K040989 · Rita Medical Systems · GEI · Apr 28, 2004 · General, Plastic Surgery
Device Facts
Record ID
K040989
Device Name
RITA SYSTEMS
Applicant
Rita Medical Systems
Product Code
GEI · General, Plastic Surgery
Decision Date
Apr 28, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
The RITA® System (RF generator and electrosurgical devices) supplies energy for use in electrosurgery and is indicated for use in percutaneous, laparoscopic, or intraoperative coagulation and ablation of soft tissue, including the partial or complete ablation of non-resectable liver lesions and the palliation of pain associated with metastatic lesions involving bone in patients who have failed or are not candidates for standard pain therapy.
Device Story
RITA System provides RF energy for soft tissue ablation/coagulation. Components: RF generator, deployable electrode device (with/without thermocouples), optional infusion pump, tubing kit, main cable, dispersive electrode, foot pedal. Used in percutaneous, laparoscopic, or intraoperative settings by clinicians. Generator monitors parameters (temperature, power, time); user sets ablation mode (constant power, constant temperature, infusion, track ablation). Infusion mode coordinates saline delivery via pump. System automatically stops energy/fluid delivery if parameters exceed limits or ablation completes. Real-time monitoring via LCD display. Benefits include targeted tissue destruction and pain palliation.
Clinical Evidence
Bench testing only. Evidence includes design verification, software validation, ablation performance testing, and sterilization validation.
Technological Characteristics
RF electrosurgical generator; deployable electrodes with optional thermocouples; saline infusion pump integration; RS-232 connectivity; LCD interface; foot pedal control; dispersive electrode return path; medical grade power supply.
Indications for Use
Indicated for percutaneous, laparoscopic, or intraoperative coagulation and ablation of soft tissue, specifically non-resectable liver lesions and palliation of pain from metastatic bone lesions in patients failing or ineligible for standard pain therapy.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
Predicate Devices
RITA Medical Systems Inc. - RITA System
Submission Summary (Full Text)
{0}------------------------------------------------
Page 1 of ②
## APR 2 8 2004
## 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS
COLL - ----------------------------------------------------------------------------------------------------------------------------
| Applicant | RITA Medical Systems, Inc.<br>967 N. Shoreline Blvd.<br>Mountain View, CA 94043 |
|--------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person | Darrin Uecker |
| Telephone Number: | (650) 314-3433 |
| Fax Number: | (650) 390-8505 |
| General<br>Provisions | Trade Name: RITA® System<br>Common/Classification Name: Electrosurgical cutting and coagulation device |
| Name of<br>Predicate | RITA Medical Systems Inc. - RITA System |
| Classification | Class II |
| Performance<br>Standards | Performance standards have not been established by the FDA under section 514 of<br>the Food, Drug and Cosmetic Act. |
| Intended Use | The RITA® System (RF generator and electrosurgical devices) supplies energy for<br>use in electrosurgery and is indicated for use in percutaneous, laparoscopic, or<br>intraoperative coagulation and ablation of soft tissue, including the partial or<br>complete ablation of non-resectable liver lesions and the palliation of pain<br>associated with metastatic lesions involving bone in patients who have failed or are<br>not candidates for standard pain therapy. |
| Device | The RITA System consists of the following components: |
| Description | RF Generator: Provides RF energy to the Device through the Main Cable. |
| | Disposable Electrode ("the Device"): Consists of a number of deployable<br>electrodes. Some or all are equipped with a thermocouple, depending on the<br>model of the Device. Some Devices are used with saline infusion ("Infusion<br>Devices") and require the use of a pump. |
| | Infusion Pump: When using the RF Generator with the Infusion Devices, an off-<br>the-shelf market-cleared infusion pump or a RITA Pump is used to deliver saline<br>through the Infusion Device during ablation. In the off-the-shelf pump, syringes<br>are connected to the Infusion Device and loaded into the Infusion Pump. In the<br>RITA Pump, a Tubing Kit is connected to the Infusion Device and loaded into the<br>RITA Pump. The RF Generator is connected to the Infusion Pump via an RS-232<br>cable. |
| | Tubing Kit: The tubing kit is designed to load into the pump head from the front<br>of the pump. On the proximal end of the tubing-set there is a PVC Bag Spike<br>fitting that allows the tubing to be attached to an IV bag. The Bag Spike fitting is<br>attached to a single length of tubing that is connected to a splitter fitting that splits<br>the single tubing into 5 separate tubes. The 5 tubes are held together with clips<br>that allow the tubes to be uniformly loaded into the pump head. The distal end on<br>the tubing-set has 5 female luers to attach to the Infusible devices. |
| | Main Cable: Connects the Device to the RF Generator. |
{1}------------------------------------------------
Page 2 of 2
Dispersive Electrode: Provides the return path for the RF energy applied by the Device. Depending on the model, the dispersive electrode may be equipped with thermocouples to measure skin temperature at the dispersive pad. Dispersive pads with temperature sensing capabilities are connected to the RF Generator at the dispersive electrode connection and at the Aux port.
Power Cord: A medical grade line cord that provides AC power to the RF Generator.
Foot Pedal: The Foot Pedal is connected to the front of the RF Generator. The Foot Pedal starts and stops the RF energy delivery when the RF Generator is in "READY" mode.
Device Description (cont.)
To use the system, the RF Generator is plugged into the wall outlet via the Power Cord. The electrosurgical device is connected to the RF Generator via the Main Cable. The Dispersive Electrode is placed on the appropriate location of the body and is connected to its port on the RF Generator. If a pump is used, it is connected to scrial port B of the RF Generator. Once the system is successfully powered up, the user can set the parameters of the ablation such as the mode of operation, the ablation time, the target temperature, and the power delivery level. One mode of operation, purge, is used to communicate with an infusion pump to prime the system. There are four modes of operation for RF energy delivery: constant power, constant temperature, infusion mode, and track ablation. In the infusion mode, the RF Generator communicates with the infusion pump to set and adjust the infusion With the electrosurgical device placed in the tissue to be ablated and its rate. electrodes deployed, RF power can be turned on. The system parameters are continuously monitored and displayed on the RF Generator. If the measured parameters arc outside the acceptable limits, the RF energy delivery and fluid delivery (if in infusion mode) automatically stops and a message appears on the liquid crystal display (LCD). The RF energy delivery and fluid delivery (if in infusion mode) also automatically ceases once the ablation is completed based on the initial user-defined parameters. RF energy can also be stopped at any point by pressing the RF on/off switch. Fluid delivery can also be stopped at any point by pressing the pump on/off switch.
Performance Data
The RITA System is subjected to design verification, software validation, ablation performance, and sterilization validation testing.
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle or bird-like figure with three stylized wing or feather shapes.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
'APR 2 8 2004
RITA Medical Systems, Inc. c/o Mr. Morten Christensen Underwriters Laboratories, Inc. 1655 Scott Boulevard Santa Clara, California 95050
Re: K040989
Trade/Device Name: RITA System Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: II Product Codc: GEI Dated: April 14, 2004 Received: April 16, 2004
Dear Mr. Christensen:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your be determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for two stated of to May 28, 1976, the enactment date of the Medical Device Amendments, or to conimered pror to they 20, 2011 11:47 accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The r ou may a news provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a dctermination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labcling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 - Mr. Morten Christensen
This letter will allow you to begin marketing your device as described in your Section 510(k) I mis letter will anow you to organization of substantial equivalence of your device to a legally premaince noticate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please if you desire specific at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small other general motimational and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Miriam C. Provost
Colia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
510(k) Number (if known): _ K 040989
Device Name:
Indications for Use:
The RITA® System (RF generator and electrosurgical devices) supplies energy for use in electrosurgery and is indicated for use in percutaneous, laparoscopic, or intraoperative coagulation and ablation of soft tissue, including:
- · the partial or complete ablation of non-resectable liver lesions and
- the palliation of pain associated with metastatic lesions involving bone in patients who have failed or are not candidates for standard pain therapy.
Prescription Use
(Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use _ (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Muriam C. Provost
Page I of I
(Division Sign-Off) Division of General, Restorative, and Neurological Devices
510(k) Number_ Ko40989
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.