TAGA VELOCITY PASSOVER HUMIDIFIER WITH HEATER, MODEL TM1000H
K040507 · Taga Medical Technologies, Inc. · BTT · Mar 4, 2004 · Anesthesiology
Device Facts
Record ID
K040507
Device Name
TAGA VELOCITY PASSOVER HUMIDIFIER WITH HEATER, MODEL TM1000H
Applicant
Taga Medical Technologies, Inc.
Product Code
BTT · Anesthesiology
Decision Date
Mar 4, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5450
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The TAGA Velocity Passover Humidifier with Velocity Humidifier Heater TM1000H is a respiratory positive airway pressure accessory intended to add moisture to the air stream gases for administration to the patient. The humidifier can be used with standard CPAP and or Bi-level Positive Airway Pressure (BiPAP) devices, which have a maximum operating pressure of 20 cm H20, and do not have automatic pressure titration capabilities.
Device Story
Device consists of a plastic passover humidifier reservoir and a separate heater plate assembly (TM1000H). Humidifier reservoir features inlet/outlet ports for connection to CPAP/BiPAP tubing; internal baffles create airflow turbulence and eddy effects over water surface to maximize evaporation. Heater plate assembly supports the reservoir; user-adjustable control knob sets plate temperature (88°F to 140°F). Device warms water in reservoir to humidify air stream delivered to patient mask. Used in home, clinic, or hospital by patients under physician prescription. Safety feature includes thermal cutoff at ~90°C to prevent overheating. Benefits include increased moisture in delivered gas for patient comfort during positive airway pressure therapy.
Clinical Evidence
Bench testing only. Performance evaluation compared relative humidity gain of the TM1000H against the predicate Devilbiss Model 9200D across various flow rates and temperature settings. Results indicated the subject device matched or exceeded the predicate's performance.
Technological Characteristics
Plastic housing; 22 mm conical connectors for CPAP/BiPAP tubing. Heater plate assembly with manual control knob (31°C-60°C range). Thermal safety cutoff at ~90°C. Electrical: 110V, 60Hz, 110W. Conforms to EN/IEC/UL 60601-1 (safety) and EN/IEC/UL 60601-1-2 (EMC).
Indications for Use
Indicated for patients requiring respiratory positive airway pressure therapy. Used with standard CPAP or Bi-level PAP devices (max 20 cm H2O) without automatic pressure titration. For single-patient use in home, physician's office, or hospital/institutional settings.
Regulatory Classification
Identification
A respiratory gas humidifier is a device that is intended to add moisture to, and sometimes to warm, the breathing gases for administration to a patient. Cascade, gas, heated, and prefilled humidifiers are included in this generic type of device.
Predicate Devices
TAGA Medical Technologies, Inc., Velocity Passover Humidifier (K031179)
Sunrise Medical HHG, Inc. Devilbiss Model 9200D heated Humidifier System (K020900)
Submission Summary (Full Text)
{0}------------------------------------------------
K040507
## 510(k) SUMMARY
The Summary of Safety and Effectiveness information on the TAGA Velocity Humidifier Heatch The Summary of Safety and Effectiveles information of I.C.R. 2007.92 and effects data available and reserveres of there spolite of fitures spolities and TM1000H is being submitted in accordance with the requirements on 21 C.S. and white of the results of fiture studies and at the time the submission was prepared, but the conclusions or recommendations set forth.
| Applicant | Gary Austin<br>TAGA Medical Technologies, Inc.<br>34675 Vokes Dr., Suite 105<br>Eastlake, Ohio 44095 |
|----------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Telephone | 440/602-8242 |
| Facsimile | 440/602-6989 |
| Date | September 19, 2003 |
| Name | TAGA Velocity Passover Humidifier with Velocity Humidifier Heater<br>TM1000H |
| Classification | Respiratory gas humidifier, 21 CFR 868.5450 |
| Predicate: | TAGA Medical Technologies, Inc., Velocity Passover Humidifier,<br>K031179 market clearance date July 17, 2003, which is used as a<br>component for the new Velocity Humidifier Heater TM1000H device and<br>the Sunrise Medical HHG, Inc. Devilbiss Model 9200D heated Humidifier<br>System, K020900, market clearance date June 03, 2002. |
| Description | The following is a summary of the Velocity Passover Humidifier<br>component for the new Velocity Humidifier Heater TM1000H device.<br><br>The non-sterile Velocity Passover Humidifier is a plastic housing<br>comprised of two (2) halves, a top and bottom, assembled together to form<br>an enclosed reservoir. An O-ring forms a seal between the two halves<br>when assembled, with a plurality of slide latches maintaining the proper<br>position and seal tensions between the two housing components. The<br>reservoir has two ports, an inlet and an outlet, on the upper portion, both<br>being 22 mm conical connectors, which allow for the connection of<br>commonly used respiratory Continuous Positive Airway Pressure (CPAP)<br>or Bi-level Positive Airway Pressure (BiPAP) flexible tubing. The inlet<br>and outlet port are clearly identified on the device. The inlet port is<br>typically attached to the pressure generating CPAP or BiPAP unit by<br>means of a short piece (12" to 24") of tubing that is supplied with the<br>humidifier. The air entering the humidifier is directed over the surface of<br>the water in the basin through a series of baffles. The design intent of the<br>baffles shape and placement is to create turbulence in the airflow over the<br>water surface. The baffles also create an eddy effect, which in turn increase<br>the duration that the air is exposed and travels across the surface of the<br>water. The combination of both of these effects maximizes the evaporation<br>process thereby elevating the humidity level of the gas before exiting the<br>device. The air exits the device through the outlet port into a second piece<br>of tubing supplied by the user that is connected the patients' mask.<br><br>The volume of water is sufficient to provide 10 hours minimum use at 70<br>°F and 25% RH ambient conditions and a patient flow rate of 90 LPM.<br>The humidifier is filled only when removed from the CPAP or BiPAP<br>systems. The Passover Humidifier is used in a horizontal position and will<br>act as a base for most marketed CPAP or BiPAP systems This feature |
{1}------------------------------------------------
## 510(k) SUMMARY con't.
The Summary of Safety and Effectiveness information on the TAGA Velocity Humidifier Heater I the Summal y of Salety and Encareaes mith the requirements of 21 C.P.R. 8807.92 and reflects data available and represented I MINOUL II being submission was prepared, but cautions of 21 Oct 16 yourse the data. The results of future studies and at the this submission won of the conclusions or recommendations set forth.
| | ensures that the water level is below the outlet port on the CPAP or BiPAP | | |
|--------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|
| | systems to eliminate the potential hazard of water reaching any CPAP or | | |
| | BiPAP electrical components. | | |
| | The following is a description for the new Velocity Humidifier Heater<br>TM1000H component of the device. | | |
| | The Velocity Humidifier Heater TM1000H consists of a plastic housing<br>with an On/Off switch, control knob, heater plate assembly and two<br>indicator LEDs. The Velocity passover humidifier rests on the heater plate<br>assembly. The heater plate assembly operates at a scale of low (88°F<br>(31°C)) to high (140°F (60°C)) as selected by the user by adjusting the<br>control knob. The Velocity humidifier heater contains a built-in safety<br>device to prevent the heater plate from reaching excessive temperatures<br>under fault conditions. If the heater plate temperature exceeds<br>approximately 90° C (194° F) power to the unit is terminated. The device<br>has an electrical rating of 110 volts 60Hz, 110 Watts. | | |
| | The device works by heating a metal plate that the plastic humidifier rests<br>on. This in turn warms the water within the humidifier. As the air flows<br>over the water, it picks up some of the moisture. As a result, the gas<br>exiting the humidifier is warmed and humidified providing a continuous<br>humid air stream gas to the patients that require continuous humid positive<br>airway pressure. The system controls the temperature of the heater plate<br>assembly only and not the water or gas temperature. | | |
| Intended Use | The TAGA Velocity Passover Humidifier with Velocity Humidifier Heater<br>TM1000H is a respiratory positive airway pressure accessory intended to<br>add moisture to the air stream gases for administration to the patient. The<br>humidifier can be used with standard CPAP and or Bi-level Positive<br>Airway Pressure (BiPAP) devices, which have a maximum operating<br>pressure of 20 cm H20, and do not have automatic pressure titration<br>capabilities. | | |
| | Environment of Use / Patient Population<br>For single patient use in home, physician's office or hospital / institutional<br>environment. | | |
| Warning: | Disconnect the air tubes prior to cleaning, water entering the CPAP unit may result in electric shock hazard or damage to the CPAP unit. Do not use bleach or chlorine based solutions to clean the humidifier or tubing. The humidifier is for single-patient use only. The Velocity Humidifier can be used with CPAP or Bi-level CPAP devices which have a maximum operating pressure of 20 cm H2O, and do not have automatic pressure titration capabilities. | | |
{2}------------------------------------------------
## 510(k) SUMMARY con't.
The Summary of Safety and Effectiveness information on the TAGA Velocity Humidifier Heater TM1000H is being submitted in accordance with the requirements of 21 C.F.R. §807.92 and reflects data available and represented at the time the submission was prepared, but caution should be exercised in interpreting the data. The results of future studies and or tests may require alterations of the conclusions or recommendations set forth.
| Caution: | Federal law (U.S.A.) restricts this device to sale by or on the order of a physician. Do not expose the tubing to direct sunlight as it may deteriorate over time. Replace the humidifier if any sign of damage to the chamber or leaking appears. Not recommended for use with auto-titrating (adjusting) CPAP device unless otherwise stated by the CPAP device. The heater is only for use with the Velocity Humidifier. |
|-----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Technological<br>Characteristics: | The Velocity Humidifier Heater TM1000H relative humidity gain matched or exceeded the Devilbiss relative humidity gain at all flows and temperature settings. |
| Standards: | The subject device conforms to the following voluntary standards:<br><br>EN/IEC/UL 60601-1 Medical electrical equipment, Part 1:<br>General requirements for safety<br><br>EN/IEC/UL 60601-1-2 Medical Electrical Equipment—Part 1:<br>General Requirements for Safety,<br>Amendment No. 2. Collateral Standard:<br>Electromagnetic Compatibility<br>Requirements and Tests |
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAR - 4 2004
Mr. Gary Austin Vice President Taga Medical Technologies 34675 Vokes Drive Suite 105 Eastlake, Ohio 44095
Re: K040507
K040507
Trade/Device Name: Taga Velocity Passover Humidifier with Heater, Model TM1000H Regulation Number: 868.5450 Regulation Name: Respiratory Gas Humidifier Regulatory Class: II Product Code: BTT Dated: February 26, 2004 Received: February 27, 2004
Dear Mr. Austin:
We have reviewed your Section 510(k) premarket notification of intent to market the device
tion the states of the institution in the closes of the sevivalent (for the We have reviewed your Section 710(t) premainly equivalially equivalent (for the referenced above and nave detimined the development of dicated predicate devices marketed in
indications for use stated in the enclosure the enclicated of the Madical Devices indications for use stated in the caciosares to registered date of the Medical Device interstate commerce prior to May 20, 1 7 7 % accordance with the provisions of
Amendments, or to devices that have been reclassified in accordance approval of a premarket Amendments, or to devices that have been features and require approval of a premaints of the Federal Food, Drug, and Cosment Act (recy market the device, subject to the general
approval application (PMA). You may, therefore, market the det include approval application (FMA). Tou may, and controls provisions of the Act include controls provisions of the Act. The general of devices, good manufacturing practice, proctice, requirements for annual regions misbranding and adulteration.
If your device is classified (see above) into either class III (Special Controls) or class III
ns affect If your device is classified (see above) into controls. Existing major regulations affecting
(PMA), it may be subject to such additional controls. Existing Title 21 Parts (PMA), it may be subject to such additional controllar connects. Title 21, Parts 800 to 898. In the Federal your device can be found in the Code of reactar regaraces.
addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that I DA's issuance of a substantial equivalence determination does not Please be advised that IPDA s Issualice of a suisitantial oque not the requirements
mean that FDA has made a determination that your device complies with other requirements
{4}------------------------------------------------
of the Act or any Federal statutes and regulations administered by other Federal agencies. or the First of all the Act's requirements, including, but not limited to: registration r out intist compry mart 807); labeling (21 CFR Part 801); good manufacturing practice and libang (2) = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = requirements as bet resir product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) Premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), 11 you dontact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Inter count Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Chiu-Lin, Ph.D.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use
510(k) Number (if known): .___________________________________________________________________________________________________________________________________________________
TAGA Velocity Passover Humidifier with Heater Device Name: _________________________________________________________________________________________________________________________________________________________________
Indications For Use:
The TAGA Velocity Passover Humidifier with Velocity Humidifier Heater TM1000H is a respiratory positive airway pressure accessory intended to add moisture to the air stream respiratory positive all way pressure accesses J michies of the used with standard CPAP and gases for administration to the patient. The namistices, which have a maximum operating
or Bi-level Positive Airway Pressure (BiPAP) devices, which have a maximum operating of Bi-level I ostuve An way I rossure (2006) - 100 mot have automatic pressure titration capabilities.
Environment of Use / Patient Population Environment of Ose / I anche I openiator.
For single patient use in home, physician's office or hospital / institutional environment.
× Prescription Use (Per 21 CFR 801 Subpart D) AND/OR
Over-The-Counter-Use _ (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
f. R.W. Webster
Division of Anesthesiology, General Hospital, Infection Control, Dental Device
510(k) Number:
Page 1 of 1
1 / 008
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.