The Medi-Stim VeriQ System is an intraoperative diagnostic system that utilizes ultrasonography to guide surgeons to successfully plan and accomplish surgical interventions. The clinical indications for the device are: - Accurate transit time blood volume and Doppler velocity flow measurements 1. during cardiovascular-, vascular-, transplantation- and neuro-surgery. - Simultaneous measurements of blood pressure, vascular resistance, interfaced 2. physiological signals and other derived parameters during these procedures. - Detection of normal and abnormal blood volume and Doppler velocity flow 3. patterns during these procedures. - Provides guidance to prepare surgical plans at the initiation of surgery and to 4. support the successful accomplishment of surgery including detection and location of vessels during surgical procedures. - Detection and quantification of the degree of stenosis in arteries by using the 5. Doppler velocity profile.
Device Story
Medi-Stim VeriQ System is an intraoperative diagnostic tool for cardiovascular, vascular, transplantation, and neuro-surgery. Device utilizes transit time and Doppler ultrasonography to measure blood volume and velocity waveforms. System inputs include ultrasonic signals from transducers and auxiliary physiological signals (e.g., blood pressure). Device processes these inputs to calculate blood volume, velocity, vascular resistance, and derived parameters; detects/quantifies arterial stenosis; and assists in vessel location. Operated by surgeons in the OR to guide surgical planning and verify intervention success. Output is displayed for real-time clinical decision-making, allowing surgeons to assess flow patterns and vessel patency during procedures.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Intraoperative diagnostic system using transit time and Doppler ultrasonography. Includes various specialized transducers (e.g., PQ, PA, PR, PI, PC, PE, PD series). Measures blood volume, velocity, pressure, and vascular resistance. Connectivity includes auxiliary physiological signal interfaces. Class II device.
Indications for Use
Indicated for patients undergoing cardiovascular, vascular, transplantation, or neuro-surgery requiring intraoperative blood volume, velocity, pressure, and vascular resistance monitoring.
Regulatory Classification
Identification
A cardiovascular blood flowmeter is a device that is connected to a flow transducer that energizes the transducer and processes and displays the blood flow signal.
{1}------------------------------------------------
K040228
APR 2 9 2004
510(k) – Medi-Stim VeriQ System 510(k) Summary Page 95
## 510(k) Summary
## 510(k) Summary
## as required by 807.92 (c)
# for Medi-Stim VeriQ System, VQ1001 -- VQ4122
## Prepared January 28th, 2004
| Submitted by: | Medi-Stim AS<br>Sandakerveien 110<br>P.O. Box 35 Grefsen<br>N-0484 Oslo<br>Norway<br>Tel. 011 47 23 05 96 60 (switch board)<br>Fax 011 47 23 05 96 61 |
|--------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Jon H. Hoem, VP Sales & Marketing, Medi-Stim USA, Inc.<br>7601 Northland Drive, Mailstop A150, Brooklyn Park,<br>MN55428<br>Tel. (763) 493 6664 direct / (612) 730 7310 cell<br>Fax (763) 391 9922 |
| Device Trade Name: | Medi-Stim VeriQ System |
| Common Name: | Medical blood volume and velocity system with blood pressure<br>and auxiliary data acquisition options. |
| Classification: | Class II |
{2}------------------------------------------------
### Predicate Devices:
Medi-Stim AS Medi-Stim Butterfly Flowmeter (K992305)
Koven Technology, Inc. Doppler Flowmeter DVM-4300T (K010452) Doppler Flowmeter ES-1000SPM Smartdop (K903026)
12125 Woodcrest Executive Drive, Suite 220 St. Louis, MO 6314 Tel: 1 (800) 521 8342 Fax: (314) 542 6020
#### Description of Device
The Medi-Stim VeriQ System is used intraoperatively to measure blood volume and velocity waveforms in arteries and veins. Transit time and Doppler ultrasonography is used to obtain these measurements.
#### Intended Use of Device
The Medi-Stim VeriQ System is an intraoperative diagnostic system that utilizes ultrasonography to guide surgeons to successfully plan and accomplish surgical interventions. The clinical indications for the device include, but are not restricted to, the following:
- 1. Accurate transit time blood volume and Doppler velocity flow measurements during cardiovascular-, vascular-, transplantation- and neuro-surgery.
- 2. Simultaneous measurements of blood pressure, vascular resistance, interfaced physiological signals and other derived parameters during these procedures.
- 3. Detection of normal and abnormal blood volume and Doppler velocity flow patterns during these procedures.
- 4. Provides guidance to prepare surgical plans at the initiation of surgery and to support the successful accomplishment of surgery including detection and location of vessels during surgical procedures.
- 5. Detection and quantification of the degree of stenosis in arteries by using the Doppler velocity profile.
{3}------------------------------------------------
### 4.3. Indications for Use Statement
The Medi-Stim VeriQ System is an intraoperative diagnostic system that utilizes ultrasonography to guide surgeons to successfully plan and accomplish surgical interventions. The clinical indications for the device are:
- Accurate transit time blood volume and Doppler velocity flow measurements 1. during cardiovascular-, vascular-, transplantation- and neuro-surgery.
- Simultaneous measurements of blood pressure, vascular resistance, interfaced 2. physiological signals and other derived parameters during these procedures.
- Detection of normal and abnormal blood volume and Doppler velocity flow ನ್ patterns during these procedures.
- Provides guidance to prepare surgical plans at the initiation of surgery and to 4. support the successful accomplishment of surgery including detection and location of vessels during surgical procedures.
- Detection and quantification of the degree of stenosis in arteries by using the 5. Doppler velocity profile.
### 4.3.1. 510(k) Indications for Use Forms
Please find the Indications for Use Forms following.
{4}------------------------------------------------
Image /page/3/Picture/2 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features a stylized eagle with three lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle. The seal is black and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
APR 2 9 2004
Medi-Stim AS c/o Mr. Jon H. Hoem Vice President, Sale & Marketing Medi-Stim USA, Inc. 7601 Northland Drive, Mailstop A150 Brooklyn Park, MN 55428
Rc: K040228 Medi-Stim VeriQ System Regulation Number: 870.2190 Regulation Name: Cardiovascular Blood Flowmeter Regulatory Class: Class II (two) Product Code: 74 DPW Dated: January 28, 2004 Received: February 2, 2004
### Dear Mr. Hoem:
We have reviewed your Section 510(k) premarket notification of intent to market the device we have loviewed your have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device microate commerce prob to that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to i cactal I ood, Drug, and Colline Cornett Co. . The general controls provisions of the Act include the general bonners for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use with the Medi-Stim VeriQ System, as described in your premarket notification:
PQ101011, PQ101012, PQ100021, PQ100031, PQ100041, PQ100051, PQ100022, PQ100032, PQ100042, PQ100052, PA100081, PA100101, PA100121, PA100161, PA100211, PA100271, PA100082, PA100101, PA100122, PA100161, PA100162, PR100251, PR100301, PR100351, PI100181, PI100141, PI100381, PI100121,
{5}------------------------------------------------
Page 2 - Mr. Jon H. Hoem
PC100181, PC103161, PC103161, PC103162, PC100141, PC100142, PC100382, PC100383, PC100122, PC100123, PC100124, PE100022, PE100032, PE100042, PE100052, PD110751, PD110752, PD120751, PD120752
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market. ·
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801, please contact the Office of Compliance at (301) 594-4591. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html". If you have any questions regarding the content of this letter, please contact Kachi Enyinna at (301) 443-8262.
Sincerely yours,
Donna R. Lochner
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Office of Device Evaluation Center for Devices and Radiological Health
Enclosures
{6}------------------------------------------------
# Indications for Use
510(k) Number (if known): K040228
Device Name: Medi-Stim VeriQ System
Indications For Use:
The Medi-Stim VeriQ System is an intraoperative diagnostic system that utilizes ultrasonography to guide surgeons to successfully plan and accomplish surgical interventions. The clinical indications for the device are:
- Accurate transit time blood volume and Doppler velocity flow measurements 1. during cardiovascular-, vascular-, transplantation- and neuro-surgery.
- Simultaneous measurements of blood pressure, vascular resistance, interfaced 2. physiological signals and other derived parameters during these procedures.
- Detection of normal and abnormal blood volume and Doppler velocity flow 3. patterns during these procedures.
- Provides guidance to prepare surgical plans at the initiation of surgery and to 4. support the successful accomplishment of surgery including detection and location of vessels during surgical procedures.
- Detection and quantification of the degree of stenosis in arteries by using the 5. Doppler velocity profile.
AND/OR -Prescription Use × (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Dona Piccione
Page 1 of 1
510(k) Number K040228
<sup>(</sup>Division Sign-Off) Division of Cardiovascular Devices
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.