← Product Code [MOS](/productcode/MOS) · K040640

# MODEL 475 - 8 CHANNEL T/R PHASED ARRAY LOWER EXTREMITY COIL (K040640)

_Igc-Medical Advances, Inc. · MOS · Mar 22, 2004 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K040640

## Device Facts

- **Applicant:** Igc-Medical Advances, Inc.
- **Product Code:** [MOS](/productcode/MOS.md)
- **Decision Date:** Mar 22, 2004
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1000
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** 3rd-Party Reviewed

## Indications for Use

Magnetic resonance imaging (MRI) of the musculoskeletal structures, soft tissue and vascular structures of the lower extremities.

## Device Story

Model 475 8 Channel T/R Phased Array Lower Extremity Coil; accessory for GE and Siemens 1.5T and 3.0T MRI systems. Device captures RF signals from lower extremity anatomy; transforms signals into 2D/3D images (proton density, T1/T2 weighted, time of flight, phase contrast). Used in clinical radiology settings; operated by MRI technicians/radiologists. Output viewed on MRI system console; assists clinicians in diagnostic assessment of musculoskeletal/vascular pathology. Provides 3x improvement in signal-to-noise ratio in peripheral regions compared to predicate; supports fast imaging protocols.

## Clinical Evidence

Bench testing only. Comparison of imaging performance parameters (FOV, SNR, image uniformity, geometric distortion, slice thickness/gap, high contrast spatial resolution) against predicate device. Results show 3x improvement in SNR in peripheral regions; no change in other performance metrics.

## Technological Characteristics

8-channel T/R phased array coil; compatible with 1.5T and 3.0T MRI systems. Materials comply with UL 94 flammability standards. Safety compliant with UL 2601-1 and IEC 601-1. Form factor designed for lower extremity imaging.

## Regulatory Identification

A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).

## Special Controls

*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

MAR 2 2 2004

K040640

# 510(k) Summary of Safety and Effectiveness

| Device Name                       | Model 475 8 Channel T/R Phased Array Lower<br>Extremity Coil                                                     |                                                                         |
|-----------------------------------|------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------|
| Applicability                     | Compatible with GE and Siemens 1.5T and 3.0T<br>MR Systems                                                       |                                                                         |
| Reason for 510(k)                 | New device                                                                                                       |                                                                         |
| Classification Name               | Magnetic Resonance Diagnostic Device                                                                             |                                                                         |
| Device Classification Panel       | Radiology                                                                                                        |                                                                         |
| Device Classification Number      | 892.1000                                                                                                         |                                                                         |
| Product Code                      | 90MOS                                                                                                            |                                                                         |
| Common Name                       | Magnetic Resonance Imaging Coil                                                                                  |                                                                         |
| Proprietary Name                  | Model 475 8 Channel T/R Phased Array Lower<br>Extremity Coil                                                     |                                                                         |
| Establishment Registration Number | 2183683                                                                                                          |                                                                         |
| Address of MFG Facility           | IGC-Medical Advances Inc.<br>10437 Innovation Drive<br>Milwaukee, WI 53226                                       |                                                                         |
| Point of Contact                  | Anthony Dietzler<br>Quality Assurance Engineer<br>(414) 258-3808 Ext. 255                                        |                                                                         |
| Classification                    | Class II                                                                                                         |                                                                         |
| Intended Uses                     |                                                                                                                  |                                                                         |
| Diagnostic Uses                   | 2D, 3D imaging, proton density, T1 and T2<br>weighted imaging. 2D, 3D time of flight, phase<br>contrast imaging. |                                                                         |
| Anatomic Regions                  | Musculoskeletal structures, soft tissue and vascular<br>structures of the lower extremities.                     |                                                                         |
| Performance Standards             | None Established under Section 514                                                                               |                                                                         |
| Voluntary Safety Standards        | UL 2601-1                                                                                                        | Medical Electrical Equipment, Part I<br>General Requirements for Safety |
|                                   | UL 94                                                                                                            | Tests for Flammability of Plastic<br>Materials                          |
|                                   | IEC 601-1                                                                                                        | General Safety Requirements for<br>Medical Electrical Equipment         |

:

{1}------------------------------------------------

### Standards

### Overview

The Radiology Devices Panel considered potential concerns regarding the safe and effective operation of Magnetic Resonance Diagnostic Devices when they recommended reclassification to Class II on July 27, 1987. After reclassification, the FDA's Center for Devices and Radiological Health (CDRH) released a draft guidance document for the content and review of Magnetic Resonance Diagnostic Device premarket notification submissions that offered clarification of these concerns. Due to considerable technological advances in MRDDs, CDRH issued an updated guidance document on November 14, 1998. The following is a summary of the information contained within this premarket notification that addresses these concerns:

The GE and Siemens 1.5T and 3.0T MRI systems, operated with the Medical Advances 8 Channel T/R Phased Array Lower Extremity Coil, are substantially equivalent to the GE Signa 1.5T system operated with the legally marketed predicate devices listed in scction 4.0, within the Class II definition of Magnetic Resonance Diagnostic Device with respect to the safety parameter action levels:

## Safety Parameters

| Maximum Static Magnetic Field:          | No change due to coil |
|-----------------------------------------|-----------------------|
| Rate of Magnetic Field Strength Change: | No change due to coil |
| RF Power Deposition:                    | Improved              |
| Acoustic Noise Levels:                  | No change due to coil |
| Biocompatibility:                       | No change             |

{2}------------------------------------------------

# Imaging Performance Parameters

| Specification Volume:             | 15-16 cm typical FOV all planes<br>(submitted device) vs. 15-20 cm<br>typical FOV all planes (predicate<br>device) |
|-----------------------------------|--------------------------------------------------------------------------------------------------------------------|
| Signal-to-Noise Ratio:            | 3 Times Improvement over predicate<br>device in peripheral regions                                                 |
| Image Uniformity:                 | No change                                                                                                          |
| Geometric Distortion:             | No change                                                                                                          |
| Slice Thickness and Gap:          | No change                                                                                                          |
| High Contrast Spatial Resolution: | No change                                                                                                          |
| Fast Imaging Protocol Compatible: | Yes                                                                                                                |

## General Safety and Effectiveness Concerns

The device contains instructions for use. It includes indications for use, precautions, cautions, contraindications, warnings and quality assurance testing. This information assures safe and effective use of the device.

## Substantial Equivalence Summary

The GE and Siemens 1.5T and 3.0T MRI systems operated with the Medical Advances 8 Channel T/R Phased Array Lower Extremity Coils addressed in this Pre-Market Notification, have the same intended use and technological characteristics as the GE Signa 1.5T system operated with the identified legally marketed predicate device. The use of these coils does not affect the GE or Siemens 1.5T and 3.0T system safety parameter specifications.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features an eagle-like emblem with three stylized lines representing the agency's mission. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the emblem.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAR 2 2 2004

IGC-Medical Advances, Inc. % Mr. Daniel W. Lehtonen Staff Engineer-Medical Devices Intertek Testing Services NA, Inc. 70 Codman Hill Road BOXBOROUGH MA 01719

Re: K040640

Trade/Device Name: Model 475: 8 Channel T/R Phased Array Lower Extremity Coil Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: 90 MOS Dated: March 4, 2004 Received: March 10, 2004

Dear Mr. Lehtonen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complics with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:

| 8xx. 1xxx                        | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4654 |
| Other                            | (301) 594-4692 |

Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Nancy C. Bugden

Nancy C. Brogdon Director, Division of Reproductive. Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Page ੀ 1

510(k) Number (if known): K C 4 0640

Model 475: 8 Channel T/R Phased Array Lower Extremity Coil Device Name:

Indications for Use:

Magnetic resonance imaging (MRI) of the musculoskeletal structures, soft tissue and vascular structures of the lower extremities.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use OR Over-The-Counter Use (Per 21 CFR 801.109) (Optional Format 1-2-96) (Division Sign-Off) Division of Reproductive, A and Radiological Devices 510(k) Number .

---

**Source:** [https://fda.innolitics.com/device/K040640](https://fda.innolitics.com/device/K040640)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
