert Components

| Proprietary Name:          | Scorpio®-Flex Posterior Stabilized Tibial Insert Components                                                                                                                                                    |
|----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Proposed Regulatory Class: | Class II                                                                                                                                                                                                       |
| Classification:            | Knee Joint; Patellofemorotibial; Polymer/metal/polymer; Semi-<br>constrained; Cemented prosthesis: 21 CFR 888.3560                                                                                             |
| Device Product Code:       | 87 JWH                                                                                                                                                                                                         |
| For Information contact:   | Karen Ariemma<br>Regulatory Affairs Specialist<br>Howmedica Osteonics Corp.<br>325 Corporate Drive<br>Mahwah, NJ 07430<br>Telephone: (201) 831-5718<br>Fax: (201) 831-6038<br>Email: karen.ariemma@stryker.com |
| Date Summary Prepared:     | June 9, 2004                                                                                                                                                                                                   |

#### Predicate Device Identification:

The Scorpio Total Knee System consists of various sizes of femoral, tibial and patellar components.

## Description of Device Modification:

This submission is intended to address a design modification to the Scorpio® Flex Posterior Stabilized Tibial Insert components. The tibial insert components have been redesigned to provide for improved hyperextension. There has been no change to the intended use of the device.

## Indications for Use:

The Scorpio -Flex Posterior Stabilized Tibial Insert components are for use in total knee arthroplasty for painful, disabling joint disease of the knee resulting from degenerative arthritis,

{1}------------------------------------------------

rheumatoid arthritis or post-traumatic arthritis; post-traumatic loss of knee joint configuration and function; moderate varus, valgus or flexion deformity in with the ligamentous structures can be returned to adequate function and stability; and/or revision of previous unsuccessful knee replacement or other procedure, ligamentous instability requiring implant bearing surface geometries with increased constraint; and absent or non-functioning posterior cruciate ligament.

## Statement of Technological Comparison:

The subject components share the same intended use and basic design concept as that of the predicate devices. A risk analysis, testing and analysis have been performed to demonstrate equivalence of the subject device to the predicate device. The testing includes range of constraint testing, analysis of tibial insert post stress and contact stress/area analysis. The results demonstrate equivalence.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Inside the circle is an abstract symbol that resembles an eagle or a bird in flight, composed of three curved lines.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

SEP 2 3 2004

Ms. Karen Ariemma Regulatory Affairs Specialist Howmedica Osteonics Corp. 325 Corporate Drive Mahwah, New Jersey 07430

Re: K041591

Trade/Device Name: Scorpio -Flex Posterior Stabilized Tibial Inserts Regulation Number: 21 CFR 888.3560 Regulation Name: Knee joint patellofemorotibial polymer semi-constrained cemented prosthesis Regulatory Class: II Product Code: JWH Dated: August 25, 2004 Received: August 26, 2004

Dear Ms. Ariemma:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set

{3}------------------------------------------------

Page 2 - Ms. Karen Ariemma

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,
Mark N Millenson

Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

510(k) Number (if known): K041159

Device Name: Scorpio®-Flex Posterior Stabilized Tibial Insert

#### Indications

- Painful, disabling joint disease of the knee resulting from degenerative arthritis, . rheumatoid arthritis or post-traumatic arthritis
- Post-traumatic loss of knee joint configuration and function .
- Moderate varus, valgus or flexion deformity in which the ligamentous structures can be . returned to adequate function and stability;
- . Revision of previous unsuccessful knee replacement or other procedure;
- . Ligamentous instability requiring implant bearing surface geometries with increased constraint;
- Absent or non-functioning posterior cruciate ligament. .

Prescription Use AND/OR (Part 21 CFR 801 Subpart D)

Over-The-Counter Use

(21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of 1

Munk N. Millerson

. Restorative, sion of Genera Neurological Devices

**510(k) Number** K041591

---

**Source:** [https://fda.innolitics.com/device/K041591](https://fda.innolitics.com/device/K041591)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
