EPI-MAX

K041059 · Dermagenics, Inc. · FRO · May 17, 2005 · SU

Device Facts

Record IDK041059
Device NameEPI-MAX
ApplicantDermagenics, Inc.
Product CodeFRO · SU
Decision DateMay 17, 2005
DecisionSESE
Submission TypeTraditional
Device ClassClass U
AttributesTherapeutic

Indications for Use

Epi-Max is indicated for use as a primary contact layer in dressing wounds such as lacerations, skin graft recipient sites, newly sutured wounds, abrasions, and minor or partial thickness burns. It may also be used as an initial layer in dressing surgical wounds with light exudates where mild medication and deoderization are desired. For use as a wound dressing to manage pressure ulcers (stages I-IV), stasis ulcers, diabetic skin ulcers, skin irritations, cuts, and abrasions.

Device Story

Epi-Max is a sterile, non-adhesive wound dressing consisting of a cellulose acetate mesh impregnated with a petrolatum-based ointment. The ointment contains polyethylene glycols (PEG 400, PEG 4000), citric acid, a proprietary saline solution, and benzoic acid. Designed to melt at skin temperature, the ointment maintains a moist wound environment and prevents adherence to the wound bed, facilitating less painful dressing removal. The device is applied by clinicians or patients as a primary contact layer for various wound types. It serves to protect the wound and manage exudate. No active electronic components or software are involved.

Clinical Evidence

No clinical data. Bench testing only, including GLP studies for cytotoxicity, irritation, and sensitization per ISO 10993 standards.

Technological Characteristics

Cellulose acetate mesh dressing impregnated with 20mg of PEG-based ointment (PEG 400, PEG 4000, citric acid, saline, benzoic acid). Melts at skin temperature. Gamma irradiated for sterility. Dimensions: 2 x 3 in2 (variable).

Indications for Use

Indicated for patients with pressure ulcers (stages I-IV), stasis ulcers, diabetic skin ulcers, skin irritations, cuts, abrasions, lacerations, skin graft recipient sites, newly sutured wounds, minor/partial thickness burns, and surgical wounds with light exudates.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ CONFIDENTIAL INFORMATION Koy/1059 # 510(k) Summary Applicant Name Dermagenics, Inc. Steve Monroe, Ph.D. 3251 Poplar Avenue, Suite 150 Memphis. TN 38111 Phone: (901) 452-2395 Fax: (901) 452-3885 Date Prepared April 22, 2004 Trade Name Epi-Max® Common Name Sterile, non-adhesive wound dressing Classification Name Dressing (General and Plastic Surgery) #### Substantial Equivalence | Predicate Device Name | 510(k) Number | |-----------------------------------------------------------|---------------| | Johnson & Johnson Petrolatum Gauze Non Adhesive Dressings | K862163 | | Kendall Xeroform Petrolatum Gauze | K973507 | | CURASALT Hypertonic Saline Dressing | K942459 | | Kerlix Zinc Saline Dressing | K944468 | #### Device Description Epi-Max is a cellulose acetate mesh wound dressing (initially in a 2 x 3 in2 size but later with smaller or larger sizes as per market demand) impregnated with 20mg of a petrolatum (polyethylene glycol-based) ointment. This ointment is a blend of PEG 400, PEG 4000, citric acid, a mild proprietary saline solution, and benzoic acid. The ointment is formulated to melt at skin temperature, helping to keep the wound moist and prevent wound-bandage adherence in order to make removal of the wound dressing less painful. Dressings are gamma irradiated for sterility. ## Indications for Use Epi-Max is indicated for use as a primary contact layer in dressing wounds such as lacerations, skin graft recipient sites, newly sutured wounds, abrasions, and minor or partial thickness burns. It may also be used as an initial layer in dressing surgical wounds with light exudates where mild medication and deoderization are desired. DERMAGENICS, INC. 4 5/17/05 168 {1}------------------------------------------------ #### Predicate Devices Epi-Max is very similar to numerous other non-adhesive petrolatum gauze dressings and wet saline dressings previously approved by the FDA, Instead of exclusively using either a petrolatum ointment or saline on the gauze, Epi-Max's ointment is a blend of USP-grade polyethylene glycols (PEGs) and a mild, proprietary saline solution buffered with USP-grade citric acid and with USP-grade benzoic acid added as a preservative. This blended ointment is designed to melt at skin temperature when applied to a wound, thus helping to promote a moist wound environment and making the dressing removal process easier and less painful. Epi-Max's indications for use are identical to the indications for use made by Kendall Healthcare Products, inc. for their Kendall Xeroform Petrolatum Gauze dressings. ### Non-clinical Performance Data Epi-Max has been found to be non-cytotoxic, not a dermal irritant, and has a 0% sensitization rating using GLP studies based upon guidelines presented in the 10993 ISO Standard. Clinical Performance Data N/A Non-clinical and Clinical Conclusions Epi-Max is safe for human use. # DERMAGENICS, INC. Image /page/1/Picture/12 description: The image contains a single digit, the number 5. The number is written in a simple, sans-serif font. The digit is centered in the image and is the only element present. 160 {2}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image contains the text "Public Health Service" in bold, black font. The text is stacked on top of each other, with "Public Health" on the top line and "Service" on the bottom line. The text appears to be extracted from a document or sign. Image /page/2/Picture/2 description: The image is a black and white seal. The seal is circular and contains an image of a bird in the center. The bird is facing to the right and has three lines that make up its body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is written around the top of the circle and "INSTITUTE OF" is written around the bottom of the circle. MAY 1 7 2005 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Steve Monroe, Ph.D. Vice President - Technology Greystone Medical Group 3251 Poplar Avenue, Suite 150 Memphis, Tennessee 38111 Re: K041059 Trade/Device Name: Epi-Max Impregnated Wound Dressing Regulatory Class: Unclassified Product Code: FRO Dated: April 22, 2005 Received: April 22, 2005 Dear Dr. Monroe: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ Page 2 - Steve Monroe, Ph.D. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled , "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, M AD Miriam C. Provost, Ph.D. Acting Director Division of General, Restorative and Neurological Devices Office of Device Évaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ K0\$\psi\$1059 # Indications for Use 510(k) Number (if known): K041059 Device Name: Epi-Max Impregnated Wound Dressing Indications For Use: For use as a wound dressing to manage pressure ulcers (stages I-IV), stasis ulcers, diabetic skin ulcers, skin irritations, cuts, and abrasions. 、 Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Restorative Logical Devices Page I of _
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...