VENOJECT LUER ADAPTER

K041358 · Terumo Europe N.V. · JKA · May 28, 2004 · Clinical Chemistry

Device Facts

Record IDK041358
Device NameVENOJECT LUER ADAPTER
ApplicantTerumo Europe N.V.
Product CodeJKA · Clinical Chemistry
Decision DateMay 28, 2004
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1675
Device ClassClass 2

Intended Use

The Terumo® Venoject Luer Adapter is a sterile, non-invasive device that permits blood specimen collection from hypodermic needles, blood vessel catheters, or blood collection systems/devices with luer fittings. When blood samples are to be obtained with a single venipuncture, the luer adapter is a conduit between the collection needle and the collection container.

Device Story

The Venoject Luer Adapter is a sterile, single-use manual device serving as a conduit for blood specimen collection. It consists of a stainless steel cannula joined to a screw connector with a male luer taper. The cannula is covered by a synthetic isoprene rubber tip that acts as a valve; when a collection container is pressed over the cannula, the tip retracts to allow blood flow; when removed, the tip extends to stop flow. The device is used in clinical settings by healthcare professionals to facilitate blood collection from needles or catheters into evacuated collection systems. It does not have direct patient contact. The device benefits patients by enabling secure, controlled blood sample collection during venipuncture.

Clinical Evidence

Bench testing only. No clinical data provided. Performance evaluated against established specifications. Biocompatibility screening tests performed with no adverse findings.

Technological Characteristics

Materials: stainless steel cannula, synthetic isoprene rubber tip. Principle: manual flow control via retractable rubber valve. Form factor: screw connector with male luer taper. Sterilization: ethylene oxide (EN 550, ISO 11135) to SAL 10^-6. Non-patient contacting.

Indications for Use

Indicated for blood specimen collection from hypodermic needles, blood vessel catheters, or luer-fitted blood collection systems. Intended for use as a conduit between a collection needle and collection container during single venipuncture procedures.

Regulatory Classification

Identification

A blood specimen collection device is a device intended for medical purposes to collect and to handle blood specimens and to separate serum from nonserum (cellular) components prior to further testing. This generic type device may include blood collection tubes, vials, systems, serum separators, blood collection trays, or vacuum sample tubes.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K041358 ### SUMMARY AND CERTIFICATION II 510(k) Summary as required by 807.92 Summary of Safety & Effectiveness Information - Proprietary Device Name II. 1. VENOJECT® Luer Adapter - Classification Name II.2. Blood Specimen Collection Device - Reason for Submission II.3. New Device #### II.4. Intended Use The Terumo Venoject Luer Adapter is a sterile, non-invasive device that permits blood specimen collection from hypodermic needles, blood vessel catheters, or blood collection systems/devices with luer fittings. When blood samples are to be obtained with a single venipuncture, the luer adapter is a conduit between the collection needle and the collection container. #### II. 2. Description The Terumo Venoject Lucr Adapter is a sterile, single use device consisting of a cannula joined to a screw connector which is connected to a male luer taper. The cannula is covered with a synthetic isoprene rubber tip for stopping blood flow. When blood is collected using an evacuated blood collecting system, the collection container is placed over the cannula, pushing the rubber tip back, allowing blood flow. When the collection container is removed, the rubber tip extends back over the cannula, stopping blood flow. The luer adapter has no direct patient contact. #### Substantial Equivalence II.6. The "Venoject Luer Adapter", manufactured by Terumo Europe N.V., submitted in this 510(k) file is substantially equivalent in intended use, description/specifications, technology/principles of operation, materials and performance to the cleared "Venoject Luer Adapter", manufactured by Terumo Corporation which is the subject of K983490. Both devices being Luer Adapters for blood specimen INTENDED USE: collection, are sterile and for single use, permitting blood specimen collection from hypodermic needles, blood vessel catheters, or blood collection systems/devices with luer fittings. For blood samples obtained by single venipuncture, the lucr adapter serves as a conduit between the needle and the collection container. {1}------------------------------------------------ | DESCRIPTION/ SPECIFICATIONS: | Both devices are made of a stainless steel cannula joined to a screw connector which is connected to a luer taper, and the cannula covered with a synthetic isoprene rubber tip.<br>Materials used for proposed device and predicate are the same. | |------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | PRINCIPLE OF TECHNOLOGY STATEMENT: | The Venoject Luer Adapter made by Terumo Europe N.V. and the Venoject Luer Adapter made by Terumo Corporation are both operated manually. | | PERFORMANCE: | The performance of the Venoject Luer Adapter has been evaluated and the device performs according to the specification. There are no significant differences in specifications and performance between proposed device and predicate. | #### II. 7. Additional Safety Information The sterility of the Venoject Lucr Adapter is assured by using a validated sterilization method qualified in accordance with EN 550: "Sterilization of Medical Devices: Validation and routine control of ethylene oxide" and ISO 11135: "Medical Devices: Validation and routine control of ethylene oxide sterilization" to a sterility assurance level (SAL) of 10° as required by EN 556-1: "Sterilization of Medical Devices -Requirements for medical devices to be designated "STERILE"Part - 1: Requirements for terminally sterilized medical devices". Because this device does not come in direct contact with the patient, only biocompatibility screening tests have been performed. The results of these tests gave no indication that additional biocompatibility testing was necessary. The expiration dating for the Venoject Luer Adapter has been established at 30 months. #### II.8. Conclusion The Venoject Luer Adapter manufactured by Terumo Europe N.V. and submitted in this 510(k) file is substantially equivalent in intended use, description, specifications, technology/principles of operation, materials and performance to the cleared Venoject Luer Adapter manufactured by Terumo Corporation which is the subject of K983490. Differences between the devices cited in this section do not raise any new issues of safety or effectiveness. {2}------------------------------------------------ #### Preparation info II.9. | Date prepared: | 05/2004 | |----------------|--------------------------------------------------------------| | Prepared by: | Mrs. M.J. Aerts - Manager Regulatory Affairs | | Prepared for: | TERUMO EUROPE N.V. | | | Researchpark Zone 2, Interleuvenlaan 40, 3001 Leuven BELGIUM | | Tel.: | 0032 / 16 / 381 353 | . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . {3}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Image /page/3/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure, with three curved lines forming its body and wings. The logo is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" in a circular arrangement. MAY 2 8 2004 Food and Drug Administration 2098 Gaither Road Rockville MD 20850 Mrs. M.J. Aerts Manager, Regulatory Affairs Terumo Europe N.V. RESEARCHPARK ZONE 2 HAASRODE Interleuvenlaan 40 B-3001 LEUVEN-BELGIE (BEGLIUM) k041358 Re: Trade/Device Name: Venoject® Luer Adapter Regulation Number: 21 CFR 862.1675 Regulation Name: Blood specimen collection device Regulatory Class: Class II Product Code: JKA Dated: May 19, 2004 Reccived: May 21, 2004 Dear Mrs. Aerts We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complics with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). {4}------------------------------------------------ ## Page 2 This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours, Seain M. Cooper, US. Div. Tean M. Cooper, MS, D.V.M. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # Use Indications for 510(k) Number (if known): 504 1358 Device Name: Venoject® Luer Adapter Indications for Use: The Terumo Venoject Luer Adapter is a sterile, non-invasive device that permits blood The Terunio Venoject Duer Natprof is a Beach () blood vessel catheters, or blood collection specifical concentify from trypoderines (1000) samples are to be obtained with a single systems/devices with fuer adapter is a conduit between the collection needle and the collection container. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) A. Roberts Cota **Division Sign-Off** Office of In Vitro Diagnostic Device Evaluation and Safety Page i of l
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...