PROFEMUR S HIP STEM

K041586 · Wrightmedicaltechnologyinc · JDL · Jul 9, 2004 · Orthopedic

Device Facts

Record IDK041586
Device NamePROFEMUR S HIP STEM
ApplicantWrightmedicaltechnologyinc
Product CodeJDL · Orthopedic
Decision DateJul 9, 2004
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3320
Device ClassClass 3
AttributesTherapeutic

Indications for Use

The PROFEMUR® S Hip Stem is indicated for use in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients with the following conditions: - non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis, ankylosis, protrusio acetabuli, and painful hip dysplasia; - inflammatory degenerative joint disease such as rheumatoid arthritis; - correction of functional deformity; and, - revision procedures where other treatments or devices have failed

Device Story

PROFEMUR® S Hip Stem is a modular orthopedic implant for total hip arthroplasty. Device features include two medial flare options, modular neck taper, triplanar proximal geometry with plasma spray, and a splined, slotted, cylindrical distal stem with grit blast. Includes lateral shoulder dimple with cross slot, plus polished distal tip and collar. Implanted by orthopedic surgeons in clinical settings to replace diseased hip joints. Provides structural support and joint articulation to reduce pain and restore function. Substantially equivalent to existing hip stem designs; relies on established orthopedic implant technology.

Clinical Evidence

No clinical data. Substantial equivalence is supported by design analysis and materials information.

Technological Characteristics

Modular hip stem featuring plasma-sprayed proximal geometry and grit-blasted distal stem. Includes modular neck taper, lateral shoulder dimple with cross slot, and polished distal tip/collar. Materials and design are consistent with established orthopedic hip prostheses.

Indications for Use

Indicated for skeletally mature patients requiring total hip arthroplasty for pain relief or improved function due to non-inflammatory degenerative joint disease (osteoarthritis, avascular necrosis, ankylosis, protrusio acetabuli, painful hip dysplasia), inflammatory degenerative joint disease (rheumatoid arthritis), functional deformity, or revision procedures.

Regulatory Classification

Identification

A hip joint metal/metal semi-constrained, with a cemented acetabular component, prosthesis is a two-part device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that consist of a femoral and an acetabular component, both made of alloys, such as cobalt-chromium-molybdenum. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # K041586 (pg 1 of 2) # 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS In accordance with the Food and Drug Administration Rule to implement provisions of the Safe Medical Devices Act of 1990 and in conformance with 21 CRF 807, this information serves as a Summary of Safety and Effectiveness for the use of the PROFEMUR® S Hip Stem. | Submitted By: | Wright Medical Technology, Inc. | |-------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Date: | July 6, 2004 | | Contact Person: | Jeanine H. Redden<br>Regulatory Affairs Specialist II | | Proprietary Name: | PROFEMUR® S Hip Stem | | Common Name: | Hip Stem | | Classification Name and Reference: | Prosthesis, hip, semi-constrained, metal/polymer, uncemented - Class II<br>21CFR 888.3350 Hip joint metal/polymer, semi- constrained, cemented prosthesis - Class II<br>Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented - Class II<br>21 CFR 888.3320 Hip joint metal/ metal semi- constrained, with a cemented acetabular component prosthesis - Class III<br>21 CFR 888.3330 Hip joint metal/ metal semi- constrained, with an uncemented acetabular component prosthesis - Class III | | Device Product Code and Panel Code: | Orthopedics/87/ LWJ, JDI, LZO, JDL, KWA | #### DEVICE INFORMATION #### A. INTENDED USE The PROFEMUR® S Hip Stem is indicated for use in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients with the following conditions: {1}------------------------------------------------ $$ \{\uparrow\downarrow C\uparrow\uparrow \in \mathcal{S}\not\subseteq\uparrow\} \quad \not\subseteq \uparrow \cup \uparrow \downarrow \uparrow $$ - non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis, 1. ankylosis, protrusio acetabuli, and painful hip dysplasia; - inflammatory degenerative joint disease such as rheumatoid arthritis; 2. - correction of functional deformity; and, 3. - revision procedures where other treatments or devices have failed 4. #### B. DEVICE DESCRIPTION The design features of the PROFEMUR® S Hip Stem are summarized below: - Offered in two medial flare options . - Modular neck taper . - Triplanar proximal geometry w/ plasma spray . - Splined, slotted, and cylindrical distal stem w/ grit blast . - Dimple with cross slot on lateral shoulder ● - Polished distal tip and collar . # C. SUBSTANTIAL EQUIVALENCE INFORMATION The indications for use and materials of the PROFEMUR® S Hip Stem are identical to the PERFECTA® RS and PROFEMUR® R Hip Stems. The design features of the PROFEMUR® S Hip Stem are substantially equivalent to the PERFECTA® RS, PROFEMUR® R, and PROFEMUR® Tapered Hip Stems. The fundamental scientific technology of the modified device has not changed relative to the predicate devices. The safety and effectiveness of the PROFEMUR® S Hip Stem are adequately supported by the substantial equivalence information, materials information, and analysis data provided within this Premarket Notification. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized human figure with outstretched arms. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 # JUL 0 9 2004 Ms. Jeanine H. Redden Regulatory Affairs Specialist II Wright Medical Technology, Inc. 5677 Airline Road Arlington, Tennessee 38002 Re: K041586 Trade/Device Name: PROFEMUR® S Hip Stem Trace Dovice Namer 21 CFR 888.3320, 21 CFR 888.3330, and 21 CFR 888.3350 Regulation Names: Hip joint metal/metal semi-constrained, with a cemented acetabular Frip Joint metal metal metal/metal semi-constrained, with an uncemented acetabular component, prosthesis; Hip joint metal/polymer semi-constrained cemented prosthesis; Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis; and Prosthesis, hip, semi-constrained, metal/polymer, uncemented Regulatory Class: III Product Codes: JDL, KWA, JDI, LZO and LWJ Dated: June 9, 2004 Received: June 14, 2004 Dear Ms. Redden: We have reviewed your Section 510(k) premarket notification of intent to market the device we never and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the encreations of the enactment date of the Medical Device Amendments, or to conimer of the to rialy 2011-11-11 in accordance with the provisions of the Federal Food, Drug, de noes that have been require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your de rice to such additional controls. Existing major regulations affecting your device can may or babyer to back frederal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean r that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must {3}------------------------------------------------ # Page 2 – Ms. Jeanine H. Redden comply with all the Act's requirements, including, but not limited to: registration and listing (21 Comply with an the 11th 3 requirements, 01); good manufacturing practice requirements as set CFK Fat 807), mocing (21 CFR Part 820); good was and if applicable, the electronic forth in the quality byticins (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) This icter will anow yourse over finding of substantial equivalence of your device to a legally premarket noticated.or. The PDF mising of cation for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific acrepliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small other general informational and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Lila Witt Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Indications for Use 510(k) Number (if known): Device Name: PROFEMUR® S Hip Stem Indications For Use: The PROFEMUR® S Hip Stem is indicated for use in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients with the following conditions: - 1. non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis, ankylosis, protrusio acetabuli, and painful hip dysplasia; - inflammatory degenerative joint disease such as rheumatoid arthritis; 2. - correction of functional deformity; and, 3. - 4. revision procedures where other treatments or devices have failed Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Division of General, Restorative, and Neurological Devices Page 1 of 1 510(k) Number_K041586 1
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...