COMPRESSAR SUPERCOMFORT DISC AND STRONGARM SUPERCOMFORT SYSTEM

K040615 · Advanced Vascular Dynamics · DXC · Jun 1, 2004 · Cardiovascular

Device Facts

Record IDK040615
Device NameCOMPRESSAR SUPERCOMFORT DISC AND STRONGARM SUPERCOMFORT SYSTEM
ApplicantAdvanced Vascular Dynamics
Product CodeDXC · Cardiovascular
Decision DateJun 1, 2004
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.4450
Device ClassClass 2
AttributesTherapeutic

Indications for Use

This device is indicated for use to provide hemostasis of the femoral vascular access site during and following catheterization or cannulation procedures.

Device Story

CompressAR SuperComfort Disc and StrongArm SuperComfort System is a mechanical vascular compression device. It consists of a compression disc with an elastomeric insert and a supporting stand. The device is used in clinical settings during and following femoral vascular catheterization. It functions by applying direct external pressure to the femoral access site, mimicking manual pressure to achieve hemostasis. The elastomeric insert increases friction against the patient's skin to maintain position. The system utilizes a two-step friction fit connection to the StrongArm stand. Healthcare providers operate the device to hold pressure, reducing the need for continuous manual compression by staff.

Clinical Evidence

Bench testing only. No clinical data provided.

Technological Characteristics

Mechanical vascular clamp. Features an elastomeric insert in the compression pad for increased skin friction. Utilizes a two-step friction fit connection to the StrongArm support stand. Non-powered, manual mechanical operation.

Indications for Use

Indicated for patients undergoing femoral vascular catheterization or cannulation procedures to assist in obtaining and maintaining hemostasis at the access site.

Regulatory Classification

Identification

A vascular clamp is a surgical instrument used to occlude a blood vessel temporarily.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ 'JUN = 1 2004 ## 510(k) Summary Herbert J. Semler 1 March 2004 Trade name - CompressAR® SuperComfort™ Disc and StrongArm™ SuperComfort™ System Common name - femoral access compression device Classification name - Clamp, Vascular This device is similar to the CompressAR® Universal System CompressAR® Comfort™ Disc, and is a modification of this predicate device. The CompressAR® SuperComfort™ Disc provides an elastomeric insert in the compression pad, and is used with a modified attachment to the CompressAR @ StrongArm™ System to provide a two-step friction fit connection. The elastomer insert contacts the patient skin providing increased friction against the skin .. The CompressAR @ SuperComfort™ Disc and StrongArm™ System is intended for use during and following femoral vascular catheterization procedures to assist in obtaining and maintaining hemostasis. The disk and supporting stand provide a mechanical means of holding external pressure at or near the site of femoral vascular access. Direct pressure is used to obtain and maintain hemostasis, in a similar fashion to direct hand pressure on the access site or at a pressure point. Testing was conducted to determine that the modified device provides mechanical clamping and holding functions, similar to the predicate device. It was concluded that the CompressAR & SuperComfort™ Disc and StrongArm™ System is equivalent to the predicate device. 2326 NW Everett St. Portland, Oregon 97210 (503) 223-2333 (503) 223-8585 Fax (800) 525-2555 {2}------------------------------------------------ Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JUN - 1 2004 Advanced Vascular Dynamics c/o Herbert J. Semler, M.D. Official Correspondent 2326 NW Everett Street Portland, OR 97210 Re: K040615 R040019 CompressAR® SuperComfort™ Disc and StrongArm™ SuperComfort™ System Regulation Number: 21 CFR 870.4450 Regulation Name: Vascular Clamp Regulatory Class: Class II (two) Product Code: DXC Dated: May 14, 2004 Received: May 14, 2004 Dear Dr. Semler: We have reviewed your Section 510(k) premarket notification of intent to market the device wo nave and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the encreativent of the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The r ou may, dicterer, mains of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Page 2 - Herbert J. Semler, M.D. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Donna R. Vorhner Image /page/2/Picture/5 description: The image contains a signature and the words "Br" and "Di". The signature is on the left side of the image and appears to be a stylized set of lines. The letters "Br" and "Di" are on the right side of the image and are stacked vertically. Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Indications for Use 510(k) Number (if known):__K040615 Device Name: CompressAR® Femoral Access Compression Device (SuperComfort) Indications For Use: This device is indicated for use to provide hemostasis of the femoral vascular access site during and following catheterization or cannulation procedures. Prescription Use _ × (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use _ (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Dring R. Vechinel (Division Sign-Off) Division of Cardiovascular Devices 510(k) Number_k040615 Page 1 of ____________________________________________________________________________________________________________________________________________________________________ .. -
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...