VASCULAR SOLUTIONS D-STAT DRY 3X3 HEMOSTATIC PAD
K040510 · Vascular Solutions, Inc. · QSX · Mar 17, 2004 · SU
Device Facts
| Record ID | K040510 |
| Device Name | VASCULAR SOLUTIONS D-STAT DRY 3X3 HEMOSTATIC PAD |
| Applicant | Vascular Solutions, Inc. |
| Product Code | QSX · SU |
| Decision Date | Mar 17, 2004 |
| Decision | SESE |
| Submission Type | Traditional |
| Device Class | Class U |
| Attributes | Therapeutic |
Indications for Use
The Vascular Solutions D-Stat Dry 3x3 Hcmostatic Pad is intended to be a topical compression bandage for use under the direction of a healthcare professional and is indicated as a trauma dressing for temporary control of moderate to severely bleeding wounds and for the control of surface bleeding from vascular access sites and percutaneous catheters or tubes.
Device Story
D-Stat Dry 3x3 Hemostatic Pad; lyophilized pad containing thrombin, sodium carboxymethylcellulose, and calcium chloride. Applied topically by healthcare professionals over bleeding source; provides physical barrier combined with manual compression. Thrombin facilitates hemostasis via enzymatic conversion of fibrinogen to fibrin, enhancing clotting cascade. Used for trauma wounds and vascular access sites. Benefits patient by accelerating clot formation and controlling bleeding.
Clinical Evidence
No clinical evaluations conducted. Bench testing only; assessed physical properties of lyophilized pad, packaging, and ability of components to clot blood.
Technological Characteristics
Lyophilized pad composed of thrombin, sodium carboxymethylcellulose, and calcium chloride. Functions as topical compression bandage. No performance standards under section 514 developed.
Indications for Use
Indicated for trauma dressing for temporary control of moderate to severely bleeding wounds and control of surface bleeding from vascular access sites and percutaneous catheters or tubes.
Regulatory Classification
Identification
To temporarily control bleeding and cover external wounds.
Predicate Devices
- Vascular Solutions Inc., D-Stat Dry Hemostatic Bandage
- On Site Gas Systems, Inc., QuikClot
- Marine Polymer Technologies, Inc. RDH Bandage
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is a symbol representing the Department of Health & Human Services USA, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The FDA logo is commonly used to indicate that a product or establishment is regulated by the FDA.
February 10, 2023
Vascular Solutions, Inc. Gregory Sachs Director of Regulatory Affairs 6464 Sycamore Court Minneapolis, Minnesota 55369
Re: K040510
Trade/Device Name: Vascular Solutions D-Stat - Dry™ 3x3 Hemostatic Pad Regulatory Class: Unclassified Product Code: QSX
Dear Gregory Sachs:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated March 17, 2004. Specifically, FDA is updating this SE Letter because FDA has better categorized your device technology under product code QSX.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Julie Morabito, OHT4: Office of Surgical and Infection Control Devices, 240-402-3839, Julie.Morabito@fda.hhs.gov.
Sincerely,
Julie A. Morabito -S
Julie Morabito, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an emblem featuring a stylized caduceus, which is a symbol often associated with healthcare and medicine.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAR 1 7 2004
Mr. Gregory W. Sachs Director of Regulatory Affairs Vascular Solutions, Inc. 6464 Sycamore Court Minneapolis, Minnesota 55369
Re: K040510
Trade/Device Name: Vascular Solutions D-Stat -- Dry™ 3x3 Hemostatic pad Regulatory Class: Unclassified Product Code: FRO Dated: February 26, 2004 Received: February 27, 2004
Dear Mr. Sachs:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments. or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (scc above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalcnce determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
Page 2 - Mr. Gregory W. Sachs
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Miriàm C. Provost
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
Indications for Use Statement
510(k) Number:
Device Name: Vascular Solutions I)-Stat-Dry™ 3x3 Hemostatic Pad
240510
Indications for Use:
The D-Stat Dry 3X3 is applied topically and is indicated as a trauma dressing for temporary control of moderate to severely bleeding wounds and for the control of surface bleeding from vascular access sites and percutaneous catheters or tubes.
Prescription Use AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IFNEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Miriam C. Provost
(Division Sign-Off Division of General, Restorative. and Neurological Devices
510(k) Number_________________________________________________________________________________________________________________________________________________________________
Page 1 of
February 26, 2004
{4}------------------------------------------------
K040510
# 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS
| Common/Usual Name: | Topical Hemostat |
|-----------------------------|------------------------------------------------------------------------------------|
| Product Trade Name: | D-Stat Dry 3x3 Hemostatic Pad |
| Classification Namc: | Unclassified<br>Product Code FRO |
| Manufacturer: | Vascular Solutions, Inc.<br>6464 Sycamore Court<br>Minneapolis, MN 55369 |
| Establishment Registration: | 2134812 |
| Contact: | Gregory W. Sachs<br>Director of Regulatory Affairs |
| Performance Standards: | No performance standards have been developed<br>under section 514 for this device. |
## Device Description:
The D-Stat Dry 3x3 Hemostatic Pad consists of a: Lyophilized pad consisting of thrombin, sodium carboxymethylcellulose and calcium chloride
The D-Stat Dry 3x3 Hemostatic Pad achieves its principal intended action (hemostasis) in the same manner as the other devices in the D-Stat Dry Product Family; it is applied directly over the source of bleeding therefore creating a physical barrier to blood flow through the application of adjunctive manual compression. The lyophilized components (thrombin, CMC, and calcium chloride) establish an environment in which a natural blood clot can build and form a physical barrier to bleeding. The thrombin facilitates hemostasis by enhancing the surface-activated clotting cascade through enzymatic cleavage and conversion of fibrinogen to fibrin.
### Intended Use:
The Vascular Solutions D-Stat Dry 3x3 Hcmostatic Pad is intended to be a topical compression bandage for use under the direction of a healthcare professional and is indicated as a trauma dressing for temporary control of moderate to severely bleeding wounds and for the control of surface bleeding from vascular access sites and percutaneous catheters or tubes.
## Summary of Non-Clinical Testing:
Testing conducted included assessments of the physical properties of the lyophilized pad and packaging and the ability of the lyophilized components to achieve its intended use
{5}------------------------------------------------
(clot blood). The results of this battery of tests confirmed the suitability of the D-Stat Dry 3x3 Hemostatic Pad for its intended use.
## Summary of Clinical Testing:
No clinical evaluations of this product for this use have been conducted.
### Predicate Devices:
The intended use of the D-Stat Dry 3x3 Hemostatic Pad is a combination of the intended use of the:
- Vascular Solutions Inc., D-Stat Dry Hemostatic Bandage .
- On Site Gas Systems, Inc., QuikClot, .
- Marine Polymer Technologies, Inc. RDH Bandage .
#### Conclusions:
The D-Stat Dry 3x3 Hemostatic Pad is substantially equivalent to Vascular Solutions Inc., D-Stat Dry Hemostatic Bandage, the On Site Gas Systems QuikClot and the Marine Polymer Technologics, Inc. RDH Bandage. The testing performed confirms that the D-Stat Dry 3x3 Hemostatic Pad will perform as intended.