The Mobile Oxygen Storage Tank (MOST) is intended to provide 93% oxygen at 50 psig nominal pressure for supplemental oxygen use only. The MOST is intended to be filled from the Expeditionary Deployable Oxygen Concentration System (E-DOCS) only. The MOST is intended for military use only.
Device Story
Mobile Oxygen Storage Tank (MOST) is a portable oxygen storage and distribution system designed for military deployment. It consists of high-pressure composite-wound brass-lined cylinders and a distribution manifold housed in a rugged, waterproof case. The system stores 10,000 liters of oxygen at 2,250 psig. It lacks internal oxygen generation, requiring filling from the Expeditionary Deployable Oxygen Concentration System (E-DOCS). The device features three output fittings, supporting up to three patients concurrently with a maximum flow capability of 225 LPM. It is transported via integrated handles. Healthcare providers use the system to provide supplemental oxygen in field environments. The device is not life-supporting or life-sustaining; failure does not result in serious health consequences. It benefits patients by providing a portable, deployable source of supplemental oxygen in austere military settings.
Clinical Evidence
Bench testing only. Performance, reliability, and safety were validated against USAF (AFMESA), DOT-CFFC, NASA, and MIL-STD-810F specifications. Testing covered purity, flow rate, mechanical integrity, and controls. No clinical data provided.
Technological Characteristics
System consists of high-pressure composite-wound brass-lined cylinders and a distribution manifold. Dimensions: 37.25" x 27.18" x 15.44"; weight: 184 lbs (fully charged). Capacity: 10,000 liters at 2,250 psig. Max flow: 225 LPM. Non-sterile. No internal oxygen generation; relies on E-DOCS for filling. Conforms to USP 24-NF19 standards for oxygen purity.
Indications for Use
Indicated for military personnel requiring supplemental oxygen. Contraindications are labeled on the device. For prescription use only.
Regulatory Classification
Identification
A portable oxygen generator is a device that is intended to release oxygen for respiratory therapy by means of either a chemical reaction or physical means (e.g., a molecular sieve).
Predicate Devices
Deployable Oxygen Concentration System (DOCS) (K020330)
Venture IOH 200 Home Fill II Complete Home Oxygen System (K003939)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows a date, "APR 3 0 2004". The month is "APR", the day is "30", and the year is "2004". The text is in a bold, sans-serif font. The text is black on a white background.
Image /page/0/Picture/1 description: The image shows a handwritten number, "K040738", in a slightly tilted, vertical orientation. The numbers are written in a simple, clear style, with each digit distinctly formed. The overall presentation suggests the number is part of a document or label, possibly for identification or tracking purposes.
Image /page/0/Picture/2 description: The image shows a stylized logo. The logo is a black triangle with a thick outline. Inside the triangle, there are two horizontal lines that create the appearance of the letters 'A' and 'C'. The logo is simple and modern.
# 510(k) Summary
## Mobile Oxygen Storage Tank (MOST)
Type of FDA Submission: Traditional 510(k)
#### Submitter Information
| Submitter's Name: | Pacific Consolidated Industries |
|-------------------|---------------------------------|
|-------------------|---------------------------------|
Submitter's Address: 3430 West Carriage Drive Santa Ana, California 92704-6412
Owner/Operator ID: 9049531
Contact Person: Tarik Naheiri
714-979-9200 (Phone) Submitter's Phone: 714-436-9150 (Fax)
Date of Preparation: March 18, 2004
| Device Name: | Mobile Oxygen Storage Tank |
|------------------------|------------------------------------------------|
| Common Trade Name: | MOST |
| Classification Name: | Portable Oxygen Generator with accessories |
| Device Classification: | 868.5440 (Class II) |
| Product Code: | CAW |
| Panel: | Anesthesiology and Respiratory Therapy Devices |
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows a black and white logo. The logo is a stylized letter 'A' with a horizontal line cutting through the middle. The horizontal line creates the appearance of the letters 'P' and 'C' within the 'A'. The logo is simple and geometric.
#### Predicate Legally Marketed Device Equivalence
Substantial equivalence is claimed to the following legally marketed predicate devices:
K020330 - Deployable Oxygen Concentration System (DOCS) K003939 - Venture IOH 200 Home Fill II Complete Home Oxygen System by Invacare Corporation
Both the DOCS and the Home Fill devices operate on the principle of using a pressure differential to strip oxygen from the ambient air using a zeolite molecular sieve in an adsorption process, as does the DOCS device.
The Home Fill II predicate includes a proprietary cylinder to be filled from an oxygen generation system. The DOCS when used with the MOST as an accessory will have the same intended use and technological characteristics as the Invacare unit.
These predicate devices use an oxygen-generating process to fill cylinders that can be transported to where the oxygen is needed. The DOCS system also has this same cylinder filling capability, but with the addition of the MOST there is another option for deploying the oxygen in the field.
#### Safe Medical Devices Act (SMDA) Statement
The oxygen supplied by the MOST is supplemental and is not considered to be life supporting or life sustaining. The failure of the device would NOT have serious health consequences to the user.
#### Description of Device
The Mobile Oxygen Storage Tank (MOST) is an oxygen storage and distribution system consisting of lightweight, high strength, high-pressure composite wound brass lined cvlinders and a distribution manifold secured within a rugged waterproof case. The MOST contains 10,000 liters of oxygen when filled to 2,250 psig. The MOST has a maximum flow capability of 225 LPM, which is equivalent to 75 LPM per Oxygen Output fitting (3 fittings per MOST).
The MOST's small footprint, relatively light weight and easy carrying system make it optimal for military deployments. Eight handles, two on each long side of the MOST and two on the top at each end are provided for ease of transportation. The MOST contains the Apparatus Kit, which includes the equipment required to distribute the oxygen to up to three patients concurrently. The exterior dimensions of the MOST are: a length of
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows a black and white logo that appears to be a stylized letter 'A'. The logo is formed by two triangles, one nested inside the other. The outer triangle is larger and more prominent, while the inner triangle is smaller and creates a negative space within the larger one. The overall design is simple, geometric, and symmetrical.
37.25", a width of 27.18", and height of 15.44." The MOST weighs approximately 184 pounds when fully charged to 2250 psig.
#### Intended Use of Device
The Mobile Oxygen Storage Tank (MOST) is intended to provide 93% oxygen at 50 psig nominal pressure for supplemental oxygen use only. The MOST is intended to be filled from the Expeditionary Deployable Oxygen Concentration System (E-DOCS) only. The MOST is intended for military use only.
## Device Labeling
The MOST, is clearly labeled "93% Oxygen".
The MOST has labels clearly showing contraindications.
The MOST is clearly labeled "For Use With EDOCS Only".
The MOST is labeled as "Rx Only"
The system is not sold or labeled as sterile.
## Comparison of Technological Characteristics
The MOST consists fundamentally of two cylinders, a manifold that connects to the EDOCS system for filling, as it does not have its own oxygen generating capability. In this lack of filling capability, the MOST is an accessory to the DOCS system and together they have the same technological characteristics as the Home Fill II systems. The MOST system is simpler and at least as safe and effective as both the DOCS and the Home Fill II systems. The technology is well established and has been used in other legally marketed products.
As with the Home Fill II system, the MOST is filled only from the DOCS system, which conforms to the USP 24-NF19 standard.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows a black and white logo that appears to be a stylized letter 'A'. The 'A' is formed by thick, bold lines, and it has a unique design where the horizontal bar is not a single line but rather a rectangular shape that connects the two sides of the 'A'. The logo has a modern and geometric feel due to its clean lines and sharp angles.
## Summary of Performance Testing
Verification and validation testing activities were conducted to establish the performance, reliability and safety characteristics of the MOST, to demonstrate performance as intended. Testing involved the following areas:
- 트 Purity
- Flow Rate
- 트 Mechanical
- 트 Controls
- . Device Performance
- . Safety
Acceptance criteria were based on USAF specifications (AFMESA), DOT-CFFC, NASA and MIL-STD-810F.
In all instances the device met all required performance criteria and functioned as intended, meeting the acceptance criteria.
## Conclusions
In summary, Pacific Consolidated Industries has demonstrated that the MOST is safe and effective. The combined testing and analysis of results provides assurance that the device meets its specifications, is safe and effective for its intended use, and is substantially equivalent to the currently marketed devices.
{4}------------------------------------------------
Image /page/4/Picture/2 description: The image is a circular seal for the Department of Health & Human Services - USA. The seal features the department's logo, which consists of three stylized human figures in profile, arranged in a stacked formation. The figures are depicted in a simple, abstract manner, with flowing lines suggesting movement and connection. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the upper perimeter of the circle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
APR 30 2004
Mr. Tarik Naheiri Director of Technology and Engineering Pacific Consolidated Industries, LLP 3430 West Carriage Drive Santa Ana, CA 92704
Re: K040738
Trade Name: Mobile Oxygen Storage Tank (MOST) Regulation Number: 21 CFR 868.5440 Regulation Name: Portable Oxygen Generator Regulatory Class: II Product Code: CAW Dated: March 18, 2004 Received: March 25, 2004
Dear Mr. Naheiri: -
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. In addition, if you wish to change or expand your current indications for use to include non-military environments, you will need to submit a new 510(k) premarket notification, and receive FDA clearance prior to marketing the device.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{5}------------------------------------------------
Page 2 - Mr. Tarik Naheiri
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Chiu-Liang, Ph.D.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows a black and white logo that appears to be a stylized letter 'A'. The logo is formed by thick, bold lines that create a triangular shape. Inside the larger triangle, there is a horizontal line that cuts across the middle, creating a smaller triangle at the top and an open space below. The overall design is simple, geometric, and symmetrical.
# Statement of Indications for Use
# Name: Mobile Oxygen Storage Tank (MOST)
The Mobile Oxygen Storage Tank (MOST) is intended to provide 93% oxygen at 50 psig The MOST Oxygon Blorago Talle (1) year (1) . The MOST is intended to be filled from the Expeditionary Deployable Oxygen Concentration System (E-DOCS) only. The MOST is intended for military use only.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Per 21 CFR 801.109)
fArh
510(k) Number.
(Division Sign-Off)
Division of Anesthesiology, General Hospital,
Infection Control, Dental Devices
PACIFIC CONSOLIDATED INDUSTRIES
5
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.