← Product Code [DQA](/productcode/DQA) · K033876

# DOLPHIN 2000 OXIMETRY SENSORS, MODELS 2020, 2040, 2060, 2070, 2321, 2323, 2341, 2342, 2343, 2344, 2220, 2260 AND 2422 (K033876)

_Dolphin Medical, Inc. · DQA · Mar 11, 2004 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K033876

## Device Facts

- **Applicant:** Dolphin Medical, Inc.
- **Product Code:** [DQA](/productcode/DQA.md)
- **Decision Date:** Mar 11, 2004
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 870.2700
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Pediatric

## Indications for Use

The Dolphin 2000™ Oximetry Sensors are indicated for use in continuous monitoring of arterial oxygen saturation and pulse rate.

## Device Story

Dolphin 2000 Oximetry Sensors are disposable or re-usable pulse oximetry probes. Disposable versions use adhesive bandages to secure emitter/detector diodes to measurement sites (finger, toe, or neonatal foot). Re-usable versions use clothespin-style clips or Y-shaped configurations with bandages to maintain contact with the patient's finger, ear, or hand. Sensors connect to compatible pulse oximeter monitors. Operation relies on optical assessment of tissue oxygenation via LEDs and photodiodes; light passes through tissue to detectors. Monitors process these signals to calculate arterial oxygen saturation (SpO2) and pulse rate. Used in clinical settings by healthcare providers to monitor patient status. Output provides real-time physiological data to assist in clinical decision-making and patient management.

## Clinical Evidence

Clinical validation performed via breathe-down protocols at the VA Hospital of Wisconsin-Milwaukee. Study involved human volunteers with SpO2 values ranging from 70-100% at rest. Accuracy was demonstrated by comparing Dolphin 2000 SpO2 values to functional SaO2 values. Data analyzed to determine ARMS (Accuracy Root Mean Square) for each probe.

## Technological Characteristics

Optical sensing via LEDs and photodiodes. Disposable sensors use adhesive tape laminates; re-usable sensors use spring-loaded clips or Y-shaped housings. Compatible with major pulse oximeter monitors. Tested to EN 60601-1 (1990), EN 60601-1-1 (1993), EN 60601-1-2 (1993), and ASTM F1415-92. Non-sterile. Biocompatible materials for prolonged skin contact.

## Regulatory Identification

An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.

## Predicate Devices

- Model 2010 ([K030952](/device/K030952.md))
- 2351 ([K012989](/device/K012989.md))
- 2352 ([K012989](/device/K012989.md))
- 2353 ([K012989](/device/K012989.md))
- 2354 ([K012989](/device/K012989.md))
- 2210 ([K032947](/device/K032947.md))
- 2421 ([K030952](/device/K030952.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

KG33876

# MAR 1 1 2004

## 14. 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS

## Dolphin 2000™ Pulse Oximetry Y Sensor 12/11/03

#### Submitter ( Consultant name and Address)

Bill Cuman 9433 S. Morning Glory Lane Highlands Ranch, CO 80130

Phone: 720-939-6482 Fax: 786-551-8221

#### Sponsor Company Name and Address and Contact Person

Dolphin Medical Inc. 12525 Chadron Avenue Hawthorne, CA 90250

Tammy Conway, QA Manager (310) 978-0516 phone: (310) 978-1816 fax:

### Manufacturing Facility Name and Address

Opto Sensors (M) Sdn. Bhd. No. 6 Jalan Angkasa Mas 1 Tabrau Industrial Estate II 81100 Johor Bahru, Malaysia

#### Common, Classification & Proprietary Names

Common Name: oximetry sensor Classification Name: oximeter Dolphin™ 2000 Oximetry Sensors Proprietary Name:

### Predicate Devices

| Sensor Model #              | Predicate<br>Model # | Predicate<br>510(k) # |
|-----------------------------|----------------------|-----------------------|
| 2020, 2040, 2060, &<br>2070 | Model 2010           | K030952               |
| 2321 & 2341                 | 2351                 | K012989               |
| 2342                        | 2352                 | K012989               |
| 2323, & 2343                | 2353                 | K012989               |
| 2344                        | 2354                 | K012989               |
| 2220 & 2260                 | 2210                 | K032947               |
| 2422, 2424, & 2426          | 2421                 | K030952               |
| Extension Cables            |                      |                       |

{1}------------------------------------------------

#### Device Description

Device Describition
The Dolphin 2000™ Oximetry Sensors are fully compatible disposable and re-usable The Dolphin 2000 ~ Oximelly Ochson's are raily of pulse oximeter monitors. They replacement sensors for use with major brands of pares while in the first and BCI Compatabile Sensors.

The disposable Dolphin 2000 Oximetry Sensors are constructed in a similar manner to The disposable Dolphin 2000 Oximelly Ochson are embedded in a laminate of tapes 
predicate devices. The emitter and detector diodes are embedded in a lands a predicate devices. The childer and development an adhesive bandage that is connected to the cable assombly. "The consortient by wrapping it around a finger backing that allows the sensor to be upplied to the End (Colphin 2000 Oximetry Sensors are of toe (measurement site). Tour sizes of aloposes, infant and neonatal application available, which are indicated for doc for duall, positive, mon-sterile for single patient use.

The re-usable Dolphin 2000 Finger Clip Oximetry Sensor is an adult-sized clothespin-The re-usable Dolphil 2005 Finger Olip Oximony Consor consists of the style clip that is placed on the one on anyon one on anyon maintained in mild
emitter and detector components mounted in opposing clip halves, mainter and consected emitter and delection components mounted in opposing only his house the optiolection in complession by a spring mings. The montain contact with the patient's finger. Clear components within contoured paids that the read to pass through the finger for the measurements.

The Re-usable Y sensor is for use on the ear, finger, hand, or neonatal foot and held in The Re-usable 1 Sensor is for use on the sensor can also be used on the adult ear with the place with a disposable bandage. The concer or enounted in a sealed pouch (same ear clip accessory." The emitter and other cructed in a Y shape. All Dolphin sensors are provided non-sterile.

#### Intended Use

Intended Use
The Dolphin 2000™ Oximetry Sensors are indicated for use in continuous monitoring of arterial oxygen saturation and pulse rate.

## Technological Characteristics Comparison

The Dolphin 2000™ Oximetry Sensors are substantially equivalent in intended use, The Dolphill 2000 - Oximony Conbore e. e commercially available oximetry sensors.

All of the Dolphin 2000™ oximetry sensors and the predicate devices operate on the All of the Dolphin 2000 - (Allinor) optical assessment of tissue oxygenation using emitters (LEDs) and detectors (photodiode).

The Dolphin 2000™ oximetry sensors are designed, configured, and manufactured for The Dulphill 2000 - Uximistry beneercially-available oximetry monitors. full companibility for use with the laboled, connections of equivalent specifications as used in the predicate devices.

The Dolphin 2000™ oximetry sensors, like the predicate devices are available in both The Dolphin 2000 - oxinstly Schools, ill.o the programs, in adult, pediatric, infant and neonatal populations.

The labeled accuracy of the Dolphin 2000 sensors is equivalent to those of the predicate devices.

{2}------------------------------------------------

### Performance Testing

#### Biocompatibility t

Biocompatibility tests, appropriate for skin-contacting devices for prolonged in Blocompaniblity tests, approphiats for onf the device components used in the assembly of the Dolphin 2000™ pulse oximetry sensors. Test results assembly of the Dolphin 2000 - paiso existed , initiant, and non-sensitizing.

#### Electrical Safety 피

Electrical Salety
The Dolphin 2000 Oximetry Sensors have been tested and found to comply with the applicable clauses of the following standards:

- Medical electrical equipment part 1: General EN 60601-1 (1990) requirements for safety
- EN 60601-1-1 (1993) Medical electrical equipment part 1: General 1 EN 60001-1-1 (1507) Modical Standard: Safety requirements for medical electrical systems
- EN 60601-1-2 (1993) Medical electrical equipment part 1: General – requirements for safety - 2. Collateral standard: Electromagnetic compatibility requirements and tests
- Standard Specification for Pulse Oximeters ASTM F1415-92 .

#### Clinical Testing

The sensors were validated in breathe-down protocols at the VA Hospital of Wisconsin - Milwaukee, (I)r. Phillip Clifford, MD.). Scientific accuracy was Wisconsin - Millindakoo, Companing Dolphin 2000 SpO2 values to functional SaO2 demorristiated by statistically oompanies as a the down protocol at rest (i.e. no motion) while values. Volunteers partionated in thing from 70-100%. Data was analyzed to determine the ARMS for each probe.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the top half of the circle. Inside the circle is a stylized image of an eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAR 1 1 2004

Dolphin Medical, Incorporated C/O Mr. Bill Curnan Regulatory Specialist Bill Curnan 9433 S. Morning Glory Lane Littleton, Colorado 80130

Rc: K033876

Trade/Device Name: Dolphin 2000 Oximetry Senors: Models 2020, 2040, 2060, 2070, 2220, 2321, 2323, 2341, 2342, 2343, and 2344 Regulation Number: 870.2700 Regulation Name: Oximeter Regulatory Class: II Product Code: DQA Dated: February 10, 2004 Received: February 12, 2004

Dear Mr. Curnan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting vour device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Mr. Curnan

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely vours,

Chiu Lin, Ph.D.

Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Attachment l

## Indications for Use

510(k) Number (if known): K033876

Device Name: Dolphin 2000 Oximetry Sensors

Indications For Use: The Dolphin 2000 Oximetry Sensors are indicated for use in continuous monitoring of arterial oxygen saturation and pulse rate.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

fthh

(Division Sign-Off) ( Cholon Oigh-Oit)
Division of Anesthesiology, General Hospital, e Malon of Anesthesiology, Gene Infection Control, Dental Devices

Page 1 of

510(k) Number: K033876

---

**Source:** [https://fda.innolitics.com/device/K033876](https://fda.innolitics.com/device/K033876)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
