CALIBRATOR FOR AUTOMATED SYSTEMS PROTEINS IN URINE/CSF

K040264 · Roche Diagnostics Corp. · JIX · Apr 2, 2004 · Clinical Chemistry

Device Facts

Record IDK040264
Device NameCALIBRATOR FOR AUTOMATED SYSTEMS PROTEINS IN URINE/CSF
ApplicantRoche Diagnostics Corp.
Product CodeJIX · Clinical Chemistry
Decision DateApr 2, 2004
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1150
Device ClassClass 2

Indications for Use

C.f.a.s. (Calibrator for automated systems) PUC (Proteins in Urine/CSF) is for use in the calibration of quantitative Roche methods on Roche clinical chemistry analyzers as specified in the enclosed value sheet.

Device Story

C.f.a.s. PUC is a liquid, single-level calibrator; buffered aqueous matrix containing human and ovine serum components. Used in clinical chemistry laboratories to calibrate Roche analyzers for albumin, total protein, and U/CSF protein assays. Provides known target values for instrument calibration; ensures accuracy of quantitative patient sample measurements. Operates via immunoturbidimetric or turbidimetric detection methods on compatible Roche/Hitachi or Cobas Integra systems.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Liquid, single-level calibrator; buffered aqueous matrix with human and ovine serum components. Traceable to CRM 470 (albumin) and NIST bovine serum albumin (total/U/CSF protein). Designed for use on Roche/Hitachi and Cobas Integra clinical chemistry analyzers.

Indications for Use

Indicated for use as a calibrator for quantitative Roche clinical chemistry assays measuring albumin, total protein, and U/CSF protein on Roche/Hitachi and Cobas Integra systems.

Regulatory Classification

Identification

A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)

Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Predicate Devices

Submission Summary (Full Text)

{0} 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY DEVICE ONLY TEMPLATE A. 510(k) Number: K040264 B. Analyte: Albumin, Total Protein, Urine/CSF (U/CSF) Protein C. Type of Test: Calibrator, Multi-Analyte Mixture D. Applicant: Roche Diagnostics E. Proprietary and Established Names: Calibrator for Automated Systems Proteins in Urine/CSF (C.f.a.s. PUC) F. Regulatory Information: 1. Regulation section: 21 CFR §862.1150 2. Classification: Class II 3. Product Code: JIX 4. Panel: Clinical Chemistry (75) G. Intended Use: 1. Indication(s) for use: "C.f.a.s. (Calibrator for automated systems) PUC (Proteins in Urine/CSF) is for use in the calibration of quantitative Roche methods on Roche clinical chemistry analyzers as specified in the enclosed value sheet." 2. Special condition for use statement(s): None noted. 3. Special instrument Requirements: Roche/Hitachi systems (for albumin and U/CSF protein) and Cobas Integra (for total U/CSF protein) H. Device Description: C.f.a.s. PUC is a liquid single-level calibrator consisting of a buffered aqueous matrix with added calibrator components to ensure optimal concentration. Albumin is detected immunoturbidimetrically, while total protein and U/CSF protein are detected turbidimetrically. {1} Page 2 of 3 # I. Substantial Equivalence Information: 1. Predicate device name(s): Calibrator for Automated Systems (C.f.a.s.) Proteins 2. Predicate K number(s): K011226 3. Comparison with predicate: Both devices are liquid, single-level calibrators. They have the same intended use and stability. They contain different analytes, and the predicate is based on stabilized human serum while the C.f.a.s. PUC is based on a buffered aqueous solution containing both human and ovine serum components. # J. Standard/Guidance Document Referenced (if applicable): See L.1.c below. # K. Test Principle: Not applicable. # L. Performance Characteristics (if/when applicable): 1. Analytical performance: a. Precision/Reproducibility: Not applicable. b. Linearity/assay reportable range: Not applicable. c. Traceability (controls, calibrators, or method): Traceability: The sponsor maintains a set of master calibrators with albumin levels traceable to European Communities' CRM 470, and total protein and U/CSF protein levels traceable to a NIST primary bovine serum albumin standard. Value Assignment: Seven external laboratories run five independent series that includes full calibration with a master calibrator and uses calibrators needing assignment as sample vials. Following outlier detection, the median is assigned as the target value. Stability: Both real-time stability studies (C.f.a.s.PUC stored at 2-8°C for 18 months) and an accelerated stability study (C.f.a.s. PUC stored at 35°C for 14 days) support a claim that unopened C.f.a.s. PUC is stable for 18 months. Open-vial stability studies support a claim that C.f.a.s. PUC is stable up to 4 weeks at 4°C. In all studies the acceptance criteria was a mean recovery of 90-110% of the reference (i.e. Day 0) material that was stored at -80°C. d. Detection limit: Not applicable. e. Analytical specificity: Not applicable. {2} f. Assay cut-off: Not applicable. 2. Comparison studies: a. Method comparison with predicate device: Not applicable. b. Matrix comparison: Not applicable. 3. Clinical studies: a. Clinical sensitivity: Not applicable. b. Clinical specificity: Not applicable. c. Other clinical supportive data (when $a$ and $b$ are not applicable): Not applicable. 4. Clinical cut-off: Not applicable. 5. Expected values/Reference range: Values of One Representative Lot | | Calibration Value (mg/L) | | | --- | --- | --- | | Analyte | HITACHI Systems | COBAS Integra | | Albumin | 441 | N/A | | Total Protein | N/A | 2080 | | U/CSF Protein | 2080 | N/A | # M. Conclusion: I recommend that the Calibrator for Automated Systems Proteins in Urine/CSF (C.f.a.s. PUC) be found substantially equivalent to the predicate.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...