DOME CX DIGITAL FLAT-PANEL DISPLAY SYSTEM, MODEL C5I
K040443 · Dome Imaging Systems, Inc. · LLZ · Jun 16, 2004 · Radiology
Device Facts
Record ID
K040443
Device Name
DOME CX DIGITAL FLAT-PANEL DISPLAY SYSTEM, MODEL C5I
Applicant
Dome Imaging Systems, Inc.
Product Code
LLZ · Radiology
Decision Date
Jun 16, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
3rd-Party Reviewed
Indications for Use
The DOME CX™ DIGITAL FLAT-PANEL DISPLAY SYSTEM™, Model C5i™ is intended to be displaying and viewing radiographs of the breast for review and analysis by trained medical practitioners.
Device Story
DOME CX Model C5i is a high-resolution active matrix liquid-crystal display (AMLCD) monitor system; designed for displaying grayscale medical images. Input via Digital Visual Interface (DVI) 1.0 from a graphics controller; utilizes integrated Cold Cathode Fluorescent Tube (CCFT) backlight. System provides 5-megapixel resolution (2048 x 2500 pixels). Used in clinical settings by trained medical practitioners for diagnostic and referral review of breast radiographs. Device does not contact patient; provides visual output for human interpretation to support clinical decision-making.
Clinical Evidence
Bench testing only. Hazard analysis performed; potential hazards classified as Minor. No clinical data provided.
Technological Characteristics
Active matrix liquid-crystal display (AMLCD) with integrated Cold Cathode Fluorescent Tube (CCFT) backlight. 5-megapixel resolution (2048 x 2500 pixels). Interface: DVI 1.0. Standalone display system; does not contact patient. No life-sustaining functions.
Indications for Use
Indicated for trained medical practitioners to display and view breast radiographs for review and analysis.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
Predicate Devices
DOME CX™ DIGITAL FLAT-PANEL DISPLAY SYSTEM™, Model C5i™ (K032202)
Submission Summary (Full Text)
{1}------------------------------------------------
КОНО443
#### JUN 1 6 2004
400 Fifth Avenue, Waltham, MA 02451-8738, USA Phone: +1-781-895-1155 Fax: +1-781-895-1133 www.planar.com
## 510(k) Summary of Safety and Effectiveness
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990.
Date Prepared:
January 19, 2004
Submitter's Information: 21 CFR 807.92(a)(1)
Planar Systems, Inc. 400 Fifth Ave. Waltham, MA 02451-8738 USA
Trade Name, Common Name and Classification: 21 CFR 807.92(a)(2)
| Trade Name: | DOME CX™ DIGITAL FLAT-PANEL DISPLAY SYSTEM |
|-----------------|--------------------------------------------|
| | Model C5i™ |
| Common Name: | Mammographic X-ray System |
| Classification: | 892.1710 |
| Name: | System, Image Processing |
Predicate Devices: 21 CFR 807. 92(a)(3)
| Device | SYSTEM, IMAGE |
|-------------------|-----------------------------------------------------------------|
| Classification | PROCESSING, |
| Name | RADIOLOGICAL |
| Regulation Number | 892.2050 |
| 510(k) Number | K032202 |
| Device Name | DOME CX™ DIGITAL FLAT-<br>PANEL DISPLAY SYSTEM™,<br>Model C5i™ |
| Applicant | Planar Systems, Inc.<br>400 Fifth Ave<br>Waltham, MA 02451-8738 |
| Product Code | 90 LLZ |
Device Description: 21 CFR 807 92(a)(4)
The DOME CX™ DIGITAL FLAT-PANEL DISPLAY SYSTEM™, Model C5i™ is a flat panel hi-resolution LCD monitor system for displaying gray scale medical images for diagnostic and referral use.
The Displays use active matrix liquid-crystal display (AMLCD) panels that generate lower The Displays use active maths liquit aryonal and consume less power than traditional CRT displays. The Displays use a common internal interface controller that connects directly displays - The Dioplay controller via a Digital Visual Interface (DVI) 1.0 interface. to a common graphios arepray stors (TFTs) control transmissive liquid-crystal elements
{2}------------------------------------------------
and use integrated Cold Cathode Fluorescent Tube (CCFT) backlight systems The and use miegrated Cold Cathodo Flabroom 7 Section 1 Section of 2018 DOME CSI uisplays five megapixolo of cattle LCD display panel, integrated drive x 2500 o-bit pixers. The system mora an external power supply. Optional connections to electronics, megrated backlight, and an oxlority, including keyboard, mouse, and calibration technology.
#### Indications for Use: 21 CFR 807 92(a)(5)
The DOME CX™ DIGITAL FLAT-PANEL DISPLAY SYSTEM™, Model C5i™ is intended to be displaying and viewing radiographs of the breast for review and analysis by trained medical practitioners.
#### Technological Characteristics: 21 CFR 807 92(a)(6)
oglour one is an image display system consisting of computer software and components. The device is an image uisplay byotont bonoloing or control any life sustaining devices. A The device does not contact the pation, nor ides ample opportunity for competent human intervention to interpret images and information being displayed.
Conclusion: 21 CFR 807 92(b)(1)
- John 21 OF C OF C OF C ( ) DIGITAL FLAT-PANEL DISPLAY SYSTEM, Model . C5i™ is substantially equivalent to the predicate device, DOME CX™ DIGITAL FLAT-PANEL DISPLAY SYSTEM, Model C5i™.
- The new and predicate devices are both 5Million Pixel Medical Flat Panel Display . The hew and predicate do need in be displaying and viewing radiographs for review and Systems intendod to be acou in title new device and predicate devices are analysis by trainou mot in the areas of technical characteristics, general function, substantially cquivalent in the arous ofference between the two devices does not affect safety or efficacy.
- The 510(k) Pre-Market Notification for the DOME CX™ DIGITAL FLAT-PANEL DISPLAY . SYSTEM™, Model C5i™ contains adequate information and data to enable FDA - CDRH to determine substantial equivalence to the predicate devices.
- The Planar Systems, Inc., DOME CX™ DIGITAL FLAT-PANEL DISPLAY SYSTEM™, . 11:01 tarar e your be manufactured by in accordance with voluntary and safety standards.
- The submission contains the results of a hazard analysis and the potential hazards have . been classified as Minor.
{3}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes forming its body and wings. The eagle is enclosed within a circle, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the circumference of the circle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
### JUN 1 6 2004
Dome Imaging Systems, Inc. % Mr. Ned Devine Responsible Third Party Official Entela, Inc. 3033 Madison Ave. SE GRAND RAPIDS MI 49548
Re: K040443 18-16-11-15
Trade/Device Name: DOME™ CX Digital Flat-Panel Display System™ Model C5iTM Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: 90 LLZ Dated: May 27, 2004 Received: June 1, 2004
Dear Mr. Devine:
We have reviewed your Section 510(k) premarket notification of intent to market the device We have reviewed your Section > Po(x) premainted in equivalent (for the indications referenced above and nave uctering a the actived predicate devices marketed in interstate for use stated in the encrosure, to regally manatiment date of the Medical Device Amendments, or to commerce prof 10 May 28, 1976, the enactined with the provisions of the Federal Food, Drug, devices that have been reclassined in according approval of a premarket approval application (PMA).
and Cosmetic Act (Act) that do not require approval of a premarket af the and Cosment Act (Act) that do not require appro the general controls provisions of the Act. The You may, therefore, market the device, subject to include registration, listing of
general controls provisions of the Act include requirements for any and general controls provisions of the rice nellations against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is classined (see above) mio existing major regulations affecting your device can
may be subject to such additional controls. Existing major regulation FDA mo may be subject to such additional controller Litter & Parts 800 to 898. In addition, FDA may be found in the Code of Poderal England Stages in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please be advised that FDA s issualled of a saccession of the requirements of the Act of the Act one mate agempl that FDA has made a determination had your do to received agencies. You must comply
or any Federal statutes and regulations administered by other Federal and CEP Pat or any Federal statures and regulations administration and listing (2) CFR Patt (2) CFR Patt with all the Act STequirements, melaaning, out her man and manufacturing as set forth in the 807), labeling (21 CFR Part 001), good handler and if applicable, the electronic product
quality systems (QS) regulation (21 CFR Part 820); and if applicable, 1000-1060 quality Systems (QS) regalation (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k)
In a legally This letter will allow you to begin marketing your antial equivalence of your device to a legally premarket notification. The PDA inding of subsition for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please
r the same of the same of a to a transmall on the same on the regulation If you desire specific advice for your device on our last ling ages on the regulation number at the top of the letter:
| 8xx.1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Additionally, for questions on the promotive and the regulation entitled, "Misbranding
Office of Compliance at (301) 594-4639. Also, please of their . Other accepted Office of Compliance at (301) 374-1057. This, pressed 377) you may obtain. Other general by reference to premarket nonmeanon - (Dr of Act may be obtained from the Division of Shall information on your responsionities and its toll-free number (800) 638-2041 or Manufacturers, International and Ochounter one of the courted.html dsmamain.html.
Sincerely yours,
Nancy C. Brydon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known): K040443
Device Name: Device Name.
DOME CX™ DIGITAL FLAT-PANEL DISPLAY SYSTEM™, Model C5i™ from Planar Systems, Inc.
Indications For Use:
The DOME CX™ DIGITAL FLAT-PANEL DISPLAY SYSTEM™, Model C5¡™ is The DOME OX - Dictirite Feing and viewing radiographs of the breast for intended to be boos by trained medical practitioners.
Prescription Use __________________ ANB/OR Over-The-Counter Use ______________________________________________________________________________________________________________ | Tooonption 01 Subpart D) (21 CFF 807 Subpart C) (Part 21 CFR 801 Subparto) (21 OF Froof Subjar 0)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concyrrence of CDRH, Office of Device Evaluation (ODE)
Luvila Segura
Page 1 of 1
(Division Sign-Cit) Division of Reproductive, Abdominal, and Radiological Device 510(k) ! Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.