SKINTACT ECG ELECTRODE

K040249 · Leonhard Lang GmbH · DRX · Feb 24, 2004 · Cardiovascular

Device Facts

Record IDK040249
Device NameSKINTACT ECG ELECTRODE
ApplicantLeonhard Lang GmbH
Product CodeDRX · Cardiovascular
Decision DateFeb 24, 2004
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.2360
Device ClassClass 2

Intended Use

Skintact® radiolucent and MRI-compatible ECG electrodes are designed for use in general electrocardiographic procedures where ECG monitoring is deemed necessary and is ordered by a physician. Such procedures include in particular patient ECG surveillance and ECG diagnosis recording. Skintact radiolucent and MRI-compatible ECG electrodes are single-use, non-sterile, disposable and are to be used on intact (uninjured) skin.

Device Story

Skintact® ECG electrodes are disposable, self-adhesive snap electrodes used for patient ECG surveillance and diagnostic recording. This version features radiolucent and MRI-compatible properties, achieved by replacing the standard snap with a carbon snap. The device consists of a carbon ABS sensor element coated with silver and a silver chloride layer, applied to the skin to capture electrical signals for transmission to an ECG monitor. Used in clinical settings by healthcare professionals; the output is a visual ECG trace on a monitor, assisting clinicians in cardiac rhythm assessment and diagnosis. The device benefits patients by enabling continuous monitoring or diagnostic testing while remaining compatible with imaging environments (X-ray/MRI).

Clinical Evidence

Bench testing only. Electrical performance verified against ANSI/AAMI EC12:2000 (AC impedance, DC offset, defibrillation recovery, noise, bias current). Biocompatibility testing (cytotoxicity, irritation, sensitization) confirmed safety. Radiolucency tested per ASTM F640-79. MRI compatibility tested per ASTM F2052-00, F2182-02a, and F2119-01. Comparison of ECG traces between carbon electrodes and predicate silver-snap electrodes demonstrated equivalent signal performance.

Technological Characteristics

Disposable, self-adhesive snap electrode. Sensor: Carbon ABS element coated with silver/silver chloride. Radiolucent and MRI-compatible design. Electrical performance meets ANSI/AAMI EC12:2000. Packaging: Water-vapour-proof, heat-sealed, aluminized pouches. No software or active electronic components.

Indications for Use

Indicated for patients requiring ECG monitoring or diagnostic recording during general electrocardiographic procedures. For use on intact skin only.

Regulatory Classification

Identification

An electrocardiograph electrode is the electrical conductor which is applied to the surface of the body to transmit the electrical signal at the body surface to a processor that produces an electrocardiogram or vectorcardiogram.

Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9. The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Electrocardiograph Electrodes.” See § 870.1(e) for availability information of guidance documents.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ FEB 2 4 2004 # 510(k) Summary # SUBMITTED ON BEHALF OF: Company Name: Leonhard Lang GmbH Archenweg 56 Address: 6020 Innsbruck Austria | Telephone: | ++ 43 / 512 / 33 4 25 7 | |------------|-------------------------| | Fax: | ++ 43 / 512 / 39 22 10 | by: Elaine Duncan, MS.M.E., RAC President, Paladin Medical, Inc. PO Box 560 Stillwater, MN 55082 715-549-6035 715-549-5380 Fax: Telephone: Elaine Duncan DATE PREPARED: CONTACT PERSON: January 30, 2004 | Trade Name: | Skintact® ECG Electrode | |----------------------|------------------------------------| | Common Name: | Disposable ECG Electrodes | | Classification Name: | Electrocardiograph (ECG) electrode | SUBSTANTIALLY EQUIVALENT TO: Skintact® radiolucent and MRI-compatible ECG Electrodes Skintact radiolucent and MRI-compatible ECG Electrodes are substantially equivalent to the Skintact® ECG Electrodes (the manufacturer's predicate devices) and equivalent to predicate ECG electrodes cleared under K971444 and K991105. DESCRIPTION of the DEVICE: Skintact® ECG Electrodes (and as also to be offered for sale under various private label tradenames) will now also be offered radiolucent and MRI-compatible. Just like the predicate electrodes, radiolucent and MRI-compatible ECG electrodes are self-adhesive, nonsterile, single-use disposable snap electrodes. The radiolucent and MRI-compatible ECG electrodes are identical in size, shape and configuration to the Skintact ECG electrodes currently marketed by Leonhard Lang, GmbH. The Skintact radiolucent and MRI-compatible ECG Electrodes are composed of the same materials as the predicate devices except the snap, which is made of carbon snap guarantees an unimpaired performance during the shelf-life of the product. All electrodes include an carbon ABS sensor element coated with silver. The silver layer is either completely or partially (in the areas in contact with the conductive gel) covered with a silver chloride layer. ## INDICATIONS FOR USE: Skintact® radiolucent and MRI-compatible ECG electrodes are designed for use in general electrocardiographic procedures where ECG monitoring is deemed necessary and is ordered by a physician. Such procedures include in particular patient ECG surveillance and ECG diagnosis recording. Skintact radiolucent and MRI-compatible ECG electrodes are single-use, non-sterile, disposable and are to be used on intact (uninjured) skin. (NO CHANGE to ORIGINAL INDICATION for USE.) {1}------------------------------------------------ #### SUMMARY of TESTING: All materials used in this new device have been cleared in the following 510(k)s: K 982521 Skintact® ECG Electrodes, S&W ECG Electrodes K 023503 Leonhard Lang Skintact® ECG Electrodes with solid adhesive gel Leonhard Lang Skintact® ECG Electrodes with KS 01 solid wet gel K 024264 Leonhard Lang Skintact® ECG Electrodes with KL 02 liquid gel K 024247 Biocompatibility testing of all materials cleared by 510(k)s confirms the materials are biocompatible and does not introduce any risks. The following testing of the different materials showed no adverse results: Cytotoxicity; Skin Irritation; Sensitization. The ANSI/AAMI EC 12:2000 "Disposable ECG electrodes" was used to define the requirements for Skintact ECG Electrodes. All electrical tests are according to ANSI/AAMI EC 12:2000. A certification to conformance EC12:2000 with this standard has been provided. The testing conducted was: AC impedance; DC offset voltage; Defibrillation overload recovery; Combined offset instability and internal noise; Bias current tolerance. The shelf life of radiolucent and MRI-compatible electrodes - was tested in real-time aging. All Skintact® ECG Electrodes (and as also to be offered for sale under various private label tradenames) are packaged in water-vapour-proofed, heat-sealed, non-transparent, aluminized pouches. Leonhard Lang has about 20 years of experience with this packaging and has met requirements for the 24 months shelf-life. No differences were required for packaging the Skintact radiolucent and MRI-compatible ECG electrodes compared to the predicate devices. The results of these tests confirm that the shelf-life of Leonhard Lang Skintact radiolucent and MRIcompatible ECG Electrodes is well inside the limits defined in ANSI/AAMI EC12-2000. Thus the conclusion that the electrical performance of the electrodes within the limits during their shelf-life of 24 months. The comparison with the predicate devices and the data from radiolucent and MRIcompatible electrodes shows similar results. The difference is negligible in the limits defined in ANSI/AAMI FC12-2000. Therefore electrical performance of the predicate devices and Skintacte radiolucent and MRI-compatible ECG electrodes is equivalent. Clinical data: Repeating EC12:2000 testing confirms equivalent data and the change in snap can not affect adhesive performance. So material change -- in this case the change in snap - does not affect wear tests and adhesive performance. Therefore these tests were not repeated. The potential effect of material change to the conducting signal was evaluated by repeating the trace testing on the carbon electrodes with different approved gels and determined that the carbon electrodes perform the same. Comparing the ECG traces between carbon electrodes with different approved gels and predicate device ECG electrodes demonstrate that carbon electrodes are equivalent to the predicate device ECG electrodes. #### Radiolucency: Radiolucency was tested according standard ASTM F640-79 "Standard Test Methods for Radiopacity of Plastics for Medical Use". Reference: "Radiopacity of ECG-Electrodes by Leonhard Lang GmbH - A study by Dr.Recheis Wolfgang, Dr.Verius Michael, Dr.Huttary Ralf, Mag.Torbica Pavle''. #### MRI-Compatibility: MRI companbility was tested according ASTM F 2052-00 "Standard Test Method for Magnetically Induced Displacement Force on Passive Implants in the Magnetic Resonance Environment", ASTM F 2182-02a "Standard Test Method for Measurement of Radio Frequency Induced Heating Near Passive Implants During Magnetic Resonance Imaging", ASTM F 2119-01 "Standard Test Method for Evaluation of MR Image Artifacts from Passive Implants". Reference: "MR-compatibility of ECG-Electrodes by Leonhard Lang GmbIl - A study by Dr.Recheis Wolfgang, Dr.Verius Michael, Dr.Huttary Ralf, Mag.Torbica Pavle". {2}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, composed of three curved lines. FEB 2 4 2004 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Leonard Lang GmbH c/o Ms. Elaine Duncan, M.S.M.E., RAC President Paladin Medical, Inc. P.O. Box 560 Stillwater, MN 55082-0560 Re: K040249 Trade Name: Skintack ECG Electrode Regulation Number: 21 CFR 870.2360 Regulation Name: Electrograph Electrode Regulatory Class: II (two) Product Code: DRX Dated: January 30, 2004 Received: February 03, 2004 ## Dear Ms. Duncan: We have reviewed your Scction 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The r ou may) dierer orovisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Page 2 - Ms. Elaine Duncan, M.S.M.E., RAC Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please n Joan the Office of Compliance at (301) 594-4648. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Dinna R. Lochner MBram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Indications for Use 510(k) Number (if known): K040249 Device Name: Skintact® radiolucent and MRI-Compatible ECG Electrodes Indications For Use: Skintact radiolucent and MRI-compatible ECG electrodes are designed for use in general electrocardiographic procedures where ECG monitoring is deemed necessary and is ordered by a physician. Such procedures include in particular patient ECG surveillance and ECG diagnosis recording. Skintact radiolucent and MRI-compatible ECG electrodes are single use, non-sterile and disposable and are to be used on intact (uninjured) skin. Prescription Use J (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) R. Sochner Jivision Sign-Off) ാത്രീംഗ് Olgh Silverscular Devices 10(k) Number K040249 Page 1 of /
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...