← Product Code [ITI](/productcode/ITI) · K041341

# GOLDEN SPYDER, MODEL GS-300 (K041341)

_Golden Technologies, Inc. · ITI · Jun 3, 2004 · Physical Medicine · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K041341

## Device Facts

- **Applicant:** Golden Technologies, Inc.
- **Product Code:** [ITI](/productcode/ITI.md)
- **Decision Date:** Jun 3, 2004
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 890.3860
- **Device Class:** Class 2
- **Review Panel:** Physical Medicine
- **Attributes:** Therapeutic

## Indications for Use

The intended use of the Golden Spyder is to provide mobility to persons limited to a seated position that are capable of operating a powered wheelchair.

## Device Story

Six-wheeled battery-powered wheelchair; welded steel frame; center-drive configuration; four-post seat mounting system. Operated via joystick controller; powered by two 12-volt batteries; includes on/off-board charger. Max speed 5-mph; 15-20 mile range; 300-lb weight capacity. Used by patients with mobility limitations; provides independent mobility.

## Clinical Evidence

No clinical data.

## Technological Characteristics

Welded steel frame; six-wheel center-drive configuration; 12-volt battery power; joystick control. Tested per ANSI/RESNA WC/01, WC/02, WC/03, WC/05, WC/10, and WC/Vol. 2-1998 for stability, braking, dimensions, obstacle climbing, and electromagnetic compatibility.

## Regulatory Identification

A powered wheelchair is a battery-operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.

## Predicate Devices

- Alante ([K011153](/device/K011153.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

JUN 0 3 2004

K041341

Page 1 of 2

## 510 (K) Summary Golden Technologies, Inc. 510 (K) Premarket Notification Golden Spyder

Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared:

Golden Technologies, Inc. 401 Bridge Street Old Forge, Pa. 18518 Phone: (800) 624-6374 Facsimile: (570) 451-7494 Contact Person: Gene Kulon Official Correspondent Date Prepared: May 18, 2004

Name of Device and Name / Address of Sponsor:

Golden Spyder

Golden Technologies, Inc. 401 Bridge Street Old Forge, Pa. 18518 Phone: (800) 624-6374 Facsimile: (570) 451-7494

Common or Usual Name: Power Wheelchair

Classification Name: Wheelchair, Powered

Comparison to Predicate Devices:

The product, which is substantially equivalent to the Golden Spyder, is the Alante, (K011153). They are both controlled by the use of a joystick controller with onboard batteries and battery charger. All safety features are equivalent.

Device Description:

{1}------------------------------------------------

Page 2 of 2

The Golden Spyder is a six wheeled battery powered wheelchair. It has a Dynamics The Golden Opyder 18 a on which is used to operate the Golden Spyder. The one-Shark of the wheelchair is made of welded steel construction. The main frame plece base of the whoclehall is made a center drive wheels are mounted on the drive also consists of a battory noor passemblies, one on the left side of the wheelchair, the sharts of separate motorigoar assemment with a removable fold down seat, which is other on the nght date it would frame via a four-post mounting system. The seat connected, to the unve and neer the center wheels to aid in traction. All seating Concentrates the ascrowing to the Golden Spyder comes standard with two easily removable 12-volt batteries and an on or off board charger. It has an operating range of 15-20 miles on a full charge. The weight capacity is 300-lbs and it has a maximum speed of 5-mph.

Intended Use:

The intended use of the Golden Spyder is to provide mobility to persons limited to a seated position that are capable of operating a powered wheelchair.

Discussion of non-clinical tests performed for determinations of substantial equivalence are as follows:

ANSI/RESNA WC/01 1990 Determination of Static Stability Testing ANSI/RESNA WC/02 1991 Determination of Dynamic Stability Testing ANSI/RESNA WC/03 1990 Determination of the Effectiveness of brakes ANSI/RESNA WC/05 1990 Determination of overall Dimensions, Mass, and Turning Space ANSI/RESNA WC/10 1990 Determination of Obstacle Climbing Ability ANSI/RESNA WC/Vol. 2-1998 Requirements and Test Methods for Electromagnetic Compatibility of Electric Powered Wheelchairs and Scooters.

Discussion of Clinical Tests Performed:

N/A

Conclusions:

The Golden Spyder has the same intended use and similar technological characteristics as the Alante. Moreover, the non-clinical testing and the predicate comparisons demonstrate that any differences in their technological characteristics do not raise any new questions of safety or effectiveness. Thus, the Golden Spyder device is substantially equivalent to the predicate device.

Discussion of Clinical Tests Performed:

N/A

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle or bird-like symbol with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the symbol.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUN 0 3 2004

Gene R. Kulon Golden Technologies, Inc. 401 Bridge Street Old Forge, Pennsylvania 18518

Re: K041341

Trade/Device Name: Golden Spyder, Wheelchair, Powered Regulation Number: 21 CFR 890.3860 Regulation Name: Powered wheelchair Regulatory Class: II Product Code: ITI Dated: May 18, 2004 Received: May 20, 2004

Dear Mr. Kulon:

We have reviewed your Section 510(k) premarket notification of intent to market the indication we nave reviewed your become 310(x) premaince is substantially equivalent (for the indications felerenced above and nave determined by marketed predicate devices marketed in interstate for use stated in the encreate) to legans intentional date of the Medical Device Amendments, or to commerce prior to May 20, 1770, the exceence with the provisions of the Federal Food. Drug. devices that have been reclassified in accessfired in a premarket approval application (PMA). and Cosment Act (Act) that do not require approvise to the general controls provisions of the Act. The Act. The r ou may, mercetore, market the device, boojes requirements for annual registration, listing of general controls provisions of the fiel labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can may be subject to suell additional controllations, Title 21, Parts 800 to 898. In addition, FDA may be found in the Out occurents concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean Frease of advised that I Dri- Sissaanse or our device complies with other requirements of the Act that I DX has made a actoring administered by other Federal agencies. You must or any I edelar statutes and regulations anchuding, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set a CI K r at 807), laboling (21 OF X Part 820), and if applicable, the electronic forul in the quality of control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) This letter will anow you to ocgin makemig your avine equivalence of your device to a legally premarket notification. The PDA midnig of substance equive and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific advice for your ac 1100 %. Also, please note the regulation entitled, the more obtain Contact the Office of Comphanes at (301) 57 160 - 1007.97). You may obtain " Misorallumg by Icierchec to premantel international in the Act from the Division of Small other general information on your responsibility of the mumber (800) 638-2041 or
Manufacturers, International and Consumer Assistance at its toll-ifremation brought (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Miriam C. Provost

Celia M. Witten, Ph.D., M.D. P Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## Indications for use

510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________

Golden Spyder, Wheelchair, Powered Device Name:

Indications for Use:

The intended use of the Golden Spyder is to provide mobility for those persons limited The intended ass of the Select capable of operating a simple hand control.

Prescription Use (Per 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use V (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Miriam C. Provost

(Division Sign-Off) Division of General, Restorative, and Neurological Devices

**510(k) Number** K041341

---

**Source:** [https://fda.innolitics.com/device/K041341](https://fda.innolitics.com/device/K041341)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
