FELCARE POWDER FREE, POLYMER COATED LATEX EXAMINATION GLOVES

K040793 · Hl Rubber Products Sdn Bhd · LYY · Apr 14, 2004 · General Hospital

Device Facts

Record IDK040793
Device NameFELCARE POWDER FREE, POLYMER COATED LATEX EXAMINATION GLOVES
ApplicantHl Rubber Products Sdn Bhd
Product CodeLYY · General Hospital
Decision DateApr 14, 2004
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.6250
Device ClassClass 1

Indications for Use

A medical glove is worn on the hand of healthcare and similar personnel to prevent contamination between healthcare personnel and the patient

Device Story

Powder-free polymer-coated latex examination glove; worn on hands of healthcare personnel; acts as physical barrier to prevent cross-contamination between patient and clinician; used in clinical settings; provides protection during patient examinations.

Clinical Evidence

Bench testing only.

Technological Characteristics

Material: Latex; Surface: Polymer coated; Form: Powder-free examination glove; Classification: Class I, 21 CFR 880.6250, Product Code LYY.

Indications for Use

Indicated for use by healthcare and similar personnel as a protective barrier to prevent cross-contamination between the wearer and the patient.

Regulatory Classification

Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the top half of the circle. In the center of the seal is an eagle-like symbol with three wave-like lines extending from its body. The seal is black and white. ## Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 APR 1 4 2004 Felda Rubber Products Sdn Bhd Mr. Kok- Kee Hon US Agent Kok-Kee Hon 6324 Meeting House Way Alexandria, Virginia 22312-1718 Re: K040793 Trade/Device Name: Powder Free Polymer Coated Latex Examination Gloves Regulation Number: 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LYY Dated: March 26, 2004 Received: March 30, 2004 Dear Mr. Hon: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If vour device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ ## Page 2 - Mr. Hon Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-5618. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## ATTACHMENT 2 ## INDICATIONS FOR USE Applicant: Felda Rubber Products Sdn. Bhd. 510(k) Number (if known): K040793 Device Name: Powder Free Polymer Coated Latex Examination Glove Indications For Use: A medical glove is worn on the hand of healthcare and similar personnel to prevent contamination between healthcare personnel and the patient Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Ke Muluy of Anesthesiology, General Hospital. ion Control, Dental Device 510(k) Number: K040793
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...