The Masimo SET® Intelli Vue Pulse Oximeter Module is intended for the continuous noninvasive monitoring of functional oxygen saturerial hemoglobin (SpO2) and pulse rate (measured by an SpO2 sensor) for adult, pediatric, and neonatal patients in hospitals and hospital-type facilities.
Device Story
The Masimo SET IntelliVue Pulse Oximeter Module is a plug-in module for Philips IntelliVue patient monitors. It uses an external SpO2 sensor to measure red and infrared light absorption in tissue. The device employs Masimo SET (Signal Extraction Technology) algorithms to decompose pulsatile absorbance signals into arterial and noise components, enabling accurate SpO2 and pulse rate calculation during patient motion and low perfusion. The module is operated by clinicians in hospital settings. Output is displayed on the host Philips monitor, providing real-time SpO2 and pulse rate data to assist in patient assessment and clinical decision-making. The device benefits patients by providing reliable monitoring even when motion artifacts or low perfusion would typically cause errors in conventional pulse oximeters.
Clinical Evidence
Clinical studies were conducted on healthy adult volunteers under induced hypoxia, motion, and low-perfusion conditions. Accuracy was validated against CO-Oximeter (SpO2) and ECG (pulse rate) references. Results demonstrated SpO2 accuracy within ±2 digits (no motion) and ±3 digits (motion) for adults/pediatrics, and ±3 digits for neonates. Pulse rate accuracy was within ±3 digits (no motion) and ±5 digits (motion). Testing followed 21 CFR Parts 50, 56, and 812.
Technological Characteristics
Microprocessor-controlled pulse oximeter module; compatible with Philips IntelliVue FMS. Uses Masimo SET signal processing algorithms. Dimensions: 36mm x 99.6mm x 97.5mm; Weight: 209g. Powered by host monitor. Operating range: 1-100% SpO2, 25-240 bpm. Connectivity: proprietary module interface. Environmental: 0°C to 55°C operating temperature.
Indications for Use
Indicated for continuous monitoring of functional arterial hemoglobin oxygen saturation (SpO2) and pulse rate in adult, pediatric, and neonatal patients. Suitable for use during motion and no-motion conditions, and for patients with good or poor perfusion in hospital and hospital-type settings.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
Predicate Devices
Masimo SET Radical Pulse Oximeter with SatShare™ and LNOP series of Sensors and Cables (K031330)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the text 'K040259' in a handwritten style. The text is rotated diagonally, with the 'K' at the bottom left and the '9' at the top right. The numbers and letters are written in a dark ink, contrasting with the white background.
APR 2 2 2004
| Submitted by: | Masimo Corporation<br>2852 Kelvin Ave<br>Irvine, CA 92614-5826<br>(714) 250-9688<br>FAX (714) 250-9686 |
|----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Company Contact: | James J. Cronin, Vice President, Regulatory Affairs/Quality Assurance |
| Date Summary Prepared: | February 2, 2004 |
| Trade Name | Masimo SET® Intellivue Pulse Oximeter Module |
| Common Name | Pulse Oximeter |
| Classification Name | Oximeter (74DQA) (870.2700) |
| Substantially Equivalent Devices | Masimo SET Radical Pulse Oximeter with SatShare™ and LNOP series of Sensors and<br>Cables 510(k) Number - K031330<br>Philips Medizin Systeme MP40, MP50, MP60, MP70 and MP90 Intellivue Patie<br>Monitor - K032858 |
The Masimo SET® IntelliVue Pulse Oximeter Module is a continuous noninvasive, arterial oxygen saturation and pulse rate montor. The Masimo SET Intellivue Module features Masimo SET algorithms in a Philips single-width Intellivae compatible module.
The following list outlines the key features and benefits of the Masimo SET IntelliVue Module
### Features and Benefits
Clinically proven Masimo SET'1M technology performance Applicable for use on neonate, pediatric and adult patients Proven for accurate monitoring in motion and low perfusion environments SpO2, pulse rate, alarm, and signal quality displays BP27 police with Philips IntelliVue models MP 40/50, MP 60/70 and MP90 must be equipped with Philips Intellivue Flexible Module Server (FMS) with at least one available module slot. Compatible with all Masimo LNOP SpO2 sensors (except LNOP DCSC)
### Intended use
The Masimo SET® Intelli Vue Pulse Oximeter Module is intended for the continuous noninvasive monitoring of functional oxygen saturerial hemoglobin (SpO2) and pulse rate (measured by an SpO2 sensor) for adult, pediatric, and neonatal patients in hospitals and hospital-type facilities.
{1}------------------------------------------------
### Indications for use
The IntelliVue Pulse Oximeter Module is indicated for the continuous nonitoring of functional organization of financed for I ne ment vice culse oximicle moduced by an Spo-sensor). The Intellivie Pulse Oximeter Module is indicated for are with adult, pediatric and neonatal patients during both motion conditions, and for patients who are well or poorly perfused in hospitals and hospital-type facilities.
### Principles of Operation
The principles of operation of the Masimo SET® Intellivue pulse oximeter Module are that oxyhemoglobin and deoxyhemoglobin I he principes of operation of the Masmo BDF - intentrade ers), the volume of arterial blood in tissue (and hence, light ative in then absorphon of real and minuted in (specifysmocraphy), and that arterio-venous shunting is highly variable and that also putin by that 0100) changed anning in paise during the pulse. Because oxyhemoglobin and deoxyhemoglobin differ in light absorption, the amount of red and infrared light absorbed by the blood is related to hemoglobin oxygen saturation. The Masimo SET® Intellivue pulse oximeter module decomposes the red and infraced pulsatile absorbance signal into an arterial signal plus a noise component and calculates the ratio of the arterial signals without noise. The ratio of the two arterial pulse-added absorbance signals and is used to find the SpO2 saturation in an empirically derived equation into the Masimo SET® Intellivue software. The values in the look-up table are based upon human blood studies against a mu the Masmo 021 - miedir as betwarers in induced hypoxia states during motion and non-motion conditions.
### Method of Operation
The Masimo SET® Intellivue pulse oximeter Module is installed in an empty slot in a compatible Philips Intelli Vue Patient I homoring System. An oximetery sensor is attached to a patient's finger and one end of a patient cable is connected to the sensor and the other end connected to the Masimo SET® Intellivue pulse oximeter Module.
The monitor will begin continuously displaying the patient's pulse rate, and SpO2 value. The practitioner can then use the I no motion wat is continuously displayed on the monitor to help assess the condition of the patient and as an aide in determining if any intervention is required by the practitioner.
Once the practitioner determines the patient no longer requires monitoring, the cable is disconnected from the oximetry sensor is removed (and disposed of if it is a single use device), and the power to the monitor is turned off.
### Power Source
The Masimo SET® Intellivue pulse oximeter module is powered by the Philips IntelliVue Patient Monitoring System.
### Specifications and Operating Ranges
| Range | | |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------|-----------------------|
| | Saturation (% SpO2) | 1% - 100% |
| | Pulse Rate (bpm) | 25 - 240 |
| | Perfusion | 0.02% - 20% |
| Accuracy | | |
| | Saturation (% SpO2) - During No Motion Conditions1 | |
| | Adults, Pediatrics | 70% - 100% ± 2 digits |
| | | 0% - 69% unspecified |
| | Neonates | 70% - 100% ± 3 digits |
| | | 0% - 69% unspecified |
| | Saturation (% SpO2) - During Motion Conditions2,3 | |
| | Adults, Pediatrics2 | 70% - 100% ± 3 digits |
| | | 0% - 69% unspecified |
| Neonates3 | | 70% - 100% ± 3 digits |
| Pulse Rate (bpm) - During No Motion Conditions1<br>Adults, Pediatric, Neonates | 25 to 240 ± 3 digits | |
| Pulse Rate (bpm) - During Motion Conditions2,3<br>Adults, Pediatric, Neonates | 25 to 240 ± 5 digits | |
| Resolution | | |
| Saturation (% SpO2)<br>Pulse Rate (bpm) | 1%<br>1 | |
| Low Perfusion Performance4 | | |
| > 0.02% Pulse Amplitude<br>and % Transmission > 5% | Saturation (% SpO2) ± 2 digits<br>Pulse Rate ± 3 digits | |
| Interfering Substances | | |
| Carboxyhemoglobin may erroneously increase readings. The level of increase is approximately equal to the<br>amount of carboxyhemoglobin present. Dyes, or any substance containing dyes, that change usual arterial<br>pigmentation may cause erroneous readings. | | |
| Isolation | | |
| Patient Leakage current<br>Dielectric Withstand (mains to patient) | Less than 100 μAmp<br>>4000 VAC | |
| Environmental | | |
| Operating Temperature<br>Storage Temperature | 32°F to + 131°F (0°C to +55°C)<br>-40°F to + 158°F (-40°C to +70°C) | |
| Relative Humidity | 5% to 95% noncondensing | |
| Operating Altitude | up to 15,000 ft | |
| Circuitry | | |
| Microprocessor controlled<br>Automatic self-test of oximeter when powered on<br>Automatic setting of parameters<br>Automatic alarm messages | | |
| Audio indicators | | |
| Controlled by Philips IntelliVue Patient Monitoring System | | |
| Physical characteristics | | |
| Dimensions:<br>Weight: | 1.4" x 3.9" x 3.8" (36 mm x 99.6 mm x 97.5 mm)<br>7.4oz. (209 g) | |
| Modes | | |
| Averaging mode:<br>Sensitivity | 5, 10, 20 seconds<br>Normal | |
{2}------------------------------------------------
### 0% - 69% unspecified
- The Masimo SET Technology with LNOP Adt sensors has been validated for no motion accuracy in human blood 1 studies on healthy adult volunteers in induced hypoxia studies in the range of 70-100% SpO2 against a laboratory cooximeter and ECG monitor. This variation equals plus or minus one standard deviation which encompasses 68% of the population.
{3}------------------------------------------------
- 2 The Masimo SET Technology with LNOP Adt sensors has been validated for motion accuracy in human blood The Husening OD - Feet.net By - I counters while performing rubbing and tapping motions, at 2 to 4 Hz at an amplitude of 1 to 2 cm and a non-repetitive motion between 1 to 5 Hz at an amplitude of 2 to 3 cm in induced hypoxia studies in the range of 70-100% SpO2 against a laboratory co-oximeter and ECG monitor. This variation equals plus or minus one standard deviation which encompasses 68% of the population.
- The specified saturation accuracy from 70% 100% for neonates is based on the results from clinical studies on adult I its specifics saturation account for the properties of fetal hemoglobin to be within ± 3 digits during both motion and no motion conditions when compared to the CO-Oximeter, and the pulse rate accuracy values for neonates to be within ± 3 digits during no motion and ± 5 digits during motion conditions when compared to the ECG.. This variation equals plus or minus one standard deviation which encompasses 68% of the population
- The Masimo SET Technology has been validated for low perfusion accuracy in bench top testing against a Biotek র্ণ Index 2 simulator and Masimo's simulator with signal strengths of greater than 0.02% and a % transmission of greater than 5% for saturations ranging from 70 to 100%. This variation equals plus or minus one standard deviation which encompasses 68% of the population.
### Environmental Testing
Applicable environmetal testing per the Reviewers Guidance for Premarket Submissions - November 1993, i.e. electrical, mechanical and environmental were performed and all tests passed.
### Nonclinical tests performed that support a determination of substantial equivalence.
The Masimo SET® Intellivue Pulse Oximeters Module was subjected to bench testing using a simulator that determined the performance accuracy of the instruments against the range of saturation and pulse rates that both devices specify.
The results of the bench testing showed that the Masimo SET® Intellivue Pulse Oximeter Module returned the same saturation accuracy values within ± 2 digits and pulse rate values within ± 3 digits when compared to the simulators used.
### Clinical tests performed that support a determination of substantial equivalence.
Clinical studies were performed using the Masimo SET® technology on healthy adult volunteer subjects during no motion and motion conditions who were subjected to a progressive induced hypoxia and measuring the arterial hemoglobin saturation value with the instruments against the arterial hemoglobin oxygen determined from arterial blood samples with a CO-Oximeter.
Clinical studies were performed using the Masimo SET® technology on healthy adult volunteer subjected to low perfusion conditions and to a progressive induced hypoxia and measuring the arterial hemoglobin saturation value with the instruments against the arterial hemoglobin oxygen determined from arterial blood samples with a CO-Oximeter.
Clinical studies were conducted following regulations under Title 21 of the Code of Federal Regulations (21CFR), Part 812 -Investigational Device Exemptions, Part 50 - Protection of Human Subjects and Part 56 - Institutional Review Boards.
The results from the clinical studies show that the Masimo SET® technology saturation accuracy values and pediatrics within ± 2 digits during no motion conditions and ± 3 digits during motion conditions when compared to the CO-Oximeter and the pulse rate accuracy values within ± 3 digits during no motion conditions and ± 5 digits during motion conditions when compared to the ECG.
The specified saturation accuracy from 70% - 100% for neonates is based on the results from clinical studies on adult volunteers and 1% was added to account for the properties of fetal hemoglobin to be within ± 3 digits during both motion and no motion conditions when compared to the CO-Oximeter, and the pulse rate accuracy values for neonates to be within ± 3 digits during no motion and ± 5 digits during motion conditions when compared to the ECG.
{4}------------------------------------------------
### Conclusions
The results of the environmetal testing demonstrated that the Masimo SET® Intellivue Pulse Oximeter Module met the requirements of Reviewers Guidance for Premarket Submissions - November 1993.
The results of the bench testing demonstrates that the Masimo SET® Intellivue Pulse Oximeters Module met its performance requirements.
The results of the clinical testing demonstrates that the Masimo SET® Intellivue Pulse Oximeters its performance requirements during no motion and motion conditions and low perfusion conditions.
The non-clinical and clinical testing performed demonstrates that the Masimo SET® Intellivue Pulse Oximeters Module is safe, effective.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle with its wings spread.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
APR 2 2 2004
Mr. James Cronin Vice President, Regulatory Affairs/Quality Assurance Masimo Corporation 2852 Kelvin Ave. Irvine, CA 92614-5826
Re: K040259
Trade/Device Name: Masimo SET Intellivue Pulsc Oximeter Module Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: DQA, DPZ Dated: February 2, 2004 Reccived: February 4, 2004
Dear Mr. Cronin:
We have reviewed your Section 510(k) premarket notification of intent to market the device wf nave roviewed your becamined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use autod in the encreations of the enactment date of the Medical Device Amendments, or to conniored prior to rial 2011a) 2011 devices that have been roomstined in assess approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The rod may, mercrove, include requirements for annual registration, listing of general controls provisions of ractice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your de rice is exassino (sional controls. Existing major regulations affecting your device can may be subject to back adden addess, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean I hat FDA has made a determination that your device complies with other requirements of the Act that I Dri has Intaliations and regulations administered by other Federal agencies. You must or any I edolia surated and sequirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
{6}------------------------------------------------
Page 2 - Mr. James Cronin
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Actily of in veur Section S product radiation control provisions (occiolis 501 Clescribed in your Section 510(k)
This letter will allow you to begin marketing your device of your device to legal This letter will anow you to begin manxoning your antial equivalence of your device to a legally premarket notification. The PDA micing of basification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific advice for your ac rise of the more the regulation entitled, Contiact the Office of Compullios at (001) of Station" (21CFR Part 807.97). You may obtain Misbrailding by terefected to premarket nother the Act from the Division of Small other geleral information on your responsible at its toll-free number (800) 638-2041 or
Manufacturers, International and Consumer Assistance at the local development treal Manufacturers, International and Consembers: http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Chris Lin, Ph.D.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
### Section 3 - Indications for Use
### 510(k) Number (if known):
Masimo SET Intellivue Pulse Oximeter Module Device Name:
#### Indications For Use:
The Masino SET® Intellivue Pulse Oximeter Module is indicated for the continuous nonitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (measured by an SpO2 sensor). The Masimo SET® Intellivue Pulse Oximeter is indicated for use with adult, pediatric, and neonatal patients during both no motion and motion conditions, and for patients who are well or poorly perfused in hospitals and hospital-type facilities.
Prescription Use _ X (Per 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use J (Per 21 CFR 807 Subpart C)
### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NE
Concurrence of CDRH, Office of Device Evaluation (ODE)
seth
(Owision Sign-Off)
Division of Anesthesiology, General Hospital,
Infection Control, Dental Devices
Int., Dental Devices
510(k) Number: K040756
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.