ACCU-STAT HOME DRUG TEST CUP FOR MAIJUANNA (THC), COCAINE (COC), AMPHETAMINE (AMP), METHAMPHETAMINE (MAMP), ECSTASY

K041221 · Accu-Stat Diagnostics, Inc. · LDJ · Jul 13, 2004 · Clinical Toxicology

Device Facts

Record IDK041221
Device NameACCU-STAT HOME DRUG TEST CUP FOR MAIJUANNA (THC), COCAINE (COC), AMPHETAMINE (AMP), METHAMPHETAMINE (MAMP), ECSTASY
ApplicantAccu-Stat Diagnostics, Inc.
Product CodeLDJ · Clinical Toxicology
Decision DateJul 13, 2004
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.3870
Device ClassClass 2

Intended Use

The Accu-Stat™ Drugs of Abuse Home Test Cup for Marijuana (THC), Cocaine (COC), Amphetamine (AMP), Methamphetamine (mAMP), Ecstasy (MDMA), Opiates (OPI), and Phencyclidine (PCP), is a screening test for the rapid detection of three to seven of the above listed drugs in a variety of combinations human urine. The designated cut-off concentrations for these drugs are as follows: Marijuana at 50 ng/ml, Cocaine at 300 ng/ml, Amphetamine at 1000 ng/ml, Methamphetamine at 1000 ng/ml, Ecstasy at 500 ng/ml, Opiates at 2000 ng/ml, and Phencyclidine at 25 ng/ml. The tests are intended for over-the-counter (OTC) consumer use as the first step in a two-step process to provide consumers with information concerning the presence of the above stated drugs or their metabolites in a urine sample. Information, along with the materials for shipping a portion of the urine specimen to the laboratory for confirmation testing of a preliminary positive result, the second step in the process, is provided.

Device Story

Two-chamber plastic test cup; contains 3-5 lateral flow immunochromatographic test strips in isolated chamber. User introduces urine via plastic key; majority of specimen remains in main chamber for potential laboratory confirmation. Visually read by consumer; colored line indicates presence/absence of drug based on competitive binding principle. Procedural control line confirms proper sample volume and wicking. Intended for OTC home use as preliminary screen; includes materials for shipping urine to lab for confirmatory testing. Provides rapid detection of drugs at specified cut-off concentrations; clinical judgment required for interpretation of preliminary positive results.

Clinical Evidence

Consumer study performed with 619 unique tests using spiked urine samples (0%, 50%, 75%, 125%, 150%, 200% of target concentrations). Correct interpretation rates ranged from 97% to 99.5% across all drug types. Incorrect interpretations were primarily associated with samples near the cutoff (±25%).

Technological Characteristics

One-step rapid chromatographic immunoassay; competitive binding principle. Device consists of antibody-coated particles and immobilized drug conjugates. Form factor is an integrated test cup. No external energy source required; relies on capillary action. Qualitative visual readout.

Indications for Use

Indicated for OTC consumer use for rapid qualitative screening of 2-6 drugs (THC, COC, AMP, mAMP, MDMA, OPI, PCP) in human urine. Preliminary positive results require confirmatory testing via alternative chemical methods (e.g., GC/MS).

Regulatory Classification

Identification

A cannabinoid test system is a device intended to measure any of the cannabinoids, hallucinogenic compounds endogenous to marihuana, in serum, plasma, saliva, and urine. Cannabinoid compounds include delta-9-tetrahydrocannabinol, cannabidiol, cannabinol, and cannabichromene. Measurements obtained by this device are used in the diagnosis and treatment of cannabinoid use or abuse and in monitoring levels of cannabinoids during clinical investigational use.

Special Controls

*Classification.* Class II (special controls). A cannabinoid test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # KO41221 ## JUL 1 3 2004 #### Section D ### 510(k) Summary | As required by 21 CFR 807.92 ] #### I. Submitter: - A. Name: Accu-Stat Diagnostics, Inc. - B. Address: 11 Orchard Road, Suite # 108, Lake Forest, CA 92630 | C. Phone and Fax Numbers: | Phone | 949-916-7560 | |---------------------------|-------|--------------| | | Fax | 949-916-7595 | - D. Contact Person: James G. Barrons - Date of Preparation of this Summary: 6 May 2004 II. - III. Trade Name: Accu-Stat™ Drugs of Abuse Home Test Cup for Marijuana (THC), Cocaine (COC), Amphetamine (AMP), Methamphetamine (mAMP), Ecstasy (MDMA), Opiates (OPI), and Phencyclidine (PCP) - IV. Common Name: At home rapid screening test for Marijuana (THC), Cocaine (COC), Amphetamine (AMP), Methamphetamine (mAMP), Ecstasy (MDMA), Opiates (OPI), and Phencyclidine (PCP) in urine. - V. Classification Name: Immunoassay for the qualitative detection of drugs of abuse in urine. - VI. The Marketed Products to Which Equivalence is claimed: The Accu-Stat™Home Drug Test for Marijuana (THC), Cocaine (COC), Amphetamine (AMP), Methamphetamine (mAMP), Ecstasy (MDMA), Opiates (OPI), and Phencyclidine (PCP) is substantially equivalent to the Phamatech At Home™ Drug Test for Cup for Amphetamine, Methamphetamine (Ecstasy), Marijuana, Cocaine, and Opiates and other commercially available drug screening tests that qualitatively measure the presence of target drugs or metabolites by visual color one-step immunoassay technology. - VII. Statement of Intended Use Compared to Other Products: The intended use of the Accu-Stat™ Home Drug Test Cup for Marijuana, Cocaine, Amphetamine, Opiates, Methamphetamine, Ecstasy, and Phencyclidine is substantially equivalent to the above listed products and is a single unit screening test for the rapid detection of from three to seven of the listed drugs in human urine. The designated cut-off concentrations for these drugs are as follows: Marijuana at 50 ng/ml, Cocaine at 300 ng/ml, Amphetamine at 1000 ng/ml, Methamphetamine at 1000 ng/ml, Ecstasy at 500 ng/ml, Opiates at 2000 {1}------------------------------------------------ ng/ml, and Phencyclidine at 25 ng/ml. The tests are intended for over-thecounter (OTC) consumer use as the first step in a two-step process that includes confirmatory testing of preliminary positive results. Information, along with the materials for shipping a portion of the urine specimen to the laboratory is provided. - Discussion of Technological Characteristics: The Accu-Stat™ Drugs of VIII. Abuse Home Test Cup for Marijuana, Cocaine, Amphetamine, Methamphetamine, Ecstasy, Opiates, and Phencyclidine, like other commercially available drug screening tests, qualitatively measures the presence or absence of THC, COC, AMP, mAMP, MDMA, OPI, PCP and their metabolites in urine, using a one step, rapid chromatographic immunoassay which operates under the principle of competitive binding. Drugs. which may be present in the urine specimen, compete against the drug conjugate for binding sites on the antibody. During testing, a urine specimen migrates upward by capillary action. Marijuana, if present in the urine specimen below 50 ng/ml, and the other drugs being tested for, if below the cut-off levels stated in Attachment 3 Table 2, will not saturate the binding sites of the antibody coated particles in the test device. The antibody coated particles will then be captured by immobilized marijuana, cocaine, or other listed drug conjugate and a visible colored line will show up in the test line region. The colored line will not form in the test line region if the marijuana level is above the 50 ng/ml because it will saturate all the binding sites of antimarijuana antibodies. The same holds true for cocaine and the other drugs if the level is above the cut-off. It will saturate all the binding sites of anticocaine (or other drug) antibodies and therefore the colored line will not form in the test region. A drug-positive urine specimen will not generate a colored line in the test line region because of drug competition, while a drug-negative urine specimen will generate a line in the test region because of the absence of a drug competition. To serve as a procedural control, a colored line will always appear at the control line region if the test has been performed properly. Examples of predicate devices include the First Check® Home Drug Tests and the Phamatech At Home Drug Tests using a single or a multi-drug display. The consumer studies using the Accu-Stat™ Home Drug Test Cup for Marijuana, Cocaine, Amphetamine, Methamphetamine Ecstasy, Opiates. and Phencyclidine demonstrates that the test exhibits excellent overall performance in the hands of lay users. The data supports the conclusion that the consumer can use the Accu-Stat™ Home Drug Tests to obtain immediate, preliminary information regarding the possible use of THC, COC, AMP, mAMP, MDMA, OPI, and PCP. - Safety and Effectiveness: Because the Accu-Stat™ Home Drug Test Cup for IX. Marijuana (THC), Cocaine (COC), Amphetamine (AMP), Methamphetamine, {2}------------------------------------------------ (mAMP), MDMA, Opiates (OPI), and Phencyclidine (PCP) are identical to the ACON Laboratories One Step Multi-Drug Multi-Line Screen Test Card with Integrated Cup that is legally marketed under K031759 for professional use, and because no special skills, training, education, or licensing is required to collect a urine specimen and activate the test, there is no issue regarding the safety or effectiveness of the product to perform its intended function, i.e., to screen urine for the presence or absence of THC, COC, AMP, mAMP, MDMA, OPI, or PCP and their metabolite(s). Because the labeling of the Accu-Stat™ Home Drug Test Cup for Marijuana (THC), Cocaine (COC), Amphetamine (AMP), Methamphetamine (mAMP), Ecstasy (MDMA), Opiates (OPI), and Phencyclidine (PCP) is substantially equivalent to a variety of rapid screening tests currently in commercial distribution, including the Phamatech At Home™ Drug Test Cup, First Check® Home Drug Test, the Advantage Diagnostics Home Drug Tests, and there have been no reports of consumer inability to follow instructions or interpret results over the many months these products have been purchased, it should be concluded that the product can be used effectively by the lay user. {3}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines representing its body and wings. The logo is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged in a circular fashion around the emblem. Food and Drug Administration 2098 Gaither Road Rockville MD 20850 JUL 1 3 2004 Mr. James G. Barrons President Accu-Stat Diagnostics 11 Orchard Road Suite 108 Lake Forest, CA 92630 Re: k041221 Trade/Device Name: Accu-Stat™ Drugs of Abuse Home Test Cup for Marijuana (THC), Cocaine (COC), Amphetamine (AMP), Methamphetamine (mAMP), Ecstasy (MDMA), Opiates (OPI), and Phencyclidine (PCP) Regulation Number: 21 CFR 862.3870 Regulation Name: Cannabinoid test system Regulatory Class: Class II Product Code: MVO, LDJ, DIO, DKZ, LAF, DJG, LCM Dated: May 6, 2004 Received: May 10, 2004 Dear Mr. Barrons: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). {4}------------------------------------------------ #### Page 2 This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours, Jean M. Cooper MS, DVM. Jean M. Cooper, MS, D.V.M. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ #### Section A #### Indications for Use #### 510 (k) Number (if known): K041221 Device Name: Accu-Stat™ Drugs of Abuse Home Test Cup for Marijuana (THC), Cocaine (COC), Amphetamine (AMP), Methamphetamine (mAMP), Ecstasy (MDMA), Opiates (OPI), and Phencyclidine (PCP) #### Indications for Use: The Accu-Stat™ Drugs of Abuse Home Test Cup for Marijuana (THC), Cocaine (COC), Amphetamine (AMP), Methamphetamine (mAMP), Ecstasy (MDMA), Opiates (OPI), and Phencyclidine (PCP), is a screening test for the rapid detection of three to seven of the above listed drugs in a variety of combinations human urine. The designated cut-off concentrations for these drugs are as follows: Marijuana at 50 ng/ml, Cocaine at 300 ng/ml, Amphetamine at 1000 ng/ml, Methamphetamine at 1000 ng/ml, Ecstasy at 500 ng/ml, Opiates at 2000 ng/ml, and Phencyclidine at 25 ng/ml. The tests are intended for over-the-counter (OTC) consumer use as the first step in a two-step process to provide consumers with information concerning the presence of the above stated drugs or their metabolites in a urine sample. Information, along with the materials for shipping a portion of the urine specimen to the laboratory for confirmation testing of a preliminary positive result, the second step in the process, is provided. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The Counter Use X (21 CFR 807 Subpart C) ### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED.) Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) Division Sign-Off Office of In Vitro Diagno Device Evaluation an 510(k) k041122 Page 1 of 1
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...