← Product Code [GEX](/productcode/GEX) · K034030

# APOGEE ELITE (K034030)

_Cynosure, Inc. · GEX · Feb 9, 2004 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K034030

## Device Facts

- **Applicant:** Cynosure, Inc.
- **Product Code:** [GEX](/productcode/GEX.md)
- **Decision Date:** Feb 9, 2004
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The Cynosure Apogee Elite laser is indicated for permanent hair reduction and the treatment of vascular lesions, pigmented lesions, and wrinkles.

## Device Story

Cynosure Apogee Elite Laser is a dual-wavelength medical laser system (755 nm Alexandrite and 1064 nm Nd:YAG). Device delivers laser energy to target tissues for dermatological procedures. Operated by clinicians in professional settings via finger or foot switch. System utilizes selective photothermolysis to target melanin in hair follicles or vascular/pigmented structures. Output facilitates hair reduction, coagulation/hemostasis of vascular lesions, and treatment of benign cutaneous lesions and wrinkles. Benefits include non-invasive management of aesthetic and dermatological conditions.

## Clinical Evidence

No clinical data provided; bench testing only.

## Technological Characteristics

Dual-wavelength laser system: 755 nm (Alexandrite crystal rod) and 1064 nm (Nd:YAG). Dimensions: 113x53x84 cm; Weight: 45 kg. Electrical: 230 VAC, 50-60 Hz, single phase. Activation via finger or foot switch.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- Cynosure Apogee TKS
- Cynosure Acclaim

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

FEB - 9 2004

:

FEB - 9 2004 K0340 30 1 of 1

## 510(K) Summary

| Submitter:                    | Cynosure, Inc.<br>10 Elizabeth Drive<br>Chelmsford, MA 01824                                                                                                                                                                                                                                                                                                                                              |
|-------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact:                      | George Cho<br>Senior Vice President of Medical Technology                                                                                                                                                                                                                                                                                                                                                 |
| Date Summary Prepared:        | December 23, 2003                                                                                                                                                                                                                                                                                                                                                                                         |
| Device Trade Name:            | Cynosure Apogee Elite Laser                                                                                                                                                                                                                                                                                                                                                                               |
| Common Name:                  | Medical Laser System                                                                                                                                                                                                                                                                                                                                                                                      |
| Classification Name:          | Instrument, surgical, powered, laser<br>79-GEX<br>21 CFR 878.4810                                                                                                                                                                                                                                                                                                                                         |
| Equivalent Device:            | Cynosure Apogee TKS, Cynosure Acclaim                                                                                                                                                                                                                                                                                                                                                                     |
| Device Description:           | Cynosure Apogee Elite laser (755nm), having an Alexandrite crystal<br>rod as the lasing medium, with a wavelength of 755 nm. The 1064 nm<br>wavelength is generated from the Nd:YAG laser head.<br>Laser activation is by either finger or foot switch. Overall weight of<br>the laser is 45 Kg, and the size is 113x53x84 cm (HxWxD).<br>Electrical requirement is 230 VAC, 50A, 50-60 Hz, single phase. |
| Intended Use:                 | The Cynosure Apogee Elite laser is indicated for permanent hair<br>reduction and the treatment of vascular lesions, pigmented lesions, and<br>wrinkles.                                                                                                                                                                                                                                                   |
| Comparison:                   | The Cynosure Apogee Elite laser is substantially equivalent to the<br>predicate devices, with the same principle of operation, the same<br>wavelength and essentially the same power range and the same<br>indications for uses.                                                                                                                                                                          |
| Nonclinical Performance Data: | none                                                                                                                                                                                                                                                                                                                                                                                                      |
| Clinical Performance Data:    | none                                                                                                                                                                                                                                                                                                                                                                                                      |
| Conclusion:                   | The Cynosure Apogee Elite laser is another safe and effective device<br>for permanent hair reduction and treatment of vascular lesions,<br>pigmented lesions and wrinkles                                                                                                                                                                                                                                 |
| Additional Information:       | none                                                                                                                                                                                                                                                                                                                                                                                                      |

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines representing its body and wings. The eagle is facing right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

FEB - 9 2004

Mr. George Cho Senior Vice President of Medical Technology Cynosure, Inc. 10 Elizabeth Drive Chelmsford, Massachusetts 01824

Re: K034030 Trade/Device Name: Cynosure Apogee Elite Laser Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: GEX Dated: December 24, 2003 Reccived: December 29, 2003

Dear Mr. Cho:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.

{2}------------------------------------------------

Page 2 - Mr. Goorge Cho

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Sincerely yours,

Mark N. Mulkern

Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

510(k) Number (if known):

K034030

Device Name: _ Cynosure Apogee Elite Dermatological Laser

Indications For Use:

## 755 nm:

The Cynosure Apogee Elite Dermatological Laser is indicated for stable long-term, or permanent hair reduction. Permanent hair reduction is defined as long-term stable reduction in the number of hair regrowth after a treatment regime. It is used for all skin types (Fitzpatrick I - VI) including tanned skin.

It is also indicated for the treatment of vascular lesions, benign pigmented lesions, and wrinkles.

1064 nm:

The Cynosure Apogee Elite Dermatological laser is intended for the coagulation and hemostasis of benign vascular lesions such as, but not limited to, port wine stains, hemangiomas, warts, telangiectasia, rosacea, venus lake, leg veins, spider veins and poikiloderma of civatte; and treatment of benign cutaneous lesions such as warts, scars, striae and psoriasis. The laser is also intended for the treatment of benign pigmented lesions such as, but not limited to, lentigos (age spots), solar lentigos (sun spots), café' au lait macules, seborrheic keratoses, nevi, chloasma, verrucae, skin tags, keratoses, tattoos (significant reduction in the intensity of black and/or blue/black tattoos) and plaques.

The laser is also indicated for the treatment of wrinkles such as, but not limited to, periocular and perioral wrinkles.

Additionally, the laser is indicated for the removal of unwanted hair, for the stable long term, or permanent, hair reduction through selective targeting of melanin in hair follicles, and for the treatment of pseudofolliculitis barbae (PFB).

## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

*Manch n Melkun*

ston Sign-Off
ion of Gerners

K034030

Image /page/3/Picture/13 description: The image shows the words "Prescription Use" in a simple, sans-serif font. The text is left-aligned and appears to be part of a larger document or form. The words are printed in black ink on a white background.

Over-The- Counter-Use

---

**Source:** [https://fda.innolitics.com/device/K034030](https://fda.innolitics.com/device/K034030)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
