← Product Code [GEI](/productcode/GEI) · K040855

# INSTRUMED LAPAROSCOPE AND ACCESSORIES (K040855)

_Instrumed International, Inc. · GEI · Apr 16, 2004 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K040855

## Device Facts

- **Applicant:** Instrumed International, Inc.
- **Product Code:** [GEI](/productcode/GEI.md)
- **Decision Date:** Apr 16, 2004
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4400
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic, 3rd-Party Reviewed

## Indications for Use

Instrumed Laparoscopes and Accessories are intended to be used by qualified physicians to provide access, illumination and visualization of internal structures and for manipulating soft tissue (grasping, cutting, coagulating, dissecting, and suturing) in a wide variety of diagnostic and therapeutic laparoscopic/urologic closed and minimally invasive procedures.

## Device Story

Instrumed Laparoscopes and Accessories provide illumination and visualization of internal structures during minimally invasive surgery. The system includes laparoscopes for viewing and monopolar laparoscopic instruments for tissue manipulation, including grasping, cutting, coagulating, dissecting, and suturing. Used in clinical settings by qualified physicians, the device facilitates diagnostic and therapeutic laparoscopic and urologic procedures. The laparoscope provides the visual field, while the instruments are used to perform surgical tasks. The device is intended to assist surgeons in performing procedures with minimal invasiveness, potentially reducing patient trauma and recovery time compared to open surgery.

## Clinical Evidence

No clinical data provided. Substantial equivalence is based on design, material, and technical characteristic comparisons to the predicate device. Performance data mentioned refers to bench testing confirming safety and effectiveness.

## Technological Characteristics

The device consists of laparoscopes and monopolar laparoscopic instruments. It utilizes optical components for illumination and visualization and mechanical components for tissue manipulation. The device is designed for use in laparoscopic and urologic procedures. Materials are stated to be identical to the predicate device. No specific software, energy source (other than external electrosurgical connection for monopolar instruments), or connectivity features are described.

## Regulatory Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

## Predicate Devices

- Gimmi Laparoscope ([K012660](/device/K012660.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows a sequence of alphanumeric characters, specifically "K040855". The characters are written in a bold, sans-serif font, and they appear to be handwritten or stylized. The number zero appears twice in the sequence.

APR 1 6 2004

## Summary of Safety and Effectiveness

| 1.              | <b>Submission Applicant &amp; Correspondent</b>                                                                                                                                                                                                                                                                                                            |            |                                                     |              |                                          |                 |              |      |              |       |                         |                 |           |
|-----------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|-----------------------------------------------------|--------------|------------------------------------------|-----------------|--------------|------|--------------|-------|-------------------------|-----------------|-----------|
|                 | <table><tr><td>Name:</td><td>Instrumed International, INC.</td></tr><tr><td>Address:</td><td>626 Cooper Court<br/>Schaumburg, IL 60173</td></tr><tr><td>Phone:</td><td>847-908-0292</td></tr><tr><td>Fax:</td><td>847-908-0293</td></tr><tr><td>Mail:</td><td>jwalsh@instrumedinc.biz</td></tr><tr><td>Contact person:</td><td>Jim Walsh</td></tr></table> | Name:      | Instrumed International, INC.                       | Address:     | 626 Cooper Court<br>Schaumburg, IL 60173 | Phone:          | 847-908-0292 | Fax: | 847-908-0293 | Mail: | jwalsh@instrumedinc.biz | Contact person: | Jim Walsh |
| Name:           | Instrumed International, INC.                                                                                                                                                                                                                                                                                                                              |            |                                                     |              |                                          |                 |              |      |              |       |                         |                 |           |
| Address:        | 626 Cooper Court<br>Schaumburg, IL 60173                                                                                                                                                                                                                                                                                                                   |            |                                                     |              |                                          |                 |              |      |              |       |                         |                 |           |
| Phone:          | 847-908-0292                                                                                                                                                                                                                                                                                                                                               |            |                                                     |              |                                          |                 |              |      |              |       |                         |                 |           |
| Fax:            | 847-908-0293                                                                                                                                                                                                                                                                                                                                               |            |                                                     |              |                                          |                 |              |      |              |       |                         |                 |           |
| Mail:           | jwalsh@instrumedinc.biz                                                                                                                                                                                                                                                                                                                                    |            |                                                     |              |                                          |                 |              |      |              |       |                         |                 |           |
| Contact person: | Jim Walsh                                                                                                                                                                                                                                                                                                                                                  |            |                                                     |              |                                          |                 |              |      |              |       |                         |                 |           |
| 2.              | <b>Device name &amp; Classification panel</b>                                                                                                                                                                                                                                                                                                              |            |                                                     |              |                                          |                 |              |      |              |       |                         |                 |           |
|                 | <table><tr><td>Trade name</td><td>Laparoscope &amp; Monopolar<br/>laparoscopic instruments</td></tr><tr><td>Common name:</td><td>Laparoscope &amp; Accessories</td></tr><tr><td>Classification:</td><td>876.1500</td></tr></table>                                                                                                                         | Trade name | Laparoscope & Monopolar<br>laparoscopic instruments | Common name: | Laparoscope & Accessories                | Classification: | 876.1500     |      |              |       |                         |                 |           |
| Trade name      | Laparoscope & Monopolar<br>laparoscopic instruments                                                                                                                                                                                                                                                                                                        |            |                                                     |              |                                          |                 |              |      |              |       |                         |                 |           |
| Common name:    | Laparoscope & Accessories                                                                                                                                                                                                                                                                                                                                  |            |                                                     |              |                                          |                 |              |      |              |       |                         |                 |           |
| Classification: | 876.1500                                                                                                                                                                                                                                                                                                                                                   |            |                                                     |              |                                          |                 |              |      |              |       |                         |                 |           |
| 3.              | <b>Substantial Equivalence</b>                                                                                                                                                                                                                                                                                                                             |            |                                                     |              |                                          |                 |              |      |              |       |                         |                 |           |
|                 | Substantial Equivalence is claimed to the following devices<br>Gimmi Laparoscope<br>Manufactured by<br>Gimmi GmbH<br>Registered in the FDA Database under the following number: K012660                                                                                                                                                                    |            |                                                     |              |                                          |                 |              |      |              |       |                         |                 |           |

:

{1}------------------------------------------------

| Description of device<br>Instrumed Laparoscopes provide illumination and visualization in diagnostic<br>procedures and in conjunction with laparoscopic instruments (accessories) operative<br>laparoscopic procedures |  |                                                                                                                                                                                |  |                                                              |  |  |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--------------------------------------------------------------|--|--|
|                                                                                                                                                                                                                        |  |                                                                                                                                                                                |  |                                                              |  |  |
|                                                                                                                                                                                                                        |  |                                                                                                                                                                                |  |                                                              |  |  |
|                                                                                                                                                                                                                        |  |                                                                                                                                                                                |  |                                                              |  |  |
|                                                                                                                                                                                                                        |  |                                                                                                                                                                                |  |                                                              |  |  |
|                                                                                                                                                                                                                        |  |                                                                                                                                                                                |  |                                                              |  |  |
|                                                                                                                                                                                                                        |  |                                                                                                                                                                                |  |                                                              |  |  |
|                                                                                                                                                                                                                        |  |                                                                                                                                                                                |  |                                                              |  |  |
|                                                                                                                                                                                                                        |  |                                                                                                                                                                                |  |                                                              |  |  |
|                                                                                                                                                                                                                        |  |                                                                                                                                                                                |  | and the control control of the control of the control of     |  |  |
|                                                                                                                                                                                                                        |  | 11 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 |  |                                                              |  |  |
|                                                                                                                                                                                                                        |  |                                                                                                                                                                                |  | and the control control control controlled in the controlled |  |  |
|                                                                                                                                                                                                                        |  |                                                                                                                                                                                |  |                                                              |  |  |
|                                                                                                                                                                                                                        |  |                                                                                                                                                                                |  |                                                              |  |  |

- 11:4

{2}------------------------------------------------

| 5. | Intended Use & Indications                                                                                                                                                                                                                                                                                                                                                              |
|----|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|    | Instrumed Laparoscopes and Accessories are intended to be used by qualified<br>physicians to provide access, illumination and visualization of internal structures<br>and for manipulating soft tissue (grasping, cutting, coagulating, dissecting, and<br>suturing) in a wide variety of diagnostic and therapeutic laparoscopic/urologic closed<br>and minimally invasive procedures. |

:

.

:

:

{3}------------------------------------------------

| 6. | Technical Comparison                                                                                                                                                                                                                                                                                  |
|----|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|    | The Instrumed product is similiar to the Gimmi product in terms of design and technical characteristics. The Laparoscope and the respective laparoscopic instruments are identical to the Gimmi products.                                                                                             |
|    | The Instrumed product is also similiar to the Gimmi products in terms of the used material.                                                                                                                                                                                                           |
|    | In terms of performance data, the presented data that was conducted on the Instrumed instruments shows in its results that the Laparoscope and the laparoscopic instruments are absolutely safe and effective for its intended use and do not raise any questions regarding safety and effectiveness. |

:

:

{4}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol resembling an abstract human figure or a caduceus, composed of three curved lines.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

APR 1 6 2004

Instrumed International, Inc. c/o Mr. Stefan Preiss TÜV America, Inc. 1775 Old Highway 8 New Brighton, Minnesota 55112

Re: K040855

Trade/Device Name: Instrumed Laparoscopes and Accessories Regulation Number: 21 CFR 878.4400, 876.1500 Regulation Name: Electrosurgical cutting and coagulation device and accessories; Endoscope and accessories Regulatory Class: II Product Code: GEI, GCJ Dated: March 25, 2004 Received: April 2, 2004

Dear Mr. Preiss:

We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed prodicate devices marketed in interstate for ass saled in the encreased in the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (sce above) into either class II (Special Controls) or class III (PMA), it rr your device to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the clectronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{5}------------------------------------------------

Page 2 - Mr. Stefan Preiss

This letter will allow you to begin marketing your device as described in your Section 510(k) I ma let notification. The FDA finding of substantial equivalence of your device to a legally promatics notificated with a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincercly yours,

Sincerely yours,

Mark M. Millburn

Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

## Indications for Use

510(k) Number (if known): _

Device Name: Instrumed Laparoscopes and Accessories

Indications for Use:

Instrumed Laparoscopes and Accessories are intended to be used by qualified physicians to provide access, illumination an i visualization of internal structures and for manipulating soft tissue (grasping, cutting, coagulating, dissecting, and suturing) in a wide variety of diagnostic and herapeutic laparoscopic/urologic closed and minimally invasive procedures.

Mark N. Millican

Division of General, Restorative, and Neurological Devices

510(k) Number KC40855

Prescription Use
(Part 21 CFR 801 Subpart D)

ANI /OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW TH S LINE-CONTINUE ON ANOTHER PAGE

OF NEEDED)

Concurrence of CDRH, Off ce of Device Evaluation (ODE)

---

**Source:** [https://fda.innolitics.com/device/K040855](https://fda.innolitics.com/device/K040855)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
