DIAPHARMA FACTOR X KIT

K040362 · Diapharma Group, Inc. · GJT · Jul 12, 2004 · Hematology

Device Facts

Record IDK040362
Device NameDIAPHARMA FACTOR X KIT
ApplicantDiapharma Group, Inc.
Product CodeGJT · Hematology
Decision DateJul 12, 2004
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 864.7290
Device ClassClass 2

Indications for Use

The DiaPharma Factor X Kit is an in vitro diagnostic test kit for the quantitative determination of Factor X activity in human citrated plasma.

Device Story

DiaPharma Factor X Kit is an in vitro diagnostic test for quantitative measurement of Factor X activity in human citrated plasma. Used in clinical laboratory settings by trained personnel. Provides healthcare providers with Factor X activity levels to monitor patients on oral anticoagulant therapy (warfarin), particularly when INR is unreliable due to conditions like lupus inhibitors, and to screen for Factor X deficiencies. Results assist clinicians in managing anticoagulation therapy and diagnosing coagulation disorders.

Clinical Evidence

No clinical data provided; substantial equivalence is based on bench testing and performance characteristics of the in vitro diagnostic assay.

Technological Characteristics

In vitro diagnostic test kit for Factor X activity. Quantitative assay performed on human citrated plasma. Standard laboratory coagulation testing methodology.

Indications for Use

Indicated for quantitative determination of Factor X activity in human citrated plasma. Useful for monitoring patients on oral anticoagulant therapy (warfarin) where baseline PT values are prolonged and INR is unreliable (e.g., patients with lupus inhibitors), and for screening for Factor X deficiencies.

Submission Summary (Full Text)

{1} 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY DEVICE ONLY TEMPLATE A. 510(k) Number: K040362 B. Purpose for Submission: To seek clearance of a Factor X (FX) Assay C. Analyte: Factor X D. Type of Test: Quantitative chromogenic assay E. Applicant: DiaPharma Group Inc. F. Proprietary and Established Names: DiaPharma Factor X Kit G. Regulatory Information: 1. Regulation section: 21 CFR 864.7290, Factor Deficiency Test 2. Classification: II 3. Product Code: GGP 4. Panel: 81 Hematology H. Intended Use: 1. Intended use(s): The DiaPharma Factor X Kit is an in vitro diagnostic test kit for the quantitative determination of Factor X activity in human citrated plasma. 2. Indication(s) for use: Factor X activity is useful for monitoring patients on oral anticoagulant therapy (warfarin) where baseline PT values may be prolonged and INR results are not reliable, such as in OAC patients with lupus inhibitors. The DiaPharma FX kit provides health care providers with a tool for clinical decision making. The DiaPharma Factor X Kit is also useful for screening for factor X deficiencies. 3. Special condition for use statement(s): {2} Page 2 of 4 4. Special instrument Requirements: I. Device Description: Each DiaPharma Factor X Kit consists of enough reagents for ~400 microplate determinations and ~200 automated-method determinations. Each kit contains: chromogenic substrate, Russell Viper Venom, Calcium Chloride, and a tris buffer containing Polybrene® to neutralize heparin. J. Substantial Equivalence Information: 1. Predicate device name(s): Diagnostic Stago Stago Factor X Deficient Plasma 2. Predicate K number(s): K933441 3. Comparison with predicate: | Similarities | | | | --- | --- | --- | | Item | Device | Predicate | | Intended Use | For the quantitative determination of Factor X activity in human citrated plasma | same | | Test Availability | Professional use | same | | Result Interpretation | % Normal | same | | Type of Test | Quantitative | same | | Differences | | | | Item | Device | Predicate | | Indications for use | For the detection of possible clotting factor deficiencies in the extrinsic coagulation pathway, and to monitor patients receiving anticoagulant therapy who have developed lupus antibodies | A general screening procedure for the detection of possible clotting factor deficiencies in the extrinsic coagulation pathway | | Reagent/Measurement System | Lyophilized reagent (Factor X chromogenic substrate and Russell’s Viper Venom) that requires reconstitution; measurement system may be microtiter plate or automated analyzer | Lyophilized reagent (thromboplastin) that requires reconstitution; measurement system by automated analyzer in optical or mechanical modes. | | Methodology | Chromogenic | Clotting | K. Standard/Guidance Document Referenced (if applicable): {3} Page 3 of 4 L. Test Principle: The method is based on a two-stage principle. Factor X in the sample is activated in the presence of calcium and Russell Viper Venom to activated Factor X (FXa). The generated FXa hydrolyses the chromogenic substrate and liberate the chromogenic group, pNA. The color is read spectrophotometrically, and is proportional to the FX activity in the sample. A standard curve is required for each Factor X assay "run". Five standards (calibrators) representing levels of 100%A, 50%, 25%, and 10% (percents of normal) must be included in each run. For standardization, the laboratory may use commercially supplied hemostasis calibration plasma, or may use a pool of calibrated fresh frozen normal plasma. M. Performance Characteristics (if/when applicable): 1. Analytical performance: a. Precision/Reproducibility: Six samples were assayed in one FX kit run in replicates of 10 using a microplate. The samples were two controls (one normal and one prolonged), two patient samples with normal levels, and two patients with prolonged levels. Precision Data | FX Concentration (%) | CV Within Run | | --- | --- | | Normal (98 – 107%) | 1.9 – 12.0% | | Abnormal (23 – 32%) | 4.7 – 19.6% | The central tendency for within-run precision of the DiaPharma Factor X kit is approximately 7 %CV. b. Linearity/assay reportable range: 10-130 % of normal c. Traceability (controls, calibrators, or method): d. Detection limit: 10% of normal e. Analytical specificity: Plasma blanks are recommended for icteric, lipemic and hemolyzed samples. The Polybrene® contained in the buffer will neutralize heparin concentrations up to 30 U/ml f. Assay cut-off: 2. Comparison studies: a. Method comparison with predicate device: A 2 site clinical study (n=155) comparing the DiaPharma Factor X Kit to the predicate device using Diagnostica STA Neoplastine reagent (K922565) on the STA Compact (K961579). 95 samples were analyzed at site 1, 60 samples at site 2. A linear regression was performed, yielding: $y = 0.76 \cdot \text{FX Clotting} + 21.77$ , $R2 = 0.90$ {4} Confidence Interval for Slope & Intercept: | FX Clotting (%) | | FX Chromogenic (%) | | | --- | --- | --- | --- | | Confidence Level (95.0%) | 8.25 | Confidence Level (95.0%) | 6.65 | | | Lower 95% | Upper 95% | | --- | --- | --- | | Intercept | 18.77 | 24.78 | | X variable | 0.72 | 0.80 | ![img-0.jpeg](img-0.jpeg) # b. Matrix comparison: # 3. Clinical studies: a. Clinical sensitivity: b. Clinical specificity: c. Other clinical supportive data (when $a$ and $b$ are not applicable): Clinical studies were done with a total of 241 individuals (92 normals + 149 OAC patients) across three sites. Citrated plasma samples from each individual were assayed by the FX Kit and by the routine prothrombin method at each site. The data were compared by log-log linear regression, and showed a correlation coefficient ("r") of 0.89. # 4. Clinical cut-off: # 5. Expected values/Reference range: DiaPharma Factor X has been evaluated in a study with 65 subjects the samples were tested according to standard procedure, and the range of results (90% limits) was $130\%$ to $59\%$ . # N. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...