THROMBELASTOGRAPH (TEG) PLATELET MAPPING ASSAY

K041502 · Haemoscope Corp. · JOZ · Sep 1, 2004 · Hematology

Device Facts

Record IDK041502
Device NameTHROMBELASTOGRAPH (TEG) PLATELET MAPPING ASSAY
ApplicantHaemoscope Corp.
Product CodeJOZ · Hematology
Decision DateSep 1, 2004
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 864.5700
Device ClassClass 2

Indications for Use

The TEG® Platelet Mapping Assay is intended to assess platelet function in patients who have received platelet inhibiting drugs such as aspirin, clopidogrel, abciximab, tirofiban, or eptifibatide.

Device Story

TEG Platelet Mapping Assay consists of blood modifiers, ADP and AA platelet agonists, and ActivatorF; used with heparinized blood samples on Thrombelastograph (TEG) Hemostasis Analyzer. Device measures inhibition of platelet function. Used in clinical settings by professionals. Output provides assessment of platelet aggregation reduction in patients on anti-platelet therapy. Healthcare providers use output to monitor drug efficacy and patient coagulation status.

Clinical Evidence

Clinical studies included 500 samples from adults (18-85) comparing TEG to optical aggregometry (r=0.948). Sensitivity/specificity for clopidogrel and aspirin were evaluated in 34 subjects; clopidogrel sensitivity 80%, specificity 86%; aspirin sensitivity 100%, specificity 92% (p≤.001). Additional studies on 10 normal volunteers (aspirin) and 67 cardiac patients (clopidogrel) showed statistically significant (p<.001) reductions in MA values and increased % platelet inhibition post-dosing.

Technological Characteristics

System uses mechanical sensing of clot formation (TEG MA parameter). Components include two-column instrument, computer interface, and software. Consumables: disposable cups and pins. Reagents: ADP, arachidonic acid, Activator F. Connectivity: computer interface module. Calibration: electronic. Sterilization: not specified. Software: hazard analysis and development processes reviewed.

Indications for Use

Indicated for assessing platelet function in patients receiving platelet-inhibiting drugs (aspirin, clopidogrel, abciximab, tirofiban, eptifibatide).

Regulatory Classification

Identification

An automated platelet aggregation system is a device used to determine changes in platelet shape and platelet aggregation following the addition of an aggregating reagent to a platelet-rich plasma.

Predicate Devices

Submission Summary (Full Text)

{0} 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY DEVICE AND INSTRUMENT TEMPLATE A. 510(k) Number: K041502 B. Purpose for Submission: To change the name of a previously cleared analyzer, and to seek clearance for a new assay on that analyzer C. Analyte: Platelet aggregation D. Type of Test: Aggregation E. Applicant: Haemoscope Corporation F. Proprietary and Established Names: Thrombelastograph® (TEG®) Hemostasis Analyzer Thrombelastograph® Platelet Mapping Assay G. Regulatory Information: 1. Regulation section: 21 CFR 864.5700 2. Classification: Class II 3. Product Code: JOZ 4. Panel: 81 Hematology H. Intended Use: 1. Intended use(s): The TEG® Platelet Mapping Assay is intended to assess platelet function in patients who have received platelet inhibiting drugs such as aspirin, clopidogrel, abciximab, tirofiban, or eptifibatide. {1} Page 2 of 7 2. **Indication(s) for use:** The TEG Hemostasis System is a non-invasive diagnostic instrument designed to monitor and analyze the coagulation state of a blood sample in order to assist in the assessment of patient clinical hemostasis conditions. The TEG analyzer is indicated for use with adult patients where an evaluation of their blood coagulation properties is desired. Coagulation evaluations are commonly used to assess clinical conditions such as post-operative hemorrhage and d/or thrombosis during and following cardiovascular surgery, organ transplantation, trauma, and cardiology procedures. 3. **Special condition for use statement(s):** 4. **Special instrument Requirements:** For use with the Thromboelastograph® (TEG®) Hemostasis Analyzer (formally called Thromboelastograph® (TEG®) Coagulation Analyzer) I. **Device Description:** The TEG® Hemostasis System consists of a two-column TEG instrument, a computer interface module, and software. In addition, Haemoscope offers consumables needed to run samples on the/ted system, including disposable sample cups and pin, biological controls, sample modifiers, and pipetting and maintenance supplies. The TEG Platelet Mapping Assay consists of a set of platelet agonists- adenosine diphosphate (ADP) and arachidonic acid (AA). When used on a heparinized blood sample, ADP and AA together with Activator F, measure the inhibition of platelet function (TEG MA parameter) J. **Substantial Equivalence Information:** 1. **Predicate device name(s):** a. Chrono-log Corporation Optical Platelet Aggregation Systems (K771198, K830749, K940792) b. Chrono-log Corporation Platelet Aggregation Reagents (K850593, K922800) c. Haemoscope Corporation TEG 5000 Series Analyzer (K002177) 2. **Predicate K number(s):** {2} Page 3 of 7 3. Comparison with predicate: | Similarities | | | | | | --- | --- | --- | --- | --- | | Item | Device | Predicate a | Predicate b | Predicate c | | Intended Use | Assess platelet function in patients who have received platelet inhibiting drugs | Perform platelet aggregation tests is PRP | To confirm normal platelet function and to diagnose platelet dysfunctions | Provide quantitative and qualitative indication of the hemostasis state of whole blood, PRP, or PPP | | User Interface | User –selectable | Same | Same | Same | | Differences | | | | | | --- | --- | --- | --- | --- | | Item | Device | Predicate a | Predicate b | Predicate c | | Matrix | Whole blood | PRP | Whole blood | Whole blood, PRP, or PPP | | Platelet function measurement method | Uses the TEG MA parameter | Asses the increase in light transmittance of the PRP sample following addition of an agonist | Assess the increase in light transmittance of the PRP sample following addition of an agonist | | K. Standard/Guidance Document Referenced (if applicable): L. Test Principle: A heparinized patient sample is split. One aliquot containing only Activator F is run to yield a TEG MA that is an expression of fibrin only, $\mathrm{MA}_{\mathrm{FIBRIN}}$. Another aliquot measures platelets activated without thrombin and with the addition of ADP or AA plus Activator F. The ADP aliquot generates the TEG $\mathrm{MA}_{\mathrm{ADP}}$ value, while the AA aliquot yields the $\mathrm{MA}_{\mathrm{AA}}$ value. A standard kaolin-activated aliquot is analyzed to measure total possible platelet activation and yields TEG $\mathrm{MA}_{\mathrm{THROMBIN}}$. The presence of platelet-inhibiting drugs is reflected in a reduction in the TEG MA values, and platelet inhibition then is a derived percentage based on MA without activating agents, and is computed separately for ADP- and TxA2 receptor inhibition. Therefore, $\% \mathrm{MA}$ reduction = $$ 100 - \left[ \left\{ \mathrm{MA}_{\mathrm{P}} - \mathrm{MA}_{\mathrm{FIBRIN}} / \mathrm{MA}_{\mathrm{THROMBIN}} - \mathrm{MA}_{\mathrm{FIBRIN}} \right\}^{*100} \right] $$ {3} Page 4 of 7 # M. Performance Characteristics (if/when applicable): 1. Analytical performance: a. Precision/Reproducibility: Evaluated according to NCCLS EP 15-P. Eight replicates were run simultaneously, and the experiment was repeated four times. Within-run precision of the MA parameter was 9%, total precision was 10%. b. Linearity/assay reportable range: Linearity was assessed following NCCLS EP6-P guidelines. Different volumes of abciximab were added to blood samples from volunteers who were not on any drugs prior to testing. TEG MA parameter showed linearity from 0 μL to 3 μL of abciximab c. Traceability (controls, calibrators, or method): d. Detection limit: e. Analytical specificity: Interference factors assessment was carried out according to NCCLS EP7-P for thrombin. Results demonstrated that heparin concentrations from 4 IU/ml upward were sufficient to inhibit thrombin. The package insert cautions the user to use sufficient amounts of heparin in the blood sample to ensure that all the thrombin is suppressed (i.e. vacutainer tubes containing &gt; 14.5 IU heparin/ml of blood should be used in phlebotomy) and that the sample would be analyzed within two hours of blood draw. f. Assay cut-off: 2. Comparison studies: a. Method comparison with predicate device: 500 samples were obtained from adults between the ages of 18 and 85. The patient population included healthy volunteers, patients undergoing cardiopulmonary bypass surgery, and patients undergoing cardiac catheterization. Each sample was divided into two, one for TEG analysis, and one for optical aggregometry (Chrono-log). Both assays were conducted according to the manufacturers recommended instructions. Regression analysis was performed between the two data sets (r=0.948, r²=.899, y = 0 +1.2888x. {4} Page 5 of 7 b. Matrix comparison: 3. Clinical studies: a. Clinical sensitivity: Defined as the percentage of patients who are not on anti=platelet drugs and achieved a percentage of platelet inhibition ≤ 20%. 34 subjects were tested using the TEG Platelet Mapping Assay as labeled. Subjects were tested before and after clopidogrel or aspirin dosing. | Parameter | Clopidogrel | Aspirin | | --- | --- | --- | | Sensitivity | 80% | 100% | | X² | 14.275 | 25.108 | | DF | 1 | 1 | | p | ≤.001 | ≤.001 | b. Clinical specificity: Clinical specificity is defined as the percentage of patients who are on anti-platelet drugs and achieved a percentage of platelet inhibition &gt;20%. Clinical studies have shown that values below 20% reflect resistance to this class of drugs. | Parameter | Clopidogrel | Aspirin | | --- | --- | --- | | Specificity | 86% | 92% | | X² | 14.275 | 25.108 | | DF | 1 | 1 | | p | ≤.001 | ≤.001 | c. Other clinical supportive data (when a and b are not applicable): Aspirin Study TEG MA measurements were taken from 10 normal volunteers before and after a 325 mg aspirin dosing. Results show a statistically significant difference (p &lt; .001) between the mean MA values before and after receiving aspirin. Results showed that the MA mean in the presence of aspirin was reduced by 86.5% due to platelet inhibition by aspirin. The percent of platelet inhibition was also analyzed, and demonstrated a statistically significant difference (p &lt; .001) between the % inhibition before receiving aspirin and after receiving aspirin. Results show that the % platelet inhibition after aspirin dosing increased by 94.3%. {5} Page 6 of 7 # Clopidogrel study TEG MA measurements were taken from 67 cardiac patients both with and without 75 mg qd clopidogrel dosing. The group consisted of patients being treated for various cardiac conditions, who were either had or had not been given a loading dose of clopidogrel (300 mg) followed by daily dosing of 75 mg for various durations. Results show a statistically significant difference (p &lt; .001) between the mean MA values for patients with and without clopidogrel. Results showed that the MA mean in the presence of clopidogrel was reduced by 48.3% due to platelet inhibition by clopidogrel. 4. Clinical cut-off: 5. Expected values/Reference range: For the TEG Platelet Mapping Assay, a patient serves as a “subject-specific” reference. Reference values for % platelet inhibition are individualized, in that a patient serves as his own control, providing both the lower and upper limits of activation, against which his inhibition is measured. N. Instrument Name: Thromboelastograph® (TEG®) Hemostasis Analyzer (formally called Thromboelastograph® (TEG®) Coagulation Analyzer) O. System Descriptions: 1. Modes of Operation: Manual, and simultaneous 2. Software: FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types: Yes ☐ X ☐ or No ☐ 3. Sample Identification: Manual input or bar code 4. Specimen Sampling and Handling: Samples are manually mixed and loaded into TEG cup wells and loaded onto instrument for testing 5. Assay Types: 4 coagulation parameters (R, K, α, MA) 6. Reaction Types: Clotting, aggregation 7. Calibration: Electronic calibration 8. Quality Control: {6} Page 7 of 7 Mechanical and electronic function checks Calibration and calibration verification Level I and Level II biological controls P. Other Supportive Instrument Performance Characteristics Data Not Covered In The "L. Performance Characteristics" Section Of The SE Determination Decision Summary: Q. Conclusion: The submitted material in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...