← Product Code [FOZ](/productcode/FOZ) · K040427

# FIBREX CATHETER PATENCY DEVICE (K040427)

_Medical Device Technologies, Inc. · FOZ · May 3, 2004 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K040427

## Device Facts

- **Applicant:** Medical Device Technologies, Inc.
- **Product Code:** [FOZ](/productcode/FOZ.md)
- **Decision Date:** May 3, 2004
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5200
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Therapeutic, 3rd-Party Reviewed

## Indications for Use

The Fibrex™ Catheter Patency Device is intended to disrupt fibrin sheath formation on both the external and internal surface of vascular access catheters, thus restoring functional patency.

## Device Story

Tri-axial delivery catheter assembly; includes outer catheter, inner guide structure, and inner stripping coil. Stripping coil composed of shape-memory/super-elastic Nitinol; preformed helical shape wraps around vascular catheter exterior to mechanically strip fibrin build-up. Hand assembly controls guide catheter movement and stripping coil insertion. Device couples to luer connector of venous lumen of existing vascular catheter. Used by clinicians to restore patency in vascular access catheters; benefits patient by maintaining catheter function and avoiding replacement. No electronic or software components.

## Clinical Evidence

Animal study conducted to evaluate fibrin stripping capabilities. Bench testing included ring retention and tensile tests. Results indicate performance comparable to predicate device.

## Technological Characteristics

Tri-axial delivery catheter assembly. Materials: Polyimid/PTFE composite (outer catheter), nylon (inner dual lumen tube), Nitinol (stripping coil/wire), gold (distal marking bands). Mechanical operation via manual hand assembly. No energy source, software, or connectivity.

## Regulatory Identification

An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.

## Predicate Devices

- En-Snare™ Endovascular Snare and Catheter ([K021606](/device/K021606.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Ko40427

MAY - 3 2004

Rev .: April 23, 2004

| 510(k) SUMMARY           |                                                                                   |            |
|--------------------------|-----------------------------------------------------------------------------------|------------|
| APPLICANT:               | Medical Device Technologies, Inc.<br>3600 SW 47th Avenue<br>Gainesville, FL 32608 |            |
| CONTACT:                 | Karl Swartz<br>Quality Assurance Manager                                          |            |
| TELEPHONE:               | (352)338-0440<br>fax (352)338-0662                                                |            |
| TRADE NAMES:             | Fibrex™ Catheter Patency Device                                                   |            |
| COMMON NAME:             | Intravascular therapeutic catheter                                                |            |
| CLASSIFICATION NAME:     | Catheter,intravascular,therapeutic,short, Regulation Number<br>880.5200           |            |
| PRODUCT CODE:            | FOZ                                                                               |            |
| PANEL:                   | General Hospital                                                                  |            |
| SUBSTANTIAL EQUIVALENCE: |                                                                                   |            |
| Company Name             | Product Name                                                                      | 510(k) No. |

En-Snare™ Endovascular Snare and Catheter K021606 Medical Device Technologies

### DESCRIPTION OF DEVICE:

The Fibrex™ Catheter Patcncy Device is a tri-axial delivery catheter assembly comprised of an outer catheter storing an inner guide structure and an inner stripping coil. The inner guide structure is for directing and guiding the stripping coil through the catheter to the free end of the catheter and back up around the exterior surface of the vascular catheter. The stripping coil is formed of a shape memory or super-elastic material, wherein the stripping coil is preformed to assume a helical shape wrapping about the outer surface of the vascular catheter in a manner providing for the stripping of build-up from the outer surface of the vascular catheter when the stripping coil is moved relative to the vascular catheter.

{1}------------------------------------------------

Korlove 1

# Rev .: April 23, 2004

The tri-axial delivery catheter assembly is attached to an injection molded plastic hand assembly that controls the movement of the guide catheter and subsequent insertion of the stripping coil. The delivery catheter assembly is adapted for selectively coupling to a luer connector secured to the venous lumen of a traditional vascular catheter.

The outer catheter of the Fibrex™ device is made from a biocompatible Polyimid/PTFF. composite. The inner dual lumen tube is made from nylon. A Nitinol wire is housed in one lumen of the guide curve catheter to provide directional memory for the actuation and operation of the Nitinol stripping coil . Both the outer catheter and the guide catheter contain distal marking bands made from gold.

# INDICATIONS FOR USE:

The Fibrex™ Catheter Patency Device is intended to disrupt fibrin sheath formation on both the external and internal surface of vascular access catheters, thus restoring functional patency.

# FUNCTIONAL & SAFETY TESTING:

The Fibrex™ Catheter Patency Device was subjected to an animal study to evaluate the fibrin stripping capabilities of the device. It was also subjected to ring retention, and tensile tests. The results of the testing indicated that they are comparable to the predicate device.

### TECHNICAL COMPARISON:

The following attributes of the Fibrex™ Catheter Patency Device were examined and found to be comparable to the predicate device:

- Intended use 1.
- 2. French Sizes
- Length 3.
- র্বা Lumens
- Distal end configuration ડ .
- Intended anatomical location of distal end 6.
- Proximal end configuration 7.
- 8. Materials
- ் Labeling

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle or bird symbol, composed of three curved lines, representing the department's mission of protecting the health of all Americans and providing essential human services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the bird symbol.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAY - 3 2004

Medical Device Technologies, Inc. c/o Mr. Karl Swartz Quality Assurance Manager 3600 SW 47th Avenue Gainsville, FL 32608

K040427 Re: FibrexTM Catheter Patency Device Regulation Number: 21 CFR 880.5200 Regulation Name: Intravascular catheter Regulatory Class: Class II Product Code: FOZ Dated: February 19, 2004 Received: February 19, 2004

### Dear Mr. Swartz:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Mr. Karl Swartz

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4648. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address

http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Dmina R. Vochner

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

{4}------------------------------------------------

Keyour7

Page 1 of 1

510(k) Number (if known): K040427

Device Name: Fibrex™ Catheter Patency Device

Indications for Use:

The Fibrex™ Catheter Patency Device is intended to disrupt fibrin sheath formation on both the external and internal surface of vascular access catheters, thus restoring functional patency.

Prescription Use __ X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Duna D. Vachner

(Division Sign-Off) Division of Cardiovascular Devices

Page ___ of ___

510(k) Number_k040477

---

**Source:** [https://fda.innolitics.com/device/K040427](https://fda.innolitics.com/device/K040427)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).