ATRION EPIDURAL ANESTHESIA KIT WITH ATRION NEEDLEVISE SHARPS SECURING SYSTEM
K041158 · Atrion Medical Products, Inc. · CAZ · Jun 16, 2004 · Anesthesiology
Device Facts
Record ID
K041158
Device Name
ATRION EPIDURAL ANESTHESIA KIT WITH ATRION NEEDLEVISE SHARPS SECURING SYSTEM
Applicant
Atrion Medical Products, Inc.
Product Code
CAZ · Anesthesiology
Decision Date
Jun 16, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5140
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Atrion Epidural Anesthesia Kit is intended for use in the administration of anesthetic agents into the epidural space. It is recommended that the epidural catheter be removed or replaced every 72 hours. The NeedleVISE™ sharps securing system is intended for use in sterile anesthesiology procedure trays as a sharps securing device for the contaminated specialized needles that do not have integral engineered sharps safety devices.
Device Story
Atrion Epidural Anesthesia Kit provides components for repeated epidural anesthetic administration; includes NeedleVISE™ sharps securing system for needle disposal. NeedleVISE™ device, housed in sterile procedure tray, allows one-handed, upright securing of contaminated needles; locks used sharps into internal metal retaining clip within plastic housing. Used by healthcare professionals in sterile field; eliminates need for external sharps containers during procedure. Benefits include enhanced sharps control and safety for clinicians. Kit includes Povidone solution (regulated as drug).
Clinical Evidence
Bench testing only. Mechanical property testing performed on NeedleVISE™ component to assess impact, stability, puncture, and leak resistance. Clinical utility evaluated via field trials assessing ease of use.
Technological Characteristics
Kit contains standard anesthesia components and NeedleVISE™ sharps securing system. NeedleVISE™ consists of plastic housing with internal metal retaining clip for locking sharps. Device is disposable, single-patient use, and intended for sterile field. No specific ASTM standards cited. Non-electronic, mechanical device.
Indications for Use
Indicated for patients requiring administration of anesthetic agents into the epidural space. The NeedleVISE™ component is indicated for use by healthcare professionals in sterile anesthesiology procedure trays to secure contaminated specialized needles lacking integral safety devices.
Regulatory Classification
Identification
An anesthesia conduction kit is a device used to administer to a patient conduction, regional, or local anesthesia. The device may contain syringes, needles, and drugs.
Predicate Devices
Cambridge Marketing, Inc. Needle Barn Stik Stopper® (K010038)
Devon Industries, Inc. Point-Lok® Needle Protection Device (K946289)
B. Braun Epidural Kit with Soft Tip Catheter (K971233)
{1}------------------------------------------------
Atrion Medical Products, Inc. 1426 Curt Francis Road Post Office Box 564 Arab, AL 35016 Tel 256 586 1580 Fax 256 586 8529
K 041158
JUN 1 7 2004
Image /page/0/Picture/3 description: The image shows the logo for Atrion Medical. The logo is in black and features a stylized "A" with a swoosh extending from the left side. The word "Atrion" is written in bold, sans-serif font. Below the word "Atrion" is the word "Medical" in a smaller font.
#### 11. 510(k) SUMMARY
| Date of Preparation: | April 30, 2003 |
|----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name: | Atrion Epidural Anesthesia Kit with NeedleVISE™ sharps<br>securing system |
| Common Name: | Anesthesia Conduction Kit |
| Classification Name: | Anesthetic Conduction Kit 21 CFR 868.5140, ProCode 73 CAZ<br>Accessory to single lumen hypodermic needle<br>21 CFR 880.5570, ProCode 80 FMI |
| Manufacturer: | Atrion Medical Products, Inc.<br>1426 Curt Francis Road<br>Arab, AL 35016 |
| Contact: | Mr. Dan Clark,<br>Atrion Medical Products, Inc.<br>1426 Curt Francis Road<br>Arab, AL 35016<br>Telephone: (256) 586-1580, Fax: (256) 586-8529 |
| Predicate: | Cambridge Marketing, Inc. Needle Barn Stik Stopper® (K010038)<br>Devon Industries, Inc. Point-Lok® Needle Protection Device<br>(K946289)<br>B. Braun Epidural Kit with Soft Tip Catheter (K971233)<br>Portex Continuous Epidural Anesthesia Tray (K802065) |
#### Device Description:
The Atrion Epidural Anesthesia Kit with NeedleVISE™ Sharps Securing System consists of components necessary for conducting repeated administration of anesthetic agents in the epidural space. The kit also contains the NeedleVISE™ sharps securing system which is a opidural opade, single patient procedure use, disposable device used for securing the specialized needles found in sterile anesthesiology procedural trays.
As provided to the user, the NeedleV/SE™ sharps securing system is retained in an upright position in the sterile procedure tray to allow for stability and one-hand insertion of sharps during a procedure. The healthcare professional using only one hand can render the sharps harmless while remaining with the patient in the sterile field. The purpose of the NeedleVISE™ sharps while rounsvitem is to provide a means to make engineered sharps control protection available within the sterile field, for all instances in which introduction of contaminated sharps containers is not acceptable.
{2}------------------------------------------------
Atrion Medical Products, Inc. 1426 Curt Francis Road Post Office Box 564 Arab, AL 35016 Tel 256 586 1580 Fax 256 586 8529
Image /page/1/Picture/1 description: The image shows the logo for Atrion Medical. The logo consists of the word "Atrion" in a bold, sans-serif font, with a stylized line extending from the left side of the "A" to the upper left corner of the image. Below the word "Atrion" is the word "Medical" in a smaller, italicized font.
### Intended Use:
The Atrion Epidural Anesthesia Kit is intended for use in the administration of anesthetic agents into the epidural space. It is recommended that the epidural catheter be removed or replaced every 72 hours.
The NeedleVISE™ sharps securing system is intended for use in sterile anesthesiology procedure trays as a sharps securing device for the contaminated specialized needles that do not have integral engineered sharps safety devices.
### Technological Characteristics:
All the components used in the Atrion Epidural Anesthesia Kit with NeedleVISE™ Sharps Securing System are legally marketed devices. The inclusion of these components in the kit will have the same implications as their current usage. The NeedleVISE™ sharps securing system has similar materials as the predicate products, but acts by locking the used sharp into a metal retaining clip contained within the plastic housing.
### Summary of Testing:
As the only new component to the kit, testing was conducted on the NeedleVISE™ sharps securing system. The tests included mechanical property testing to assess impact, stability, puncture and leak resistance. The results showed the NeedleVISE™ device to be substantially equivalent to the predicate sharps securing devices. Clinical utility was evaluated in field trials designed to assess ease of use.
{3}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird with three stylized wing-like shapes.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUN 1 7 2004
Mr. Dan Clark Vice President Regulatory and Quality Atrion Medical Products, Incorporated 1426 Curt Francis Road Arab, Alabama 35016
Re: K041158
R041156
Atrion Epidural Anesthesia Kit with NeedleVISE™ sharps securing system Regulation Number: 868.5140 Regulation Name: Anesthesia Conduction Kit Regulatory Class: II Product Code: 73 CAZ Dated: April 30, 2004 Received: May 3, 2004
Dear Mr. Clark:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your becarement the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the encreativent of the enactment date of the Medical Device Amendments or to conninerec prof to May 20, 1978, the ecordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act). You may, therefore, market the device, subject to the general and Cosmette Feel (und Free), a same , you are responsible to determine that the medical devices controls provisions of rions of rion been determined as substantially equivalent under you use as contiponents in the mess (Section 510(k) of the act), or were on the market prior to the premarket non from on proves of the Medical Device Amendments. Please note: If you May 26, 1770, the onaoment and on bells (i.e., unfinished) and further process (e.g., sterilize) purchase your de ree componenee are including these components in your kit. The general you must submit a new by Act include requirements for annual registration, listing of devices, Controls provisions of the Ired labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it your device is elassifica nal controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean I lease oe advised that I Dr other mation that your device complies with other requirements of the Act that I DI has intatutes and regulations administered by other Federal agencies. You must
{4}------------------------------------------------
comply with all the Act's requirements, including, but not limited to: registration and listing (21 Comply with an the 11et 61ed by and 801); good manufacturing practice requirements as set CTN in the quality systems (21 CFR Part 820); and if applicable, the electronic form in the quality 355cms (Sections 531-542 of the Act); 21 CFR 1000-1050.
In addition, we have determined that your device kit contains Povidone Solution which is subject to regulation as a drug.
Our substantially equivalent determination does not apply to the drug component of your device. Our substantially equivalent determinenter for Drug Evaluation and Research before marketing we recommend your first commonent. For information on applicable Agency requirements for marketing this drug, we suggest you contact:
Director, Division of Drug Labeling Compliance (HFD-310) Center for Drug Evaluation and Research Food and Drug Administration 5600 Fishers Lane Rockville, Maryland 20857 (301) 594-0101
This letter will allow you to begin marketing your device as described in your Section 510(k) I this icher will anow you to begin mailing of substantial equivalence of your device to a legally premarket notification: "The results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation, please contact the Office If you desire specific advise for your do, please note the regulation entitled, "Misbranding by of Comphanee to premarket notification'(21 CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll free number (800) 638-2041 or (301) 443-6597, or at its Internet address http://www.fda.gov/dsma/dsmamain.html.
Sincerely yours,
Chiu Lin, Ph.D.
Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{5}------------------------------------------------
## 12. INDICATIONS FOR USE STATEMENT
# Indications for Use
510(k) Number (if known): Ko YUSS ___
Device Name: Atrion Epidural Anesthesia Kit with Atrion NeedleVISE™ sharps securing system
Indications For Use:
The Atrion Epidural Anesthesia Kit is intended for use in the administration of anesthelic agents of The Athon Epidural Anesthesia Nr is intendoulter as the as a looks as a removed or replace every 72 hours.
The NeedleVISE™ sharps securing system is intended for use in sterile anesthesiology The NeedleVISE™ Sharps Securing System is mondonation device for the contaminated procedure thays within the Stenle "hold" as "a sheered sharps safety devices.
Prescription Use (Per 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Lilia M. Pinto
on Sign-Off vision of Anesthesiology, General Hospit. Infection Control, Dental
510(k) Number:
Page 1 of
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.