K040009 · The Binding Site, Ltd. · DFH · Mar 11, 2004 · Immunology
Device Facts
Record ID
K040009
Device Name
FREELITE HUMAN KAPPA AND LAMBDA KITS
Applicant
The Binding Site, Ltd.
Product Code
DFH · Immunology
Decision Date
Mar 11, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.5550
Device Class
Class 2
Indications for Use
This kit is intended for the quantitation of kappa free light chains in serum and urine on the Dade Behring Nephelometer II (BN™II). Measurement of the various amounts of the different types of light chains aids in the diagnosis and monitoring of multiple myeloma, lymphocytic neoplasms, Waldenstrom's macroglobulinemia, amyloidosis, light chain deposition disease and connective tissue disorders such as systemic lupus erythematosus.
Device Story
The FREELITE™ Human Kappa or Lambda Free Kit is an in vitro diagnostic assay for quantifying free light chains in serum and urine. The device utilizes latex-enhanced nephelometric or turbidimetric technology; polystyrene beads are coated with monospecific antibodies. When the sample is added, antigen-antibody reactions form insoluble immune complexes, increasing light scatter. The intensity of scattered light is proportional to the antigen concentration. The assay is performed on automated clinical chemistry analyzers (Dade Behring BN II, Beckman Coulter IMMAGE, or Roche HITACHI series). Healthcare providers use the quantitative results to assist in diagnosing and monitoring plasma cell dyscrasias and related autoimmune conditions. The primary change in this submission is the transition of the latex reagent from a freeze-dried to a liquid format and the addition of an amyloidosis clinical claim.
Clinical Evidence
No clinical data provided in the document; substantial equivalence is based on technological characteristics and intended use.
Technological Characteristics
Latex-enhanced nephelometric/turbidimetric immunoassay. Reagents consist of polystyrene beads coated with monospecific antibodies. Energy source is the light beam of the clinical analyzer. Analyzers used: Dade Behring BN II, Beckman Coulter IMMAGE, Roche HITACHI 911/912/917/Modular P. Liquid reagent format. Quantitative measurement of light scatter.
Indications for Use
Indicated for the quantitation of kappa or lambda free light chains in serum and urine to aid in the diagnosis and monitoring of multiple myeloma, lymphocytic neoplasms, Waldenstrom’s macroglobulinemia, amyloidosis, light chain deposition disease, and systemic lupus erythematosus.
Regulatory Classification
Identification
An immunoglobulin (light chain specific) immunological test system is a device that consists of the reagents used to measure by immunochemical techniques both kappa and lambda types of light chain portions of immunoglobulin molecules in serum, other body fluids, and tissues. In some disease states, an excess of light chains are produced by the antibody-forming cells. These free light chains, unassociated with gamma globulin molecules, can be found in a patient's body fluids and tissues. Measurement of the various amounts of the different types of light chains aids in the diagnosis of multiple myeloma (cancer of antibody-forming cells), lymphocytic neoplasms (cancer of lymphoid tissue), Waldenstrom's macroglobulinemia (increased production of large immunoglobulins), and connective tissue diseases such as rheumatoid arthritis or systemic lupus erythematosus.
Submission Summary (Full Text)
{0}
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
A. 510(k) Number:
k040009
B. Analyte:
Kappa free light chains and lambda free light chains
C. Type of Test:
Nephelometric or turbidimetric, quantitative
D. Applicant:
The Binding Site, Ltd.
E. Proprietary and Established Names:
FREELITE™ Human Kappa Free Kit for use on the Dade Behring Nephelometer™ II
FREELITE™ Human Lambda Free Kit for use on the Dade Behring Nephelometer™ II
FREELITE™ Human Kappa Free Kit for use on the Beckman Coulter IMMAGE™
FREELITE™ Human Lambda Free Kit for use on the Beckman Coulter IMMAGE™
FREELITE™ Human Kappa Free Kit for use on the HITACHI 911/912/917/Modular P
FREELITE™ Human Lambda Free Kit for use on the HITACHI 911/912/917/Modular P
F. Regulatory Information:
1. Regulation section:
21 CFR §866.5550, Immunoglobulin (light chain specific) immunological test system
2. Classification:
Class II
3. Product Code:
DFH, Kappa, antigen, antiserum, control
DEH, Lambda, antigen, antiserum, control
4. Panel:
IM (82)
G. Intended Use:
The amyloidosis claim has been added to the intended use of the FREELITE™ Human Kappa Free kit and the FREELITE™ Human Lambda Free kit for use on the Beckman Coulter IMMAGE™ and the Hitachi 911/912/917/Modular P. No change was made to the FREELITE™ Human Kappa Free kit and the FREELITE™ Human Lambda Free kit for use on the Dade Behring Nephelometer™ II.
The intended use for the FREELITE™ Human Kappa or Lambda Free Kit is as follows:
The Binding Site FREELITE™ Human Kappa or Lambda Free Kit is intended for the quantitation of kappa free light chains or lambda free light chains in serum and urine on the Dade Behring Nephelometer II (BN™ II) or Beckman Coulter IMMAGE or Roche HITACHI 911, HITACHI 912, HITACHI 917 and Modular P. Measurement of the various amounts of the different types of light chains aids in the diagnosis and monitoring of multiple myeloma, lymphocytic neoplasms, Waldenstrom’s macroglobulinemia, amyloidosis, light chain deposition disease and connective tissue diseases such as systemic lupus erythematosus.
1. Indication(s) for use:
{1}
Page 2 of 8
Same as intended use.
2. Special condition for use statement(s):
For prescription use only.
3. Special instrument Requirements:
Dade Behring Nephelometer II (BNII), Beckman Coulter IMMAGE, Roche HITACHI 911, 912, 917 and Modular P analyzers
## H. Device Description:
The FREELITE™ Human Kappa or Lambda Free Kit consists of the following: latex reagent (polystyrene beds coated with monospecific antibody), a standard, a high control, a low control and supplementary reagent.
## I. Substantial Equivalence Information:
1. Predicate device name(s):
FREELITE Human Kappa Free Kit for use on the Dade Behring Nephelometer™ II
FREELITE Human Lambda Free Kit for use on the Dade Behring Nephelometer™ II
FREELITE Human Kappa Free Kit for use on the Beckman Coulter IMMAGE™
FREELITE Human Lambda Free Kit for use on the Beckman Coulter IMMAGE™
FREELITE Human Kappa Free Kit for use on the HITACHI 911/912/917/Modular P
2. Predicate K number(s):
k003669, k003671, k023009, k023131, k031016
3. Comparison with predicate:
The purpose for this submission was to provide data to support changing the format of the latex reagent from freeze dried to liquid for the Dade Behring BNII and Beckman IMMAGE FREELITE™ Human Kappa Free kits and FREELITE™ Human Lambda Free Kits and the addition of the amyloidosis claims to the Beckman IMMAGE and Roche HITACHI FREELITE™ kits.
## J. Standard/Guidance Document Referenced (if applicable):
None referenced.
## K. Test Principle:
The FREELITE™ Human Kappa Free kits and FREELITE™ Human Lambda Free kits are nephelometric or turbidimetric assays. The test sample is added to a solution containing the appropriate antibody in a reaction vessel. A beam of light is passed through the vessel and as the antigen-antibody reaction proceeds, the light passing through is increasingly scattered as insoluble immune complexes are formed. The amount of immune complex formed is proportional to the antigen concentration in the test sample. In nephelometry, the light scatter is monitored by measuring the light intensity at an angle away from incident light whereas in turbidimetry, light scatter is by measuring the increase in intensity of the incident light beam. A series of calibrators of known antigen concentration are assayed to construct a calibration curve which will be used for determining the antigen concentration of test samples.
## L. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
The precision studies were performed at the Binding Site with the liquid latex reagents. Studies for each analyzer were performed at different times using different reagent lots. The kappa light chain assays used three concentrations of Level 1 whereas the lambda light
{2}
Page 3 of 8
chain assays used Level 1 and two other positive samples. The samples were assayed in replicates of 10 for the within-run precision and in singlicate on 10 separate runs for the between-run precision. Results were summarized below:
BN II
| Within-run precision - Kappa | | | | Within-run precision - Lambda | | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| Sample | Level 1 | Level 2 | Level 3 | Sample | Level 1 | Level 2 | Level 3 |
| Mean (mg/L) | 51.06 | 32.76 | 13.56 | Mean (mg/L) | 71.84 | 21.89 | 15.12 |
| SD (mg/L) | 2.16 | 1.58 | 0.42 | SD (mg/L) | 3.28 | 1.08 | 1.21 |
| %CV | 4.23 | 4.84 | 3.10 | %CV | 4.8 | 5.2 | 8.4 |
| Between-run precision - Kappa | | | | Between-run precision - Lambda | | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| Sample | Level 1 | Level 2 | Level 3 | Sample | Level 1 | Level 2 | Level 3 |
| Mean (mg/L) | 54.66 | 32.17 | 12.02 | Mean (mg/L) | 71.71 | 24.20 | 18.43* |
| SD (mg/L) | 3.84 | 2.56 | 0.71 | SD | 5.06 | 1.08 | 1.40 |
| %CV | 7.4 | 8.38 | 6.27 | %CV | 7.45 | 4.72 | 8.08 |
*One outlier sample (13.1 mg/L) was excluded from the calculation. If included, the mean±SD and % CD are 17.9±2.08 mg/L and 11.6% respectively.
IMMAGE
| Within-run precision - Kappa | | | | Within-run precision - Lambda | | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| Sample | Level 1 | Level 2 | Level 3 | Sample | Level 1 | Level 2 | Level 3 |
| Mean (mg/L) | 57.1 | 27.11 | 18.3 | Mean (mg/L) | 98.78 | 46.79 | 31.83 |
| SD (mg/L) | 1.09 | 0.88 | 0.43 | SD (mg/L) | 2.25 | 1.81 | 1.93 |
| %CV | 2.00 | 3.41 | 2.48 | %CV | 2.4 | 4.08 | 6.4 |
| Between-run precision - Kappa | | | | Between-run precision - Lambda | | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| Sample | Level 1 | Level 2 | Level 3 | Sample | Level 1 | Level 2 | Level 3 |
| Mean (mg/L) | 57.59 | 27.5 | 20.09 | Mean (mg/L) | 101.62 | 64.52 | 33.85 |
| SD (mg/L) | 3.51 | 1.89 | 2.38 | SD (mg/L) | 2.76 | 5.75 | 2.16 |
| %CV | 6.42 | 7.23 | 12.47 | %CV | 2.86 | 9.39 | 6.74 |
b. Linearity/assay reportable range:
A serum sample containing high concentrations of free kappa and lambda was serially diluted and each dilution was assayed at the normal sample dilution of 1:100 for BNII and 1:10 for IMMAGE. The samples were also run at 1:20 for BNII, 1:5 for IMMAGE Lambda and 1:1 for both IMMAGE Kappa and Lambda to confirm linearity at the low and high regions of the measuring range. All studies were performed at the Binding Site. No 95% CI for the slope or intercept was provided but was calculated by the reviewer.
BNII
Kappa - Linearity at 1:100 gave y = 1.0052x - 0.1154 (r² = 0.9955). The 95% CI for slope and the intercept were (0.9532 to 1.0542) and (-1.3597 to 1.1767) respectively (9 samples)
{3}
Page 4 of 8
Linearity at 1:20 gave y = 1.0423x + 3.6932 (r² = 0.9834).
The 95% CI for the slope and the intercept were (0.9215 to 1.1631) and (-6.0518 to 13.4383) respectively (9 samples)
The combined data gave y = 1.0774x - 0.2537 (r² = 0.9882) with 95% CI of slope (1.0151 to 1.1397) and the intercept (-3.9761 to 3.4687).
The assay measuring range is 6-190 mg/L
Lambda - Linearity at 1:100 gave y = 0.9839x + 0.0893 (r² = 0.9769). The 95% CI for the slope and the intercept were (0.8487 to 1.1192) and (-6.8080 to 6.9866) respectively (9 samples)
Linearity at 1:20 gave y = 1.0353x -7.3410 (r² = 0.9985). The 95% CI for the slope and the intercept were (0.9959 to 1.0747) and (-14.3463 to -0.3357) respectively (8 samples)
The combined data gave y = 1.0151x - 2.7463 (r² = 0.9979) with 95% CI of the slope and the intercept (0.9894 to 1.0408) and (-6.0199 to 0.5726) respectively.
The assay measuring range is 8-260 mg/L
## IMMAGE
Kappa - Linearity at 1:10 gave y = 0.9826x - 1.0617 (r² = 0.9938). The 95% CI for slope and the intercept were (0.9134 to 1.0518) and (-3.4615 to 1.3381) respectively (9 samples)
Linearity at 1:1 gave y = 1.0402x - 9.1018 (r² = 0.9977). The 95% CI for the slope and the intercept were (0.9491 to 1.1314) and (-19.9493 to 1.7456) respectively. (5 samples)
The combined data gave y = 1.0065x - 3.2801 (R² = 0.9962) with 95% CI of slope (0.9672 to 1.0458) and the intercept (-6.2814 to -0.2789).
The assay measuring range 6-180 mg/L
Lambda - Linearity at 1:10 gave y = 1.0678x - 4.0516 (r² = 0.9951). The 95% CI for the slope and the intercept were (1.0006 to 1.1350) and (-7.6061 to -0.4981) respectively (9 samples)
Linearity at 1:5 gave y = 1.0348x + 0.8347 (r² = 0.9962). The 95% CI for the slope and the intercept were (0.9709 to 1.0986) and (-5.2347 to 6.9040) respectively (10 samples)
Linearity at 1:1 gave y = 1.0308x - 15.4130 (r² = 0.9782). The 95% CI for the slope and the intercept were (0.9491 to 1.1314) and (-80.2294 to 49.4033) respectively (4 samples).
{4}
Page 5 of 8
The combined data gave y = 1.0077x - 1.3177 (r² = 0.9764) with 95% CI of the slope and the intercept (0.9365 to 1.0788) and (-7.8189 to 5.1835) respectively. The assay measuring range is 8-270 mg/L
The combined linearity data for both analytes on the BNII and IMMAGE were also reanalyzed using NCCLS EP6 and found to be acceptable.
c. Traceability (controls, calibrators, or method):
No reference standards or method available. For each assay kit, a calibrator, a low and a high positive control are included.
d. Detection limit (functional sensitivity):
Not performed for the purpose of this submission.
e. Analytical specificity:
Interference was assessed by spiking high concentrations of triglycerides (1930 formazine turbidity units), hemoglobin (5g/L) and bilirubin (200 mg/L) into aliquots of a test serum sample containing known concentrations of free kappa and free lambda light chains. In addition, saline was used as control for hemoglobin and bilirubin and chyle blank for triglycerides. All assays were run in triplicate. Results of spiked in samples were compared to control samples to determine % interference. No significant interference was observed with the BNII or the IMMAGE assays.
f. Assay cut-off:
Not performed for the purpose of this submission
2. Comparison studies:
a. Method comparison with predicate device:
The modified kappa and lambda kits were compared with the previously cleared kits. All normal and patient samples were from Birmingham. The normal samples were collected from a blood transfusion service and the patient samples were from a university hospital clinical immunology laboratory. All samples were stored frozen at -20°C prior to use.
BNII
Kappa - 95 serum samples consisted of 21 normal, 8 SLE and 66 myeloma samples were analyzed. The concentration of kappa free light chains ranged from 0.12 mg/L to 2615 mg/L with a mean of 165.57 mg/L and a median of 7.66 mg/L. Linear regression analysis yielded y = 0.9818x + 4.5206 and the correlation coefficient (r) was 0.9956. No 95% CI for the slope or the y-intercept was provided. As calculated by the reviewer, the 95% CI for the slope is 0.9627 to 1.0009 and for the intercept, -4.6485 to 13.6072.
{5}
Page 6 of 8
Lambda - 54 samples consisted of 5 normal, 6 SLE and 43 myeloma samples were assayed. The concentration of lambda free light chains ranged from 2.15 mg/L to 3455 mg/L with a mean of 442.65 mg/L and a median of 17.03 mg/L. Linear regression yielded y = 0.9266x + 9.5556 and the correlation coefficient (r) was 0.9752. Similar to kappa, no 95% CI for the slope and intercept was provided. As calculated by the reviewer, the 95% CI for the slope is 0.8681 to 0.9851 and for the intercept, -47.5224 to 66.5807.
## IMMAGE
Kappa - 63 serum samples consisted of 21 normal, 2 SLE and 40 myeloma samples were analyzed. The concentration of kappa free light chains ranged from 1.6 mg/L to 2198 mg/L with a mean of 155.49 mg/L and a median of 6.83 mg/L. Linear regression analysis yielded y = 1.0314x + 2.6792 and the correlation coefficient (r) was 0.9805. No 95% CI for the slope or the y-intercept was provided. As calculated by the reviewer, the 95% CI for the slope is 0.9785 to 1.0843 and for the intercept, -19.962 to 25.3215.
Lambda - 65 samples consisted of 21 normal, 2 SLE and 42 myeloma samples were assayed. The concentration of lambda free light chains ranged from 2.4 mg/L to 3738 mg/L with a mean of 425.13 mg/L and a median of 11.89 mg/L. Linear regression yielded y = 0.8503x -11.42 and the correlation coefficient (r) was 0.9651. Similar to kappa, no 95% CI for the slope and intercept was provided. As calculated by the reviewer, the 95% CI for the slope is 0.7922 to 0.9083 and for the intercept, -62.82 to 39.95.
To demonstrate equivalent performance between the IMMAGE and BNII FREELITE Kappa and Lambda Free kits, the following comparison studies were performed:
Kappa - Samples for this study consisted of 21 normal samples, 2 SLE samples and 34 myeloma samples. The normal samples were from a blood transfusion service and the clinical samples from a clinical laboratory. Using the BNII kit, the mean kappa free light chain concentration was 126 mg/L (ranged from 0.08 mg/L to 2330 mg/L) as compared to a mean concentration of 140.5 mg/L (ranged from 0.69
{6}
Page 7 of 8
mg/L to 2460 mg/L) with the IMMAGE kit. The difference between the sample means was not statistically significant. Linear regression analysis yielded y = 1.07x + 5.89 and the correlation coefficient (r) was 0.99. There was no statistical analysis of the slope or y-intercept. As calculated by the reviewer, the 95% confidence interval for the y-intercept was -12.66 to 24.45 and for the slope, 1.02 to 1.11. The standard error with y (IMMAGE) for a given x (BNII) was 66 mg/L.
Lambda - Samples for this study consisted of 21 normal samples, 2 SLE samples and 29 myeloma samples. The normal samples were from a blood transfusion service and the clinical samples from a clinical laboratory. Using the BNII kit, the mean lambda free light chain concentration was 296.5 mg/L (ranged from 0.78 mg/L to 1980 mg/L) as compared to a mean concentration of 296.6 mg/L (ranged from 0.69 mg/L to 2420 mg/L) with the IMMAGE kit. The difference between the sample means was not statistically significant. Linear regression analysis yielded y = 0.97x + 10.08 and the correlation coefficient (r) was 0.97. There was no statistical analysis of the slope or y-intercept. As calculated by the reviewer, the 95% confidence interval for the y-intercept was -34.64 to 54.79 and for the slope, 0.90 to 1.05. The standard error with y (IMMAGE) for a given x (BNII) was 140 mg/L.
No new comparison study was performed for the HITACHI kits since these kits have demonstrated equivalent performance to the BNII kits in previous submissions.
b. Matrix comparison:
Since the sample matrices are the same, .no comparison study is needed.
3. Clinical studies:
a. Clinical sensitivity:
Not applicable.
b. Clinical specificity:
Not applicable.
c. Other clinical information
Not applicable
4. Clinical cut-off:
Not applicable.
5. Expected values/Reference range:
Not applicable for the purpose of this submission.
{7}
Page 8 of 8
M. Conclusion:
Based on the review of the information provided in this 510 (k) submission, the modified devices are Substantially Equivalent to the predicate devices regulated under 21 CFR §866.5555, Immunoglobulin (light chain specific) immunological test system (Class II, product code DFH, Kappa, antigen, antiserum, control and DEH, Lambda, antigen, antiserum, control, Immunology Device Panel 82).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.