VARELISA B2 GLYCOPROTEIN I IGG ANTIBODIES, MODELS 18748 AND 18796

K040449 · Pharmacia Deutschland GmbH · MSV · May 11, 2004 · Immunology

Device Facts

Record IDK040449
Device NameVARELISA B2 GLYCOPROTEIN I IGG ANTIBODIES, MODELS 18748 AND 18796
ApplicantPharmacia Deutschland GmbH
Product CodeMSV · Immunology
Decision DateMay 11, 2004
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.5660
Device ClassClass 2

Indications for Use

The Varelisa ß2-Glycoprotein I IgG Antibodies EIA kit is designed for the semiquantitative and qualitative determination of ß2-glycoprotein I IgG antibodies in serum or plasma. The presence of B2-glycoprotein I antibodies can be used in conjunction with clinical findings and other laboratory tests to aid in the diagnosis of thrombotic disorders related to the primary Antiphospholipid Syndrome or occurring secondary to systemic lupus erythematosus (SLE) or other autoimmune diseases.

Device Story

Indirect noncompetitive enzyme immunoassay (EIA) for semiquantitative/qualitative detection of ß2-glycoprotein I IgG antibodies in human serum or plasma. Device utilizes microplate strips coated with human purified ß2-glycoprotein I antigen. Patient sample added; specific antibodies bind to antigen. Enzyme-labeled conjugate added, binding to antigen-antibody complex. Substrate added; color formation rate proportional to antibody concentration. Performed in clinical laboratory setting by trained personnel. Output is optical density (OD) measurement used to determine antibody presence/concentration. Results assist clinicians in diagnosing thrombotic disorders related to Antiphospholipid Syndrome or secondary to autoimmune conditions like SLE. Benefits include standardized diagnostic support for autoimmune-related thrombosis.

Clinical Evidence

Bench testing only. Comparison study performed using positive, equivocal, and negative sera, externally defined calibrators, and samples from healthy subjects. Data demonstrates performance consistent with medical literature and comparability to predicate device. No clinical prospective/retrospective trials reported.

Technological Characteristics

Indirect noncompetitive enzyme immunoassay (EIA). Components: microplate strips coated with human purified ß2-glycoprotein I antigen, calibrators, controls, enzyme-labeled conjugate, substrate, stop solution, buffers. Manual or automated plate processing. Standalone diagnostic test system.

Indications for Use

Indicated for patients suspected of having thrombotic disorders related to primary Antiphospholipid Syndrome or secondary to systemic lupus erythematosis (SLE) or other autoimmune diseases.

Regulatory Classification

Identification

A multiple autoantibodies immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoantibodies (antibodies produced against the body's own tissues) in serum and other body fluids. Measurement of multiple autoantibodies aids in the diagnosis of autoimmune disorders (disease produced when the body's own tissues are injured by autoantibodies).

Predicate Devices

Submission Summary (Full Text)

{0} 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY DEVICE ONLY TEMPLATE A. 510(k) Number: k040449 B. Purpose of Submission: New device C. Analyte: $\beta 2$-Glycoprotein I IgG antibodies D. Type of Test: Semiquantitative and qualitative ELISA E. Applicant: Pharmacia Deutschland GmbH/Sweden Diagnostics (Germany) GmbH F. Proprietary and Established Names: Varelisa B2-Glycoprotein I IgG Antibodies EIA Kit G. Regulatory Information: 1. Regulation section: 21 CFR 866.5660, Multiple Autoantibodies Immunological System 2. Classification: Class II 3. Product Code: MSV, System, Test, Autoantibodies, $\beta 2$-Glycoprotein I ($\beta 2$-GPI) 4. Panel: Immunology (82) H. Intended Use: 1. Intended use(s): The Varelisa B2-Glycoprotein I IgG Antibodies Kit is designed for the semiquantitative and qualitative determination of $\beta 2$-glycoprotein I IgG antibodies in serum or plasma. 2. Indication(s) for use: The presence of $\beta 2$-glycoprotein I antibodies can be used in conjunction with clinical findings and other laboratory tests to aid in the diagnosis of thrombotic disorders related to the primary Antiphospholipid Syndrome or occurring secondary to systemic lupus erythematosis (SLE) or other autoimmune diseases. 3. Special condition for use statement(s): Prescription use only. {1} Page 2 of 6 4. Special instrument Requirements: None. I. Device Description: The Varelisa B2-Glycoprotein I IgG Antibodies kit consists of an ELISA strip wells (12 strips X 8 wells) coated with purified human β2-glycoprotein I antigen. It also contains ancillary reagents such as calibrators, positive and negative controls, wash buffer, sample diluent, antibody-chromogen conjugate, enzyme substrate, stop solution for enzyme reaction, and a frame to hold ELISA strip wells. Also included are a package insert and a control certificate. J. Substantial Equivalence Information: 1. Predicate device name(s): INOVA QUANTA Lite B2 GPI IgG 2. Predicate K number(s): k970551 3. Comparison with predicate: | Similarities | | | | --- | --- | --- | | Item | Device | Predicate | | Intended Use | Semiquantitative and qualitative determination of β2-glycoprotein I IgG antibodies in serum or plasma for use in conjunction with clinical findings and other laboratory tests to aid in the diagnosis of thrombotic disorders related to primary Antiphospholipid Syndrome or occurring secondary to systemic lupus erythematosis (SLE) or other autoimmune diseases. | Qualitative detection of β2-GPI IgG antibodies in serum for use in conjunction with clinical findings and other laboratory tests to aid in the diagnosis of certain autoimmune disease thrombotic disorders, such as those secondary to systemic lupus erythematosis (SLE) or other lupus-like thrombotic diseases. | | Antigen | Human affinity-purified β2-glycoprotein I | β2-glycoprotein I, unknown source | | Conjugate | Anti-human IgG Antibody Conjugate, Horseradish peroxidase | Same | | Assay Principle | Indirect noncompetitive enzyme immunoassay | Same | | Sample dilution | 1:101 | Same | {2} Page 3 of 6 | Differences | | | | --- | --- | --- | | Item | Device | Predicate | | Specimen | Serum and plasma | Serum | | Calibrators | Six prediluted calibrators | Five prediluted calibrators | | Assay evaluation | OD cut-off | Decision point | ## K. Standard/Guidance Document Referenced (if applicable): None ## L. Test Principle: The Varelisa B2-Glycoprotein I IgG Antibodies EIA Kit is an indirect noncompetitive enzyme immunoassay. The wells of the microplate are coated with human purified $\beta 2$-glycoprotein I antigen. Antibodies specific for $\beta 2$-glycoprotein I present in the applied patient sample bind to the antigen and are detected by an enzyme labeled secondary antibody complex, which leads to the formation of an enzyme labeled conjugate-antibody-antigen complex. The enzyme labeled antigen-antibody complex converts the added substrate to form a colored solution. The rate of color formation from the chromogen (substrate) is a function of the amount of conjugate complexed with the bound antibody and thus is proportional to the initial concentration of the respective antibodies in the patient sample. The antibody concentration is determined by comparison to a calibration curve run with the assay and is reported in U/mL. ## M. Performance Characteristics (if/when applicable): ### 1. Analytical performance: #### a. Precision/Reproducibility: Five serum samples (low, medium, high, equivocal, negative) were selected from a serum bank. The samples were diluted 1:101 and measured in 5 runs, with 4 replicates per run. Standards and controls were analyzed in triplicate. Specified target values were: variance "within" < 12%; "between" < 15%; "total" < 15%. The specifications were met for each sample, with the following variances: | Sample ID | | Run 1 | Run 2 | Run 3 | Run 4 | Run 5 | Mean (U/mL) | Variance | | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | | | | | | | | Within | Between | total | | ZSB 20429 | Mean (u/mL) CV% | 73.5 2.69 | 74.8 0.72 | 75.2 1.39 | 79.8 3.15 | 74.5 4.45 | 75.6 | 2.81 | 2.92 | 4.06 | | ZSB 13214 | Mean (u/mL) CV% | 37.7 2.15 | 40.8 2.89 | 39.6 2.28 | 42.5 2.58 | 40.1 1.70 | 40.1 | 2.37 | 4.20 | 4.82 | | ZSB 19162 | Mean (u/mL) CV% | 21.4 3.26 | 22.1 1.97 | 21.4 1.11 | 22.6 3.48 | 21.9 2.50 | 21.9 | 2.63 | 1.91 | 3.25 | | negative | Mean (u/mL) CV% | 7.4 8.6 | 6.8 9.4 | 7.0 8.3 | 7.0 21.4 | 8.0 5.2 | 7.2 | 11.7 | 2.9 | 12.0 | | equiv. | Mean (u/mL) | 17.6 | 17.6 | 17.2 | 19.8 | 19.2 | 18.3 | 8.2 | 4.8 | 9.5 | {3} Page 4 of 6 b. Linearity/assay reportable range: Three samples selected from a serum bank were diluted 1:1, 2:3, 1:2, 1:4, 1:8, 1:16, and 1:32 in Sample Diluent. Standards, Controls and each dilution were measured in triplicate. Specifications were that observed/expected units ratio would not differ by more than ± 20% for at least 3 successive dilutions of each tested sample. The dilutions met the criteria and were considered linear. Recovery of analyte was assayed using two samples from a serum bank. Each sample was tested in six replicates to determine pre-existing antibodies. Each sample was then diluted 1:101 and spiked with 1/10 volume of Standards S1 to S6 and measured in duplicate. Acceptance criteria were that recovery would be 100 ± 20%. In all cases for both samples, the acceptance criteria were met. c. Traceability (controls, calibrators, or method): Not traceable to external sources. The consistency of the calibration material is given by a testing procedure comprising several internal target values that must be met. Each new production of calibration material is adapted to the corresponding master calibration material. d. Detection limit: Sample Buffer was diluted according to Directions for Use and measured 56 times on one plate. Standards and Controls were analyzed in four replicates. Two dilutions of the first standard point, S2 (1:2, 1:4) were run in four replicates. Analytical sensitivity was calculated as the mean of the optical densities of the sample buffer plus 3SD and expressed in U/mL. The discrimination value D, proves the ability to discriminate between the lowest standard point and the background, was performed as follows: $$ D = \frac {\eta_ {B} - \eta_ {A}}{\sqrt {\left(\sigma_ {B} ^ {2} - \sigma_ {A} ^ {2}\right)}} $$ Where: A is Sample Buffer; B is Standard S2; ηA, ηB are mean OD's; σA, σB are SD's. Specifications were that analytical sensitivity should be below 1 U/mL, the mean + 3SD of sample buffer should be < standard point S2, and the discrimination value should be > 2.0. The mean +3SD was 0.016 OD, which corresponded to analytical sensitivity of 0.04 U/mL. The other criteria were met. e. Analytical specificity: Interference was tested against potentially interfering substances found in blood: bilirubin, hemoglobin, chyle, and rheumatoid factor. {4} Page 5 of 6 Three samples from a serum bank were diluted 1:101 and spiked with different super-normal amounts of interferent. Spiked samples were analyzed in triplicate. Acceptance criteria were that spiked samples should show no more than 20% variation from unspiked sample. | Additives (final sample concentration) | I | II | III | IV | V | | --- | --- | --- | --- | --- | --- | | Bilirubin F (mg/dL) | 0.0 | 4.8 | 9.5 | 14.3 | 19.0 | | Bilirubin C (mg/dL) | 0.0 | 5.5 | 11.0 | 16.5 | 22.0 | | Chyle (Units) | 0.0 | 357.5 | 715.0 | 1072.5 | 1430.0 | | Hemoglobin (mg/dL) | 0.0 | 122.5 | 245.0 | 367.5 | 490.0 | | RF (IU/mL) | 0.0 | 98.0 | 294.0 | 490.0 | - | All samples met acceptance criteria and showed no significant influence on the test results. Additional interference tests on negative and equivocal sera with the same interferents demonstrated a negligible effect of the interferents on the equivocal samples, and a more pronounced effect on the negative samples. The negative samples, however, remained negative, so the interference was judged to be not important. f. Assay cut-off: The semi-quantitative cut-points were determined by measuring 432 samples from apparently healthy Caucasian donors, equally distributed by sex and age. Diluted samples, standards, and controls were analyzed in duplicate. Specifications called for the mean +2SD to be smaller than the lower limit of the equivocal range, and the mean +3SD to be smaller than the upper limit of the equivocal range. The following values were selected for negative, equivocal, and positive: <10 U/mL = negative 10-15 U/mL = equivocal >15 U/mL = positive The specifications were fulfilled and no difference was detected between sexes or ages. 2. Comparison studies: a. Method comparison with predicate device: The following samples were used in comparing the new device with 42 samples from a serum bank at Pharmacia, representing SLE, lupus-like, unknown clinical diagnosis, and healthy individuals. The results were compared semiquantitatively (Varelisa) vs qualitatively {5} Page 6 of 6 (QUANTA Lite) in a 6 field analysis. 37 of 42 sera agreed qualitatively (pos/neg) for overall agreement of 88.1%. Three sera were equivocal by the Varelisa assay and either positive (1) or negative (2) by the qualitative predicate. Two sera were positive by Varelisa and negative by the predicate. | Varelisa B2-Glycoprotein antibodies assay Lot SF18 | n = 42 | INOVA QUANTA Lite Lot 170242 | | | --- | --- | --- | --- | | | | Positive | Negative | | | positive | 20 | 0 | | | equivocal | 1 | 1 | | | negative | 1 | 17 | b. Matrix comparison: Twenty blood donor samples in the positive and negative range of the assay were tested in serum, citrate plasma, heparin plasma, and EDTA plasma, in the negative range of the assay. No significant differences were noted for any of the matrices. 3. Clinical studies: a. Clinical sensitivity: Not applicable b. Clinical specificity: Not applicable c. Other clinical supportive data (when a and b are not applicable): Not applicable 4. Clinical cut-off: Not applicable 5. Expected values/Reference range: Expected value for normal healthy individuals is negative, or <10 U/mL. In a study of 432 healthy Caucasians, the following values were obtained: N: 432 Mean: 1.6 U/mL Mean +2SD 8.8 U/mL Median: 0.9 U/mL 95th percentile 4.5 U/mL N. Conclusion: The submitted information in this premarket notification is complete, and supports a substantial equivalence decision.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...