K041242 · Candela Corp. · GEX · Nov 22, 2004 · General, Plastic Surgery
Device Facts
Record ID
K041242
Device Name
CANDELA SMOOTHBEAM LASER SYSTEM
Applicant
Candela Corp.
Product Code
GEX · General, Plastic Surgery
Decision Date
Nov 22, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
For use in dermatology: incision, excision, ablation, and vaporization with hemostasis of soft tissue, Treatment of back acne, Treatment of atrophic acne scars, Treatment of facial wrinkles, Treatment of mild to moderate acne vulgaris, New indication : Treatment of Sebaceous Hyperplasia
Device Story
Candela Smoothbeam Laser System is a 1450 nm diode laser for dermatological procedures. System components include power supply, optical delivery system, software control system, and Dynamic Cooling Device (DCD). DCD delivers cryogen spray via handpiece nozzle to cool skin, minimizing thermal damage and pain during treatment. Device produces circular beams on skin. Operated by clinicians in a clinical setting; users select parameters via control panel. Safety interlocks protect patients and operators. Laser energy targets soft tissue for ablation/vaporization or acne/wrinkle treatment. Clinical benefit derived from controlled thermal effects on skin tissue.
Clinical Evidence
No new clinical trials were conducted for this submission. Safety and effectiveness for the expanded indication (sebaceous hyperplasia) are supported by published clinical data using the predicate Smoothbeam Laser, which demonstrated safety and efficacy for similar dermatological conditions.
Technological Characteristics
1450 nm diode laser; continuous wave; embedded processor control; Dynamic Cooling Device (cryogen spray); safety interlock systems; conforms to 21 CFR 1040 (Laser Performance Standard), UL 2601 (Electrical Safety), and EN 60601-1-2 (EMC).
Indications for Use
Indicated for dermatology patients requiring soft tissue incision, excision, ablation, or vaporization; treatment of back acne, atrophic acne scars, facial wrinkles, mild to moderate acne vulgaris, and sebaceous hyperplasia.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
Candela Smoothbeam 1450 nm Diode Laser System (K013825, K030834)
Submission Summary (Full Text)
{0}------------------------------------------------
K041242
## NOV 2 2 2004
#### General Information:
This 510(k) is to provide notification of substantial equivalence for the Candela Smoothbeam Laser System, I mis 10(x) is to provide notheaton of succements veal notended for use in the treatment of facial acne
| Submitted by: | Candela Corporation<br>530 Boston Post Road<br>Wayland, MA 01778-1886 |
|--------------------|---------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Lorraine Calzetta Patrovic |
| Date prepared: | April 23, 2004 |
| Classification: | Class II (21 CFR § 878.4810 Laser Surgical Instrument for use in General<br>and Plastic Surgery and in Dermatology) |
| Common Name: | Dermatology Laser, Smoothbeam Laser System |
| Predicate Devices: | Candela Smoothbeam 1450 nm Diode Laser System (K013825 ,K030834) |
#### Description:
The Diode laser is a Continuous Wave, diode medical laser, controlled by an embedded processor, to be used I Is Doc tase is a continued of wrinkles. The Candela Smoothbeam Laser System is comprised of a power supply, optical delivery system, software control system and Dynamic Cooling Device. The laser output power supply, optical defreery of claring cosmation of the produces circular beams on the skin. The Dynamic Clooling Device, provides a short burst of cryogen spray during the laser treatment. The cryogen is delivered via a hose to a nozzle located in the handpiece. The Dynamic Cooling Device functions to cool the skin during the a hose to a nozine risunizing thermal damage to skin during laser treatment and reducing pain associated with laser treatment.
The Candela Smoothbeam Laser System is equipped with safety interlock systems to protect patients and operations. IJsers of the device, make selections from a control panel to regulate operation during the laser treatment.
#### Testing:
As a laser product, the Smoothbeam Laser System is required to conform and does conform to the Laser Performance Standard (21 CFR 1040). In addition, the device will conform to the UL 2601 Electrical Safety Standard and with the Harmonized Standard EN 60601-1-2, Part 2 established by the European Community.
#### Safety and Effectiveness Information:
The modality of action for the treatment of sebaceous hyperplasia is similar to that of previously cleared indications in the treatment of back acne and the treatment of inflammatory acne vulgaris. Published clinical data, using t he predicate Smoothbeam Laser, cleared for use in the market, produced results that demonstrate that the Smoothbeam Laser is safe and effective for the expanded indication, - the treatment of sebaceous hyperplasia
### Summary of Substantial Equivalence:
The Candela Smoothbeam Laser System described in this submission is identical to the currently marketed Candela Smoothbeam Laser System based on operating principles, materials, mechanism of action, design, construction and intended use. Candela believes that the Smoothbeam Laser is substantially equivalent to the predicate device.
Image /page/0/Picture/14 description: The image shows the characters 'C', a colon, and the number '1'. The characters are written in a simple, slightly rough font. The colon is positioned between the 'C' and the '1', indicating a possible label or ratio.
{1}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized emblem featuring a symbol that resembles a bird or a stylized human figure with outstretched arms. The emblem is composed of three curved lines that create a sense of movement and dynamism.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV 2 2 2004
Ms. Lorraine Calzetta Patrovic Manager, Regulatory Affairs Candela Corporation 530 Boston Port Road Wayland, Massachusetts 01778-1886
Re: K041242
Trade/Device Name: Candela Smoothbeam Laser System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: GEX Dated: August 24, 2004 Received: August 26, 2004
Dear Ms. Patrovic:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate 10 rate exaction to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) rms reter notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
R. Mark A. Millmann
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known):
Device Name: Candela Smoothbeam Laser ( aka Candela MID IR Laser)
Indications For Use:
The Candela Smoothbeam Laser is indicated for the following uses :
For use in dermatology: incision, excision, ablation, and vaporization with hemostasis of soft tissue,
Treatment of back acne
Treatment of atrophic acne scars
Treatment of facial wrinkles
Treatment of mild to moderate acne vulgaris
## New indication : Treatment of Sebaceous Hyperplasia
Prescription Use _X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Mark N. Mellem
Page 1 of 1
Division of General, Restorative, and Neurological Devices
510(k) Number K041242
02:3
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.