← Product Code [LLZ](/productcode/LLZ) · K040486

# MEDIPACS (K040486)

_Neobit Co., Ltd. · LLZ · Mar 11, 2004 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K040486

## Device Facts

- **Applicant:** Neobit Co., Ltd.
- **Product Code:** [LLZ](/productcode/LLZ.md)
- **Decision Date:** Mar 11, 2004
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Software as a Medical Device, 3rd-Party Reviewed

## Indications for Use

MEDIPACS™ is a device that receives digital images and data from various sources (i.e. CT scanners, MR scanners, ultrasound systems, R/F Units, computed & direct radiographic devices, secondary capture devices, scanners, imaging gateways, etc.). Images and data can be captured, stored, communicated, processed and displayed within the system and or across computer networks at distributed locations.

## Device Story

MEDIPACS is a Picture Archiving and Communication System (PACS) for handling, storing, and displaying digital medical images and patient data. Input sources include CT, MR, ultrasound, R/F units, and radiographic devices. The system transmits data over high-speed networks to centralized storage, allowing simultaneous access by multiple users at distributed locations. Operated by trained professionals (physicians, nurses, technicians) in clinical environments, the system facilitates image interpretation and reporting. It does not contact patients or control life-sustaining equipment; clinical decisions remain the responsibility of the interpreting physician.

## Clinical Evidence

No clinical data provided; bench testing only. The submission includes a hazard analysis and confirms compliance with voluntary standards.

## Technological Characteristics

PACS software for image management and display. Non-contact device. Networked architecture for data transmission and storage. Software level of concern: minor. No patient-contacting materials.

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- Marotech, Inc. PACS ([K012844](/device/K012844.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

MEDIGAPS

## MAR 1 1 2004

K040486
Page 1 of 2

51C(k) Summary of Safety and Effectiveness

This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990.

Date Prepared: December 22, 2003

Submitter's Information: 21 CFR 807.92(a)(1) Mr. Koo Chang Woon Director of R&D NEOBIT Co. Ltd. 7h BITVILLE 1237-33 Seocho-dong Deoch-ku Seoul, 137-072 Korea Tel: 82-2-598-7708 Fax: 82-2-598-7758

Trade Name, Common Name and Classification: 21 CFR 807.92(a)(2) Trade Name: MEDIPACS™ Picture Archiving Communications System Common Name: Device Classification: 892.2050 System, Image Processing Name:

Predicate Device: 21 CFR 807. 92(a)(3)

| Manufacturer:                    | Marotech, Inc.<br>7F Boryung B/D,<br>66-21 Wonnam-Dong, Jongro-Gu,<br>Seoul, 110-750, Korea |
|----------------------------------|---------------------------------------------------------------------------------------------|
| 510(k) Number:                   | K012844                                                                                     |
| Decision Date:                   | 11/08/2001                                                                                  |
| Decision:                        | Substantially Equivalent                                                                    |
| Panel Code device reviewed by:   | Radiology                                                                                   |
| Panel Code device classified by: | Radiology                                                                                   |
| Product Code:                    | LLZ                                                                                         |
| Device Classification Name:      | SYSTEM, IMAGE PROCESSING, RADIOLOGICAL                                                      |
| Regulation Number:               | Class II - 892.2050                                                                         |

Device Description: 21 CFR 807 92(a)(4)

MEDIPACS™ handles and displays various objects in a Picture Archive and Communication System (PACS) environment. These objects can be images, requests, patients, examination etc. The system transmits digital electronic images and generates reports over a high-speed network to centralized storage. After transmission, patient information and images are available throughout the facility to many users simultaneously.

{1}------------------------------------------------

Pérez López

### Indications for Use: 21 CFR 807 92(a)(5)

MEDIPACS™ is a device that receives digital images and data from various sources (i.e. CT scanners, MR scanners, ultrasound systems, R/F Units, computed & direct radiographic devices, secondary capture devices, scanners, imaging gateways, etc.). Images and data can be stored, communicated, processed and displayed within the system and or across computer networks at distributed locations. Typical users of this system are trained professionals, including but not limited to physicians, nurses, and technicians.

### Technological Characteristics: 21 CFR 807 92(a)(6)

The device does not contact the patient, nor does it control any life sustaining devices. A physician, providing ample opportunity for competent human intervention interprets images and information being displayed and printed.

#### Conclusion: 21 CFR 807 92(b)(1)

The 510(k) Pre-Market Notification for MEDIPACS™ contains adequate information and data to enable FDA - CDRH to determine substantial equivalence to the predicate device.

MEDIPACS™ has been and will be manufactured in accordance with the voluntary standards listed in the enclosed voluntary standard survey. The submission contains the results of a hazard analysis and the "Level of Concern for potential hazards have been classified as "minor".

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features a stylized depiction of an eagle with its wings spread, along with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged in a circular pattern around the eagle. The seal is in black and white.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

NEOBIT Co., Ltd. % Mr. Ned Devine Responsible Third Party Entela, Inc. 3033 Madison Ave. SE GRAND RAPIDS MI 49548 Re: K040486

Trade/Device Name: MEDIPACSTM System Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: 90 LLZ Dated: February 25, 2004 Received: February 25, 2004

Dear Mr. Devine:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050

MAR 1 1 2004

{3}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:

| 8xx. 1xxx                        | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4654 |
| Other                            | (301) 594-4692 |

Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours.

Nancy C. Brigdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

Page 2

{4}------------------------------------------------

# Indications for Use

510(k) Number (if known):

Device Name: MEDIPAC™ System

Indications For Use:

MEDIPACS™is a device that receives digital images and data from various sources (i.e. CT scanners, MR scanners, ultrasound systems, R/F Units, computed & direct radiographic devices, secondary capture devices, scanners, imaging gateways, etc.).

Images and data can be captured, stored, communicated, processed and displayed within the system and or across computer networks at distributed locations.

Prescription Use ___X__ (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use_ (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Nancy C. Hogdon

(Division Sign-Off Division of Reproductive, Abdom and Radiological Devices 510(k) Number

---

**Source:** [https://fda.innolitics.com/device/K040486](https://fda.innolitics.com/device/K040486)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
