← Product Code [JAQ](/productcode/JAQ) · K040704

# MODIFICATION TO HENSCHKE HDR CERVIX APPLICATOR AND HILARIS/NORI HDR CERVIX APPLICATOR (K040704)

_Mick Radio-Nuclear Instruments, Inc. · JAQ · Jun 18, 2004 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K040704

## Device Facts

- **Applicant:** Mick Radio-Nuclear Instruments, Inc.
- **Product Code:** [JAQ](/productcode/JAQ.md)
- **Decision Date:** Jun 18, 2004
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.5700
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Therapeutic

## Indications for Use

The applicators presented in this 510(k) notification have been developed to function as Remote After-Loader sealed sources in the intracavitary treatment of cancer of the vagina and cervix.

## Device Story

HDR compatible Tandem and Ovoid Applicators; used in high dose rate (HDR) brachytherapy for cervical and vaginal cancer. Device acts as a delivery system to link to an HDR unit, allowing remote positioning of radioactive sources within the treatment volume. Enables precise dosimetry and stable source delivery. Used in clinical settings by healthcare professionals. Benefits patient by facilitating targeted radiation therapy to cancerous tissue.

## Clinical Evidence

No clinical data; bench testing only.

## Technological Characteristics

Remote controlled radionuclide applicator system. Materials are identical to predicate devices. Mechanical design facilitates coupling with HDR afterloading units for source positioning. Class II device (21 CFR 892.5700).

## Regulatory Identification

A remote controlled radionuclide applicator system is an electromechanical or pneumatic device intended to enable an operator to apply, by remote control, a radionuclide source into the body or to the surface of the body for radiation therapy. This generic type of device may include patient and equipment supports, component parts, treatment planning computer programs, and accessories.

## Predicate Devices

- Henschke Applicator ([K871217](/device/K871217.md))
- Hilaris/Nori Applicator ([K891377](/device/K891377.md))
- Cervix Applicator ([K033371](/device/K033371.md))

## Reference Devices

- Vaginal Applicator Set, Shielded ([K001544](/device/K001544.md))
- HDR Tandem/Ring Applicator with Rectal Retractor ([K011657](/device/K011657.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

JUN 1 8 2004

Page 1 of 2
K040704

# Summary of Safety and Effectiveness Compliance with 513 (i) of the Federal Food, Drug and Cosmetic Act

March 15, 2004

#### General Provisions 1.

| Common/Usual Name: | Remote Controlled Radionuclide Applicator System:                             |
|--------------------|-------------------------------------------------------------------------------|
| Proprietary Name:  | Modified Henschke HDR Cervix Applicator<br>Hilaris/Nori HDR Cervix Applicator |

Applicant Name and Address:

Mick Radio-Nuclear Instruments, Inc. 521 Homestead Avenue Mount Vernon, New York 10550

#### 2. Name of Predicate Devices:

- (1)

| Model and Manufacturer                                        | K Number |
|---------------------------------------------------------------|----------|
| Henschke Applicator; Mick Radio-Nuclear Instruments, Inc.     | K871217  |
| Hilaris/Nori Applicator; Mick Radio-Nuclear Instruments, Inc. | K891377  |
| Cervix Applicator; Varian Medical Systems, Inc.               | K033371  |

Any statement made in conjunction with this submission regarding substantial equivalence to any other product only relates to whether the product can be lawfully marketed without pre-market approval or reclassification and is not to be interpreted as an admission or used as evidence in patent infringement litigation. As the Commissioner of the FDA has indicated, " ... a determination of substantial equivalence under the Federal Food, Drug, and Cosmetic Act relates to the fact that the product can be lawfully marketed without pre-market approval or reclassification. This determination is not intended to have any bearing whatsoever on the resolution of patent infringement suits." 42 Fed. Reg. 42,520 et seg. (1977).

{1}------------------------------------------------

Page 2 of 2
K040704

### 3. Classification

This device is classified as a class II device according to 21 CFR 892.5700 .

#### Performance Standards 4.

Performance standards for applicators for remote controlled afterloading brachytherapy have not been established by the FDA under Section 514 of the Food, Drug and Cosmetic Act.

#### Intended Use and Device Description ડ.

The Mick Radio-Nuclear Instruments, Inc. HDR compatible Tandem and Ovoid Applicators are intended for use in HDR Brachytherapy as described for the predicate device (Cervix Applicator, Varian Medical Systems, K033371). The design of these systems is the similar to that of the predicate devices listed below.

| Predicate Device        | Manufacturer                         | K Number |
|-------------------------|--------------------------------------|----------|
| Henschke Applicator     | Mick Radio-Nuclear Instruments, Inc. | K871217  |
| Hilaris/Nori Applicator | Mick Radio-Nuclear Instruments, Inc. | K891377  |

The delivery of radiation therapy to cervix via high dose rate Brachytherapy requires the ability to properly link the applicator to the HDR unit so that the radioactive source can be remotely positioned in the applicator along with the ability to properly localize the sealed sources at the treatment volume providing precise dosimetry and then a stable delivery system from which treatment can be administered. The Mick Radio-Nuclear HDR compatible Tandem and Ovoid Applicators are designed to act as such a device and utilize the same design as previously cleared for other applicators manufactured by Mick Radio-Nuclear Instruments (Vaginal Applicator Set, Shielded, K001544, HDR Tandem/Ring Applicator with Rectal Retractor, K011657).

#### 6. Biocompatibility

No new issues of biocompatibility are raised with regard to this device.

#### 7. Summary of Substantial Equivalence

This device is similar in design and construction, utilizes the identical materials, and has the same intended use and performance characteristics to the predicate devices. No new issues of safety or effectiveness are introduced by using this device.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three bars representing its wings. The eagle is positioned to the right of a circular text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES USA".

Food and Drug Administration 9200 Eorporate Boulevard Rockville MD 20850

JUN 1 8 2004

Mr. Felix Mick President Mick Radio-Nuclear Instruments, Inc. 521 Homestead Avenue MOUNT VERNON NY 10550

Re: K040704 Trade/Device Name: HDR Compatible Tandem and Ovoid Applicators Regulation Number: 21 CFR 892.5700 Regulation Name: Remote controlled radionuclide applicator system Regulatory Class: II Product Code: 90 JAQ Dated: March 15, 2004 Received: March 22, 2004

Dear Mr. Mick:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding ar.d adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act that is a regulations and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:

| 8xx.1xxx                         | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4654 |
| Other                            | (301) 594-4692 |

Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Nancy C. Brogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## Indications for Use

510(k) Number: ~~To be assigned~~ K040704

Device Name: HDR Compatible Tandem and Ovoid Applicators

Indications for Use:

The applicators presented in this 510(k) notification have been developed to function as The applicators presonted in this 310(t), hemote After-Loader sealed sources in the intracavitary treatment of cancer of the vagina and cervix.

Concurrence of CDRH, Office of Device Evaluation (ODE)

| Prescription Use: | ✓ |
|-------------------|---|
|-------------------|---|

Over-The Counter Use: ____ (Per 21 CFR 801.109)

tion Use: V or

Daniel R. Gleason

(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number. K04070

HDR Compatible Tandem and Ovoid Applicators - 510(k)

---

**Source:** [https://fda.innolitics.com/device/K040704](https://fda.innolitics.com/device/K040704)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
