The intended use of the CERAGEM-C Thermal Massager is to provide muscle relaxation therapy by delivering heat and soothing massage. Additionally, the Ceragem-C Thermal Massager provides radiant infrared heat for, - temporary relief of minor muscle and joint pain, and stiffness - the temporary relief of minor joint pain associated with arthritis - the temporary increase in local circulation where applied - relaxation of muscles
Device Story
CERAGEM-C Thermal Massager is an electrically powered, motorized physical therapy table. Device delivers muscle relaxation therapy via massage rollers and radiant infrared heat. Input: electrical power. Transformation: motorized carriage moves jade massage rollers under a padded cushion along the torso; internal light bulbs heat rollers to emit infrared heat (5-20 microns); additional heat provided by Epoxy Carbon Panels. Output: mechanical massage and radiant heat applied to user. Used in clinical or home settings; operated by user. Healthcare providers use output to facilitate muscle relaxation and pain relief. Benefits include temporary relief of minor muscle/joint pain, stiffness, arthritis-related pain, and increased local circulation.
Clinical Evidence
Bench testing only. No clinical data provided. Safety and effectiveness supported by comparison of functional characteristics and adherence to international medical electrical equipment safety standards (UL 60601-1, CAN/CSA C22.2 No. 601.1-M90, EN 60601-1, IEC 60601-1, IEC 60601-2-38).
Indicated for individuals seeking temporary relief of minor muscle and joint pain, stiffness, minor joint pain associated with arthritis, and for the temporary increase in local circulation and muscle relaxation.
Regulatory Classification
Identification
A multi-function physical therapy table is a device intended for medical purposes that consists of a motorized table equipped to provide patients with heat, traction, and muscle relaxation therapy.
{0}------------------------------------------------
JUL 2 8 2004
#### 510(k) SUMMARY XI.
Pursuant to 513(i)(3)(A) of the Food, Drug and Cosmetic Act, Ceragem Co., Ltd is required to submit with this Premarket Notification either an "…adequate summary of any information respecting safety and effectiveness or state that information will be made available upon request of any person." Ceragem Co., Ltd. chooses to submit a summary of the safety and effectiveness information. The summary is as follows:
| Submitter's Name | CERAGEM Co., Ltd.<br>177-14, Osaejdabg-ri, Seongger-eup, Cheonan-si,<br>Chungcheongnam-do, Korea<br>Contact Person: Kevin Park<br>Tel: 213.480.7851 |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|
| Date Prepared: | December 31, 2003 |
| Proprietary Name: | CERAGEM-C Thermal Massager |
| Classification Name: | Multi-Function Therapy Table (Class II), 21 CFR 890.5880 |
| Predicate Device: | Migun HY5000 Thermassage Energy Product (K032449)<br>Migun Medical Instrument Co., Ltd. |
### Product Description:
The CERAGEM-C Thermal Massager is an electrically powered motorized multi-functional physical therapy table. Its intended use is to provide muscle relaxation therapy by delivering heat and soothing The massage function is delivered by massage rollers mounted on an independent carriage massage. underneath a pad on the table torso section. The heat function is delivered by two components (1) heated jade massage rollers mounted together on the moving carriage and (2) heated Epoxy Carbon Panels. In both cases, the radiant infrared heat is emitted. Together, the massage action and heated pressure points apply light pressure as well as radiant infrared heat to the user.
{1}------------------------------------------------
# Statement of Intended Use Compared to Predicate Device:
The intended use of the CERAGEM-C Thermal Massager is the same as the above-identified predicate device. It is intended to provide muscle relaxation therapy by delivering heat and soothing massage. Additionally, the CERAGEM-C Thermal Massager provides radiant infrared heat for,
- ¢ temporary relief of minor muscle and joint pain, and stiffness
- 0 the temporary relief of minor joint pain associated with arthritis
- ¢ the temporary increase in local circulation where applied
- ◊ relaxation of muscles
## Discussion of Technological Characteristics:
Both the CERAGEM-C Thermal Massager and the predicate device use massage rollers and heat to provide muscle relaxation therapy - The CERAGEM-C emits topical radiant infrared heat in the range of 5-20 microns from the jade massage rollers, heated by light bulbs located inside the rollers. The ,在第一次 predicate device also emits topical heat through light bulbs located in the jade massage rollers in the range of 5-20 microns. Both the CERAGEM-C device and the predicate provide the massage compone by attaching the jade rollers to a carriage mounted under a padded cushion that traverses the torso CERAGEM-C device also transmils heat through the three Epoxy Carbon Panels. While the predicate device does not include Epoxy Carbon Panels, the maximal heat of the CERAGEM-C remains that of the predicate. Additionally both the CERAGEM-C and the predicate device have and meet the following standards:
Underwriters Laboratories Standard UL 60601-1:1997 200 Edition, "Medical Electrical Equipment, 1. Part 1: General Requirements for Safety" including Amendments 1 and 2:
CAN/CSA C22.2 No. 601.1-M90, "Medical Equipment Part 1: General Requirements for Safety," 2. including C22.2 No. 601.1 S1-94, Supplement No. 1-94 to CAN/CSA-C22.2 No. 601.1-M90.
3. EN 60601-1 (1990), "Medical Electrical Equipment Part 1, General Requirements for Safety" including Amendments A1 and A2;
IEC 60601-1, 200 Edition (1998) "Medical Equipment- General Requirements for Safety + A1:91 4. and A2:95"; and
IEC 60601-2-38, 1st Edition (1996) "Medical Electrical Equipment - Part 2: Particular 5. Requirements for the Safety of Electrically Operated Beds."
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the word "CERAGEM" in bold, black letters. The letters are slightly slanted to the right. There is a circled R symbol to the right of the word.
### Conclusion:
Based on comparison of the CERAGEM-C Thermal Massager to the predicate device, we conclude that the CERAGEM-C Thermal Massager has the same intended use, indications for use, and intended population, and similar functional and performance characteristics. The addition of the three Epoxy Carbon Panels does not raise new safety or effectiveness issues since the temperature never exceeds the maximal temperature of the predicate device. Other visual distinctions do not impact safety or effectiveness.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is a black and white circular logo for the Department of Health & Human Services - USA. The logo features the department's emblem, which is a stylized depiction of a human figure embracing a globe. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the upper perimeter of the circle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUL 28 2004
Ceragem Company, Ltd. c/o Ms. Suzan Onel Kirkpatrick and Lockhart LLP 1800 Massachusetts Avenue, NW Suite 200 Washington, DC 20036
Re: K040031
Trade/Device Name: Ceragem-C Thermal Massager Regulation Number: 21 CFR 890.5880 Regulation Names: Multi-function Physical Therapy Table Regulatory Class: Class II Product Code: JFB Dated: April 28, 2004 Received: April 29, 2004
Dear Ms. Suzan Onel:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
{4}------------------------------------------------
#### Page 2 - Ms. Suzan Onel
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594- 1308. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Mark A. Milken
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
**CERAGEM**®
Page ol
510(k) Number (if known): CERAGEM-C Thermal Massager Device Name:
Indications For Use:
The intended use of the CERAGEM-C Thermal Massager is to provide muscle relaxation therapy by delivering heat and soothing massage. Additionally, the Ceragem-C Thermal Massager provides radiant infrared heat for,
� temporary relief of minor muscle and joint pain, and stiffness
◊ the temporary relief of minor joint pain associated with arthritis
0 the temporary increase in local circulation where applied
0 relaxation of muscles
Prescription Use AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 807 Subpart C)
(Please Do NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
*Manh M*
**(Division Sign-Off)**
Division of General, Restorative, and Neurological Devices
(Optional Format 3-10-98)
510(k) Number_K 040 o 3/
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.