← Product Code [MPO](/productcode/MPO) · K040044

# HEEL SNUGGLER, MODEL 99047 (K040044)

_Respironics, Inc. · MPO · Jan 27, 2004 · Physical Medicine · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K040044

## Device Facts

- **Applicant:** Respironics, Inc.
- **Product Code:** [MPO](/productcode/MPO.md)
- **Decision Date:** Jan 27, 2004
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 890.5710
- **Device Class:** Class 1
- **Review Panel:** Physical Medicine
- **Attributes:** Therapeutic, Pediatric, 3rd-Party Reviewed

## Indications for Use

The Respironics Hecl Snuggler infant heel warmer is intended for use whenever circulation needs to be stimulated in the infant heel in order for blood sampling to occur.

## Device Story

Heel Snuggler is a disposable, butterfly-shaped chemical heat pack used to stimulate circulation in an infant's heel for blood sampling. The device consists of a packet containing a non-toxic solution and a catalyst disk. Activation occurs when the user flexes the internal catalyst disk, initiating an exothermic chemical reaction that generates warmth. The device features adhesive-ended straps that secure the packet around the infant's foot in a boot-like configuration, with adhesive applied only to the packet, not the skin. Perforations allow for easy removal after use. It is intended for clinical use by healthcare providers to facilitate blood collection procedures.

## Clinical Evidence

Bench testing only. Design verification tests were performed based on risk analysis and product requirements; all tests met acceptance criteria.

## Technological Characteristics

Butterfly-shaped disposable packet containing non-toxic solution and a metal catalyst disk. Exothermic chemical reaction principle. Features adhesive-ended straps for securing to infant foot. Perforated design for removal. Non-powered.

## Regulatory Identification

A hot or cold disposable pack is a device intended for medical purposes that consists of a sealed plastic bag incorporating chemicals that, upon activation, provides hot or cold therapy for body surfaces.

## Predicate Devices

- DeRoyal Industries Infant Heel Warmer ([K954716](/device/K954716.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

JAN 2 7 2004

Image /page/0/Picture/2 description: The image shows a handwritten text that appears to be a combination of letters and numbers. The text includes 'K040044' prominently displayed at the top, with the number '1072' written below it. The handwriting is somewhat stylized, with the numbers and letters varying in size and thickness, giving the text a unique, informal appearance.

Image /page/0/Picture/3 description: The image shows a logo for Respironics. The logo consists of the word "RESPIRONICS" in a bold, sans-serif font, with the website address "www.respironics.com" underneath. Above the word "RESPIRONICS" is a triangular shape containing a stylized image that resembles a flower or a three-leaf clover. The background of the image is a grainy, textured pattern.

#### SECTION 9

#### 510(K) SUMMARY

| Date of Submission                                      | December 15, 2003                                                                                                                                               |
|---------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Official Contact / Address<br>of Manufacturing facility | Zita A. Yurko<br>Manager, Regulatory Affairs<br>Respironics, Inc.<br>1001 Murry Ridge Lane<br>Murrysville, PA 15668<br>Phone: 724-387-4120<br>Fax: 724-387-4216 |
| Proprietary Name                                        | Heel Snuggler                                                                                                                                                   |
| Common/Usual Name                                       | Infant Heel Warmer (Chemical Heat Pack)                                                                                                                         |
| Classification Reference                                | 21 CFR 890.5710                                                                                                                                                 |
| Classification                                          | Class I                                                                                                                                                         |
| Appropriate Classification Panel                        | Physical Medicine                                                                                                                                               |
| Product Code                                            | MPO                                                                                                                                                             |
| Predicate Device                                        | DeRoyal Industries Infant Heel Warmer (K954716)                                                                                                                 |
| Reason for submission                                   | Acquisition of Children's Medical Ventures resulted in 510(k) filing                                                                                            |

. .

#### Intended Use/Indications for Use

The Respironics Hecl Snuggler infant heel warmer is intended for use whenever circulation needs to be stimulated in the infant heel in order for blood sampling to occur.

.

{1}------------------------------------------------

K040044
20F2

Image /page/1/Picture/2 description: The image shows a logo with the word "RESPIRONICS" in a sans-serif font. Above the word is a triangular shape containing a stylized figure. Below the word "RESPIRONICS" is the text "www.respironics.com".

### Patient Population/Environment of Use

The Heel Snuggler is disposable and for single patient use only.

#### Substantial Equivalence

This traditional 510(k) submittal demonstrates that the Heel Snuggier is substantially equivalent to : • DeRoyal Industries, Inc. Infant Heel Warmer (K954716).

Design verification tests were performed on the Heel Snuggier as a result of the risk analysis and II :: product requirements. All tests were verified to meet the required acceptance criteria. Respironics has determined that the differences between the Heel Snuggler infant heel warmer and the predicate infant heel warmer have no impact on the safety and effectiveness of the Heel Snuggler and that al hazards were successfully mitigated.

#### Device Description

The Heel Snuggler is a butterfly shaped packet that is filled with a non-toxic solution and a catalyst disk. Two straps with adhesive ends are attached to one side of the packet for securing the heel warmer to the infant's foot. Immediately following activation of the user rests the bottom of the infant's foot on the larger portion of the packet and holds the smaller portion of the packet to the ankle. The straps with adhesive tape on the end are then fitted across the infant's foot and ankle and secured to the opposite side of the packet. The adhesive portion of the strap is affixed to the packet only, not the infant's skin. This forms a boot shape around the infant's foot. The heel warmer are perforated, which enables the user to easily tear the straps and remove the heel warmer from the infant's foot after use.

When the catalyst disk inside the packet is flexed by the catalyst reachs with the solution a ri initiates a chemical exothermic reaction, which produces warmth.

End of Section.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS) of the United States. The seal features a circular design with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is an emblem depicting a stylized eagle with three wings, symbolizing health, services, and humanity.

Public Health Service

JAN 2 7 2004

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Respironics, Inc. C/o Mr. Ned Devine Entela, Inc 3033 Madison Avenue SE Grand Rapids, Michigan 49548

Re: K040044

Trade/Device Name: Heel Snuggler Infant Hecl Warmer Regulation Number: 21 CFR 890.5710 Regulation Name: Hot or cold disposable pack Regulatory Class: I Product Code: MPO Dated: January 12, 2004 Received: January 12, 2004

Dear Mr. Devine:

We have reviewed your Scetion 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisling of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.

{3}------------------------------------------------

#### Page 2 - Mr. Ned Devine

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-1308. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely vours.

sincerely yours.

Mark M. Mulhearn

Celia M. Witten, PhD, MD Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

### Indications for Use

510(k) Number (if known): _K040044

Device Name: ___Heel Snuggler Infant Hecl Warmer______________________________________________________________________________________________________________________________

Indications for Use:

## Intended Use/Indications for Use

The Respironics Heel Snuggler infant heel warmer is intended for use whenever circulation needs to be stimulated in the infant heel in order for blood sampling to occur.

# Patient Population/Environment of Use

The Heel Snuggler is disposable and for single patient use only.

Mark N. Mulkerson

Prescription Use __ X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

---

**Source:** [https://fda.innolitics.com/device/K040044](https://fda.innolitics.com/device/K040044)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
