OASIS DRY MOUTH RELIEF DISCS OR OASIS MOUTH MOISTURIZING DISCS

K041563 · Glaxosmithkline Consumer Healthcare, L.P. · LFD · Aug 27, 2004 · DE

Device Facts

Record IDK041563
Device NameOASIS DRY MOUTH RELIEF DISCS OR OASIS MOUTH MOISTURIZING DISCS
ApplicantGlaxosmithkline Consumer Healthcare, L.P.
Product CodeLFD · DE
Decision DateAug 27, 2004
DecisionSESE
Submission TypeTraditional
Device ClassClass U
AttributesTherapeutic

Indications for Use

Under the supervision of a healthcare professional, OASIS® dry mouth relief discs have been formulated for the relief of chronic and temporary xerostomia (dry mouth), which may be a result of disease such as Sjogren's Syndrome, oral inflammation, medication, chemo or radiotherapy, stess or aging. OTC labeling stipulates that OASIS® mouth moisturizing discs have been specially formulated for the relief of dry mouth symptoms such as difficulties in swallowing, speech, and changes in taste. These symptoms may be brought on by disease, stress, aging or medication.

Device Story

OASIS® is a dissolvable oral disc containing a lubricating polymer and flavoring agents. Designed for symptomatic relief of xerostomia (dry mouth). Used by patients in home or clinical settings; administered orally as needed. Device dissolves slowly in the mouth to provide moisture and lubrication, addressing symptoms like swallowing/speech difficulty and taste changes. Benefits include improved oral comfort and function for patients suffering from dry mouth due to disease, medication, or therapy.

Clinical Evidence

Bench testing only. Functional and performance evaluations were conducted to assess safety and effectiveness; results were satisfactory.

Technological Characteristics

Oral disc containing lubricating polymer and flavoring agents. Non-sterile. Ready-to-use. Formulated for slow dissolution in the oral cavity.

Indications for Use

Indicated for relief of chronic and temporary xerostomia (dry mouth) and associated symptoms (swallowing/speech difficulty, taste changes) in patients with conditions including Sjogren's Syndrome, oral inflammation, medication side effects, chemo/radiotherapy, stress, or aging.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K041563 # AUG 2 7 2004 #### H. 510(k) Summary - 1. Submission Applicant & Correspondent: | Name: | GlaxoSmithKline Consumer Healthcare, L.P. | | | |-----------------|----------------------------------------------------|--|--| | Address: | 1500 Littleton Road | | | | | Parsippany, NJ 07054-3884 | | | | Phone No .: | (973) 889-2566 | | | | Contact Person: | Anthony Amitrano, U.S. Director Regulatory Affairs | | | | | anthony.g.amitrano@gsk.com | | | | 2. Name of Device: | OASIS® | |-------------------------------|---------------------------------------------------------------------| | Trade/Proprietary/Model Name: | OASIS® Dry Mouth Relief Discs OR<br>OASIS® Mouth Moisturizing Discs | | | 3. Devices to Which New Device is Substantially Equivalent: | |---------------------------|-------------------------------------------------------------| | Gebauer Company: | Salivart cleared in 510(k) K981693 | | Sinclair Pharmaceuticals: | SST cleared in 510(k) K023046 | | Sinclair Pharmaceuticals: | Salinum or Oraclair cleared in 510(k) K024148 | - 4. Device Description: OASIS® is a disc that is allowed to dissolve slowly in the mouth. The product contains a lubricating polymer, with other ingredients to produce a pleasant flavored disc. The discs are presented in containers of various counts. {1}------------------------------------------------ 5. Intended Use of the Device: ### Rx: Under the supervision of a healthcare professional, OASIS® dry mouth relief discs have been formulated for the relief of chronic and temporary xerostomia (dry mouth), which may be a result of disease such as Sjogren's Syndrome, oral inflammation, medication, chemo or radiotherapy, stess or aging. #### Over-the-counter: OTC labeling stipulates that OASIS® mouth moisturizing discs have been specially formulated for the relief of dry mouth symptoms such as difficulties in swallowing, speech, and changes in taste. These symptoms may be brought on by disease, stress, aging or medication. - 6. Summary of the Technological Characteristics of the Device Compared to the Predicate Devices: OASIS® has the same intended use/indications as the predicate devices Gebauer Company Salivart, Sinclair Pharmaceuticals SST, and Sinclair Pharmaceuticals Salinum or Oraclair and is described in the following table. {2}------------------------------------------------ | Product Name | OASIS | Salivart | SST | Salinum or<br>Oraclair | |---------------------------------|-------------------------------------------|-------------------------------------------|--------------------------------------------------|-------------------------------------------| | Method of Use | Ready to use | Ready to use | Ready to use | Ready to use | | # of<br>Applications<br>per Day | Take as<br>needed | Take as<br>needed | Take as<br>needed up to<br>16 tablets per<br>day | Take as<br>needed | | Claim | Symptomatic<br>treatment of<br>xerostomia | Symptomatic<br>treatment of<br>xerostomia | Symptomatic<br>treatment of<br>xerostomia | Symptomatic<br>treatment of<br>xerostomia | | Area of Use | Oral cavity | Oral cavity | Oral cavity | Oral cavity | | Disease State | Xerostomia | Xerostomia | Xerostomia | Xerostomia | | Type of<br>Product | Disc | Solution | Tablet | Solution | | Presentation | Non sterile | Non sterile | Non sterile | Non sterile | ## 7. Tests and Conclusions: Functional and performance evaluation has been conducted to assess the safety and effectiveness of OASIS® discs. All results are satisfactory. {3}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three bars representing the human services aspect. The eagle is positioned to the right of the circular text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 AUG 2 7 2004 Mr. Anthony G. Amitrano Director, U.S. Regulatory Affairs GlaxoSmithKline Consumer Healthcare, L.P. 1500 Littleton Road Parsippany, New Jersey 07054-3884 Re: K041563 K041505 Trade/Device Name: OASIS® Dry Mouth Relief Discs OR Mouth Moisturizing Discs Regulation Number: Unclassified Regulation Name: None Regulatory Class: None Product Code: LFD Dated: June 9, 2004 Received: June 11, 2004 Dear Mr. Amitrano: We have reviewed your Section 510(k) premarket notification of intent to market the device We have reviewed your becases on and the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in indications for use stated in the ensiesary, to the enactment date of the Medical Device Interstate connineree prior to ria, 25, 20, 20, 20, 2017 11:42 pm accordance with the provisions of Amendinents, or to devroos that have a ct (Act) that do not require approval of a premarket the rederal Pood, Drug, and Coment , therefore, market the device, subject to the general approval application (1 Mr.). - The general controls provisions of the Act include Controls provisions of the rion of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III If your device is classified (boo as a 107.07) in a controls. Existing major regulations affecting (FMA), it may of subjoct to back adding the casing the 21, Parts 800 to 898. In your device can be found in the Seas nouncements concerning your device in the Federal Register. {4}------------------------------------------------ Page 2 -- Ms. Amitrano Please be advised that FDA's issuance of a substantial equivalence determination does not Please be advised that ITDA s Issualled of a backance complies with other requirements mean that FDA has made a deceminations administered by other Federal agencies. of the Act or any Federal statutes and regulations administered by other Federal status in of the Act of any rederal statues and reguirements . including, but not limited to: registration . You must comply with all the Act's requirements . . . . . . . . . . . . . . You must comply with an the Tec s requirements) in CFR Part 801); good manufacturing practice and listing (21 CFR Part 807), laboling (21 CFR Part 820), and if requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820), and if requirements as set forth in the quality Systems (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 5 10(k) I his letter will anow you to begin malibating , butantial equivalence of your device to a a nd the premarket notification. The PDA initing of substants of the results and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), If you desire specific advice for your de room at (301) 594-4613. Also, please note the regulation N please contact the Office of Comphanos as (50%) - 10% - 10% Part 807.97). You entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the may obtain other general miorination on your itsp. I spistance at its toll-free Division of Binan 2001 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ## Indications for Use 510(k) Number (if known): K041563 OASIS® Dry Mouth Relief Discs OR Mouth Moisturizing Discs Device Name: Indications For Use: Under the supervision of a healthcare professional, OASIS® dry mouth relief discs have been formulated for the relief of chronic and temporary xerostomia (drv mouth), which may be a result of diseases such as Sjogren's Syndrome, oral inflammation, medication, chemo or radiotherapy, stress or aging. OASIS® mouth moisturizing discs have been formulated for relief of dry mouth symptoms such as difficulties in swallowing, speech and changes in taste. These symptoms may be brought on by disease, stress, aging or medication. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Susan Runge Page 1 of (Division Sign-Off) (Division of Anesthesiology, General Hospital, Infection Control, Dental Devices 510(k) Number: K041561
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...