K040289 · Cynosure, Inc. · GEX · Apr 30, 2004 · General, Plastic Surgery
Device Facts
Record ID
K040289
Device Name
SMARTLITE D FREQUENCY DOUBLED ND:YAG LASER
Applicant
Cynosure, Inc.
Product Code
GEX · General, Plastic Surgery
Decision Date
Apr 30, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The SmartLite D laser device is intended for ablation, incision, excision, vaporization and coagulation of soft tissue, in ophthalmology, ENT, and dermatology. In addition, the laser is intended for application in dentistry including: Frenectomy . Treatment of Oral Mucous Cysts . Treatment of Benign Vascular Lesions . Photocoagulation of Superficial Vessels . Vaporization of Superficial Blood Vessels or Lymphs Containing Vessels ● Treatment of Superficial Tongue Lesions . Tissue Management and Hemostasis for Crown and Bridge Impressions ◆ Incision and Drainage for Abscess . Gingivoplasty / Gingivectomy . Hemostasis during Dental Procedurcs . Operculectomy (Operculotomy) . Excisional Biopsy . Free Gingival Graft (Adjunct) . Vestibuloplasty . Soft Gutta Percha . Treatment of Canker sores, Herpetic Lesions, and Aphthous Ulcers ● Laser-assisted Bleaching / Whitening of the Teeth .
Device Story
SmartLite D is a frequency-doubled Nd:YAG laser (532 nm wavelength) utilizing a Nd:YAG crystal rod as the lasing medium. Device is operated by a physician or dentist via footswitch. It performs ablation, incision, excision, coagulation, and vaporization of soft tissue. The laser energy is delivered to the target site to achieve the desired surgical or therapeutic effect. The device is a compact unit (10 kg, 27x26x36 cm) intended for clinical use. It provides a surgical tool for various soft tissue procedures and dental applications, including tooth whitening.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Frequency-doubled Nd:YAG laser; 532 nm wavelength; Nd:YAG crystal rod medium; footswitch activation; 230 VAC, 15A, 50-60 Hz power; 10 kg weight; 27x26x36 cm dimensions.
Indications for Use
Indicated for ablation, incision, excision, vaporization, and coagulation of soft tissue in ophthalmology, ENT, dermatology, and dentistry (including specific oral lesions, gingival procedures, and tooth whitening).
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
Fisma Elite Family of Lasers, frequency doubled Nd:YAG
Fisma Dental 200
Fisma Dental 300
Fisma Dental 400
Submission Summary (Full Text)
{0}------------------------------------------------
APR 3 0 2004
K 040289
## 510(K) Summary
| Submitter: | Cynosure, Inc.<br>10 Elizabeth Drive<br>Chelmsford, MA 01824 |
|-------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | George Cho<br>Senior Vice President of Medical Technology |
| Date Summary Prepared: | February 5, 2004 |
| Device Trade Name: | SmartLite D Laser |
| Common Name: | Medical Laser System |
| Classification Name: | Instrument, surgical, powered, laser<br>79-GEX<br>21 CFR 878.48 |
| Equivalent Device: | Fisma Elite Family of Lasers, frequency doubled Nd:YAG |
| | Fisma Dental 200, Dental 300, Dental 400 |
| Device Description: | SmartLite D is a frequency doubled Nd:YAG laser, having a Nd:YAG<br>crystal rod as the lasing medium. It is a laser with a wavelength of<br>532 nm. |
| | Laser activation is by footswitch. Overall weight of the laser is 10 Kg<br>and the size is 27x26x36 cm (HxWxD). |
| | Electrical requirement is 230 VAC, 15A, 50-60 Hz, single phase. |
| Intended Use: | The SmartLite D is indicated for ablation, incision, excision,<br>coagulation and vaporization of tissue in ophthalmology, dermatology,<br>ENT & dentistry. |
| Comparison: | The SmartLite D laser has an equivalent indication for uses, the same<br>principle of operation, the same wavelength and essentially the same<br>pulse energy range as the predicate devices. |
| Nonclinical Performance Data: | none |
| Clinical Performance Data: | none |
| Conclusion: | The SmartLite D laser is another safe and effective device for the<br>intended uses. |
| Additional Information: | none |
{1}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized representation of a human figure embracing a globe, symbolizing the department's mission to protect and promote the health and well-being of Americans.
Public Health Service
APR 3 0 2004
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. George Cho Senior Vice President of Medical Technology Cynosure, Inc. 10 Elizabeth Drive Chelmsford, Massachusetts 01824
Re: K040289 Trade/Device Name: SmartLite D Laser Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: GEX Dated: February 5, 2004 Received: February 6, 2004
Dear Mr. Cho:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your becaon 910(x) proxice is substantially equivalent (for the indications felerenced above and nave uctorinmed the end predicate devices marketed in interstate for use stated in the enclosure) to regard cate of the Medical Device Amendments, or to commerce proof to May 20, 1976, the eccordance with the provisions of the Federal Food, DINEA devices that have tocer recuire approval of a premarket approval application (PMA). alla Cosmetic Act (Tec) that as novements with the general controls provisions of the Act. The 1 ou may, dierelore, market the act include requirements for annual registration, listing of general controls provincies practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into cither class II (Special Controls) or class III (PMA), it n your device to such additional controls. Existing major regulations affecting your device can may oe sunyeet to back as Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean r lease be as made a determination that your device complics with other requirements of the Act than + Direcal statutes and regulations administered by other Federal agencies. You must or any I with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set OFF rate 8075 https://www.bg (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
## Page 2 - Mr. George Cho
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Sincerely yours,
Mark A. Melkersen
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
04028 510(k) Number (if known):
Device Name: SmartLite D Laser
Indications For Use: The SmartLite D laser device is intended for ablation, incision, excision, vaporization and coagulation of soft tissue, in ophthalmology, ENT, and dermatology. In addition, the laser is intended for application in dentistry including:
- Frenectomy .
- Treatment of Oral Mucous Cysts .
- Treatment of Benign Vascular Lesions .
- Photocoagulation of Superficial Vessels .
- Vaporization of Superficial Blood Vessels or Lymphs Containing Vessels ●
- Treatment of Superficial Tongue Lesions .
- Tissue Management and Hemostasis for Crown and Bridge Impressions ◆
- Incision and Drainage for Abscess .
- Gingivoplasty / Gingivectomy .
- Hemostasis during Dental Procedurcs .
- Operculectomy (Operculotomy) .
- Excisional Biopsy .
- Free Gingival Graft (Adjunct) .
- Vestibuloplasty .
- Soft Gutta Percha .
- Treatment of Canker sores, Herpetic Lesions, and Aphthous Ulcers �
- Laser-assisted Bleaching / Whitening of the Teeth .
Prescription Use (Part 21 CFR 801 Subpart D) OR
Over-The-Counter Use (Part 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE
Mark A. Millman
(Division S Division of General. Restorative, and Neurological Levices
**510(k) Number** K040284
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.