# MODIFICATION TO ENCOMPASS (K040896)

_Heartlab, Inc. · LLZ · Jul 16, 2004 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K040896

## Device Facts

- **Applicant:** Heartlab, Inc.
- **Product Code:** [LLZ](/productcode/LLZ.md)
- **Decision Date:** Jul 16, 2004
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Software as a Medical Device

## Indications for Use

Encompass™ is a picture archiving and communications system intended to be used as a networked cardiovascular information management system. Encompass™ is software comprised of modular software programs that run on standard "off-the-shel?" personal coonyters and servers running the Windows 2000/XP operating system. Encompass™ is image data surage and display software that accepts DICOM (Digital Imaging and Communications in Medicine) data from laboratories, which support DICOM standard imaging transfer. The system provides the capability to; consolidate images generated by equipment from multiple OEM vendors, view images, enter clinical findings while viewing the associated images, perform digital subtraction, create graphical representation of coronary arteries, perform quantitative measurements on both cath and ultrasound images, perform quantitative analysis on cath images, generate and on voview patient reports with additional measurement and report writer capabilities and provides accessible digital image archive. Encompass™ is a scalable network system designed to service customers ranging in size from small departments (with 2 or 3 users) to large hospital networks (with tens of users).

## Device Story

Encompass is a networked cardiovascular information management system (PACS) running on standard Windows-based PCs/servers. It ingests DICOM-compliant images from various modalities (x-ray angiography, ultrasound, IVUS, CT, nuclear medicine, MRI). Software modules allow clinicians to consolidate, view, and manipulate images (brightness, contrast, zoom, digital subtraction, playback control). It enables quantitative measurements on cath/ultrasound images, graphical representation of coronary/vascular arteries, and optical character recognition (OCR) of echo measurements. Users enter clinical findings and generate reports. Used in clinical settings by physicians and medical staff to support diagnostic decision-making and patient data management. System is scalable from small departments to large hospital networks.

## Clinical Evidence

Bench testing only. Verification and validation activities were performed, including risk analysis, requirement reviews, design reviews, unit-level module verification, system integration testing, and final acceptance testing. Results demonstrated that predetermined acceptance criteria were met.

## Technological Characteristics

Software-based PACS running on standard off-the-shelf Windows 2000/XP PCs/servers. Connectivity via DICOM standard. Complies with 21 CFR 1020.10 (monitors), 21 CFR 1040.10 (fiber optics), and ACR/NEMA data compression standards. Supports JPEG compression (ISO/IEC 10918-1).

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- Model Impax ([K022292](/device/K022292.md))
- EchoNet ([K954860](/device/K954860.md))
- Inturis Suite ([K994210](/device/K994210.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# JUL 1 6 2004

# 510(K) Summary

This is a summary of 510(k) safety and effectiveness information is being submitted in accordance with the SMDA 1990 and 21 CFR 807.92.

#### DATE:

5 April 2004

#### SUBMITTER:

Heartlab Inc. One Crosswind Road Westerly, RI 02891 Phone: (401) 596-0592 Fax: (401) 596-8562

### CONTACT PERSON:

Richard Petrocelli Tel No: (401) 596-0592

## IDENTIFICATION OF THE PRODUCT

TRADE NAME: COMMON NAME: CLASSIFICATION NAME: Encompass™ Cardiac Network Image Processing System, LLZ

## SUBSTANTIAL EQUIVALENCE INFORMATION:

Encompass is considered comparable and substantially equivalent to the following predicate devices currently in commercial distribution:

Model Impax (k022292) EchoNet (k954860) Inturis Suite (k994210) Manufacturer Agfa Corp. Lockheed Martin Philips Medical Systems

{1}------------------------------------------------

#### DEVICE DESCRIPTION:

Encompass™ is a picture archiving and communications system intended to be used as a networked cardiovascular information management system. Encompass™ is software comprised of modular software programs that run on standard "off-the-shelf" personal computers and servers running the Windows 2000/XP operating system. Encompass™ is image data storage and display software that accepts DICOM (Digital Imaging and Communications in Medicine) data from laboratories, which support DICOM standard imaging transfer. The system provides the capability to; consolidate images generated by equipment from multiple OEM vendors, view images, enter clinical findings while viewing the associated images, perform digital subtraction, create graphical representation of coronary arteries, perform quantitative measurements on both cath and ultrasound images, perform quantitative analysis on cath images, generate and review patient reports with additional measurement and report writer capabilities and provides accessible digital image archive. Encompass™ is a scalable network system designed to service customers ranging in size from small departments (with 2 or 3 users) to large hospital networks (with tens of users). The original core functionality is detailed below:

- 1. Review of x-ray angiography, ultrasound, intravascular ultrasound, computed tomography (CT), nuclear medicine, and magnetic resonance imaging (MRI) images.
- 2. Compare images from different studies on one or two monitors, regardless of modality.
- 3. Perform report data entry and view the report.
- 4. Print, save as mpeg/avi, save as bitmap, and copy to clipboard any image(s).
- 5. Perform standard image processing such as brightness, contrast, gamma, sharpen, window/level, invert, pan, zoom, and digital subtraction.
- 6. Perform standard stop, start, single frame advance and reverse, previous/next image playback, and various standard clinical presentations.
- 7. Control the speed of playback.
- 8. Use a supported jog wheel to control playback.
- 9. Play/repeat a subset of a loop with user defined begin and endpoints.
- 10. Search sources for studies based on demographic information.
- 11. Copy any study or subset of study from any supported source to any supported destination.
- 12. Write DICOM study to CD/DVD for interchange.
- 13. Calibrate the monitor to a SMPTE pattern.

{2}------------------------------------------------

## DESCRIPTION OF CHANGE OR MODIFICATION:

Encompass™ has been modified to permit the following software functionality:

- 1. Create graphical representation of coronary and vascular arteries.
- 2. Perform quantitative measurements on clinical images, and store the results in the database.
- 3. Perform quantitative analysis on cath images and store the results in the database.
- 4. Perform optical character recognition of echo measurements from ultrasound images and store the results in the database.
- 5. Supports the HIPAA directives by allowing capabilities with regards to patient data security, access and privacy concerns.

#### INTENDED USE:

The intent of this device is to provide diagnostic quality image review, image archive, analysis and measurement capabilities, and findings report generation and report review capabilities. Encompass™ is indicated for use by the physician to aid in diagnosis, and by medical professionals whenever they would require or desire access to medical images and patient demographic information.

#### STANDARDS:

Encompass™ is designed in accordance with product safety and performance requirements set forth in the following standards:

- 1. Digital Imaging and Communications in Medicine (DICOM)
- 2. 21 CFR 1020.10 Video Monitor Performance Requirements
- 3. 21 CFR 1040.10 Fiber optic communications Performance
- 4. Society of Motion Picture and Television Engineers (SMPTE)
- 5. ACR/NEMA Data Compression Standard
- 6. Underwriters Laboratories (U.L.) Standard No. 544 for Medical and Dental Equipment
- 7. ISO/IEC 10918-1 Digital Compression and Coding Continuous-Tone Still Images (JPEG)

{3}------------------------------------------------

## SUMMARY OF DESIGN CONTROL ACTIVITIES:

The risk analysis used to assess the impact of the modifications was based upon the previous risk analysis and incorporated the software modifications indicated in this submission.

The following quality assurance design control measures were applied to the development of the Encompass™ product:

- 1. Risk Analysis
- 2. Requirement Reviews
- 3. Design Reviews
- Testing on unit level (Module verification) 4.
- 5. Integration testing (System verification)
- 6. Final acceptance testing (Validation)
- 7. Performance testing

## COMPARISON OF TECHNOLOGICAL CHARACTERISTICS

The modifications to the Encompass™ product do not affect the intended use or alter the fundamental scientific technology of the device. The only device modification was to software. There are no labeling changes that affect the intended use of the device. The device modifications raise no new issues of safety or effectiveness.

All products operate on commercially available computer systems.

All comply with DICOM standard for communications with other imaging system components.

## DECLARATION OF CONFORMITY (807.87(g) / 21 CFR 820.30)

All verification and validation activities were performed by the designated individual(s), and the results demonstrated that the predetermined acceptance criteria were met.

Daniel M. Kpoto
Signature

Daniel Reifsteck, VP of Engineering Operations Typed Name

## 4-5-2004

*Dated*

Dated

{4}------------------------------------------------

The manufacturing facility is in conformance with the design control procedure requirements as specified in 21 CFR 820.30, and the records are available for review.

Jois Katler

Signature

Lori Kahler, Quality Assurance Manager Typed Name

4-5-2001

Dated

### CONCLUSIONS:

The principles of operation of the Encompass system are substantially equivalent to the currently marketed products. This system poses no added risk to safety.

This concludes this Special 510(k): Device Modification (k003562).

Richard Petrocelli, President Heartlab, Inc

Attachment: Indications For Use

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular arrangement of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES USA". Inside the circle is an abstract image of an eagle with its wings spread.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUL 1 6 2004

Ms. Lori Kahler Quality Assurance Manager Heartlab Cardiac Solutions One Crosswind Road WESTERLY RI 02891

Re: K040896

Trade/Device Name: EncompassTM Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II

Product Code: 90 LLZ Dated: May 14, 2004 Received: May 17, 2004

Dear Ms. Kahler:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{6}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:

| 8xx.1xxx                         | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4654 |
| Other                            | (301) 594-4692 |

Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely vours,

Nancy C. Brydon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows the logo for Heartlab Cardiac Solutions. The word "Heartlab" is in large, bold, white letters on a black background. Below "Heartlab" is the phrase "CARDIAC SOLUTIONS" in smaller, white, block letters, also on a black background. The phrase is contained within a black rectangle.

### Attachment

510(k) Number: Ko4 0 891 Device Name: Encompass™

#### Indications For Use:

Encompass™ is a picture archiving and communications system intended to be used as a networked cardiovascular information management system. Encompass™ is software comprised of modular software programs that run on standard "off-the-shel?' personal coonyters and servers running the Windows 2000/XP operating system. Encompass™ is image data surage and display software that accepts DICOM (Digital Imaging and Communications in Medicine) data from laboratories, which support DICOM standard imaging transfer. The system provides the capability to; consolidate images generated by equipment from multiple OEM vendors, view images, enter clinical findings while viewing the associated images, perform digital subtraction, create graphical representation of coronary arteries, perform quantitative measurements on both cath and ultrasound images, perform quantitative analysis on cath images, generate and on voview patient reports with additional measurement and report writer capabilities and provides accessible digital image archive. Encompass™ is a scalable network system designed to service customers ranging in size from small departments (with 2 or 3 users) to large hospital networks (with tens of users).

Prescription Use

Nancy C Brogdon
(Division Sign-Off)

(Division Sign-Off) Division of Reproductive. Abdominal and Radiological Devic 510(k) Number

---

**Source:** [https://fda.innolitics.com/device/K040896](https://fda.innolitics.com/device/K040896)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).