RT 138 & RT 141 DUAL-HEATED NEONATAL BREATHING CIRCUITS

K034026 · Fisher & Paykel Healthcare, Ltd. · BZE · Jan 12, 2005 · Anesthesiology

Device Facts

Record IDK034026
Device NameRT 138 & RT 141 DUAL-HEATED NEONATAL BREATHING CIRCUITS
ApplicantFisher & Paykel Healthcare, Ltd.
Product CodeBZE · Anesthesiology
Decision DateJan 12, 2005
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.5270
Device ClassClass 2
AttributesTherapeutic, Pediatric

Intended Use

The dual-heated breathing circuits are intended as conduits of breathing gas for ventilation of patients, and to maintain the temperature of humidified inspired gas. The RT236 is used for flow rates between 0.3 and 4 L/min, and the RT138 and RT141 neonatal breathing circuits are intended to deliver humidified breathing gases for administration to an infant patient. Gases available for medical use do not contain sufficient moisture and may damage or irritate the respiratory tract, or desiccate secretions of patients whose supraglottic airways have been bypassed. This may be indicated for patients requiring mechanical ventilation, positive pressure breathing assistance, or general medical gases. These gases may be delivered by facemask or through bypassing the upper airways, for example use of an endotracheal tube.

Device Story

Dual heated neonatal breathing circuits (RT138, RT141, RT235, RT236) function as conduits within respiratory humidification systems. Inspiratory and expiratory limbs deliver humidified gas to infant patients and remove expired gas. Integrated heater wires in both limbs minimize condensate formation. RT138/RT236 designed for low flow (0.3–4 L/min); RT141/RT235 for high flow (>4 L/min). Used in clinical settings by healthcare providers to prevent respiratory tract damage or secretion desiccation caused by dry medical gases. Output is the delivery of temperature-controlled, humidified gas to the patient, supporting mechanical ventilation or positive pressure assistance.

Clinical Evidence

Bench testing only. Testing included material evaluation per ISO 10993-1, inspiratory and expiratory limb performance testing, and electrical/thermal safety testing. No clinical data provided.

Technological Characteristics

Dual-heated breathing circuits with heater wires in both inspiratory and expiratory limbs. Materials evaluated per ISO 10993-1. Designed for specific flow rate ranges (0.3-4 L/min or >4 L/min). Electrical and thermal safety features integrated for condensate management.

Indications for Use

Indicated for infant patients requiring mechanical ventilation, positive pressure breathing assistance, or general medical gases, including those with bypassed upper airways (e.g., endotracheal tube) or facemask delivery.

Regulatory Classification

Identification

A breathing system heater is a device that is intended to warm breathing gases before they enter a patient's airway. The device may include a temperature controller.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ 1/12/05 K034026 Fisher & Paykel Healthcare Limited 15 Maurice Paykel Place, East Tamaki P O Box 14 348, Panmure Auckland, New Zealand Telephone: +64 9 574 0100 Facsimile: +64 9 574 0158 Website: www.fphcare.com #### Summary of safety and effectiveness 1.3 | Contact person | Adele Bindon | |---------------------|----------------------------------------------------------------------------------------------| | Date prepared | '22 December 2003 | | Trade name | RT138 Dual Heated Neonatal Breathing Circuit<br>RT141 Dual Heated Neonatal Breathing Circuit | | Common name | Heated Breathing Circuit | | Classification name | Breathing System Heater (21 CFR 868.5270) | | Predicate device | RT130 Neonatal Breathing Circuit K020332<br>RT131 Neonatal Breathing Circuit K020332 | 12.3 {1}------------------------------------------------ #### Description of device: The RT138 and RT141 dual heated neonatal breathing circuits are classified as 'Breathing System Heater' according to 21 CFR §868.5270. Infant breathing circuits form part of the respiratory humidification system in which the inspiratory limb delivers humidified gas to the patient and the expiratory limb carries the expired gas away from the patient. Heater wires in the inspiratory and expiratory limb minimises the formation of condensate. The RT138 is for gas flows of 0.3 - 4 L/min and the RT141 is for gas flows greater than 4 L/min. #### Intended use: The RT138 and RT141 neonatal breathing circuits are intended to deliver humidified breathing gases for administration to an infant patient. Gases available for medical use do not contain sufficient moisture and may damage or irritate the respiratory tract, or desiccate secretions of patients whose supraglottic airways have been bypassed. This may be indicated for patients requiring mechanical ventilation, positive pressure breathing assistance, or general medical gases. These gases may be delivered by facemask or through bypassing the upper airways, for example use of an endotracheal tube. ### Technological characteristics summary: The RT138 and RT141 are dual-heated breathing circuits, compared to the predicate devices (RT130 and RT131) which are single-heated breathing circuits. This means that there is a heater wire in both the inspiratory and expiratory limb. The inspiratory limb is identical in all aspects to that of the predicate device. The expiratory limb has different performance characteristics to that of the predicate, reflecting the addition of a heater wire. The intended use of the RT138 and RT141 is the same as the predicate devices. #### Summary of testing: The following testing has been performed on the RT138 and RT141: - . Materials All materials used in the RT138 and RT141 have either been evaluated according to tests outlined in ISO 10993-1; or have been previously used in a predicate device; or a justification for minimal testing has been provided - . Inspiratory and expiratory limb performance - . Electrical and thermal safety {2}------------------------------------------------ ### Conclusions demonstrating safety, effectiveness and performance: The testing carried out for the RT138 and RT141 indicates that it meets design and performance functional requirements. This information indicates that the RT138 and RT141 are equivalent to or better than the predicate devices in terms of safety, effectiveness and performance. {3}------------------------------------------------ ## DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is circular and contains an image of a bird in flight. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" is arranged around the perimeter of the circle. #### Public Health Service JAN 1 2 2005 Food and Drug Administration 200 Corporate Boulevard Rockville MD 20850 Ms. Adele Bindon Regulatory Affairs Engineer Fisher & Paykel Healthcare, Limited 15 Maurice Paykel Place, East Tamaki PO Box 14 348, Panmure Auckland, New Zealand Re: K034026 Trade/Device Name: RT236 and RT235 Dual Heated Neonatal Breathing Circuits Regulation Number: 868.5270 Regulation Name: Breathing System Heater Regulatory Class: II Product Code: BZE Dated: December 23, 2004 Received: December 28, 2004 Dear Ms. Bindon: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ #### Page 2 - Ms. Bindon Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours. Chi-Lien, Ph.D. Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # 1.2 Indications for use statement K034026 510(k) Number Respiratory Humidifier Device Name The dual-heated breathing circuits are intended as conduits of breathing gas for ventilation of patients, and to maintain the temperature of humidified inspired gas. The RT236 is used for flow rates between 0.3 and 4 L/min, and the RT235 is for flow rates greater than 4 L/min, for infant patients. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) Please do not write below this line - continue on another page if needed Concurrence of CDRH, Office of Device Evaluation (ODE) Page 1 of 1
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...